Respiratory InsufficiencyNCT04355754

Ventil Device Usefulness in Mechanically Ventilated ICU Patients

A study testing Ventil - a gas flow divider in people with Respiratory Insufficiency, run by Medical University of Gdansk at 1 site in Poland.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

During Covid-19 pandemic many patients require mechanical ventilation due to disastrous impact of SARS-CoV-2 on lungs. In several countries there is a shortage of ICU beds and ventilators. Critically ill patients are treated outside ICUs. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Respiratory Insufficiency
What is being tested?
Ventil - a gas flow divider
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
40 participants
Where
1 sitePoland
Study dates
15 Apr 2020to 31 Dec 2020
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • ICU patients who require mechanical ventilation

Who cannot join

  • Patients who require complex modes of ventilation will not be recruited
  • The sudy in an individual patient be discontinued if case of:
  • Respiratory Criteria:
  • episodes of desaturation <90% (in pts without COPD) without reversibel reason
  • need for FiO2 increase by 10%
  • need for switch to other than CMV mode of ventillation
Show all 14Show fewer
  • need for neuromucular blockade or for deepen sedation because of assynchrony between patient and venilator
  • cummulation of CO2>45 mm Hg (in pts without COPD) not responding to the increase of minute ventilation for 30 minutes
  • Pplat >30 cmH2O
  • Circulatory criteria:
  • New haemodynamic disturbances that cannot be explaned by other reasons
  • ↑ or ↓of BP by 20%
  • ↑ or ↓of HR by 20%
  • clinically important heart rhythm disturbances

Trial locations and contacts

1 site in 1 country.

Radoslaw Owczuk, prof contact

+48583493270

r.owczuk@gumed.edu.pl

Magdalena A Wujtewicz, dr contact

+48583493270

magwuj@gumed.edu.pl

Poland1 site, 1 recruiting
  • Medical University of Gdansk

    Gdansk, 80-214

    Magdalena A Wujtewicz, +48583493357, magwuj@gumed.edu.pl

    Radoslaw Owczuk, 605408140, r.owczuk@gumed.edu.pl

    Recruiting

Other trials in Poland

About this study

Official title: Assesment of Usefulness of Ventil Device for Mechanical Ventilation in ICU Patients

During Covid-19 pandemic many patients require mechanical ventilation due to disastrous impact of SARS-CoV-2 on lungs. In several countries there is a shortage of ICU beds and ventilators. Critically ill patients are treated outside ICUs. Doctors are facing ethical dilemmas who they should treat with ventilation, who should receive ventilator and who should but will not. In ICUs or step down units or in nursery homes there are also patients beyond hope treated - very often they are dependent on mechanical ventilation. Some attempts to invent a device that could replace complex machines in patients with anticipated poor outcome have been made. Ventil was used in clinical scenarios for separate lung ventilation with good effect. As a flow divider it has a potential to ventilate 2 patients at the same time. In the study Ventil will ventilate one patient and instead of the second there will be an artificial lung. Tidal volumes, minute ventilation, PEEP set and final will be checked. Ppeak, Pmean, Pplat, Cdyn, airway resistance, EtCO2, Sat O2, HR, SAP, DAP will be monitored every 2 hrs, as well as blood-gas analysis (every 8 hrs).

Read the detailed description

During Covid-19 pandemic many patients require mechanical ventilation due to disastrous impact of SARS-CoV-2 on lungs. In several countries there is a shortage of ICU beds and ventilators. Critically ill patients are treated outside ICUs. Doctors are facing ethical dilemmas who they should treat with ventilation, who should receive ventilator and who should but will not.

Before pandemic in ICUs there had always been patients who required mechanical ventilation because of extrapulmonary reasons as well as palliative cases or those in vegetative condition. Most of these patients require just a simple ventilator, not a sophisticated mode of ventilation. Some attempts to invent a device that could replace complex machines in patients with anticipated poor outcome have been made. Ventil was used in clinical scenarios for separate lung ventilation with good effect. VENTIL device, a flow divider, theoretically allows for independent, fully automated synchronous ventilation of 2 patients with use of only one respirator. In the shortage of respirators (ex. terrorist attack, natural disasters) device allows also to ventilate in classical system two patients using single respirator.

Ventil - independent lung ventilation system was constructed by engineers from Nalecz Institute of Biocybernetics and Biomedical Engineering of Polish Academy of Science. Ventil was tested in the clinical scenario - it's safety had been confirmed. Several years ago the working prototypes of the device, after approval of Ethical Committee, was tested in about 150 patients, who were ventilated with independent synchronous lung ventilation and had been found useful.

The idea of the study is to check the usefulness of the device (modern version) in ICU patients who can be ventilated with volume -controlled mode of ventilation as an attempt to use single device for ventilation of 2 patients. At this moment, according to several scientific societies, sharing mechanical ventilators should not be attempted because it cannot be done safely with current equipment. However, it is possible that using a designed flow divider will allow safe ventilation in patients without needs for complicated modes of ventilation. Then maybe it will be possible to release some ventilators and to use them in patients in severe condition, mainly in the era of extreme ventilator shortage. In the study Ventil will ventilate one patient and instead of the second, there will be an artificial lung. Tidal volumes, minute ventilation, PEEP set and final will be checked. Ppeak, Pmean, Pplat, Cdyn, airway resistance, EtCO2, Sat O2, HR, SAP, DAP will be monitored (every 2 hrs), as well as blood-gas analysis (every 8 hrs)

Being tested
Ventil - a gas flow divider
Keywords
critically ill patient, intensive care unit, mechanical ventilation, pandemic
Medical subject headings
Respiratory Insufficiency

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Ventil - a gas flow dividerDevice

    Checking ventilation parameters set and obtained and clinical parameters while ventilating the patient and an artificial lung simultaneously, using inspiratory flow divider.

Study groups in NCT04355754
Study groupRoleWhat they receive
mechanically ventilated patientsExperimentalAdult ICU patients who are mechanically ventilated and who do not require complex modes of ventilation. A designated flow divider (Ventil) will be used to divide inspiratory gas flow from ventilator in two separate streams - one to the patient and the second to the artificial lung
Receives: Device: Ventil - a gas flow divider

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Number of cases in which it was necessary to stop using Ventil and to step- back to ventilation without this flow divider

    Ventil will be removed from the patient-ventilator circiuit in case of episodes of desaturation <90% (in pts without COPD) without reversibel reason; need for FiO2 increase by 10%; need for switch to other than CMV mode of ventillation need for neuromucular blockade or for deepen sedation because of assynchrony between patient and venilator cummulation of CO2>45 mm Hg (in pts without COPD) not responding to the increase of minute ventilation for 30 minutes; if Pplat >30 cmH2O; in case of new haemodynamic disturbances that cannot be explaned by other reasons; in case of increase or decrease of BP by 20%; increase or decrease of HR by 20%; in case of occurence of clinically important heart rhythm disturbances

    Measured: 48 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted21 Apr 2020
  2. Study start15 Apr 2020
  3. Primary completionMay 2020 (estimated)
  4. Study completion31 Dec 2020 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Model description
The flow divider is connected to the ventilator. Half of the fresh gas flows to the patient, hallf to the artificial lung ( 1:1 ratio). For adequate ventilation tidal volume set on the ventilator is then doubled. Chosen respiratory and circulatory parameters are beeing noted.
Masking
None (open label)
Enrolment
40 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04355754
Sponsor's protocol ID
Ventil-COVID-19
Status last verified
Apr 2020
First submitted
16 Apr 2020
Last updated
27 Apr 2020
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

A study testing Ventil - a gas flow divider in people with Respiratory Insufficiency, run by Medical University of Gdansk at 1 site in Poland.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 27 Apr 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Poland. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Radoslaw Owczuk, prof (+48583493270, r.owczuk@gumed.edu.pl), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Number of cases in which it was necessary to stop using Ventil and to step- back to ventilation without this flow divider (time frame: 48 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 27 Apr 2020.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in
Clinical Trial NCT04355754 | AzTrials