Validation of a Diagnostic Test for Glucose-6-phosphate Dehydrogenase (G6PD) Deficiency in Anti-coagulated Blood

A study testing SD Biosensor STANDARD G6PD Test, Pointe Scientific Test Kit and HemoCue System in people with G6PD Deficiency, run by PATH at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The primary objective of the study is to assess the accuracy of the SD Biosensor STANDARD™ point-of-care (POC) G6PD test in measuring G6PD activity and classifying results compared to a reference assay and across repeated measurements in capillary samples. Read more about this study

Who is it for?
Ages 18 to 65 yearsHealthy volunteers can take part
Conditions the study is for
G6PD Deficiency
What is being tested?
SD Biosensor STANDARD G6PD Test, Pointe Scientific Test Kit and HemoCue System
Could I receive a placebo?
No, everyone receives an active treatment
How many people
229 participants
Where
1 siteUnited States
Study dates
12 Aug 2019to 31 Jan 2020
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 65 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Must communicate an understanding of the study protocol.
  • Must be able to provide written consent to undergo screening and provide medical history.
  • Participants must be afebrile and in general good health in the opinion of the investigator as determined by vital signs, medical history, and physical examination.
  • Black/African-American, by self-report.

Who cannot join

  • Blood transfusion in the past 3 months by self-report

Trial locations and contacts

1 site in 1 country.

United States1 site
  • BSC corporation

    Reading, Pennsylvania, 19602

Other trials in United States

About this study

Official title: Validation of a Diagnostic Test for Glucose-6-phosphate Dehydrogenase Deficiency: Diagnostic Accuracy and Repeatability in Capillary Samples

The primary objective of the study is to assess the accuracy of the SD Biosensor STANDARD™ point-of-care (POC) G6PD test in measuring G6PD activity and classifying results compared to a reference assay and across repeated measurements in capillary samples.

Read the detailed description

Cross-sectional diagnostic accuracy study with 250 volunteer participants. The clinic will recruit and consent adult study participants. Clinic staff will draw venous blood and obtain finger stick capillary blood samples. Clinic staff will perform the investigational SD Biosensor STANDARD™ G6PD test for glucose-6-phosphate dehydrogenase (G6PD) deficiency and a HemoCue® hemoglobin test on finger stick capillary blood. An anti-coagulated venous blood sample will be sent to a Clinical Laboratory Improvement Amendments (CLIA) certified lab for G6PD reference testing by the gold standard assays: G6PD measurement by spectrophotometry using the Pointe Scientific G6PD reference assay and hemoglobin measurement by a hematology analyzer.

Individuals identified as G6PD deficient or intermediate by the reference test will be notified of their results by the clinic and referred to their physician for follow-up.

This study includes a nested repeatability study and a nested sample stability study. In the nested repeatability study up to 20 consented participants will provide 4 additional finger stick samples. Clinic staff will perform the SD Biosensor STANDARD™ G6PD test to assess the repeatability of the test in capillary blood over 8 G6PD and hemoglobin measurements on up to 3 different instruments.

In the nested sample stability study up to 8 consented participants will provide an additional venous blood draw sample to be tested at additional time points to document stability of the samples over time when tested by the SD Biosensor STANDARD™ G6PD Test.

Conditions
G6PD Deficiency
Being tested
SD Biosensor STANDARD G6PD Test, Pointe Scientific Test Kit, HemoCue System
Keywords
G6PD Deficiency
Medical subject headings
Glucosephosphate Dehydrogenase Deficiency

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • SD Biosensor STANDARD G6PD TestDiagnostic test

    Also called SD Biosensor G6PD Test

    The SD Biosensor G6PD Analyzer is designed to measure the quantitative determination of total hemoglobin concentration and G6PD enzymatic activity in fresh human whole blood specimens based on reflectometry assays in a point-of-care setting. The test is intended to aid in the identification of people with G6PD deficiency. The test is currently not licensed for use in the US and is considered an investigational product.

  • Pointe Scientific Test KitDiagnostic test

    The Pointe Scientific test kit will serve as the reference assay to assess G6PD activity. Its intended use is for the quantitative, kinetic determination of G6PD in blood at 340 nm.

  • HemoCue SystemDiagnostic test

    The HemoCue hemoglobin (Hb) 201+ system is designed for quantitative point-of-care whole blood hemoglobin determination in primary care using a specially designed analyzer, the HemoCue Hb 201+ Analyzer, and specially designed microcuvettes, the HemoCue Hb 201+Microcuvettes.

Study groups in NCT04054661
Study groupRoleWhat they receive
G6PD Diagnostic TestingOtherParticipants provided whole blood samples as well as fingerstick capillary blood samples. At the clinic site lab, study staff conducted the SD Biosensor STANDARD point-of-care G6PD test and the point-of-care HemoCue hemoglobin test on both finger stick blood and whole blood samples. At the reference laboratory, G6PD activity was measured from whole blood samples using the Pointe Scientific G6PD reference assay and hemoglobin was measured using a hematology analyzer.
Receives: Diagnostic Test: SD Biosensor STANDARD G6PD Test; Diagnostic Test: Pointe Scientific Test Kit; Diagnostic Test: HemoCue System

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Sensitivity of SD Biosensor STANDARD G6PD Test for Identifying G6PD Deficient Individuals

    For the purposes of this study, an individual was considered G6PD deficient if they tested positive in the Pointe Scientific assay. A true positive was defined as a participant with ≤ 30% of normal G6PD activity in circulating venous blood determined by the Pointe Scientific test.

    Diagnostic sensitivity of the SD Biosensor STANDARD G6PD test is defined as the percentage of participants who tested positive for G6PD deficiency on the reference test who were identified by the test assay as positive, calculated as:

    Number of participants who were both test and true positive / (Number of participants who were test and true positive + Number of participants who were test negative but true positive [ie false negative]) * 100%.

    Sensitivity of the SD Biosensor STANDARD G6PD test was calculated from both venous and capillary blood samples.

    Measured: All samples were collected on study day 1

  2. Sensitivity of SD Biosensor STANDARD G6PD Test for Identifying Women With Intermediate G6PD Activity

    To investigate the performance of the SD Biosensor STANDARD G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, assay sensitivity was determined at a G6PD activity threshold of 70%. A true positive for intermediate G6PD activity in females was defined as G6PD activity between 30 and 70% of normal in circulating venous blood as determined by the Pointe Scientific test.

    Sensitivity of the SD Biosensor STANDARD G6PD test is the percentage of women who tested positive for intermediate G6PD activity on the reference test who were identified by the test assay as positive, calculated as:

    Number of women who were both test and true positive / (Number of women who were test and true positive + Number of women who were test negative but true positive [ie false negative]) * 100%.

    Sensitivity of the SD Biosensor STANDARD G6PD test for identifying females with intermediate G6PD activity was calculated from both venous and capillary blood samples.

    Measured: All samples were collected on study day 1

  3. Specificity of SD Biosensor STANDARD G6PD Test for Identifying G6PD Deficient Individuals

    For the purposes of this study, an individual was considered G6PD deficient if they tested positive by the Pointe Scientific assay. A true negative was defined as > 30% of normal G6PD activity in circulating venous blood as determined by the Pointe Scientific test.

    Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as:

    Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative [ie false positive]) * 100%.

    Specificity of the SD Biosensor STANDARD G6PD test was calculated from both venous and capillary blood samples.

    Measured: All samples were collected on study day 1

  4. Specificity of SD Biosensor STANDARD G6PD Test for Identifying Women With Intermediate G6PD Activity

    To investigate the performance of the SD Biosensor STANDARD G6PD test to distinguish females with intermediate G6PD activity from females with normal activity, the specificity was determined at a G6PD activity threshold of 70%. A true negative for intermediate G6PD activity in females was defined as G6PD activity > 70% of normal in circulating venous blood as determined by the Pointe Scientific test.

    Specificity is the percentage of participants who tested negative for G6PD deficiency on the reference test who were identified as negative using the test assay, calculated as:

    Number of participants who were both test and true negative / (Number of participants who were test and true negative + Number of participants who were test positive but true negative [ie false positive]) * 100%.

    Specificity of the SD Biosensor STANDARD G6PD test was calculated from both venous and capillary blood samples.

    Measured: All samples were collected on study day 1

Secondary outcomes (8)
  1. Accuracy Between the SD Biosensor STANDARD G6PD Test Assay and the Pointe Scientific Test Kit

    Accuracy between the SD Biosensor STANDARD G6PD test assay and the Pointe Scientific G6PD reference assay was calculated as the percent agreement between both G6PD tests, ie, the percentage of participants with the same diagnosis according to both tests (either deficient, intermediate, or normal G6PD levels).

    Accuracy of the SD Biosensor STANDARD G6PD test was calculated from both venous and capillary blood samples.

    Measured: All samples were collected on study day 1

  2. Accuracy Between the SD Biosensor STANDARD G6PD Test Measure of Hemoglobin and the Reference Hemoglobin Test

    Accuracy between the SD Biosensor STANDARD G6PD test measure of hemoglobin and the hemoglobin reference assay was calculated as the percentage agreement between both tests, ie, the percentage of participants with the same diagnosis according to both tests (either non/mild anemia, moderate anemia or severe anemia).

    Measured: All samples were collected on study day 1

  3. Repeatability Substudy: G6PD Levels in Capillary Blood Samples Measured Using the SD Biosensor G6PD Test, by Operator and Analyzer

    Capillary blood samples were collected in duplicate from four different fingers on the participants hands, for a total of eight samples. Within each duplicate, a sample was tested on 1 of 2 SD Biosensor G6PD instruments by 1 of 2 operators. Overall there were 3 instruments used in the testing and 3 operators.

    Measured: All samples were collected on study day 1 and tested immediately

  4. Repeatability Substudy: Hemoglobin Levels in Capillary Blood Samples Measured Using the SD Biosensor STANDARD Hemoglobin Test, by Operator and Analyzer

    Capillary blood samples were collected in duplicate from four different fingers on the participants hands, for a total of eight samples. Within each duplicate, a sample was tested on 1 of 2 SD Biosensor instruments by 1 of 2 operators. Overall there were 3 instruments used in the testing and 3 operators.

    Measured: All samples were collected on study day 1 and tested immediately

  5. Sample Stability Substudy: G6PD Levels Measured Over Time After Storage at Room Temperature

    Each participant in the sample stability substudy underwent an additional blood draw in 3 different anti-coagulants which were stored at room temperature (15-30°C/59-86°F). The freshly collected venous whole blood samples were tested at the clinic site lab using the SD Biosensor STANDARD G6PD test within 1 hour of collection for baseline (0 hour) and at 4 additional time points, up to 2 days (48 hours) after collection. The anti-coagulants included:

    • ethylenediaminetetraacetic acid (EDTA)-anti-coagulated blood
    • heparin-anti-coagulated blood
    • acid citrate dextrose (ACD)-anti-coagulated blood

    Measured: Blood samples were collected on day 1. Samples were tested immediately (< 1 hour) and at 2, 4, 6, and 24 hours after collection.

  6. Sample Stability Substudy: G6PD Levels Measured Over Time After Refrigerated Storage

    Each participant in the sample stability substudy underwent an additional blood draw in 3 different anti-coagulants which were stored in a refrigerator (2-8°C/25-46°F). The freshly collected venous whole blood samples were tested at the clinic site lab using the SD Biosensor STANDARD G6PD test within 1 hour of collection for baseline (0 hour) and at 5 additional time points, up to 7 days (168 hours) after collection. The anti-coagulants included:

    • ethylenediaminetetraacetic acid (EDTA)-anti-coagulated blood
    • heparin-anti-coagulated blood
    • acid citrate dextrose (ACD)-anti-coagulated blood

    Measured: Blood samples were collected on day 1. Samples were tested immediately (< 1 hour) and at 24, 48, 72, 96, and 168 hours after collection.

  7. Sample Stability Substudy: Hemoglobin Levels Measured Over Time After Storage at Room Temperature

    Each participant in the sample stability substudy underwent a blood draw in 3 different anti-coagulants which were stored at room temperature (15-30°C/59-86°F). Hemoglobin levels were measured from the freshly collected venous whole blood samples at the clinic site lab using the SD Biosensor STANDARD hemoglobin test within 1 hour of collection for baseline (0 hour) and at 4 additional time points, up to 24 hours after collection. The anti-coagulants included:

    • ethylenediaminetetraacetic acid (EDTA)-anti-coagulated blood
    • heparin-anti-coagulated blood
    • acid citrate dextrose (ACD)-anti-coagulated blood

    Measured: Blood samples were collected on day 1. Samples were tested immediately (< 1 hour) and at 2, 4, 6, and 24 hours after collection.

  8. Sample Stability Substudy: Hemoglobin Levels Measured Over Time After Refrigerated Storage

    Each participant in the sample stability substudy underwent an additional blood draw in 3 different anti-coagulants which were stored in a refrigerator (2-8°C/25-46°F). Hemoglobin levels were measured from the freshly collected venous whole blood samples at the clinic site lab using the SD Biosensor STANDARD hemoglobin test within 1 hour of collection for baseline (0 hour) and at 5 additional time points, up to 7 days (168 hours) after collection. The anti-coagulants included:

    • ethylenediaminetetraacetic acid (EDTA)-anti-coagulated blood
    • heparin-anti-coagulated blood
    • acid citrate dextrose (ACD)-anti-coagulated blood

    Measured: Blood samples were collected on day 1. Samples were tested immediately (< 1 hour) and at 24, 48, 72, 96, and 168 hours after collection.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 Aug 2019
  2. Study start12 Aug 2019
  3. Primary completion31 Jan 2020
  4. Study completion31 Jan 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
229 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04054661
Sponsor's protocol ID
1416844
Status last verified
Aug 2021
First submitted
12 Aug 2019
Last updated
9 Nov 2021
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No: Not planning to share IPD to other researchers

Publications

  • Pal S, Myburgh J, Bansil P, Hann A, Robertson L, Gerth-Guyette E, Ambler G, Bizilj G, Kahn M, Zobrist S, Manis MR, Styke NA, Allan V, Ansbro R, Akingbade T, Bryan A, Murphy SC, Kublin JG, Layton M, Domingo GJ. Reference and point-of-care testing for G6PD deficiency: Blood disorder interference, contrived specimens, and fingerstick equivalence and precision. PLoS One. 2021 Sep 20;16(9):e0257560. doi: 10.1371/journal.pone.0257560. eCollection 2021.

    PubMed 34543346

Questions about this trial

What is this clinical trial studying?

A study testing SD Biosensor STANDARD G6PD Test, Pointe Scientific Test Kit and HemoCue System in people with G6PD Deficiency, run by PATH at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 9 Nov 2021. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 5 Oct 2021). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Sensitivity of SD Biosensor STANDARD G6PD Test for Identifying G6PD Deficient Individuals (time frame: All samples were collected on study day 1). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT04054661, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 9 Nov 2021.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in