Respiratory ComplicationRRNCT04292639
Vital USA Respiratory Rate Validation In Adults
A study testing The Vital Detect Pulse Oximeter in people with Respiratory Complication, run by Vital USA, Inc. at 1 site in United States.
Trial summary
Vital USA is dedicated to developing and applying innovative electronic medical solutions that improve patient care in multiple clinical settings. Vital USA is a manufacturer of a multi-parameter patient monitor that is intended to be used in a variety of medical settings including hospitals, medical offices and… Read more about this study
- Who is it for?
- Ages 18 to 52 yearsHealthy volunteers can take part
- Conditions the study is for
- Respiratory Complication
- What is being tested?
- The Vital Detect Pulse Oximeter
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 20 participants
- Where
- 1 siteUnited States
- Study dates
- 17 Dec 2019to 20 Dec 2019
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 52 years
- All sexes
- Healthy volunteers accepted
- Adult
Who may be able to join
- Subject must have the ability to understand and provide written informed consent
- Subject is adult over 18 years of age
- Subject must be willing and able to comply with study procedures and duration
- Subject is a non-smoker
- Male or female of any race
Who cannot join
- Subject is considered as being morbidly obese (defined as BMI >39.5)
- Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
- Subjects with known respiratory conditions such as:
- uncontrolled / severe asthma,
- flu,
- pneumonia / bronchitis,
Show all 34Show fewer
- shortness of breath / respiratory distress,
- respiratory or lung surgery,
- emphysema, COPD, lung disease
- Subjects with self-reported heart or cardiovascular conditions such as:
- have had cardiovascular surgery
- Chest pain (angina)
- heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia
- previous heart attack
- blocked artery
- unexplained shortness of breath
- congestive heart failure (CHF)
- history of stroke
- transient ischemic attack
- carotid artery disease
- myocardial ischemia
- myocardial infarction
- cardiomyopathy
- Self-reported health conditions as identified in the Health Assessment Form
- diabetes,
- uncontrolled thyroid disease,
- kidney disease / chronic renal impairment,
- history of seizures (except childhood febrile seizures),
- epilepsy,
- history of unexplained syncope,
- recent history of frequent migraine headaches,
- recent head injury within the last 2 months,
- Cancer / chemotherapy
- Other known health condition, should be considered upon disclosure in health assessment form
Trial locations and contacts
1 site in 1 country.
United States1 site
Clinimark Laboratory Services
Louisville, Colorado, 80027
About this study
Official title: Final Test Report for Vital USA Respiratory Rate Validation In Adults
Vital USA is dedicated to developing and applying innovative electronic medical solutions that improve patient care in multiple clinical settings. Vital USA is a manufacturer of a multi-parameter patient monitor that is intended to be used in a variety of medical settings including hospitals, medical offices and patients' homes.
Read the detailed description
Purpose of Clinical Investigation Respiratory Rate is an important physiological measurement in the healthcare setting. End Tidal Carbon Dioxide (EtCO2) is considered to be the highest level of safety and accuracy in Respiratory Rate monitoring by the American Society of Anesthesiology and the American Association of Respiratory Care.
The purpose of this study was to conduct a Respiratory Rate accuracy validation comparing the Vital Detect to an FDA cleared End Tidal Carbon Dioxide monitor Reference Standard (GE Datex-Ohmeda).
Respiratory Rate is an important physiological measurement in the healthcare setting. End Tidal Carbon Dioxide (EtCO2) is considered to be the highest level of safety and accuracy in Respiratory Rate monitoring by the American Society of Anesthesiology and the American Association of Respiratory Care.
The purpose of this study was to conduct a Respiratory Rate accuracy validation comparing the Vital Detect to an FDA cleared End Tidal Carbon Dioxide monitor Reference Standard (GE Datex-Ohmeda).
- Conditions
- Respiratory Complication
- Being tested
- The Vital Detect Pulse Oximeter
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
The Vital Detect Pulse OximeterDevice
Also called VitalDetect
The Vital Detect is a noninvasive device designed for spot-checking physiological parameters, such as Noninvasive Blood Pressure, Oxygen Saturation, Pulse Rate, Body Temperature, and Respiratory Rate. This device is a finger-cuff technology which is applicable for use at-home or in healthcare facilities in/on individuals 18 years of age and older. It may be used by the individual themselves or an operator. The Vital Detect is not intended for continuous monitoring or for use with high frequency surgical equipment. It is non transit-operable. The Vital USA Vital Detect is composed of a sensor head mounted on a base connected to a phone or computer via Vital Detect App which is free to download.
| Study group | Role | What they receive |
|---|---|---|
| Respiratory Rate | Experimental | The purpose of this study is to conduct a Respiratory Rate accuracy validation comparing the Vital USA Vital Detect to an FDA cleared End Tidal Carbon Dioxide monitor Reference Standard (GE Datex-Ohmeda). This report documents exclusively the results of the Respiratory Rate accuracy performance for the Vital USA Vital Detect. Receives: Device: The Vital Detect Pulse Oximeter |
What is being measured
The primary outcome is the main question the study is designed to answer.
Respiratory Rate Accuracy Validation for the Vital USA Vital Detect Monitor.
The primary objective of this study was to compare the accuracy of device under test for the measurement of respiratory rate. The respiratory rate was measured simultaneously with EtCO2 (Reference) and the Vital USA Vital Detect monitor (Device Under Test).The endpoint of interest was accuracy as measured by the Accuracy root-mean-square (ARMS) difference between the Device Under Test (DUT) and the reference respiratory rate system (Ref) for all stable respiratory periods.
Measured: 20 Seconds
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted3 Mar 2020
- Study start17 Dec 2019
- Primary completion20 Dec 2019
- Study completion20 Dec 2019
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Respiratory ComplicationStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Model description
- The Vital Detect is a noninvasive device designed for spot-checking physiological parameters, such as Noninvasive Blood Pressure, Oxygen Saturation, Pulse Rate, Body Temperature, and Respiratory Rate. This device is a finger-cuff technology which is applicable for use at-home or in healthcare facilities in/on individuals 18 years of age and older. It may be used by the individual themselves or an operator. The Vital Detect is not intended for continuous monitoring or for use with high frequency surgical equipment. It is non transit-operable. The Vital USA Vital Detect is composed of a sensor head mounted on a base connected to a phone or computer via Vital Detect App which is free to download.
- Masking
- None (open label)
- Enrolment
- 20 participants (actual)
Sponsor and investigators
- Lead sponsor
- Vital USA, Inc.
- Sponsor type
- Industry (pharma, biotech, devices)
- Collaborators
- Clinimark, LLC
- Responsible party
- Sponsor
- Investigators
- Arthur Cabrera, MD, Principal investigator, Clinimark, LLC
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT04292639
- Sponsor's protocol ID
- PR 2019-326
- Acronym
- RR
- Status last verified
- May 2020
- First submitted
- 21 Feb 2020
- Last updated
- 28 May 2020
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- Yes
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- No
Study at a glance
From the registry record, last updated by the sponsor on 28 May 2020. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Final Test Report for Vital USA Respiratory Rate Validation In Adults
- ClinicalTrials.gov ID
- NCT04292639
- Acronym / sponsor ID
- RR / PR 2019-326
- Conditions
- Respiratory Complication
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 20 participants (actual)
- Lead sponsor
- Vital USA, Inc.Industry (pharma, biotech, devices)
- Collaborators
- Clinimark, LLC
- Study start
- 17 Dec 2019 (actual)
- Primary completion
- 20 Dec 2019 (actual)
- Study completion
- 20 Dec 2019 (actual)
- First posted
- 3 Mar 2020 (actual)
- Results
- Posted 28 May 2020
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Vital USA is dedicated to developing and applying innovative electronic medical solutions that improve patient care in multiple clinical settings. Vital USA is a manufacturer of a multi-parameter patient monitor that is intended to be used in a variety of medical settings including hospitals, medical offices and patients' homes.
Detailed description
Purpose of Clinical Investigation Respiratory Rate is an important physiological measurement in the healthcare setting. End Tidal Carbon Dioxide (EtCO2) is considered to be the highest level of safety and accuracy in Respiratory Rate monitoring by the American Society of Anesthesiology and the American Association of Respiratory Care.
The purpose of this study was to conduct a Respiratory Rate accuracy validation comparing the Vital Detect to an FDA cleared End Tidal Carbon Dioxide monitor Reference Standard (GE Datex-Ohmeda).
Respiratory Rate is an important physiological measurement in the healthcare setting. End Tidal Carbon Dioxide (EtCO2) is considered to be the highest level of safety and accuracy in Respiratory Rate monitoring by the American Society of Anesthesiology and the American Association of Respiratory Care.
The purpose of this study was to conduct a Respiratory Rate accuracy validation comparing the Vital Detect to an FDA cleared End Tidal Carbon Dioxide monitor Reference Standard (GE Datex-Ohmeda).
Study design
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
- Model description
- The Vital Detect is a noninvasive device designed for spot-checking physiological parameters, such as Noninvasive Blood Pressure, Oxygen Saturation, Pulse Rate, Body Temperature, and Respiratory Rate. This device is a finger-cuff technology which is applicable for use at-home or in healthcare facilities in/on individuals 18 years of age and older. It may be used by the individual themselves or an operator. The Vital Detect is not intended for continuous monitoring or for use with high frequency surgical equipment. It is non transit-operable. The Vital USA Vital Detect is composed of a sensor head mounted on a base connected to a phone or computer via Vital Detect App which is free to download.
| Arm | Type | Receives |
|---|---|---|
| Respiratory Rate | Experimental | Device: The Vital Detect Pulse OximeterThe purpose of this study is to conduct a Respiratory Rate accuracy validation comparing the Vital USA Vital Detect to an FDA cleared End Tidal Carbon Dioxide monitor Reference Standard (GE Datex-Ohmeda). This report documents exclusively the results of the Respiratory Rate accuracy performance for the Vital USA Vital Detect. |
Who may qualify
- Ages
- 18 to 52 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult
Key inclusion criteria (5 listed)
- Subject must have the ability to understand and provide written informed consent
- Subject is adult over 18 years of age
- Subject must be willing and able to comply with study procedures and duration
- Subject is a non-smoker
- Male or female of any race
Key exclusion criteria (34 listed)
- Subject is considered as being morbidly obese (defined as BMI >39.5)
- Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
- Subjects with known respiratory conditions such as:
- Subjects with self-reported heart or cardiovascular conditions such as:
- Self-reported health conditions as identified in the Health Assessment Form
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| The Vital Detect Pulse OximeterAlso: VitalDetectThe Vital Detect is a noninvasive device designed for spot-checking physiological parameters, such as Noninvasive Blood Pressure, Oxygen Saturation, Pulse Rate, Body Temperature, and Respiratory Rate. This device is a finger-cuff technology which is applicable for use at-home or in healthcare facilities in/on individuals 18 years of age and older. It may be used by the individual themselves or an operator. The Vital Detect is not intended for continuous monitoring or for use with high frequency surgical equipment. It is non transit-operable. The Vital USA Vital Detect is composed of a sensor head mounted on a base connected to a phone or computer via Vital Detect App which is free to download. | Device | Respiratory Rate |
Primary outcome
Respiratory Rate Accuracy Validation for the Vital USA Vital Detect Monitor.
Time frame: 20 Seconds
The primary objective of this study was to compare the accuracy of device under test for the measurement of respiratory rate. The respiratory rate was measured simultaneously with EtCO2 (Reference) and the Vital USA Vital Detect monitor (Device Under Test).The endpoint of interest was accuracy as measured by the Accuracy root-mean-square (ARMS) difference between the Device Under Test (DUT) and the reference respiratory rate system (Ref) for all stable respiratory periods.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 28 May 2020.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Individual participant dataSharing planned: No
Related trials for Respiratory Complication
RecruitingPhase 1/2
Study of INBRX-106 and INBRX-106 in Combination With Pembrolizumab (Keytruda®) in Subjects With Locally Advanced or Metastatic Solid Tumors (Hexavalent OX40 Agonist)
Solid Tumor, Non-Small Cell Lung Cancer
Active, not recruiting
CT for Personalized Mechanical Ventilation
Ventilator-Induced Lung Injury, Sepsis Syndrome
Not yet recruiting
nCPAP-NAVA-Backup Versus Standard nCPAP for Severe Apnea in Preterm Infants ≤28 Weeks
Apnea of Prematurity
Recruiting
The Vanguard Study: Testing a New Way to Screen for Cancer
Bladder Carcinoma, Breast Carcinoma
Recruiting
Optimization of Saturation Targets And Resuscitation Trial (OptiSTART)
Premature Infants, Bronchopulmonary Dysplasia
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Arthur Cabrera, MD principal investigator
Clinimark, LLC
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing The Vital Detect Pulse Oximeter in people with Respiratory Complication, run by Vital USA, Inc. at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 5 inclusion and 34 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 28 May 2020. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 28 May 2020). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Respiratory Rate Accuracy Validation for the Vital USA Vital Detect Monitor. (time frame: 20 Seconds). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 28 May 2020.