WaveLight® EX500 Excimer Laser System for the Correction of Myopia Using InnovEyes™ in Conjunction With InnovEyes™ Sightmap

A study testing WaveLight EX500 excimer laser system with InnovEyes sightmap, LASIK and InnovEyes sightmap in people with Myopia, run by Alcon Research at 9 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to evaluate the safety and effectiveness of the WaveLight EX500 excimer laser system for the correction of myopia with and without astigmatism using InnovEyes in conjunction with InnovEyes sightmap. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Myopia
What is being tested?
WaveLight EX500 excimer laser system with InnovEyes sightmap, LASIK and InnovEyes sightmap
Could I receive a placebo?
No, everyone receives an active treatment
How many people
272 participants
Where
9 sitesUnited States
Study dates
9 Mar 2020to 9 Nov 2022
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Myopia up to and including -11.00 D with or without astigmatism up to -4.50 D, with MRSE no more than -12.00 D
  • Best corrected photopic distance visual acuity of 20/20 or better
  • Uncorrected photopic distance visual acuity of 20/40 or worse
  • Stable refraction (within ± 0.50 D) as determined by MRSE for a minimum of 12 months prior to surgery

Who cannot join

  • History or evidence of active or inactive corneal disease or retinal vascular disease, keratoconus or glaucoma (or suspect)
  • Previous intraocular or corneal surgery
  • Intent to have monovision treatment
  • Other protocol-specified inclusion and exclusion criteria may apply.

Trial locations and contacts

9 sites in 1 country.

9 sites in 1 country

United States9 sites
  • Alcon Investigative Site

    Chevy Chase, Maryland, 20815-7018

  • Alcon Investigative Site

    Dallas, Texas, 75243

  • Price Vision Group

    Indianapolis, Indiana, 46260

  • Alcon Investigative Site

    Kansas City, Missouri, 64154

  • Alcon Investigative Site

    Mt. Pleasant, South Carolina, 29464

  • Alcon Investigative Site

    Omaha, Nebraska, 68135

  • Alcon Investigative Site

    San Diego, California, 92122

  • Alcon Investigative Site

    Sioux Falls, South Dakota, 57108

  • Alcon Investigative Site

    St Louis, Missouri, 63017

Other trials in United States

About this study

Official title: Pivotal Trial of WaveLight® EX500 Excimer Laser System for the Correction of Myopia With and Without Astigmatism Using InnovEyes™ in Conjunction With InnovEyes™ Sightmap

The purpose of this study is to evaluate the safety and effectiveness of the WaveLight EX500 excimer laser system for the correction of myopia with and without astigmatism using InnovEyes in conjunction with InnovEyes sightmap.

Read the detailed description

Qualified subjects will receive treatment in both eyes and be followed for 1 year. Subjects will be asked to attend a total of 9 visits (Screening, Day 0 Surgery, Day 1, Week 1, Month 1, Month 3, Month 6, Month 9, and Year 1). The primary endpoints will be assessed at refractive stability, defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at 3 months after surgery when compared with the 1-month visit) that meets all the protocol-specified stability requirements. Total expected duration of subject participation is approximately 1 year. This study will be conducted in the United States.

Conditions
Myopia
Being tested
WaveLight EX500 excimer laser system with InnovEyes sightmap, LASIK, InnovEyes sightmap
Keywords
LASIK, Ray Tracing, Myopia, Astigmatism
Medical subject headings
Myopia, Astigmatism

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • WaveLight EX500 excimer laser system with InnovEyes sightmapDevice

    FDA approved stationary scanning-spot excimer laser system used during refractive surgery for the treatment of myopia, myopic astigmatism, hyperopia, hyperopic astigmatism, and mixed astigmatism. For this clinical study, the WaveLight EX500 software will be upgraded to include the InnovEyes treatment.

  • LASIKProcedure or surgery

    Procedure that reshapes the front part of the eye (cornea) so that light traveling through it is properly focused on the back part of the eye (retina). The procedure will be performed using the WaveLight EX500 excimer laser system and data obtained from the InnovEyes sightmap.

  • InnovEyes sightmapDevice

    Non-contact ophthalmic diagnostic device that captures images of the eye and collects ocular data preoperatively to generate a patient-specific ablation profile for refractive correction surgery

Study groups in NCT04219891
Study groupRoleWhat they receive
WaveLight EX500 excimer laser systemExperimentalLaser-assisted in situ keratomileusis (LASIK) surgery using the WaveLight EX500 excimer laser with InnovEyes software in conjunction with InnovEyes sightmap
Receives: Device: WaveLight EX500 excimer laser system with InnovEyes sightmap; Procedure: LASIK; Device: InnovEyes sightmap

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Percentage of Eyes With Uncorrected Distance Visual Acuity (UCDVA) of 20/40 or Better (in Eyes With Preoperative Best Corrected Distance Visual Acuity (BCDVA) of 20/20 or Better) at Refractive Stability

    Visual acuity (VA) was assessed using Early Treatment of Diabetic Retinopathy Study (ETDRS) charts at a distance of 4 meters and reported in Snellen, where 20/20 indicates normal distance eyesight. Preoperatively, VA was assessed with correction in place. Postoperatively, VA was assessed with no correction in place. Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at Month 3 after surgery when compared with the Month 1 visit) that meets all the protocol-specified stability requirements. Refractive stability was considered to be achieved at the Month 3 visit. A higher percentage indicates a better outcome. Both eyes contributed data to this analysis. No hypothesis testing was pre-specified for this endpoint. This is a co-primary endpoint.

    Measured: Preoperative Visit (Day -30 to Day -1), Month 1, Month 3

  2. Percentage of Eyes With Manifest Refractive Spherical Equivalent (MRSE) Within +/- 0.50 Diopter (D) at Refractive Stability

    A manifest refraction (manual refraction) was performed using a phoropter. The spherical equivalent was calculated by adding the sum of the sphere power with half of the cylinder power and reported in diopters. Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at Month 3 after surgery when compared with the Month 1 visit) that meets all the protocol-specified stability requirements. Refractive Stability was considered to be achieved at the Month 3 visit. A higher percentage indicates a better outcome. Both eyes contributed data to this analysis. No hypothesis testing was pre-specified for this endpoint. This is a co-primary endpoint.

    Measured: Month 1, Month 3

  3. Percentage of Eyes With MRSE Within +/- 1.00 D at Refractive Stability

    A manifest refraction (manual refraction) was performed using a phoropter. The spherical equivalent was calculated by adding the sum of the sphere power with half of the cylinder power and reported in diopters. Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at Month 3 after surgery when compared with the Month 1 visit) that meets all the protocol-specified stability requirements. Refractive stability was considered to be achieved at the Month 3 visit. A higher percentage indicates a better outcome. Both eyes contributed data to this analysis. No hypothesis testing was pre-specified for this endpoint. This is a co-primary endpoint.

    Measured: Month 1, Month 3

  4. Percentage of Eyes That Achieve Refractive Stability Assessed as Change From Baseline in MRSE of Equal to or Less Than 1.0 D

    A manifest refraction (manual refraction) was performed using a phoropter. The spherical equivalent was calculated by adding the sum of the sphere power with half of the cylinder power and reported in diopters. Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at Month 3 after surgery when compared with the Month 1 visit) that meets all the protocol-specified stability requirements. Refractive stability was considered to be achieved at the Month 3 visit. A higher percentage indicates a better outcome. Both eyes contributed data to this analysis. No hypothesis testing was pre-specified for this endpoint. This is a co-primary endpoint.

    Measured: Preoperative Visit (Day -30 to Day -1), Month 1, Month 3

  5. Percentage of Eyes With BCDVA Worse Than 20/40 (With BCDVA 20/20 or Better Preoperatively) at Refractive Stability

    VA was assessed using ETDRS charts at a distance of 4 meters with correction in place. BCDVA was measured in Snellen, where 20/20 indicates normal distance eyesight. Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at Month 3 after surgery when compared with the Month 1 visit) that meets all the protocol-specified stability requirements. Refractive stability was considered to be achieved at the Month 3 visit. A lower percentage indicates a better outcome. Both eyes contributed data to this analysis. No hypothesis testing was pre-specified for this endpoint. This is a co-primary endpoint.

    Measured: Day -30 to -1 (preoperative), up to Month 3

  6. Percentage of Eyes With BCDVA Loss of 2 Lines (10 Letters) or More From Preoperative Visit at Refractive Stability

    VA was assessed using ETDRS charts at a distance of 4 meters with correction in place and measured in logarithm Minimum Angle of Resolution, where each line on the letter chart represents 0.1 Snellen and 0.0 logMAR equates to 20/20 Snellen, or normal distance eyesight. Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at Month 3 after surgery when compared with the Month 1 visit) that meets all the protocol-specified stability requirements. Refractive stability was considered to be achieved at the Month 3 visit. A lower percentage indicates a better outcome. Both eyes contributed data to this analysis. No hypothesis testing was pre-specified for this endpoint. This is a co-primary endpoint.

    Measured: Day -30 to -1 (preoperative), up to Month 3

  7. Percentage of Eyes With Increase of Manifest Refractive Astigmatism Greater Than 2.00 Diopter (D) of Absolute Cylinder Compared to Preoperative Visit at Refractive Stability

    Manifest refractive astigmatism is the cylindrical portion of the subjective refraction. It is measured with ETDRS charts at a distance of 4 meters. Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at Month 3 after surgery when compared with the Month 1 visit) that meets all the protocol-specified stability requirements. Refractive stability was considered to be achieved at the Month 3 visit. A lower percentage indicates a better outcome. Both eyes contributed data to this analysis. No hypothesis testing was pre-specified for this endpoint. This is a co-primary endpoint.

    Measured: Day -30 to -1 (preoperative), up to Month 3

  8. Percentage of Eyes With Non-flap Related Ocular Serious Adverse Events at Refractive Stability

    An adverse event was defined as any untoward medical occurrence, unintended disease or injury, or untoward clinical signs (including abnormal laboratory findings) in subjects, users or other persons, whether or not related to the investigational medical device (test product). Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at Month 3 after surgery when compared with the Month 1 visit) that meets all the protocol-specified stability requirements. Refractive stability was considered to be achieved at the Month 3 visit. A lower percentage indicates a better outcome. Both eyes contributed data to this analysis. No hypothesis testing was pre-specified for this endpoint. This is a co-primary endpoint.

    Measured: Up to Month 3

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted7 Jan 2020
  2. Study start9 Mar 2020
  3. Primary completion9 Nov 2022
  4. Study completion9 Nov 2022

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
272 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04219891
Sponsor's protocol ID
RFP911-C001
Status last verified
Mar 2025
First submitted
3 Jan 2020
Last updated
6 Apr 2025
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing WaveLight EX500 excimer laser system with InnovEyes sightmap, LASIK and InnovEyes sightmap in people with Myopia, run by Alcon Research at 9 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 6 Apr 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 9 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 6 Apr 2025). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Percentage of Eyes With Uncorrected Distance Visual Acuity (UCDVA) of 20/40 or Better (in Eyes With Preoperative Best Corrected Distance Visual Acuity (BCDVA) of 20/20 or Better) at Refractive Stability (time frame: Preoperative Visit (Day -30 to Day -1), Month 1, Month 3). The study lists 8 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 6 Apr 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in