Preterm InfantNCT04241263

Neonatal Wireless Monitoring System for Intensive Care

A study testing Wireless vital sign monitoring - comparative and Wireless vital sign monitoring - usability in people with Preterm Infant and Infant, Newborn, run by Cambridge University Hospitals NHS Foundation Trust at 1 site in United Kingdom.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

Babies who require intensive care need to have their heart rate, oxygen levels and temperature measured continuously using specially designed monitors. These monitors allow the clinical team to constantly assess the baby and see if they are becoming unwell. Read more about this study

Who is it for?
Ages Up to 6 monthsHealthy volunteers can take part
What is being tested?
Wireless vital sign monitoring - comparative and Wireless vital sign monitoring - usability
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
20 participants
Where
1 siteUnited Kingdom
Study dates
1 Mar 2020to 31 Jul 2021
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Up to 6 months
  • All sexes
  • Healthy volunteers accepted
  • Child

Who may be able to join

  • Babies being cared for on the NICU at Addenbrooke's Hospital (ii) Written, informed parental consent

Who cannot join

  • Known to have an intolerance to silicone or silicone gel adhesives. (ii) Has a pacemaker, or other implanted electronic device

Trial locations and contacts

1 site in 1 country.

Kathryn Beardsall, MRCP contact

01223746791

kb274@cam.ac.uk

United Kingdom1 site, 1 recruiting
  • Cambridge University Hospitals NHS Foundation Trust

    Cambridge, Cambridgeshire, CB2 0QQ

    Kathryn Beardsall, kb274@medschl.cam.ac.uk

    Recruiting

Other trials in United Kingdom

About this study

Official title: Feasibility Study of a Neonatal Wireless Transmission System (NeWTS)

Babies who require intensive care need to have their heart rate, oxygen levels and temperature measured continuously using specially designed monitors. These monitors allow the clinical team to constantly assess the baby and see if they are becoming unwell. However, these monitors each have separate wires to connect them to the baby. Another important part of care in these babies is being able to allow parents to hold their babies, getting them out of their incubators to be cuddled. This is called Kangaroo care and has been shown to impact on long term health for both babies and parents.

We have designed a system to free babies of the wires that tether them to the monitors. This system has been tested in adults and we now want to test them on the babies.

Read more

There are two study parts with two different questions:

  1. How good is the data? This will compare the new wireless system with the existing system. Babies will have both standard wired monitors and the new wireless monitor on at the same time so we can compare the data to see how good is the quality of the data collected on the new system.
  2. What do parents and staff think about the wireless system during Kangaroo care? Babies who no longer need monitoring (but who previously had been monitored with the wired system), will have the wireless system attached and the parent will take their baby from the cot for Kangaroo Care. After the baby has been placed back in the cot, we will conduct interviews with the parent and the nurse to find out what they thought of the new system.

We hope this will help the investigators to understand about how good the data is we collect and how we might be able to improve the system.

Read the detailed description

Summary of study design

This is a single-center study recruiting from the neonatal intensive care unit (NICU) at Addenbrooke's Hospital, Cambridge. There are 2 study parts:

  1. Quantitative assessment of the accuracy of the new system in comparison with standard clinical monitoring systems
  2. Qualitative Assessment of parent and staff views of caring for babies with the new system in comparison with conventional clinical monitoring systems

Quantitative study arm Study population Up to 24 participants will be recruited from the Addenbrooke's Hospital NICU. The participants will be recruited from three groups, defined by the BAPM Categories of Care The population size has been selected to include sufficient babies to collect a reasonable amount of data to ascertain, with confidence, our defined outcomes. Furthermore, the data should come from across the range of variables seen in babies requiring intensive care while limiting exposure of the infants at most risk. The study will be performed in a stepwise fashion, starting with those infants requiring lower dependency care.

Inclusion criteria

  • Babies being cared for on the NICU at Addenbrookes Hospital
  • Written, informed parental consent Exclusion criteria
  • Known to have an intolerance to silicone or silicone gel adhesives.
  • Has a pacemaker or other implanted electronic device.

Planned procedures Participants will have the newly developed wireless sensors for ECG, pulse oximetry, and temperature attached adjacent to the current clinical, wired sensors. Once the application sites have been identified and prepared, the sensors will be placed on the baby by an appropriately trained individual. The wireless receiver will connect to a separate patient monitor to log the comparative data. This study monitor will be covered and silenced such that data will not be visible to the clinical team, and it will not be possible to confuse the study monitor from the 'clinical' monitor. The sensors will be in place for a period of up to 24 hours. After this period, the study sensors will be removed, and the site inspected by a member of the research team and scored using the standard tissue viability scoring system already in use on the unit. The sensors will be disposed of in keeping with the unit and Trust policy.

After the study intervention, a questionnaire will be given to the nurse caring for the baby to determine their views on the wireless system.

Data collection Data will be collected from both the study monitor and the 'clinical' monitor. The data are formed of two types: parameter data and waveform data. Parameter data has a two-second interval. The waveform data has a period of approximately 4ms.

Qualitative study arm

Study population Up to 10 participants will be recruited from the Addenbrooke's Hospital. This is in line with other studies with similar cohorts and balances the need to obtain a range of views and experiences, with the imposition on parents' time in such an environment.

Inclusion criteria

  • BAPM level transitional care babies who do not clinically require vital sign monitoring.
  • The baby has had vital sign monitoring at some point during their stay on the unit.
  • Written, informed parental consent Exclusion criteria
  • Known to have an intolerance to silicone or silicone gel adhesives.
  • Has a pacemaker or other implanted electronic medical device.

Planned procedure Babies who are not indicated for vital sign monitoring will have the wireless monitoring system attached. Once the application sites have been identified and prepared, the sensors will be placed on the baby by an appropriately trained individual. The wireless receiver will connect to the study monitor to log the vital sign data, and all alarms will be silenced. The monitor will be covered and silenced such that it will not impact on clinical care.

The baby will be transferred from the cot/incubator to the parent for kangaroo care as per standard NICU procedure. The parent and baby will remain in contact for as long as the parent is comfortable, and the baby remains stable and in keeping with unit policy and procedures. The baby will then be transferred back to the cot, and monitoring will be removed.

Semi-structured interviews lasting approximately 30 minutes will then be conducted with, (i) the nurse, and (ii) the parent(s). These interviews will focus on previous experiences with vital sign monitoring systems, and their experience of using the new wireless system.

Being tested
Wireless vital sign monitoring - comparative, Wireless vital sign monitoring - usability
Keywords
Patient monitoring, Medical device, Wireless
Medical subject headings
Premature Birth, Birth Weight

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Wireless vital sign monitoring - comparativeDevice

    Participants who currently have vital sign monitoring will, in addition, have the wireless system attached. We will then collect up to 24 hours of vital sign data from both systems, which will allow us to make direct comparisons between them.

  • Wireless vital sign monitoring - usabilityDevice

    Babies will be recruited who do not currently require vital sign monitoring. The wireless system will be attached to these babies by nurses who have been trained in its use. A parent will then be asked to perform Kangaroo Care for as long as they feel comfortable. After the baby is placed back in its cot, and the sensors have been removed, we will conduct (separate) interviews with the nurse and the parent on the neonatal intensive care unit (NICU).

Study groups in NCT04241263
Study groupRoleWhat they receive
Comparative assessment of data qualityOtherQuantitative comparison between the current wired system and the new wireless system
Receives: Device: Wireless vital sign monitoring - comparative
UsabilityOtherQualitative assessment of the new wireless system
Receives: Device: Wireless vital sign monitoring - usability

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean (SD) difference in ECG waveform data

    comparison of data between wired (standard care) and wireless system

    Measured: Up to 24 hours of monitoring data per participant

Secondary outcomes (2)
  1. Impact on care: qualitative interviews with parents and staff

    Qualitative interviews

    Measured: immediately post intervention

  2. Impact on care: qualitative questionnaire with parents and staff

    Questionnaire

    Measured: immediately post intervention

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted27 Jan 2020
  2. Study start1 Mar 2020
  3. Primary completion31 Jul 2021 (estimated)
  4. Study completion31 Jul 2021 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Non-randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
20 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04241263
Sponsor's protocol ID
A095011
Status last verified
Jan 2021
First submitted
12 Jan 2020
Last updated
12 Jan 2021
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Wireless vital sign monitoring - comparative and Wireless vital sign monitoring - usability in people with Preterm Infant and Infant, Newborn, run by Cambridge University Hospitals NHS Foundation Trust at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 12 Jan 2021. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Kathryn Beardsall, MRCP (01223746791, kb274@cam.ac.uk), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean (SD) difference in ECG waveform data (time frame: Up to 24 hours of monitoring data per participant). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 12 Jan 2021.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in