Truenat™ HCV RNA Assay Evaluation

A study testing Molbio Truenat™ HCV assay (the investigational product) and Abbott RealTime HCV assay in people with Hepatitis C, run by Foundation for Innovative New Diagnostics, Switzerland at 5 sites in 5 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Molbio Diagnostics Ltd. (India) has developed the Truelab™ Real Time quantitative PCR system that is widely used in India for diagnostics of tuberculosis (TB). The system consists of two portable machines and two microfluidic cartridges and can be used in point-of-care settings. Read more about this study

Who is it for?
Ages 16 years and olderHealthy volunteers can take part
Conditions the study is for
Hepatitis C
What is being tested?
Molbio Truenat™ HCV assay (the investigational product) and Abbott RealTime HCV assay
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
1,330 participants
Where
5 sitesin 5 countries
Study dates
12 Jan 2020to 20 Jul 2021
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 16 years and older
  • All sexes
  • Healthy volunteers accepted
  • Child, Adult, Older adult

Who may be able to join

  • Individuals reached the age of an adult as defined in the country that are able to understand the scope of the study and provide informed consent will be invited to take part in the study. In order to meet the study objectives, four different population groups will be considered:
  • HCV SEROPOS: Individuals at risk of having HCV infection based on positive HCV serology test results, without history of HCV treatment
  • Documented positive result of HCV serology test
  • HCV RISK: Individuals with unknown serology status at risk of HCV infection based on past and/or current exposure to risk factors
  • Past and/or current exposure to one of the high risk factors as defined by WHO and CDC guidelines (Appendix 1)
  • HCV TREAT: Individuals diagnosed with chronic HCV infection who initiated or completed the anti-HCV treatment with direct acting antivirals (DAA) presenting at the clinical site for treatment monitoring or test of cure (i.e. sustained virological response)
  • Initiated on DAA treatment (regardless of type of DAA regimen) within 12 months prior to the enrolment to the study
  • BLOOD DONORS: Healthy blood donors free of HBV, HCV and HIV infections, as determined by standard blood bank protocols

Who cannot join

  • Previously enrolled in the study
  • Unwilling or unable to provide required volume of fingerstick blood
  • Unwilling or unable to provide required volume of venous whole blood
  • Positive results in HBV, HCV and HIV tests routinely used in the blood bank to assess blood products safety

Trial locations and contacts

5 sites in 5 countries.

5 sites in 5 countries

Denmark1 site
  • Aarhus University Hospital

    Aarhus, 8200

Other trials in Denmark

Ethiopia1 site
  • St Paul's Hospital Millennium Medical College

    Addis Ababa

Other trials in Ethiopia

Georgia1 site
  • National Centre for Diseases Control

    Tbilisi

Other trials in Georgia

Thailand1 site
  • HIV-NAT Thai Red Cross AIDS Research Centre

    Bangkok, 10330

Other trials in Thailand

Ukraine1 site
  • International Charitable Foundation Alliance for Public Health

    Kyiv, 03150

Other trials in Ukraine

About this study

Official title: Multicentre Clinical Performance Evaluation of the Molbio Diagnostics Truenat™ HCV RNA Assay

Molbio Diagnostics Ltd. (India) has developed the Truelab™ Real Time quantitative PCR system that is widely used in India for diagnostics of tuberculosis (TB). The system consists of two portable machines and two microfluidic cartridges and can be used in point-of-care settings. The manufacturer has recently developed a new assay that detects HCV RNA, the Truenat™ HCV RNA assay. It is a simple two-step assay for RNA extraction and amplification with a total turnaround time of 60 min, using whole blood, plasma and serum as sample types. Most importantly, the assay can be performed from a drop of capillary blood eliminating the need for venous access and blood draw and increasing its usability in the settings where phlebotomy service are not available. To date, validation of the assay was performed using archived plasma specimens and contrived whole blood specimens. FIND aims to conduct a multicentre evaluation to assess the assay's sensitivity, specificity and quantitative accuracy in freshly collected whole blood, plasma and serum specimens from target populations. The evaluation aims to gather performance data in line with the requirements set forth in the Common Technical Specifications 2009/886/EC (CTS) of the CE In Vitro Diagnostics Medical Devices Directive 98/79/CE (CE-IVDD), as well as the World Health Organization (WHO) Technical Specification Series 10 (draft) (TSS-10) for In vitro diagnostic (IVDs) medical devices used for the qualitative and quantitative detection of HCV RNA.

Read the detailed description

Even though hepatitis C virus (HCV) infection can now be cured, globally, fewer than 20% of the 71 million living with HCV are aware of their status; diagnosis remains a major challenge. While serology screening tests that detect the presence of antibodies against HCV, can be performed in decentralized settings, these tests cannot distinguish between resolved (about 25% of seropositive cases) and active infection. Therefore, positive serology test results require further confirmation by laboratory-based methods that detects the viral particle. The two-step testing process results in high drop-off in confirmatory testing among HCV seropositive individuals, especially in low- and middle-income countries (LMIC) where laboratory infrastructure is lacking. Providing access to confirmatory testing in decentralized settings has a high potential to increase the coverage of HCV testing services in LMIC.

Molbio Diagnostics Ltd. (India) has developed the Truelab™ Real Time quantitative PCR system that is widely used in India for diagnostics of tuberculosis (TB). The system consists of two portable machines and two microfluidic cartridges and can be used in point-of-care settings. The manufacturer has recently developed a new assay that detects HCV RNA, the Truenat™ HCV RNA assay. It is a simple two-step assay for RNA extraction and amplification with a total turnaround time of 60 min, using whole blood, plasma and serum as sample types. Most importantly, the assay can be performed from a drop of capillary blood eliminating the need for venous access and blood draw and increasing its usability in the settings where phlebotomy service are not available. To date, validation of the assay was performed using archived plasma specimens and contrived whole blood specimens. FIND aims to conduct a multicentre evaluation to assess the assay's sensitivity, specificity and quantitative accuracy in freshly collected whole blood, plasma and serum specimens from target populations. The evaluation aims to gather performance data in line with the requirements set forth in the Common Technical Specifications 2009/886/EC (CTS) of the CE In Vitro Diagnostics Medical Devices Directive 98/79/CE (CE-IVDD), as well as the World Health Organization (WHO) Technical Specification Series 10 (draft) (TSS-10) for In vitro diagnostic (IVDs) medical devices used for the qualitative and quantitative detection of HCV RNA. The data will be published as a peer-reviewed report and made available to the test manufacturer to support regulatory submissions.

Conditions
Hepatitis C
Being tested
Molbio Truenat™ HCV assay (the investigational product), Abbott RealTime HCV assay
Medical subject headings
Hepatitis C

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Molbio Truenat™ HCV assay (the investigational product)Device

    The Molbio Truenat™ HCV assay (the investigational product) is a quantitative chip-based Real Time Reverse Transcription Polymerase Chain Reaction (qRT-PCR) tests for the detection of HCV genomic RNA from human whole blood, EDTA plasma and serum. The Truenat™ HCV assay runs on the Truelab™ Uno, Truelab™ Duo or Truelab™ Quadro Dx. Results of testing by Truenat™ HCV assay will not be used to make any clinical decisions, and should not be communicated to study participants.

  • Abbott RealTime HCV assayOther

    Plasma specimens will be also tested on Abbott RealTime HCV assay that is approved for HCV diagnostics use by countries' authorities. Only the results of Abbott RealTime HCV assay may be used to make clinical decisions.

Study groups in NCT04236973
Study groupRoleWhat they receive
clinical performance of the Truenat™ HCV assayExperimentalThe study will be conducted in different geographical regions and populations and is designed to meet requirements of the Common Technical Specifications 2009/886/EC (CTS) of the CE-IVDD and WHO TSS-10 (draft) for IVDs medical devices used for the qualitative and quantitative detection of Hepatitis C RNA.
Receives: Device: Molbio Truenat™ HCV assay (the investigational product)
comparison CE-IVD marked reference assay armActive comparatorPlasma specimens from the participants will be tested on Abbott RealTime HCV assay that is approved for HCV diagnostics use by countries' authorities.
Receives: Other: Abbott RealTime HCV assay

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Point estimates of clinical sensitivity and specificity

    Point estimates (with 95% confidence intervals) of clinical sensitivity and clinical specificity of the Truenat™ HCV assay in fingerstick blood and plasma measured against Abbott RealTime HCV assay in plasma. Sensitivity and specificity of the Truenat™ HCV assay will be obtained for each sample type separately, using Abbott RealTime HCV assay as a reference method.

    Measured: From day 1 (enrolment) up to day 30 (completion of the reference testing)

  2. Accuracy of quantification of HCV viral load level in fingerstick blood and plasma by the Truenat™ HCV assay

    To evaluate the quantitative performance of the Truenat™ HCV assay in fingerstick blood and plasma specimens, assay results for each specimen type within the linear range will be compared with the results of Abbott RealTime HCV assay performed from a paired plasma specimen. The data will be visualized with the use of scatter plots, and analyzed using Deming regression. The resulting slope of the regression fir will be reported, with 95% confidence intervals.

    In addition to the linear regression analysis, a Bland-Altman plot will also be generated.

    The percentage of samples with results >0.5 Log IU/mL apart in quantitation between the test and the reference method will be reported.

    Measured: From day 1 (enrolment) up to day 30 (completion of the reference testing)

Secondary outcomes (3)
  1. Level of agreement between result outputs of the Truenat™ HCV assay performed from different specimens.

    Percentage of agreement between result outputs of the Truenat™ HCV assay performed from different specimen types will be calculated, with 95% confidence intervals based on Wilson's score method, as the percentage of the number of samples with identical results over the total number of samples tested. In addition, the Cohen's k coefficient of agreement will also be calculated as a measure of agreement between different sample types.

    Measured: From day 1 (enrolment) up to day 30 (completion of the reference testing)

  2. Accuracy of quantification of HCV viral load levels in serum and venous whole blood specimens by the Truenat™ HCV assay

    Evaluation of the correlation of HCV viral load levels determined by the Truenat™ HCV assay in different specimen types, Truenat™ HCV assay results for each specimen type within the linear range will be compared with the results of the same assay in another specimen type. The data will be visualized with the use of scatter plots, and analyzed using Deming regression.

    Measured: From day 1 (enrolment) up to day 30 (completion of the reference testing)

  3. Operational usability of the Truenat HCV assay

    Assessment of operational characteristics including but not limited to rate of invalid results, total hands on time, description of operator experience with the assays through the appraisal sheet.

    Measured: From the enrolment of the first participant to the completion of all study procedures for the last enrolled participant and up to 12 months

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted22 Jan 2020
  2. Study start12 Jan 2020
  3. Primary completion20 Jul 2021
  4. Study completion20 Jul 2021

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Non-randomised
Intervention model
Sequential
Masking
None (open label)
Enrolment
1,330 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04236973
Sponsor's protocol ID
HC014
Status last verified
Jul 2021
First submitted
15 Jan 2020
Last updated
26 Jul 2021
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study testing Molbio Truenat™ HCV assay (the investigational product) and Abbott RealTime HCV assay in people with Hepatitis C, run by Foundation for Innovative New Diagnostics, Switzerland at 5 sites in 5 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 8 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 26 Jul 2021. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 5 sites in Denmark, Ethiopia, Georgia, Thailand, Ukraine. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Point estimates of clinical sensitivity and specificity (time frame: From day 1 (enrolment) up to day 30 (completion of the reference testing)). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 26 Jul 2021.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in