Premature InfantNCT04546568

SCO2T (Servo vs. Servo)

A study of Premature Infant and Oxygen Therapy, run by University of Edinburgh at 1 site in United Kingdom.

Terminated

Stopped early. The study ended before it was planned to and will not resume.

Official record
Search all trials

Trial summary

Most premature babies require oxygen therapy. There is uncertainty about what oxygen levels are the best. The oxygen levels in the blood are measured using a monitor called a saturation monitor and the oxygen the baby breathes is adjusted to keep the level in a target range. Read more about this study

Who is it for?
Ages 2 days and olderHealthy volunteers can take part
Conditions the study is for
Premature Infant, Oxygen Therapy, Hypoxia
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
11 participants
Where
1 siteUnited Kingdom
Study dates
10 Aug 2020to 3 May 2023
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 2 days and older
  • All sexes
  • Healthy volunteers accepted
  • Child, Adult, Older adult

Who may be able to join

  • Infants born at less than 30 weeks gestation
  • Infants greater than 48 hours of age
  • Infants who are receiving supplementary oxygen
  • Person with parental responsibility able to give consent

Who cannot join

  • Congenital anomalies that would prevent targeting SpO2 to 90-95% (e.g. cardiac defects)
  • Clinical condition of an infant would impair accurateTcPO2 measurement (e.g. impaired perfusion or requirement of inotropic or vasopressor support)

Trial locations and contacts

1 site in 1 country.

United Kingdom1 site
  • The Simpson Centre for Reproductive Health, Royal Infirmary Edinburgh

    Edinburgh, City Of Edinburgh, EH16 4SA

Other trials in United Kingdom

About this study

Official title: Servo Controlled Oxygen Targeting (SCO2T): Servo Versus Servo

Most premature babies require oxygen therapy. There is uncertainty about what oxygen levels are the best. The oxygen levels in the blood are measured using a monitor called a saturation monitor and the oxygen the baby breathes is adjusted to keep the level in a target range. Although there is evidence that lower oxygen levels maybe harmful, it is not known how high they need to be for maximum benefit. Very high levels are also harmful. Saturation monitors are not very good for checking for high oxygen levels. For this a different kind of monitor, called a transcutaneous monitor, is better.

Keeping oxygen levels stable is usually done by nurses adjusting the oxygen levels by hand (manual control). There is also equipment available that can do this automatically (servo control). It is not known which is best.

Read more

Studies of automated control have shown that infants spend more time within their intended target oxygen saturation range. These have not included measurements of transcutaneous oxygen. There are no previous studies directly comparing automated respiratory devices.

The investigators aim to show the transcutaneous oxygen levels as well as the oxygen saturation levels when babies have their oxygen adjusted using two automated (servo) control devices delivering nasal high flow.

For a period of 12 hours each baby will have their oxygen adjusted automatically using each devices for 6 hours respectively. The investigators will compare the range of oxygen levels that are seen between the two respiratory devices.

Read the detailed description

Presently oxygen is titrated against saturation (SpO2) by manual adjustment. Automated or servo-control systems have been developed that result in tighter control of SpO2 and more time spent in the intended target range. These systems are already in clinical use. Automated systems produce quite large fluctuations in fraction of inspired oxygen (FiO2) in order to keep SpO2 in range. It is possible that this could result in short periods of high or low oxygen tension (PO2) that are undetectable using saturation monitoring. Studies to date have examined the effects of manual and automated (servo) oxygen targeting on SpO2 but not on transcutaneous oxygen tension (TcPO2). There are no studies directly comparing two automated systems.

There is a need to determine the achieved SpO2 and TcPO2 distributions associated with the use of different automated control systems as a first step in planning future trials. When this is measured over a small number of hours it is not anticipated that this would have an influence on clinical outcome.

This study is a prospective, single centre, randomised crossover trial of two automated (servo) control devices - IntellO2 (Vapotherm, USA) versus Leoni plus CLAC (Löwenstein Medical, Germany) - delivering nasal high flow employing automated oxygen titration. Each infant will act as their own control. Infants born at less than 30 weeks gestation, greater than 48 hour of age and receiving supplementary oxygen will be eligible for inclusion.

The study will be undertaken in the Neonatal Unit at the Simpson Centre for Reproductive Health at the Royal Infirmary of Edinburgh.

Total study time is 12 hours for each infant. Infants will be randomised to commence on either automated (servo) control using either Leoni plus CLAC or IntellO2, Vapotherm. SpO2 (range 90-95%) will be continuously monitored as per normal standard of care. A second pulse oximetry probe will be place for servo control input.

Additional monitoring will be carried out as shown below:

  1. TcPO2 monitoring
  2. FiO2 monitoring
  3. Heart rate monitoring (used to validate SpO2 readings)
  4. Arterial gas sampling (only if conducted by the direct care team as part of the routine care of the infant; no extra blood samples will be taken as part of the study)

FiO2 will be adjusted by the respiratory support devices which have integrated automated oxygen control, set to maintain a SpO2 target range of 90-95%. The IntellO2 device uses Precision Flow technology (IntellO2, Vapotherm, USA). By means of a modified closed-loop algorithm, the devise uses Masimo pulse oximetry to target a user-set SpO2 value. The Leoni plus CLAC (Closed-Loop Automated oxygen Control) ventilator (Leoni plus, Löwenstein Medical, Germany) similarly uses MasimoSET (Signal Extraction Technology) to target SpO2 and automate oxygen control. Both devices have a signal averaging time of 8 seconds (Masimo, Irvine, USA). The Leoni plus CLAC algorithm is set to a 30 seconds wait time between adjustments, allowing up to 120 automated adjustments/hour.

SpO2 readings will be downloaded directly from the multiparameter patient monitor. SpO2 will be measured using a Phillips MX500 multiparameter monitor (Phillips, Germany, CE 0366). TcPO2 will be measured using a SenTec Digital Monitoring System with OxiVent sensor (SenTec AG, Switzerland, European patent No. 1535055, CE 0120). Both monitors are routinely used in clinical practice. Transcutaneous data will be recorded contemporaneously and the site of the transcutaneous probe will be rotated on each infant every 2 hours. Control of sensor temperature and application duration are designed to meet all applicable standards and this monitoring device is use routinely in many neonatal units.

Being tested
Servo control (closed-loop automatic control of the inspiratory fraction of oxygen (FiO2)) - Leoni plus CLAC, Servo control (closed-loop automatic control of the inspiratory fraction of oxygen (FiO2)) - IntellO2 OAM
Keywords
Servo Control, Close-Loop Automated Oxygen Control
Medical subject headings
Premature Birth, Hypoxia, Hyperoxia, Obstetric Labor, Premature

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Servo control (closed-loop automatic control of the inspiratory fraction of oxygen (FiO2)) - Leoni plus CLACDevice

    FiO2 adjustments will be made by the Leoni plus CLAC (Closed-Loop Automated oxygen Control) ventilator (Leoni plus, Löwenstein Medical, Germany).

    Manual adjustments of the inspired oxygen fraction can additionally be made as per standard care.

  • Servo control (closed-loop automatic control of the inspiratory fraction of oxygen (FiO2)) - IntellO2 OAMDevice

    FiO2 adjustments will be made by the IntellO2 Oxygen Assist Module (OAM) for Precision Flow (IntellO2, Vapotherm, USA).

    Manual adjustments of the inspired oxygen fraction can additionally be made as per standard care.

Study groups in NCT04546568
Study groupRoleWhat they receive
Servo control - Leoni plus CLACExperimentalAutomated control of oxygen. The oxygen saturation target range will be set to 90-95% as per standard practice. Automated oxygen control can be overridden by manual adjustment of oxygen at any time if this is considered necessary to optimise control of oxygenation according to current clinical targets.
Receives: Device: Servo control (closed-loop automatic control of the inspiratory fraction of oxygen (FiO2)) - Leoni plus CLAC
Servo control - IntellO2 Precision Flow, VapothermActive comparatorAutomated control of oxygen. The oxygen saturation target range will be set to 90-95% (set to maintain an integral value of 93%) as per standard practice. Automated oxygen control can be overridden by manual adjustment of oxygen at any time if this is considered necessary to optimise control of oxygenation according to current clinical targets.
Receives: Device: Servo control (closed-loop automatic control of the inspiratory fraction of oxygen (FiO2)) - IntellO2 OAM

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Incidence of hyperoxia and hypoxia on transcutaneous monitoring

    To discover the percentage time spent within a TcPO2 range of 50mmHg (6.7kPa) - 80mmHg (10.7kPa) when infants are targeted to an SpO2 range of 90-95% using two automated (servo) control devices (Leoni plus CLAC and Vapotherm IntellO2) delivering nasal high flow.

    Measured: 12 hours

Secondary outcomes (8)
  1. Transcutanous oxygen variability

    To discover the variability in TcPO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using the two automated (servo) control devices delivering nasal high flow.

    Measured: 12 hours

  2. Incidence of hyperoxia and hypoxia on saturation monitoring

    To discover the percentage time spent within target SpO2 range of 90-95% when infants are targeted using the two automated (servo) control devices delivering nasal high flow.

    Measured: 12 hours

  3. Saturation variability

    To discover the variability in SpO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using the two automated (servo) control devices delivering nasal high flow.

    Measured: 12 hours

  4. Fraction of inspired oxygen variability

    To discover the variability in FiO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using the two automated (servo) control devices delivering nasal high flow.

    Measured: 12 hours

  5. Pooled frequency histogram of TcPO2

    To generate a pooled frequency histogram of percentage time at a TcPO2 of below 30mmHg, 30-39.9mmHg, 40-49.9mmHg, 50-59.9mmHg, 60-69.9mmHg, 70-79.9mmHg, and 80mmHg and above for infants targeted to an SpO2 range of 90-95% using the two automated (servo) control devices delivering nasal high flow.

    Measured: 12 hours

  6. Pooled frequency histogram of SpO2

    To generate a pooled frequency histogram of percentage time at each SpO2 point between 80 - 100% for infants targeted to an SpO2 range of 90-95% using the two automated (servo) control devices delivering nasal high flow.

    Measured: 12 hours

  7. Pooled frequency histogram of FiO2

    To generate a pooled frequency histogram of the cumulative frequency at a FiO2 of 0.21-0.3, 0.31-0.4, 0.41-0.5, 0.51-0.6, 0.61-0.7, 0.81-0.9 and 0.91-1.0 for infants targeted to an SpO2 range of 90-95% using the two automated (servo) control devices delivering nasal high flow.

    Measured: 12 hours

  8. Desaturations

    To discover the frequency, duration and depth of desaturations and the area (change in PO2 versus time) above and below the set PO2 threshold for infants targeted to an SpO2 range of 90-95% using the two automated (servo) control devices delivering nasal high flow.

    Measured: 12 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted14 Sep 2020
  2. Study start10 Aug 2020
  3. Primary completion3 May 2023
  4. Study completion3 May 2023

Why it stopped: Failed to meet recruitment target

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Crossover
Model description
This is a randomised cross-over study of servo-controlled oxygen targeting in premature infants, with infants acting as their own controls
Masking
None (open label)
Enrolment
11 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04546568
Sponsor's protocol ID
AC20034
Status last verified
May 2022
First submitted
4 Sep 2020
Last updated
2 Jun 2023
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study of Premature Infant and Oxygen Therapy, run by University of Edinburgh at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: stopped early, as of the sponsor's last update on 2 Jun 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study was stopped early (reason given: Failed to meet recruitment target). Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Incidence of hyperoxia and hypoxia on transcutaneous monitoring (time frame: 12 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 2 Jun 2023.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in