Respiratory FailureNCT04030208
Evaluating Safety and Efficacy of Umbulizer in Patients Requiring Intermittent Positive Pressure Ventilation
A study testing Umbulizer and Traditional Mechanical Ventilator in people with Respiratory Failure and Respiratory Insufficiency, run by Umbulizer LLC at 1 site in Pakistan.
Trial summary
Randomized, control crossover study comparing Umbulizer's efficacy to traditional mechanical ventilators Read more about this study
- Who is it for?
- Ages 18 to 85 years
- Conditions the study is for
- Respiratory Failure, Respiratory Insufficiency, Respiratory Arrest
- What is being tested?
- Umbulizer and Traditional Mechanical Ventilator
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 24 participants
- Where
- 1 sitePakistan
- Study dates
- 4 Sep 2018to 31 Dec 2019
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 85 years
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- 1) Patients needing mechanical ventilation
Who cannot join
- Patients with significant airway resistance (i.e. severe chronic obstructive pulmonary disease or asthma) where clinical investigator is concerned about gas trapping or barotrauma
- Patients with very low lung compliance (i.e. severe acute respiratory distress syndrome or restrictive disease) where clinical investigator is concerned about problems delivering sufficient tidal volume or end expiratory pressure
- Patients who experienced a myocardial infarction within the last 6 weeks
- Patients who are determined by a clinical investigator to be too hemodynamically unstable to be enrolled
Trial locations and contacts
1 site in 1 country.
About this study
Randomized, control crossover study comparing Umbulizer's efficacy to traditional mechanical ventilators
Read the detailed description
Mechanical ventilators cost tens of thousands of dollars, rendering them difficult for low and middle income countries to procure. Given chronic shortages of these devices in places like Pakistan, patients are often ventilated manually with Ambu Bags for hours or days. This method is not suitable for long term ventilation and is extremely life threatening.
Umbulizer seeks to reduce preventable deaths due to ventilator scarcity by providing a reliable alternative. The device is accurate, low-cost, and portable, and is an ideal solution for treating patients with respiratory diseases in low-resource settings. The primary objective of this study is to test efficacy and determine non-inferiority of this device in comparison to ventilation provided by traditional ventilators. Secondary objectives of this study include:
- Testing safety of the device for use over a predetermined period of time
- Evaluating patient comfort during ventilation with this device
- Assessing ease of use of this device for doctors and medical staff
- Being tested
- Umbulizer, Traditional Mechanical Ventilator
- Keywords
- respiratory disease, low resource settings, mechanical ventilation, ventilator performance, comparative study, patient safety, ease of use
- Medical subject headings
- Respiratory Insufficiency, Apnea, Respiration Disorders
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
UmbulizerDevice
Umbulizer is a reliable, user-friendly, and portable system that provides consistent air to the patient by quantitatively delivering a pre-determined tidal volume, inhale/exhale (I/E) ratio, peak pressure, and breathing rate (BPM). It can be configured to deliver either controlled ventilation or assist-controlled ventilation.
Traditional Mechanical VentilatorDevice
A medical device designed to artificially perform breathing functions for patients who are unable to adequately perform these functions on their own.
| Study group | Role | What they receive |
|---|---|---|
| Sequence A | Other | If patient is randomly assigned to Sequence A, they will be placed on a traditional ventilator for the first period. This period will be 1 hour in duration. Delivered tidal volume, respiratory rate (RR), and peak pressure will be recorded throughout the period using the ventilator. Heart rate (HR), Blood Pressure (BP), and oxygen (O2) saturation measurements will be recorded every 5 minutes. Arterial blood gas (ABG) data will be taken at t = 1 hour if the patient is stable on the traditional ventilator. ABGs may be taken more frequently if the patient is destabilizing. Next, the study participant will be transitioned to the Umbulizer to begin Period 2. This period will also be 1 hour in duration. Delivered tidal volume, RR, and peak pressure will be recorded throughout the period using Umbulizer. HR, BP, O2 saturation measurements will be taken every 5 minutes. ABG data will be taken at t = 30 minutes and 1 hour. Receives: Device: Umbulizer; Device: Traditional Mechanical Ventilator |
| Sequence B | Other | If patient is randomly assigned to Sequence B, they will be shifted to the Umbulizer for the first period. This period will be 1 hour in duration. Delivered tidal volume, RR, and peak pressure will be recorded throughout the period using Umbulizer. HR, BP, O2 saturation measurements will be recorded every 5 minutes. ABG data will be taken at t = 30 minutes and 1 hour. Next, the study participant will be transitioned to a traditional ventilator to begin Period 2. This period will also be 1 hour in duration. Delivered tidal volume, RR, and peak pressure will be recorded throughout the period using the ventilator. HR, BP, O2 saturation measurements will be recorded every 5 minutes. ABG data will be taken at t = 1 hour if the patient is stable on the traditional ventilator. ABGs may be taken more frequently if the patient is destabilizing. Receives: Device: Umbulizer; Device: Traditional Mechanical Ventilator |
What is being measured
The primary outcome is the main question the study is designed to answer.
Partial Pressure of Carbon Dioxide (PaCO2)
PaCO2 in mmHg at t = 1 hr (from Arterial Blood Gas or ABG readings) when patient is on Umbulizer compared to PaCO2 at t = 1 hr when patient is on a traditional mechanical ventilator
Measured: 2 hours
pH
pH at t = 1 hr (from ABG readings) when patient is on Umbulizer compared to pH at t = 1 hr when patient is on a traditional mechanical ventilator
Measured: 2 hours
Secondary outcomes (1)
Number of Adverse Events
Number of adverse events on Umbulizer compared to on a traditional ventilator
Measured: 2 hours
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted23 Jul 2019
- Study start4 Sep 2018
- Primary completion30 Nov 2019
- Study completion31 Dec 2019
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Respiratory FailureStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Supportive care
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- Single blind
- Who is masked
- Outcomes assessor
- Enrolment
- 24 participants (actual)
Sponsor and investigators
- Lead sponsor
- Umbulizer LLC
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Kamran Chima, MBBS, Principal investigator, Services Hospital
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT04030208
- Sponsor's protocol ID
- 1
- Status last verified
- Aug 2020
- First submitted
- 9 Jul 2019
- Last updated
- 6 Aug 2020
- Results posted
- No
- Data monitoring committee
- Yes
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- Undecided
Study at a glance
From the registry record, last updated by the sponsor on 6 Aug 2020. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Evaluating Safety and Efficacy of Umbulizer in Patients Requiring Intermittent Positive Pressure Ventilation
- ClinicalTrials.gov ID
- NCT04030208
- Acronym / sponsor ID
- 1
- Keywords
- respiratory disease, low resource settings, mechanical ventilation, ventilator performance, comparative study, patient safety, ease of use
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 24 participants (actual)
- Lead sponsor
- Umbulizer LLCIndustry (pharma, biotech, devices)
- Study start
- 4 Sep 2018 (actual)
- Primary completion
- 30 Nov 2019 (actual)
- Study completion
- 31 Dec 2019 (actual)
- First posted
- 23 Jul 2019 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Randomized, control crossover study comparing Umbulizer's efficacy to traditional mechanical ventilators
Detailed description
Mechanical ventilators cost tens of thousands of dollars, rendering them difficult for low and middle income countries to procure. Given chronic shortages of these devices in places like Pakistan, patients are often ventilated manually with Ambu Bags for hours or days. This method is not suitable for long term ventilation and is extremely life threatening.
Umbulizer seeks to reduce preventable deaths due to ventilator scarcity by providing a reliable alternative. The device is accurate, low-cost, and portable, and is an ideal solution for treating patients with respiratory diseases in low-resource settings. The primary objective of this study is to test efficacy and determine non-inferiority of this device in comparison to ventilation provided by traditional ventilators. Secondary objectives of this study include:
- Testing safety of the device for use over a predetermined period of time
- Evaluating patient comfort during ventilation with this device
- Assessing ease of use of this device for doctors and medical staff
Study design
- Primary purpose
- Supportive care
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- Single blind (outcomes assessor)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Sequence A | Other | Device: Umbulizer, Device: Traditional Mechanical VentilatorIf patient is randomly assigned to Sequence A, they will be placed on a traditional ventilator for the first period. This period will be 1 hour in duration. Delivered tidal volume, respiratory rate (RR), and peak pressure will be recorded throughout the period using the ventilator. Heart rate (HR), Blood Pressure (BP), and oxygen (O2) saturation measurements will be recorded every 5 minutes. Arterial blood gas (ABG) data will be taken at t = 1 hour if the patient is stable on the traditional ventilator. ABGs may be taken more frequently if the patient is destabilizing. Next, the study participant will be transitioned to the Umbulizer to begin Period 2. This period will also be 1 hour in duration. Delivered tidal volume, RR, and peak pressure will be recorded throughout the period using Umbulizer. HR, BP, O2 saturation measurements will be taken every 5 minutes. ABG data will be taken at t = 30 minutes and 1 hour. |
| Sequence B | Other | Device: Umbulizer, Device: Traditional Mechanical VentilatorIf patient is randomly assigned to Sequence B, they will be shifted to the Umbulizer for the first period. This period will be 1 hour in duration. Delivered tidal volume, RR, and peak pressure will be recorded throughout the period using Umbulizer. HR, BP, O2 saturation measurements will be recorded every 5 minutes. ABG data will be taken at t = 30 minutes and 1 hour. Next, the study participant will be transitioned to a traditional ventilator to begin Period 2. This period will also be 1 hour in duration. Delivered tidal volume, RR, and peak pressure will be recorded throughout the period using the ventilator. HR, BP, O2 saturation measurements will be recorded every 5 minutes. ABG data will be taken at t = 1 hour if the patient is stable on the traditional ventilator. ABGs may be taken more frequently if the patient is destabilizing. |
Who may qualify
- Ages
- 18 to 85 years
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (1 listed)
- 1) Patients needing mechanical ventilation
Key exclusion criteria (4 listed)
- Patients with significant airway resistance (i.e. severe chronic obstructive pulmonary disease or asthma) where clinical investigator is concerned about gas trapping or barotrauma
- Patients with very low lung compliance (i.e. severe acute respiratory distress syndrome or restrictive disease) where clinical investigator is concerned about problems delivering sufficient tidal volume or end expiratory pressure
- Patients who experienced a myocardial infarction within the last 6 weeks
- Patients who are determined by a clinical investigator to be too hemodynamically unstable to be enrolled
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| UmbulizerUmbulizer is a reliable, user-friendly, and portable system that provides consistent air to the patient by quantitatively delivering a pre-determined tidal volume, inhale/exhale (I/E) ratio, peak pressure, and breathing rate (BPM). It can be configured to deliver either controlled ventilation or assist-controlled ventilation. | Device | Sequence A, Sequence B |
| Traditional Mechanical VentilatorA medical device designed to artificially perform breathing functions for patients who are unable to adequately perform these functions on their own. | Device | Sequence A, Sequence B |
Primary outcome
Partial Pressure of Carbon Dioxide (PaCO2)
Time frame: 2 hours
PaCO2 in mmHg at t = 1 hr (from Arterial Blood Gas or ABG readings) when patient is on Umbulizer compared to PaCO2 at t = 1 hr when patient is on a traditional mechanical ventilator
pH
Time frame: 2 hours
pH at t = 1 hr (from ABG readings) when patient is on Umbulizer compared to pH at t = 1 hr when patient is on a traditional mechanical ventilator
1 secondary outcome
Number of Adverse Events
Time frame: 2 hours
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 6 Aug 2020.
- Individual participant dataSharing planned: Undecided
Related trials for Respiratory Failure
Enrolling by invitation
Personalized Ventilator Settings for Patients on ECMO
Acute Respiratory Distress Syndrome (ARDS), Extracorporeal Membrane Oxygenation
Not yet recruiting
Regional Lung Ventilation During Prone Position in in Intubated Children
Pediatric Acute Respiratory Failure
Not yet recruiting
Smartphone Camera Measurement of Breathing-Related Heart Rate Variability Compared With ECG in Healthy Adults
Respiratory Sinus Arrhythmia, Heart Rate Variability
Not yet recruiting
Mechanical Ventilator Adjustments and Patient Dyspnea
Respiratory Failure, Dyspnea During Mechanical Ventilation
Not yet recruitingObservational study
Clinical Genetics Branch Eligibility Screening Survey
Melanoma, Li-Fraumeni Syndrome
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Kamran Chima, MBBS principal investigator
Services Hospital
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Umbulizer and Traditional Mechanical Ventilator in people with Respiratory Failure and Respiratory Insufficiency, run by Umbulizer LLC at 1 site in Pakistan.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 1 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 6 Aug 2020. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Pakistan. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Partial Pressure of Carbon Dioxide (PaCO2) (time frame: 2 hours). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?