Assessment of the Echoendoscope for Clinical Performance

An observational study of Endosonography, run by Peking Union Medical College Hospital at 1 site in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The primary objective for this study is to compare the safety and validity of two endoscopic ultrasonography (EUS) systems (GF-UE260-ME2 and EG-UR5-S50) used for clinical performance and establish a relative EUS performance assessment system and clinical case data to provide support for the product performance… Read more about this study

Who is it for?
Ages 18 to 70 years
Conditions the study is for
Endosonography
What is being tested?
Assessment of the safety and validity of echoendoscope
How many people
130 participants
Where
1 siteChina
Study dates
1 Nov 2019to 22 Feb 2022
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 70 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: All patients referred to PUMCH Endoscopy Unit for EUS assessment will be included.

Who may be able to join

  • The patients with clinical indication for the endoscopic ultrasonography examination or diagnosis;
  • Patients volunteering to participate in this study and sign the informed consent.

Who cannot join

  • Patients do not sign the informed consent.
  • Mental disorders and non-cooperated patient
  • Severe cardiopulmonary diseases (such as severe cardiac dysfunction, severe pulmonary dysfunction and severe arrhythmia)
  • Shock or other severe patient
  • Suspected or confirmed upper digestive tract perforation or perforation acute phase
  • Acute phase of gastric and esophageal chemical burns
  • Patients do not suitable to participate in this study.

Trial locations and contacts

1 site in 1 country.

China1 site
  • Department of Gastroenterology, Peking Union Medical College Hospital

    Beijing, Beijing Municipality, 100730

Other trials in China

About this study

Official title: Prospective Comparison of Two EUS System for Clinical Performance

The primary objective for this study is to compare the safety and validity of two endoscopic ultrasonography (EUS) systems (GF-UE260-ME2 and EG-UR5-S50) used for clinical performance and establish a relative EUS performance assessment system and clinical case data to provide support for the product performance improvement.

Read the detailed description

This research is a random, opening and parallel-controlled clinical test. Tested device model: The EG-UR5 ultrasonic endoscope manufactured by SonoSscope Medical Crop. That used with the HD-500 image processor, HDL-500X light source and S50Exp ultrasonic color doppler instrument .

Compared device model: The GF-UE260-ME2 ultrasonic endoscope manufactured by Olympus and its compatible light source, image processor and ultrasonic instrument.

To scientifically compare the clinical performances of the test model and the compared model, the enrolled objects would be randomly distributed to the tested group and the compared group to conducted the examination with the tested model and the compared model respectively.

EUS will be performed in the standard fashion by one of the experienced endosonographers in the unit. Once the EUS is complete, the procedure will be stopped and the echoendoscope withdrawn from the patient.

  1. Check Points Providing optical images of the human upper gastrointesinal tract (including esophagus, stomach and duodenum) for observing, diagnosis and treatment, and for the ultrasonic check and diagnosis for the upper gastrointestinal tract and surrounding organs.
  2. Test Items 1. Ultrasonic and endoscopic image quality assessment 2. Device controllability assessment 3. System safety assessment 4. System stability assessment

The data are recorded in the CRF, including patient information, clinical assessment for the usage of the endoscope and device operation, thus to perform a comprehensive assessment whether the safety and validity of this product can meet the clinical requirements.

Conditions
Endosonography
Being tested
Assessment of the safety and validity of echoendoscope
Keywords
echoendoscope, endoscopic ultrasonography

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Assessment of the safety and validity of echoendoscopeOther

    1. Ultrasonic and endoscopic image quality assessment
    2. Device controllability assessment
    3. System safety assessment
    4. System stability assessment
Study groups in NCT04161950
Study groupRoleWhat they receive
Tested device model (EG-UR5-S50 )Tested device model: The EG-UR5 echoendoscope manufactured by SonoSscope Medical Crop. That used with the HD-500 image processor, HDL-500X light source and S50 ultrasonic processor.
Receives: Other: Assessment of the safety and validity of echoendoscope
Compared device model (GF-UE260-ME2)Compared device model: The GF-UE260 echoendoscope manufactured by Olympus and its compatible light source, image processor and ME2 ultrasonic processor.
Receives: Other: Assessment of the safety and validity of echoendoscope

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Ultrasonic Image Quality Assessment

    Check Points for Ultrasonic Image Esophagus, stomach, cholecyst, bile duct, pancreas and portal vein system Ultrasonic image quality Assessment standards Assess the ultrasonic image quality of all judgment points according to the ranking degree: excellent, well, bad. Only the assessment of all organs are excellent, the ultrasonic image quality is excellent. If one or more than one assessments are good, the ultrasonic image quality is good. If one or more than one assessments are bad, the ultrasonic image quality is bad.

    The comparison of the image qualities of the tested model and compared model is verified by the non-inferior validation method and the boundary is 10%, and the confidence interval rate differential is 95%.

    Measured: up to 2 months

  2. Endoscopic Image Quality Assessment

    Check Points for endoscopic Image Esophagus, proximal stomach, stomach body,gastric angle,antrum and duodenal bulb.

    Assess the ultrasonic image quality of all judgment points according to the ranking degree: excellent, well, bad. Only the assessment of all observed points are excellent, the endoscopic image quality is excellent. If one or more than one observation points are good, the endoscopic image quality is good. If one or more than one observation points are bad, the endoscopic image quality is bad.

    The comparison of the image qualities of the tested model and compared model is verified by the non-inferior validation method and the boundary is 10%, and the confidence interval rate differential is 95%.

    Measured: up to 2 months

Secondary outcomes (3)
  1. Device controllability Assessment

    The good ration of the device controllability is computed to assess the performance of the ultrasonic endoscope and the target value is more than 98%.

    Measured: up to 2 months

  2. System safety assessment

    Judge the device safety according to the system safety adverse events and the target rate (the estimate value) of the events is 0.

    Measured: up to 2 months

  3. System stability assessment

    Judge the device stability according to the system stability adverse events and the target rate (the estimate value) of the events is less than 5%.

    Measured: up to 2 months

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 Nov 2019
  2. Study start1 Nov 2019
  3. Primary completion31 Dec 2021
  4. Study completion22 Feb 2022

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Case only
Time perspective
Prospective
Enrolment
130 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04161950
Sponsor's protocol ID
2017YFC0109804
Status last verified
May 2025
First submitted
10 Nov 2019
Last updated
31 May 2025
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No: The individual participant data are private.

Publications

  • Roubein LB. Endoscopic ultrasonography and the malignant esophageal stricture: implications and complications. Gastrointest Endosc. 1995 Jun;41(6):613-5. doi: 10.1016/s0016-5107(95)70205-9. No abstract available.

    PubMed 7672562
  • Jenssen C, Alvarez-Sanchez MV, Napoleon B, Faiss S. Diagnostic endoscopic ultrasonography: assessment of safety and prevention of complications. World J Gastroenterol. 2012 Sep 14;18(34):4659-76. doi: 10.3748/wjg.v18.i34.4659.

    PubMed 23002335

Questions about this trial

What is this clinical trial studying?

An observational study of Endosonography, run by Peking Union Medical College Hospital at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 31 May 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Ultrasonic Image Quality Assessment (time frame: up to 2 months). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT04161950, which also lists 2 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 31 May 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in