Study of ECHELON ENDOPATH(TM) Staple Line Reinforcement

A study testing Staple Line Reinforcement in people with Gastric Procedures and Lung Resection Procedures, run by Ethicon Endo-Surgery at 8 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This prospective, single-arm, multi-center evaluation will collect clinical data in a post-market setting. The two types of procedures studied will be gastric and thoracic. Investigators will perform each procedure using the device in compliance with their standard surgical approach and the ECHELON ENDOPATH Staple… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Gastric Procedures, Lung Resection Procedures
What is being tested?
Staple Line Reinforcement
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
240 participants
Where
8 sitesUnited States
Study dates
29 Sep 2020to 8 Feb 2023
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Primary procedure (gastric or lung resection) where the ECHELON SLR is planned to be used for reinforcement of staple lines per the IFU in either a or b:
  • Gastric procedures limited to laparoscopic gastric resection, robotic laparoscopic gastric resection, partial gastrectomy, gastric wedge resection, subtotal gastrectomy, laparoscopic Roux-en-Y gastric bypass, and robotic laparoscopic gastric bypass; or
  • Lung resection procedures that include lobectomy, segmentectomy or wedge resection, and lung volume reduction surgery, and may be video assisted thoracic surgery (VATS) or open procedures;
  • Willingness to give consent and comply with all study-related evaluations and visit schedule; and
  • At least 18 years of age.

Who cannot join

  • Preoperative:
  • Physical or psychological condition which would impair study participation;
  • Body mass index (BMI) ≥ 46.0 kg/m2;
  • The procedure is a revision/reoperation for the same indication or same anatomical location;
  • A procedure where extended wound or organ support is required;
  • Any medical condition that the investigator deems could impact inflammatory or immune response;
Show all 11Show fewer
  • Concurrent treatment with medications that the investigator deems could have influence on wound healing;
  • History of hypersensitivity to polyglactin (Vicryl®), Polydioxanone (PDO or PDS), or hypersensitivity to related products (cross-allergy); or
  • Enrollment in a concurrent interventional clinical study that could impact the study endpoints.
  • Intraoperative:
  • Per surgeon discretion, presence of adhesions that could lead to an increased risk of leak occurrence at a different location than the staple line.

Trial locations and contacts

8 sites in 1 country.

8 sites in 1 country

United States8 sites
  • Our Lady of the Lake Regional Medical Center

    Baton Rouge, Louisiana, 70808

  • Atrium Health

    Charlotte, North Carolina, 28203

  • University of Virginia

    Charlottesville, Virginia, 22908

  • University of Missouri

    Columbia, Missouri, 65201

  • Henry Ford Health System

    Detroit, Michigan, 48202

  • Spectrum Health

    Grand Rapids, Michigan, 49503

  • Indiana University Health

    Indianapolis, Indiana, 46202

  • Long Island Jewish Medical Center

    New York, New York, 11040

Other trials in United States

About this study

Official title: A Prospective, Single-Arm, Multi-Center Study of the ECHELON ENDOPATH(TM) Staple Line Reinforcement Device in Gastric and Lung Resection Procedures

This prospective, single-arm, multi-center evaluation will collect clinical data in a post-market setting. The two types of procedures studied will be gastric and thoracic. Investigators will perform each procedure using the device in compliance with their standard surgical approach and the ECHELON ENDOPATH Staple Line Reinforcement instructions for use.

Read the detailed description

The primary objective of this study is to prospectively generate clinical data on device-related adverse events in a post-market setting using ECHELON ENDOPATH Staple Line Reinforcement per its instructions for use. There will be no blinding in this study. An interim analysis is planned for sample-size re-estimation.

Being tested
Staple Line Reinforcement
Keywords
Buttress, gastrectomy, gastric bypass, lung volume reduction

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Staple Line ReinforcementDevice

    ENDOPATH ECHELON Staple Line Reinforcement is used for reinforcement of staple lines during lung resection and bariatric surgical procedures, according to instructions for use.

Study groups in NCT04544865
Study groupRoleWhat they receive
Gastric and thoracic staple line reinforcementExperimentalECHELON ENDOPATH Staple Line Reinforcement is used during a gastric or thoracic procedure.
Receives: Device: Staple Line Reinforcement

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Number of Participants Who Experienced Specific Device-related Adverse Events (AEs) Through 70 Days Post Gastric Resection Procedures

    Number of participants who experienced specific device-related AEs post gastric resection procedures was reported. Specific device-related AEs were bleeding (defined as occurrence of post-surgery blood transfusion deemed related to bleeding at the staple line or return to surgery room before 70-day post-surgery follow-up visit due to bleeding deemed related to the staple line), leak (defined as occurrence of intra-surgery or post-surgery gastrointestinal leak related to the staple line as documented intra-surgery, by clinical exam, or radiographically), and stricture (defined as occurrence of stricture documented radiographically or by endoscopy along the staple line).

    Measured: From Day 0 (day of surgery) through 70 days post-surgery

  2. Number of Participants Who Experienced Specific Device-related Adverse Events Through 70 Days

    Number of participants who experienced specific device-related AEs post lung resection procedures was reported. Specific device related AEs were prolonged air leak (occurrence of intra-surgery or post-surgery lung leak related to the staple line as documented intra-surgery, by clinical exam, or radiographically) deemed related to the staple line (greater than post-surgery Day 7), and empyema (defined as purulent fluid collection in the pleural space documented radiographically, excluding chronic empyema).

    Measured: From Day 0 (day of surgery) through 70 days post-surgery

Secondary outcomes (3)
  1. Number of Study Devices Replaced During Surgery

    Number of study devices replaced during surgery due to slipping or bunching or not properly loaded onto stapler cartridge, Other - preloaded wrong color cartridge, Other- SLR added before cartridge, Other- dropped on the ground was reported.

    Measured: Day 0 (Intra-surgery)

  2. Device Usability Evaluation by Surgeon Questionnaire: Investigator's Device Experience Feedback

    Investigators were asked to answer a non-validated device questionnaire related to his/her experience using the ECHELON SLR. Questionnaire included following questions "experienced less manipulation or movement compared to previous device, experienced greater confidence the device will deliver best result compared to previous device, experienced greater confidence in simplicity of set up compared to previous device, experienced less frustration with device compared to previous device, experienced less waste compared to previous device, Likely to recommend to a colleague. Each question response was ranged from 1 to 5 where 1=strongly disagree, 2=slightly disagree,3=neutral, 4=slightly agree, 5= strongly agree, where the higher response signified good experience on the device. In this outcome measure, device questionnaire responses which were provided by selected investigators after the site's first, third, and 5th procedure were reported.

    Measured: Day 0 (Intra-surgery)

  3. Device Usability Evaluation by Surgeon Questionnaire: Investigator's Device Satisfaction Feedback

    Investigators were asked to answer a non-validated device questionnaire related to his/her experience using the ECHELON SLR. Questionnaire included following questions "satisfaction of operative flow compared to previous device and satisfaction of manipulation and repositioning compared to previous device" and each question response was ranged from 1 to 5 where 1=very dissatisfied, 2=dissatisfied, 3=neither dissatisfied nor satisfied (NDNS), 4=satisfied, 5=very satisfied, where higher response signified good satisfaction with the device. In this outcome measure, device questionnaire responses which were provided by selected investigators after the site's first, third, and fifth procedure were reported.

    Measured: Day 0 (Intra-surgery)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted10 Sep 2020
  2. Study start29 Sep 2020
  3. Primary completion17 Oct 2022
  4. Study completion8 Feb 2023

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
240 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04544865
Sponsor's protocol ID
ESC_2018_03
Status last verified
Nov 2023
First submitted
4 Sep 2020
Last updated
5 Dec 2023
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
Yes: Johnson \& Johnson Medical Device Companies have an agreement with the Yale Open Data Access (YODA) Project to serve as the independent review panel for evaluation of requests for clinical study reports and participant level data from investigators and physicians for scientific research that will advance medical knowledge and public health. Requests for access to the study data can be submitted through the YODA Project site at http://yoda.yale.edu.

Questions about this trial

What is this clinical trial studying?

A study testing Staple Line Reinforcement in people with Gastric Procedures and Lung Resection Procedures, run by Ethicon Endo-Surgery at 8 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 5 Dec 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 8 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 5 Dec 2023). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Number of Participants Who Experienced Specific Device-related Adverse Events (AEs) Through 70 Days Post Gastric Resection Procedures (time frame: From Day 0 (day of surgery) through 70 days post-surgery). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 5 Dec 2023.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in