Validation Study for the Alertgy Non-invasive Continuous Glucose Monitor (ANICGM)

A study testing Alertgy non-invasive continuous glucose monitoring device (ANICGM) in people with Type 2 Diabetes, run by The Cleveland Clinic at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Diabetes mellitus (DM) affects 30 million people in the United States. To achieve glucose control, most patients are prescribed glucose meters by their physicians. Obtaining glucose levels in this manner necessitates cleaning the fingers, attaching a lancet to a device (or simply using a lancet if a device is not… Read more about this study

Who is it for?
Ages 18 to 75 years
Conditions the study is for
Type 2 Diabetes
What is being tested?
Alertgy non-invasive continuous glucose monitoring device (ANICGM)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
15 participants
Where
1 siteUnited States
Study dates
6 Nov 2019to 30 May 2020
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 75 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Type 2 diabetes on diet or on pharmacologic treatment for diabetes
  • Hemoglobin A1c between 7.5 - 10.0%
  • Age 18 - 75 years old

Who cannot join

  • Patients taking prandial insulin
  • Fasting blood glucose by fingerstick of < 70 mg/dL or >250 mg/dL
  • Pregnancy
  • End stage renal disease
  • Decompensated or acute heart failure
  • Medications that may cause false readings with glucose meters: acetaminophen, ascorbic acid, dopamine, maltodextrin, mannitol
  • Conditions that limit the ANICGM such as lesions on the forearms
  • Other conditions that the investigator deems will affect the conduct of the study

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Cleveland Clinic

    Cleveland, Ohio, 44195

Other trials in United States

About this study

Official title: Validation Study for the Alertgy Non-invasive Continuous Glucose Monitor

Diabetes mellitus (DM) affects 30 million people in the United States. To achieve glucose control, most patients are prescribed glucose meters by their physicians. Obtaining glucose levels in this manner necessitates cleaning the fingers, attaching a lancet to a device (or simply using a lancet if a device is not available), pricking the finger with a lancet, placing a drop blood on a strip, and awaiting the readout that results after some chemical reactions. Thereafter, the lancet has to be disposed of in a safe receptacle and the finger has to be blotted to stop the blood from oozing.

The anxiety, pain, and tedious process have led researchers to develop other means of checking glucose levels. There are now continuous glucose monitoring systems (CGMS) that entail inserting a subcutaneous sensor that sends readings through a transmitter. These CGMs may or may not need calibration with a fingerstick glucose reading, and the subcutaneous sensor still has to be changed every 10 -14 days.

The Alertgy non-invasive continuous glucose monitor (ANICGM) is a device that does not entail any subcutaneous insertion of a sensor. It is strapped on to the wrist, and glucose readings are given based on subcutaneous signals. In 2001, a non-invasive device called Glucowatch Biographer was introduced that also involved subcutaneous signals without needing a subcutaneous insertion. However, for various reasons such as long calibration period and reading inaccuracies, the product did not take off. The ANICGM is a promising device that might overcome the limitations of existing and previous methods of non-invasive glucose measurement.

Read the detailed description

The ANICGM uses a weak electromagnetic field generated by its wrist band sensor to look into the body to produce a spectrum that provides a measurement of blood glucose present in the wrist area of the body. The core sensing technology used is dielectric spectroscopy, and has been shown capable of measuring blood glucose, non-invasively, in a laboratory environment. The system uses proprietary and patented dielectric materials and signal processing to enhance performance in both selectivity and sensitivity for blood glucose measurement.

Much like how a MRI uses a strong magnetic field and its interaction with the body to create a picture of what is inside, the ANICGM uses a safe weak low frequency radiofrequency field to take a picture of a person's chemical spectrum, some of it specific only to blood glucose.

The device sends and receives back signals from the wrist area that are used to generate a dielectric spectrum once every 6 seconds. The POC BG using the Accuchek Inform II will be used to provide calibration values for the ANICGM every 5 minutes in mg/dL, during the calibration process. The POC BG levels will be entered into the Alertgy laptop and a proprietary calibration program will be used to analyze the spectral data collected and will generate a calibration algorithm.

Conditions
Type 2 Diabetes
Being tested
Alertgy non-invasive continuous glucose monitoring device (ANICGM)
Keywords
glucose monitor
Medical subject headings
Diabetes Mellitus, Type 2

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Alertgy non-invasive continuous glucose monitoring device (ANICGM)Device

    Alertgy non-invasive continuous glucose monitoring device (ANICGM)

Study groups in NCT04166032
Study groupRoleWhat they receive
ANICGMExperimentalnon-invasive continuous glucose monitoring device
Receives: Device: Alertgy non-invasive continuous glucose monitoring device (ANICGM)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean Absolute Relative Difference (MARD) of Glucose Values Between the ANICGM and FDA-approved Glucose Monitoring Device

    Hypothesis is that ANICGM device will perform similarly as other blood glucose monitoring devices Mean absolute relative difference (MARD) is computed by taking the arithmetic mean of the absolute relative differences between the ANICGM system measures and the reference standard FS BG, which serves as the denominator of the calculation. The MARD is expressed as a percentage, and a lower MARD signifies better concordance between the two measurements.

    Measured: 2 time points taken over 2 separate days respectively, over a span of up to 14 days

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted18 Nov 2019
  2. Study start6 Nov 2019
  3. Primary completion4 Feb 2020
  4. Study completion30 May 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Primary purpose
Screening
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
15 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04166032
Sponsor's protocol ID
18-1385
Status last verified
Oct 2021
First submitted
13 Nov 2019
Last updated
9 Nov 2021
Data monitoring committee
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Alertgy non-invasive continuous glucose monitoring device (ANICGM) in people with Type 2 Diabetes, run by The Cleveland Clinic at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 9 Nov 2021. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 9 Nov 2021). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean Absolute Relative Difference (MARD) of Glucose Values Between the ANICGM and FDA-approved Glucose Monitoring Device (time frame: 2 time points taken over 2 separate days respectively, over a span of up to 14 days). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 9 Nov 2021.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in