Arterio-venous FistulaSEMPER FINCT04381754

Use of Implanting the Biotronik Passeo-18 Lux Drug Coated Balloon to Treat Failing Haemodialysis Arteriovenous Fistulas and Grafts.

An observational study of Arterio-venous Fistula and Dialysis Access Malfunction, run by Singapore General Hospital at 1 site in Singapore.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

The most common problem with haemodialysis arteriovenous fistulas (AVF) and arterio-venous grafts (AVG) is stenosis, which can lead to inadequate dialysis, and eventual access thrombosis. Conventional plain old balloon angioplasty is associate with high recurrence rates of stenosis and repeated interventions. Read more about this study

Who is it for?
Ages 21 to 90 years
What is being tested?
Passeo-18 Lux
Type of treatment
Medical device
How many people
100 participants
Where
1 siteSingapore
Study dates
Jun 2020to Jun 2022
How long will I take part?
12 Months
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 21 to 90 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: Patients who have a native upper limb AVF or AVG, that is currently in use for haemodialysis, which has a significant inflow or outflow stenosis. Significant stenosis defined as diameter greater than 50% compared to adjacent normal segment and/or other evidence of AVF malfunction. Malfunction defined as the AVF performing inadequate dialysis and clinical signs of a failing dialysis access. Inadequate fistula volume flow - Qa as measured by ultrasound < 500mL/min as probably inadequate. Dialysis numbers: - Qb - a significant stenosis is suggested when the pump speed is < 200ml/min, venous pressure is > 140 mmHg and/or arterial pressure > - 100mmHg. Maximum of 2 discrete stenoses (separated by > 3cm) are allowed to be included.

Who may be able to join

  • Patient aged ≥21 years and ≤90 years
  • Native AVF was created more than 2 months prior to the index procedure and had undergone 10 or more hemodialysis sessions utilizing 2 needles
  • Target lesion location had to be located between the anastomosis to the axillary-subclavian vein junction, as defined by insertion of the cephalic vein.
  • On initial fistulogram, target lesion stenosis had to be >50% on angiographic assessment and in keeping with the clinical indicator for intervention
  • Stenosis had to be < 10cm in length to allow for potential treatment with one PCB (length 12 cm) only
  • Stenosis had to be initially treated successfully with a high-pressure plain balloon prior to PCB treatment as defined by:
Show all 12Show fewer
  • No clinically significant dissection
  • No extravasation requiring treatment/stenting
  • Residual stenosis ≤20% by angiographic measurement
  • Ability to completely efface the lesion waist using the pre-dilation balloon
  • No more than one additional ("nontarget") lesions in the access circuit that had to be also successfully treated (≤30% residual stenosis) before drug elution. Separate lesion was defined by at least 3 cm in distance from the target lesion.
  • Reference vessel diameter 4mm - 8mm

Who cannot join

  • Women who were pregnant, lactating, or planning on becoming pregnant during the study
  • Subject had more than two lesions in the access circuit
  • Subject had a secondary non-target lesion that could not be successfully treated
  • Sepsis or active infection
  • Asymptomatic target lesions
  • A thrombosed access or an access with thrombosis treated ≤30 days prior to the index procedure
Show all 17Show fewer
  • Surgical revision of the access site performed, planned or expected ≤ 3 months before or after the index procedure
  • Patients who were taking immunosuppressive therapy or are routinely taking ≥ 15 mg of prednisone per day;
  • Currently participating in an another investigational drug, biologic, or device study involving sirolimus or paclitaxel
  • Contraindication to aspirin or clopidogrel usage
  • Mental condition rendering the subject unable to understand the nature, scope and possible consequences of the study, or language barrier such that the subject is unable to give informed consent
  • Uncooperative attitude or potential for non-compliance with the requirements of the protocol making study participation impractical
  • Where final angioplasty treatment requires a stent or drug eluting balloon > 8mm in diameter
  • Metastatic cancer or terminal medical condition
  • Blood coagulation disorders
  • Limited life expectancy (< 12 months)
  • Allergy or other known contraindication to iodinated media contrast, heparin or paclitaxel

Trial locations and contacts

1 site in 1 country.

Singapore1 site
  • Singapore General Hospital

    Singapore, 169856

Other trials in Singapore

About this study

Official title: Use of Implanting the Biotronik Passeo-18 Lux Drug Coated Balloon to Treat Failing Hemodialysis Arteriovenous Fistulas and Grafts

The most common problem with haemodialysis arteriovenous fistulas (AVF) and arterio-venous grafts (AVG) is stenosis, which can lead to inadequate dialysis, and eventual access thrombosis. Conventional plain old balloon angioplasty is associate with high recurrence rates of stenosis and repeated interventions. The advent of successful drug-eluting technology in the treatment of the coronary vascular bed and subsequent positive accumulating evidence in the peripheral arterial circulation has prompted the use of drug coated balloons (DCB) in the access fistula circuit for venous stenosis and in-stent restenosis. Recent studies suggest that DCBs may significantly reduce re-intervention rates on native and recurrent lesions. The restenosis process is in part or in whole the result of neo-intimal hyperplasia (NIH) and NIH is considered the main culprit in access circuit target lesion stenosis. NIH is the blood vessel's healing response to the barotrauma from the angioplasty process. A critical component of NIH is the cellular proliferative stage with mononuclear leucocytes identified as the primary inflammatory cell type involved. The rationale for drug elution is to block the NIH response with an anti-metabolite such as paclitaxel. It is important to emphasize that the role of drug elution in the treatment of vascular stenosis is not to obtain a good haemodynamic and luminal result but to preserve a good result obtained during POBA from later restenosis due to NIH and minimise reinterventions and readmissions to hospital for what is a frail population of patients.

A meta-analysis performed by Khawaja et al. seemed to suggest that DCBs conferred some benefit in terms of improving target lesion primary patency (TLPP) in AVFs. An updated meta-analysis performed by our own institution recently showed that DCB appears to be a better and safe alternative to conventional balloon angioplasty (CBA) in treating patients with HD stenosis based on 6- and 12-months primary patency and increased intervention free period.

The Passeo-18 Lux (Biotronik Asia Pacific Pte Ltd (Singapore)) drug-coated balloon (DCB) is packaged with a low dose of paclitaxel. Recent studies have shown that low dose coating of paclitaxel with this DCB is useful for preventing restenosis, decrease lumen loss and target lesion revascularization in the peripheral vasculature6 but has not been tested in the dialysis access circuit.

Being tested
Passeo-18 Lux
Medical subject headings
Arteriovenous Fistula

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Passeo-18 LuxDevice

    Also called Biotronik

    Patients with significant inflow or outflow stenosis between the anastomosis to the axillary-subclavian vein junction, as defined by the insertion of the cephalic vein, who had undergone fistuloplasty with Passeo-18 Lux

Study groups in NCT04381754
Study groupRoleWhat they receive
AVG/AVF Treated with Passeo-18 LuxPatients with failing dialysis access, treated lesions located between the anastomosis to the axillary-subclavian vein junction.
Receives: Device: Passeo-18 Lux

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. 6-month Target Lesion Primary Patency

    Patency with no re-intervention to the area 5mm proximal to, within, and 5mm distal to, the index treatment segment. TLPP ends when any of the following occur: 1) clinically driven re-intervention to the treatment segment 2) thrombotic occlusions that includes the treatment segment 3) surgical intervention that excludes the treatment segment from the access circuit 4) abandonment of the AVF/AVG due to an inability to treat the treatment segment

    Measured: 6-month post-procedure

Secondary outcomes (5)
  1. Primary Patency

    A duration of time measuring intra-access patency that starts from the date of angioplasty with Passeo-18 Lux DCB to the date of one of the following events: thrombosis, or any intervention to facilitate, maintain or re-establish patency (e.g. angioplasty)

    Measured: 12 months post-op

  2. Primary assisted patency

    Interval date of angioplasty with Passeo-18 Lux DCB until thrombosis

    Measured: 12 months post-op

  3. Secondary Patency

    A duration of time measuring intra-access patency that starts from the date of angioplasty with Passeo-18 Lux DCB to the date of vascular access abandonment

    Measured: 12 months post-op

  4. Number of reinterventions

    Measured: 12 months post-op

  5. Adverse Events

    Intraoperative/perioperative complications, infections, revision surgeries required

    Measured: 12 months post-op

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted11 May 2020
  2. Study startJun 2020 (estimated)
  3. Primary completionJan 2022 (estimated)
  4. Study completionJun 2022 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
100 participants (estimated)
Target duration
12 Months
IDs, dates and publications
ClinicalTrials.gov ID
NCT04381754
Sponsor's protocol ID
1.5
Acronym
SEMPER FI
Status last verified
May 2020
First submitted
6 May 2020
Last updated
11 May 2020
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Troisi N, Frosini P, Somma C, Romano E, Guidotti A, Dattolo PC, Ferro G, Chisci E, Michelagnoli S. Drug-coated balloons reduce the risk of recurrent restenosis in arteriovenous fistulas and prosthetic grafts for hemodialysis. Int Angiol. 2018 Feb;37(1):59-63. doi: 10.23736/S0392-9590.17.03886-X. Epub 2017 Nov 10.

    PubMed 29125264
  • Liao MT, Chen MK, Hsieh MY, Yeh NL, Chien KL, Lin CC, Wu CC, Chie WC. Drug-coated balloon versus conventional balloon angioplasty of hemodialysis arteriovenous fistula or graft: A systematic review and meta-analysis of randomized controlled trials. PLoS One. 2020 Apr 14;15(4):e0231463. doi: 10.1371/journal.pone.0231463. eCollection 2020.

    PubMed 32287283
  • Kennedy SA, Mafeld S, Baerlocher MO, Jaberi A, Rajan DK. Drug-Coated Balloon Angioplasty in Hemodialysis Circuits: A Systematic Review and Meta-Analysis. J Vasc Interv Radiol. 2019 Apr;30(4):483-494.e1. doi: 10.1016/j.jvir.2019.01.012. Epub 2019 Mar 8.

    PubMed 30857987
  • Khawaja AZ, Cassidy DB, Al Shakarchi J, McGrogan DG, Inston NG, Jones RG. Systematic review of drug eluting balloon angioplasty for arteriovenous haemodialysis access stenosis. J Vasc Access. 2016 Mar-Apr;17(2):103-10. doi: 10.5301/jva.5000508. Epub 2016 Feb 5.

    PubMed 26847736
  • Yan Wee IJ, Yap HY, Hsien Ts'ung LT, Lee Qingwei S, Tan CS, Tang TY, Chong TT. A systematic review and meta-analysis of drug-coated balloon versus conventional balloon angioplasty for dialysis access stenosis. J Vasc Surg. 2019 Sep;70(3):970-979.e3. doi: 10.1016/j.jvs.2019.01.082.

    PubMed 31445651
  • Brodmann M, Zeller T, Christensen J, Binkert C, Spak L, Schroder H, Righini P, Nano G, Tepe G. Real-world experience with a Paclitaxel-Coated Balloon for the treatment of atherosclerotic infrainguinal arteries: 12-month interim results of the BIOLUX P-III registry first year of enrolment. J Vasc Bras. 2017 Oct-Dec;16(4):276-284. doi: 10.1590/1677-5449.007317.

    PubMed 29930661

Questions about this trial

What is this clinical trial studying?

An observational study of Arterio-venous Fistula and Dialysis Access Malfunction, run by Singapore General Hospital at 1 site in Singapore.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 12 inclusion and 17 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 11 May 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Singapore. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

6-month Target Lesion Primary Patency (time frame: 6-month post-procedure). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT04381754, which also lists 6 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 11 May 2020.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in