Laparoscopic General, Gynecological and Urologic SurgeriesMEMSEVNCT04539405
Evaluation of the Sevoflurane Consumption During General Anesthesia When Using the MemsorbTM Membrane
An observational study of Laparoscopic General, Gynecological and Urologic Surgeries, run by Ciusss de L'Est de l'Île de Montréal at 1 site in Canada.
Trial summary
The study aims at determining whether replacing the classical chemical absorber Dräegersorb 800+ on Dräeger Perseus A500 machines (Dräeger, Lübeck, Germany) by the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) with the help of high-quality monitoring (BIS and NOL) and high-end… Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Laparoscopic General, Gynecological and Urologic Surgeries
- What is being tested?
- To evaluate sevoflurane consumption when using Memsorb and To evaluate sevoflurane consumption when using classical Dräegersorb
- Type of treatment
- Medical device
- How many people
- 50 participants
- Where
- 1 siteCanada
- Study dates
- 4 May 2021to 18 Feb 2022
- How long will I take part?
- 1 Day
- Phase
- Observational
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Study population: Enrollment of 50 randomized adult patients undergoing general anesthesia for laparoscopic general, gynecological or urologic surgery. Type of surgery: laparoscopic general, gynecological and urologic surgery with general anesthesia that does not require additional regional anesthesia such as an epidural analgesia.
Who may be able to join
- ASA 1-3,
- Laparoscopic general, gynecological or urologic surgery requiring general anesthesia without additional regional anesthesia technique,
- Fully consented,
- BMI < 40,
- Age > 18.
Who cannot join
- Allergy or contra-indication to any drug used in the study protocol,
- History of unstable coronary artery disease,
- Serious cardiac arrhythmia (including atrial fibrillation),
- History of substance abuse in the last 2 months prior to surgery,
- Chronic use of psychotropic and/or opioid drugs (still existing within the last month prior to surgery,
- History of psychiatric diseases,
Show all 10Show fewer
- History of refractory PONV in previous surgery,
- Allergy to any drug used in the study protocol,
- Non-scheduled surgery,
- Refusal of the patient for participation in the study.
Trial locations and contacts
1 site in 1 country.
Canada1 site
Hopital Maisonneuve Rosemont, CIUSSS de l'Est de l'Ile de Montréal
Montreal East, Quebec, H1T2M4
About this study
Official title: Evaluation of the Sevoflurane Consumption During General Anesthesia When Using the MemsorbTM Membrane and Low Gas Flow Versus the Classical Chemical Absorbents With 2L.Min-1 Gas Flow for CO2 Removal From the Anesthesia Circuit.
The study aims at determining whether replacing the classical chemical absorber Dräegersorb 800+ on Dräeger Perseus A500 machines (Dräeger, Lübeck, Germany) by the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) with the help of high-quality monitoring (BIS and NOL) and high-end ventilators (Dräeger Perseus A500 machines; Dräeger, Lübeck, Germany) that allow minimal fresh gas flow, will significantly decrease the use of sevoflurane and its related atmospheric pollution.
Read the detailed description
Assessing the impact of anesthesia practice on global warming and carbon footprint becomes part of the standard of care and is a growing concern within the anesthesia community. Global Warming Potential (GWP) is a measure of how much a given mass of greenhouse gas contributes to global warming over a specified time period. The time period used, 20 versus 100 years, might drastically change the way we see impact of anesthesia on climate changes and GWP20 of CO2 is, by definition set at "1". Inhaled anesthetics have various GWP20: 349 for sevoflurane and 3714 for desflurane. These numbers might slightly change from one to another report/study. However, GWP20 and CDE20 alone are not sufficient to evaluate the environmental impact of anesthetic gases.
Other parameters must be included in the analysis: fresh gas flow (FGF), carrier gas (air, O2, N2O) and potency of the anesthetic gas. Unfortunately, the majority of trials did not fully consider the FGF reduction and the fact that desflurane can be administered with new closed or very low-flow anesthesia circuits as opposed as the recommended 2L.min-1 that must be used for sevoflurane according to its monography when classical chemical absorbents are used by the anesthesia team. Most of the calculations were made on a purely theoretical approach that could be different from actual measurements based on a strictly monitored anesthesia practice.
For sevoflurane, the standard FGF must be set at 2L.min-1 as there is still controversy concerning impact on renal function at lower flows if classical CO2 absorbents are used. It is still not recommended to administer sevoflurane during anesthesia with a FGF lower than 2L.min-1 in Canada when classical chemical absorbents are used according to Baxter monography when classical absorbents are used on the anesthesia circuit (http://www.baxter.ca/fr_CA/assets/downloads/monographs/Sevoflurane_FR.pdf). When continuous and accurate gas monitoring and analysis is used as recommended nowadays by all GCP guidelines (see Canadian guidelines for anesthesia practice), the use of closed or semi-closed-circuit anesthesia with very low FGF might allow for a reduction of more than 80% of the anesthetic gas administration and its consequent pollution.
Moreover, there are few clinical trials looking at the sparing effect on the consumption of anesthetic gases when the depth of anesthesia is properly monitored, with the bispectral index for instance. Indeed, and because the lack of appropriate and precise technology at the time of completion of these trials, most of the studies used only end-tidal concentration in % of the anesthetic gases to argue that they had decreased the consumption of gas during surgery. This remains a very indirect assessing method based on extrapolations as opposed to direct measurements. Studying the effect of the combination of BIS/NOL indices (depth of hypnosis / depth of analgesia) monitoring and the use of our Drager ventilators with low-FGF on precise consumption in mL of each gas in a clinical environment allows to get high quality data never reported in the past. The fact this study also uses the NOL index to ensure that level of analgesia is controlled and equivalent in all groups will also reinforce the idea that what this study measures in terms of anesthetic gas consumption is based on the real need for hypnosis for each participant, and not an overconsumption of gas because of poor control of nociception and analgesia.
Recently a new device was developed to extract CO2 from the ventilation-anesthesia circuit: the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada). All details on the technology of this new membrane are given in the Appendix 1 attached to this proposal (see at the bottom of the present text).
For the last 5 years, clinical trials have been conducted in humans on the use of this new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) and compared this membrane to the classical Drägersorb 800+ CO2 absorbent (Dräger, Lübeck, Germany). After REB committee approval and investigational testing authorization (ITA) by Health Canada, Dr O. Hung (from Halifax, Canada) reported out of 200 patients (100 with MemsorbTM, 100 with DraegersorbTM; ClinicalTrials.gov Identifier: NCT03014336) comparable data regarding the end-tidal CO2 with a median of 5.1% for memsorb™ and 5.0% for control (DraegersorbTM), stable over 2h of anesthesia. Vapor consumption data did not differ significantly between the 2 groups. Anesthesiologist used fresh gas flows at their discretion (from 0.3L.min-1 to 2.7L.min-1). As a conclusion to their study, Memsorb™ was shown to remove CO2 out of an anesthesia circuit as safely and as efficiently than the classical CO2 absorbent (Draegersorb) BUT without the known limitations of chemical absorbents at very low gas flow (0.3L.min-1).
The study here aims at determining whether replacing the classical chemical absorber Dräegersorb 800+ on Dräeger Perseus A500 machines (Dräeger, Lübeck, Germany) by the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) with the help of high-quality monitoring (BIS and NOL) and high-end ventilators (Dräeger Perseus A500 machines; Dräeger, Lübeck, Germany) that allow minimal fresh gas flow, will significantly decrease the use of sevoflurane and its related atmospheric pollution.
Indeed, MemsorbTM membrane allows to safely reduce the gas flow of the Dräger A500 ventilator as low as 0.2L.min-1 for the administration of sevoflurane (as there is no chemical reaction between the memsorb membrane and the sevoflurane, see ref 16) into the breathing circuit whereas the classical DräegersorbTM has to use a minimal gas flow of 2L.min-1 as recommended by the above cited monography (as below 2L.min-1, the Draegersorb might interact with sevoflurane and produce toxic compounds).
The impact of the present study will be that it will demonstrate sevoflurane administration at 0.2L.min-1 when using the MemsorbTM membrane (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) reduces anesthesia related pollution to its minimum compared to the mandatory FGF at 2L.min-1 that MUST be used with the classical Draegersorb CO2 absorbent. As a consequence of this study, it is expected that anesthesiologists will drastically change towards the use of MemsorbTM in their clinical practice to significantly lessen the major impact they have on the environment.
The primary endpoint of this study is to show a significant decrease of at least 25% of the sevoflurane consumption when using MemsorbTM versus DrägersorbTM. This will be expressed in mL.kg-1.h-1 of surgery and the primary objective will focus on H1 of surgery, H1 starting at the time of incision (T0).
- Being tested
- To evaluate sevoflurane consumption when using Memsorb, To evaluate sevoflurane consumption when using classical Dräegersorb
- Keywords
- Sevoflurane, General anesthesia, Memsorb membrane
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
To evaluate sevoflurane consumption when using MemsorbDevice
Sevoflurane administration at 0.2L.min-1 (minimal gas flow)
To evaluate sevoflurane consumption when using classical DräegersorbDevice
Sevoflurane administration at 2L.min-1 (classical gas flow)
| Study group | Role | What they receive |
|---|---|---|
| Memsorb | M for memsorbTM group with the minimal gas flow possible (sevoflurane administration at 0.2L.min-1) with the ventilator Draeger A-500 Perseus, Receives: Device: To evaluate sevoflurane consumption when using Memsorb | |
| Dräegersorb | D for DraegersorbTM group with gas flow at 2L.min-1 (classical sevoflurane administration) with the same ventilator Draeger A-500 Perseus. Receives: Device: To evaluate sevoflurane consumption when using classical Dräegersorb |
What is being measured
The primary outcome is the main question the study is designed to answer.
To show a significant decrease of at least 25% of the sevoflurane consumption
To compare the total sevoflurane consumption when using MemsorbTM versus DrägersorbTM. This will be expressed in mL.kg-1.h-1 of surgery and the primary objective will focus on H1 of surgery, H1 starting at the time of incision (T0).
Measured: Intraoperative
Secondary outcomes (4)
To evaluate sevoflurane consumption for each hour of surgery
To evaluate sevoflurane consumption in mL.kg-1.h-1 for each hour of surgery (H2, 3, 4 etc…) in laparoscopic abdominal surgeries of more than 1hour.
Measured: Intraoperative
To evaluate the pollution induced by anesthesia at H1 of surgery and consequent hours
To evaluate the pollution induced by anesthesia in grams of CO2 when using MemsorbTM versus DrägersorbTM at H1 of surgery
Measured: Intraoperative
To evaluate the levels of expired CO2
To evaluate the levels of expired CO2 analyzed by the Dräger Perseus A500 ventilator (electronic records per seconds) throughout the surgery when using MemsorbTM versus DrägersorbTM. Mean expired CO2 for each hour of surgery will be used in mmHg
Measured: Intraoperative
To evaluate the global costs of sevoflurane consumption
To evaluate the global costs of sevoflurane consumption in both the groups and also the costs related to the use of the MemsorbTM versus DrägersorbTM. Will be given in CAN$
Measured: Intraoperative
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted7 Sep 2020
- Study start4 May 2021
- Primary completion18 Feb 2022
- Study completion18 Feb 2022
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Laparoscopic General, Gynecological and Urologic SurgeriesStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Case-control
- Time perspective
- Prospective
- Enrolment
- 50 participants (actual)
- Target duration
- 1 Day
Sponsor and investigators
- Lead sponsor
- Ciusss de L'Est de l'Île de Montréal
- Sponsor type
- University, hospital or other
- Collaborators
- DMF Medical Incorporated
- Responsible party
- Principal investigator, Philippe Richebe, Ciusss de L'Est de l'Île de Montréal
- Investigators
- Philippe PR Richebe, MD, PhD, Principal investigator, CIUSSS Est de l'île de Montréal
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT04539405
- Sponsor's protocol ID
- 2020-2083
- Acronym
- MEMSEV
- Status last verified
- Jan 2023
- First submitted
- 31 Aug 2020
- Last updated
- 23 Jan 2023
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
Study at a glance
From the registry record, last updated by the sponsor on 23 Jan 2023. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Evaluation of the Sevoflurane Consumption During General Anesthesia When Using the MemsorbTM Membrane and Low Gas Flow Versus the Classical Chemical Absorbents With 2L.Min-1 Gas Flow for CO2 Removal From the Anesthesia Circuit.
- ClinicalTrials.gov ID
- NCT04539405
- Acronym / sponsor ID
- MEMSEV / 2020-2083
- Keywords
- Sevoflurane, General anesthesia, Memsorb membrane
- Study type
- Observational
- Recruitment status
- Completed
- Enrolment
- 50 participants (actual)
- Lead sponsor
- Ciusss de L'Est de l'Île de MontréalUniversity, hospital or other
- Collaborators
- DMF Medical Incorporated
- Study start
- 4 May 2021 (actual)
- Primary completion
- 18 Feb 2022 (actual)
- Study completion
- 18 Feb 2022 (actual)
- First posted
- 7 Sep 2020 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The study aims at determining whether replacing the classical chemical absorber Dräegersorb 800+ on Dräeger Perseus A500 machines (Dräeger, Lübeck, Germany) by the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) with the help of high-quality monitoring (BIS and NOL) and high-end ventilators (Dräeger Perseus A500 machines; Dräeger, Lübeck, Germany) that allow minimal fresh gas flow, will significantly decrease the use of sevoflurane and its related atmospheric pollution.
Detailed description
Assessing the impact of anesthesia practice on global warming and carbon footprint becomes part of the standard of care and is a growing concern within the anesthesia community. Global Warming Potential (GWP) is a measure of how much a given mass of greenhouse gas contributes to global warming over a specified time period. The time period used, 20 versus 100 years, might drastically change the way we see impact of anesthesia on climate changes and GWP20 of CO2 is, by definition set at "1". Inhaled anesthetics have various GWP20: 349 for sevoflurane and 3714 for desflurane. These numbers might slightly change from one to another report/study. However, GWP20 and CDE20 alone are not sufficient to evaluate the environmental impact of anesthetic gases.
Other parameters must be included in the analysis: fresh gas flow (FGF), carrier gas (air, O2, N2O) and potency of the anesthetic gas. Unfortunately, the majority of trials did not fully consider the FGF reduction and the fact that desflurane can be administered with new closed or very low-flow anesthesia circuits as opposed as the recommended 2L.min-1 that must be used for sevoflurane according to its monography when classical chemical absorbents are used by the anesthesia team. Most of the calculations were made on a purely theoretical approach that could be different from actual measurements based on a strictly monitored anesthesia practice.
For sevoflurane, the standard FGF must be set at 2L.min-1 as there is still controversy concerning impact on renal function at lower flows if classical CO2 absorbents are used. It is still not recommended to administer sevoflurane during anesthesia with a FGF lower than 2L.min-1 in Canada when classical chemical absorbents are used according to Baxter monography when classical absorbents are used on the anesthesia circuit (http://www.baxter.ca/fr_CA/assets/downloads/monographs/Sevoflurane_FR.pdf). When continuous and accurate gas monitoring and analysis is used as recommended nowadays by all GCP guidelines (see Canadian guidelines for anesthesia practice), the use of closed or semi-closed-circuit anesthesia with very low FGF might allow for a reduction of more than 80% of the anesthetic gas administration and its consequent pollution.
Moreover, there are few clinical trials looking at the sparing effect on the consumption of anesthetic gases when the depth of anesthesia is properly monitored, with the bispectral index for instance. Indeed, and because the lack of appropriate and precise technology at the time of completion of these trials, most of the studies used only end-tidal concentration in % of the anesthetic gases to argue that they had decreased the consumption of gas during surgery. This remains a very indirect assessing method based on extrapolations as opposed to direct measurements. Studying the effect of the combination of BIS/NOL indices (depth of hypnosis / depth of analgesia) monitoring and the use of our Drager ventilators with low-FGF on precise consumption in mL of each gas in a clinical environment allows to get high quality data never reported in the past. The fact this study also uses the NOL index to ensure that level of analgesia is controlled and equivalent in all groups will also reinforce the idea that what this study measures in terms of anesthetic gas consumption is based on the real need for hypnosis for each participant, and not an overconsumption of gas because of poor control of nociception and analgesia.
Recently a new device was developed to extract CO2 from the ventilation-anesthesia circuit: the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada). All details on the technology of this new membrane are given in the Appendix 1 attached to this proposal (see at the bottom of the present text).
For the last 5 years, clinical trials have been conducted in humans on the use of this new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) and compared this membrane to the classical Drägersorb 800+ CO2 absorbent (Dräger, Lübeck, Germany). After REB committee approval and investigational testing authorization (ITA) by Health Canada, Dr O. Hung (from Halifax, Canada) reported out of 200 patients (100 with MemsorbTM, 100 with DraegersorbTM; ClinicalTrials.gov Identifier: NCT03014336) comparable data regarding the end-tidal CO2 with a median of 5.1% for memsorb™ and 5.0% for control (DraegersorbTM), stable over 2h of anesthesia. Vapor consumption data did not differ significantly between the 2 groups. Anesthesiologist used fresh gas flows at their discretion (from 0.3L.min-1 to 2.7L.min-1). As a conclusion to their study, Memsorb™ was shown to remove CO2 out of an anesthesia circuit as safely and as efficiently than the classical CO2 absorbent (Draegersorb) BUT without the known limitations of chemical absorbents at very low gas flow (0.3L.min-1).
The study here aims at determining whether replacing the classical chemical absorber Dräegersorb 800+ on Dräeger Perseus A500 machines (Dräeger, Lübeck, Germany) by the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) with the help of high-quality monitoring (BIS and NOL) and high-end ventilators (Dräeger Perseus A500 machines; Dräeger, Lübeck, Germany) that allow minimal fresh gas flow, will significantly decrease the use of sevoflurane and its related atmospheric pollution.
Indeed, MemsorbTM membrane allows to safely reduce the gas flow of the Dräger A500 ventilator as low as 0.2L.min-1 for the administration of sevoflurane (as there is no chemical reaction between the memsorb membrane and the sevoflurane, see ref 16) into the breathing circuit whereas the classical DräegersorbTM has to use a minimal gas flow of 2L.min-1 as recommended by the above cited monography (as below 2L.min-1, the Draegersorb might interact with sevoflurane and produce toxic compounds).
The impact of the present study will be that it will demonstrate sevoflurane administration at 0.2L.min-1 when using the MemsorbTM membrane (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) reduces anesthesia related pollution to its minimum compared to the mandatory FGF at 2L.min-1 that MUST be used with the classical Draegersorb CO2 absorbent. As a consequence of this study, it is expected that anesthesiologists will drastically change towards the use of MemsorbTM in their clinical practice to significantly lessen the major impact they have on the environment.
The primary endpoint of this study is to show a significant decrease of at least 25% of the sevoflurane consumption when using MemsorbTM versus DrägersorbTM. This will be expressed in mL.kg-1.h-1 of surgery and the primary objective will focus on H1 of surgery, H1 starting at the time of incision (T0).
Study design
- Observational model
- Case-control
- Time perspective
- Prospective
- Target follow-up
- 1 Day
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Memsorb | – | Device: To evaluate sevoflurane consumption when using MemsorbM for memsorbTM group with the minimal gas flow possible (sevoflurane administration at 0.2L.min-1) with the ventilator Draeger A-500 Perseus, |
| Dräegersorb | – | Device: To evaluate sevoflurane consumption when using classical DräegersorbD for DraegersorbTM group with gas flow at 2L.min-1 (classical sevoflurane administration) with the same ventilator Draeger A-500 Perseus. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
- Study population
- Enrollment of 50 randomized adult patients undergoing general anesthesia for laparoscopic general, gynecological or urologic surgery. Type of surgery: laparoscopic general, gynecological and urologic surgery with general anesthesia that does not require additional regional anesthesia such as an epidural analgesia.
Key inclusion criteria (5 listed)
- ASA 1-3,
- Laparoscopic general, gynecological or urologic surgery requiring general anesthesia without additional regional anesthesia technique,
- Fully consented,
- BMI < 40,
- Age > 18.
Key exclusion criteria (10 listed)
- Allergy or contra-indication to any drug used in the study protocol,
- History of unstable coronary artery disease,
- Serious cardiac arrhythmia (including atrial fibrillation),
- History of substance abuse in the last 2 months prior to surgery,
- Chronic use of psychotropic and/or opioid drugs (still existing within the last month prior to surgery,
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| To evaluate sevoflurane consumption when using MemsorbSevoflurane administration at 0.2L.min-1 (minimal gas flow) | Device | Memsorb |
| To evaluate sevoflurane consumption when using classical DräegersorbSevoflurane administration at 2L.min-1 (classical gas flow) | Device | Dräegersorb |
Primary outcome
To show a significant decrease of at least 25% of the sevoflurane consumption
Time frame: Intraoperative
To compare the total sevoflurane consumption when using MemsorbTM versus DrägersorbTM. This will be expressed in mL.kg-1.h-1 of surgery and the primary objective will focus on H1 of surgery, H1 starting at the time of incision (T0).
4 secondary outcomes
To evaluate sevoflurane consumption for each hour of surgery
Time frame: Intraoperative
To evaluate the pollution induced by anesthesia at H1 of surgery and consequent hours
Time frame: Intraoperative
To evaluate the levels of expired CO2
Time frame: Intraoperative
To evaluate the global costs of sevoflurane consumption
Time frame: Intraoperative
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 23 Jan 2023.
Related trials for Laparoscopic General, Gynecological and Urologic Surgeries
Not yet recruitingObservational study
Clinical Trial Protocol for Evaluating the Efficacy and Safety of Thoracoabdominal Endoscopic Surgery System in Remote Laparoscopic Surgery for Urology, General Surgery, Gynecology and Thoracic Surgery
Colorectal Cancer, Esophageal Cancer
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Philippe PR Richebe, MD, PhD principal investigator
CIUSSS Est de l'île de Montréal
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Laparoscopic General, Gynecological and Urologic Surgeries, run by Ciusss de L'Est de l'Île de Montréal at 1 site in Canada.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 5 inclusion and 10 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 23 Jan 2023. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Canada. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
To show a significant decrease of at least 25% of the sevoflurane consumption (time frame: Intraoperative). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?