Oxygen SaturationDSHWDNCT04429906

Diagnostic Study of Huami Smart Wearable Device

An observational study of Oxygen Saturation, run by Peking University First Hospital at 1 site in China.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

This is a cross-sectional diagnostic study. In this study, wrist oxygen saturation (SwO2) will be measured by Huami smart wearable device, SpO2 of pulse oxygen saturation monitor with medical device registration certificate and SaO2 of arterial blood gas analysis will be used as reference values. Read more about this study

Who is it for?
Ages 18 to 85 years
Conditions the study is for
Oxygen Saturation
What is being tested?
Huami smart wearable device
Type of treatment
Medical device
How many people
30 participants
Where
1 siteChina
Study dates
29 May 2020to 30 Jun 2020
How long will I take part?
1 Day
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 85 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: Inpatients in the Department of Respiratory and Critical Care at Peking University First Hospital

Who may be able to join

  • Inpatients in the Department of Respiratory and Critical Care at Peking University First Hospital
  • Oxygen intake or no oxygen intake when enrolled: 80%≤SaO2/SpO2≤100%
  • Age between 18 and 85 years, either gender
  • COHb<3%, MetHb<2%;
  • Sign an informed consent or sign an informed consent by a legally authorized representative;
  • Be willing and able to follow the research program.

Who cannot join

  • current smokers or those exposed to high levels of carbon monoxide.
  • critically ill patients who are in a state of shock, resulting in peripheral circulation disorders.
  • Patients with hypothermia below 35°C.
  • Patients with bilateral wrist and hand oedema, soft tissue damage to the wrist, skin incomplete and unable to wear a watch/ bracelet.
  • Vascular blood supply abnormalities of both upper limbs (occlusion, thrombosis, post-trauma, etc.), which affects the test results
  • Patients with bilateral upper extremity pigmentation affecting the test results;
  • Patients with limb mobility impairment due to severe cerebrovascular disease sequelae.
  • Patients with severe mental system disease that makes it impossible to cooperate with the researchers.

Trial locations and contacts

1 site in 1 country.

China1 site, 1 recruiting
  • Peking University First Hospital

    Beijing, Beijing Municipality, 100034

    Guangfa Wang, MD., 13810644029, wangguangfa@hotmail.com

    Jiping Liao, PhD., 13521714181, colorfulwing01@163.com

    Recruiting

Other trials in China

About this study

Official title: Evaluation of Accuracy of Blood Oxygen Saturation Detection Function of Huami Smart Wearable Device

This is a cross-sectional diagnostic study. In this study, wrist oxygen saturation (SwO2) will be measured by Huami smart wearable device, SpO2 of pulse oxygen saturation monitor with medical device registration certificate and SaO2 of arterial blood gas analysis will be used as reference values. A total of 180 group data will be measured in 30~50 patients with SaO2/ SpO2 between 80%∽100%. The investigators aim to evaluate the accuracy of blood oxygen saturation detection function of huami smart wearable device.

Read the detailed description

This is a cross-sectional diagnostic study. 30~50 inpatient aged 18~85y with SaO2/ SpO2 between 80%∽100% will be recruited from the Department of Respiratory and Critical Care of Peking University First Hospital. 180 group data ( 37 of SaO2/ SpO2 between 80%∽90%, 143 of SaO2/ SpO2 between 90%∽100%) will be measured, with an average of 4-6 groups per patient. Some clinical information such as demographic data, laboratory tests, comorbidity, smoking habit and will be collected. Patients have their pulse oximetry measured multiple times using a pulse oximeter/tabletop electrocardiograph with a medical device registration certificate, and arterial blood gas analysis will be performed if required. While monitoring the pulse oximetry, a Huami smart wearable device is worn on the ipsilateral wrist to determine blood oxygen saturation. Primary outcome is the wearable device oximetry accuracy Arms (root mean square value of the SwO2-RefSpO2 difference) using a medical device registration certificate pulse oximetry monitor/tabletop ECG monitor as a reference..

The study protocol has been approved by the Peking University First Hospital Institutional Review Board (IRB) (2020-114). Any protocol modifications will be submitted for the IRB review and approval.

Being tested
Huami smart wearable device
Keywords
Oxygen saturation Smart wearable device Accuracy

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Huami smart wearable deviceDevice

    Wear the pulse-oxygen monitoring finger set of the medical monitor while wearing the Huami Smart Wearable device on the ipsilateral wrist according to the instructions for use. The medical monitor displays a steady pulse oximetry level for at least 30 seconds, starts the first measurement, slides the main dial page of the wearable device to the "Oxygen Saturation" measurement interface, clicks the measurement button to start the single measurement of blood oxygen saturation. Record the pulse oximetry and pulse rate values measured at the same time by the medical monitor and the Huami Smart Wearable Pulse Oximetry Device, respectively. After an interval of 30 seconds, repeat the above steps to start the second measurement and record the measured value. After an interval of 30 seconds, repeat the above steps to start the third measurement and record the measured value. The average of three measurements was taken as data for the test group.

Study groups in NCT04429906
Study groupRoleWhat they receive
test groupWear the pulse-oxygen monitoring finger set of the medical monitor while wearing the Huami Smart Wearable device on the ipsilateral wrist according to the instructions for use. The medical monitor displays a steady pulse oximetry level for at least 30 seconds, starts the first measurement, slides the main dial page of the wearable device to the "Oxygen Saturation" measurement interface, clicks the measurement button to start the single measurement of blood oxygen saturation. Record the pulse oximetry and pulse rate values measured at the same time by the medical monitor and the Huami Smart Wearable Pulse Oximetry Device, respectively. After an interval of 30 seconds, repeat the above steps to start the second measurement and record the measured value. After an interval of 30 seconds, repeat the above steps to start the third measurement and record the measured value. The average of three measurements was taken as data for the test group.
Receives: Device: Huami smart wearable device
control group APulse oximetry monitor/desktop ECG monitor with a medical device registration certificate was selected as reference device A. The mean value of qualified pulse oximetry measured by pulse oximetry monitor/desktop ECG monitor within 2 minutes of each successive measurement of the Huami Smart Wearable Device was used as control group A measurement.
control group BA carbon monoxide blood gas analyzer (CO-oximeter) was selected as reference device B. Arterial blood was sampled and arterial oxygen saturation (SaO2) was obtained from the blood gas analyzer as a control group B measurement.

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Wearable device oximetry accuracy Arms

    Record the pulse oximetry and pulse rate values measured at the same time by the medical monitor and the Huami Smart Wearable Pulse Oximetry Device, respectively. After an interval of 30 seconds, repeat the above steps to start the second measurement and record the measured value. After an interval of 30 seconds, repeat the above steps to start the third measurement and record the measured value. The average of three measurements was taken as data for the test group.

    Pulse oximetry monitor/desktop ECG monitor with a medical device registration certificate was selected as reference device A. The mean value of qualified pulse oximetry measured by pulse oximetry monitor/desktop ECG monitor was used as control group A measurement.

    Wearable device oximetry accuracy Arms (root mean square value of the SwO2-RefSpO2 difference) using a medical device registration certificate pulse oximetry monitor/tabletop ECG monitor as a reference.

    Measured: 30mins

Secondary outcomes (5)
  1. Wearable device blood oxygen saturation accuracy Arms

    Record the pulse oximetry and pulse rate values measured by the Huami Smart Wearable Pulse Oximetry Device. After an interval of 30 seconds, repeat the above steps to start the second measurement and record the measured value. After an interval of 30 seconds, repeat the above steps to start the third measurement and record the measured value. The average of three measurements was taken as data for the test group.A carbon monoxide blood gas analyzer (CO-oximeter) was selected as reference device B. Arterial blood was sampled and arterial oxygen saturation (SaO2) was obtained from the blood gas analyzer as a control group B measurement.Wearable device blood oxygen saturation accuracy Arms (expressed as root mean square value of the SwO2-SaO2 difference) using a carbon monoxide - blood gas analyzer (CO-oximeter) as reference.

    Measured: 30mins

  2. Accuracy of blood oxygen saturation of wearable devices when SpO2/SaO2 is at 90%-100%.

    When SpO2/SaO2 is at 90%-100%,wearable device oximetry accuracy Arms (root mean square value of the SwO2-RefSpO2 difference) using a medical device registration certificate pulse oximetry monitor/tabletop ECG monitor as a reference;wearable device blood oxygen saturation accuracy Arms (expressed as root mean square value of the SwO2-SaO2 difference) using a carbon monoxide - blood gas analyzer (CO-oximeter) as reference.

    Measured: 30mins

  3. Accuracy of blood oxygen saturation of wearable devices when SpO2/SaO2 is at 80%-89%.

    when SpO2/SaO2 is at 80%-89%,wearable device oximetry accuracy Arms (root mean square value of the SwO2-RefSpO2 difference) using a medical device registration certificate pulse oximetry monitor/tabletop ECG monitor as a reference;wearable device blood oxygen saturation accuracy Arms (expressed as root mean square value of the SwO2-SaO2 difference) using a carbon monoxide - blood gas analyzer (CO-oximeter) as reference.

    Measured: 30mins

  4. Correlation analysis of parameters that have an impact on the accuracy of blood oxygen saturation of wearable devices.

    When the test result is negative, evaluate the parameters that have an impact on the accuracy of blood oxygen saturation of wearable devices.

    Measured: 30mins

  5. The variability of blood oxygen saturation of wearable devices tested three times.

    Record the pulse oximetry and pulse rate values measured at the same time by the medical monitor and the Huami Smart Wearable Pulse Oximetry Device, respectively. After an interval of 30 seconds, repeat the above steps to start the second measurement and record the measured value. After an interval of 30 seconds, repeat the above steps to start the third measurement and record the measured value. Evaluate the variability of blood oxygen saturation of wearable devices tested three times.

    Measured: 30mins

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted12 Jun 2020
  2. Study start29 May 2020
  3. Primary completion30 Jun 2020 (estimated)
  4. Study completion30 Jun 2020 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Other
Time perspective
Cross-sectional
Enrolment
30 participants (estimated)
Target duration
1 Day
IDs, dates and publications
ClinicalTrials.gov ID
NCT04429906
Sponsor's protocol ID
2020114-0604
Acronym
DSHWD
Status last verified
Jun 2020
First submitted
4 Jun 2020
Last updated
12 Jun 2020
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Olsen M, Zeitzer JM, Richardson RN, Davidenko P, Jennum PJ, Sorensen HBD, Mignot E. A Flexible Deep Learning Architecture for Temporal Sleep Stage Classification Using Accelerometry and Photoplethysmography. IEEE Trans Biomed Eng. 2023 Jan;70(1):228-237. doi: 10.1109/TBME.2022.3187945. Epub 2022 Dec 26.

    PubMed 35786544

Questions about this trial

What is this clinical trial studying?

An observational study of Oxygen Saturation, run by Peking University First Hospital at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 12 Jun 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Each site lists its own contact: choose the nearest one under Trial locations. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Wearable device oximetry accuracy Arms (time frame: 30mins). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT04429906, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 12 Jun 2020; AzTrials last refreshed its copy on 31 Oct 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in