Aerosol ContainmentNCT04864236

An Isolation Device to Contain Aerosol During Aerosol Generating Procedures in the Operating Room

A study testing Novel isolation device to contain aerosol in people with Aerosol Containment and Aerosol Generating Procedure, run by John Shin at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to test the safety and efficacy of a new isolation device to contain aerosol during aerosol-generating procedures in the operating room. Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
What is being tested?
Novel isolation device to contain aerosol
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
79 participants
Where
1 siteUnited States
Study dates
29 Jun 2021to 1 Sep 2021
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Elective scheduled surgery (not emergent)
  • Planned general anesthesia requiring intubation

Who cannot join

  • Patients refusing or unable to consent for any reason, including claustrophobia or inability to cooperate
  • Patients predicted to require the use of fiberoptic bronchoscopy for intubation

Trial locations and contacts

1 site in 1 country.

United States1 site
  • UCLA Medical Center

    Los Angeles, California, 90095

Other trials in United States

About this study

Official title: A Randomized Controlled Trial to Test a Novel Negative Pressure Isolation Device During Aerosol Generating Procedures in the Operating Room

The purpose of this study is to test the safety and efficacy of a new isolation device to contain aerosol during aerosol-generating procedures in the operating room.

Read the detailed description

Patients with SARS-CoV-2 infections may require procedures that generate infectious aerosols (suspension of tiny particles or droplets in the air). Aerosols are believed to be the primary method of spread of SARS-CoV-2, as well as many other communicable diseases. In the health care setting, the providers who are in close proximity to patients during these aerosol-generating procedures (ie. Intubation (tube in your throat to help you breathe), noninvasive positive pressure ventilation (breathing support administered through a face or nasal mask) methods such as continuous positive airway pressure (CPAP) or bilevel positive airway pressure (BiPAP), bronchoscopy (thin tube through nose or mouth into the lungs), high flow nasal cannula (oxygen delivery through the nose at a very high flow rate)) are at risk for becoming infected with SARS-CoV-2. The current standard of care to prevent the spread of virus to health care workers involves the use of N-95 respirators (particulate filtering facepiece respirators), which have been intermittently in short supply throughout the current COVID-19 pandemic. In addition, even when there are adequate supplies of N-95 respirators, health care workers have been reusing the respirators multiple times, which may result in decreased efficacy and potential viral exposure. In order to provide enhanced protection to health care workers during aerosol-generating procedures, the study investigators have created a device, named SLACC (Suction-assisted Local Aerosol Containment Chamber). SLACC has the ability to isolate the atmosphere around the patient's upper airway (nose, mouth, throat) via negative pressure in a manner analogous to isolation rooms used to house patients with communicable diseases spread by aerosol (e.g. tuberculosis). Furthermore, SLACC is inexpensive, portable, and relatively simple to manufacture.

The SLACC uses an external suction source to maintain a negative pressure micro-atmosphere around a patient's head (the source of infectious aerosols). The components of SLACC include a clear polymer frame, a flexible drape, and integrated sealed arm sleeves to protect the hands and forearms of the health care provider. The structure is flexible during use, transparent, can be rapidly assembled, and easily collapsible for removal and transport. It is assembled using adhesive-backed polymer sections, allowing the structure to fold from a flat shipping/storage conformation into a free-standing structure. Once deployed with the patient's head inside the chamber, the drape is secured across the open side of the chamber with adhesive and covers the patient's torso. Many other containment devices have been proposed during the COVID-19 pandemic, but none have been demonstrated to contain aerosols effectively. With external suction applied, the investigators believe SLACC vastly improves upon the aerosol containment ability of other devices. The investigators have tested SLACC in a simulated setting and recently published the results in a high-quality, peer-reviewed journal (PMID: 32541251).

The investigators propose to test the safety and efficacy of SLACC in containing contamination during aerosol-generating procedures in the operating room through a randomized controlled trial. The investigators have identified a method of objectively quantifying health care worker exposure to respiratory particles and plan to compare exposure to the number and size of respiratory particles when SLACC is used compared to the current standard of care (no device used).

Being tested
Novel isolation device to contain aerosol
Keywords
health care worker exposure

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Novel isolation device to contain aerosolDevice

    The novel isolation device is called SLACC (Suction-assisted Local Aerosol Containment Chamber). SLACC has the ability to isolate the atmosphere around the patient's upper airway (nose, mouth, throat) via negative pressure in a manner analogous to isolation rooms used to house patients with communicable diseases spread by aerosol (e.g. tuberculosis).

    The SLACC uses an external suction source to maintain a negative pressure micro-atmosphere around a patient's head (the source of infectious aerosols). The components of SLACC include a clear polymer frame, a flexible drape, and integrated sealed arm sleeves to protect the hands and forearms of the health care provider. The structure is flexible during use, transparent, can be rapidly assembled, and easily collapsible for removal and transport. Once deployed with the patient's head inside the chamber, the drape is secured across the open side of the chamber with adhesive and covers the patient's torso.

Study groups in NCT04864236
Study groupRoleWhat they receive
Intervention groupExperimentalThis group will undergo airway management in the operating room as part of the anesthesia for surgery in the presence of the novel isolation device.
Receives: Device: Novel isolation device to contain aerosol
Control groupNo interventionNot described

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Particle Count Measurements During Intubation

    particle counts will be measured using a commercially-available particle counter at two locations: at the level of the anesthesiology provider's face (at the patient's head), and at the level of the assistant provider's face (at the patient's side)

    Measured: particle counts will be measured continuously from the time of patient's entry into the operating room until the patient has been intubated for surgery, approximately 15 minutes total

Secondary outcomes (5)
  1. Time to Intubation

    defined as time from the first entry of laryngoscope into pt's oropharynx to the time of confirmed successful tracheal intubation

    Measured: during intubation

  2. Total Number of Intubation Attempts

    total number of intubation attempts during airway management

    Measured: during intubation

  3. Pre-operative Airway Assessment (Mallampati Score)

    Mallampati score, can be given a score of 1-4 (whole numbers only), score of 4 indicates higher likelihood of difficult intubation and score of 1 indicates lower likelihood of difficult intubation

    Measured: pre-operative assessment

  4. Laryngoscopy Grade View (Cormack-Lehane) Obtained During Intubation

    Cormack-Lehane grade assigned by anesthesiologist during intubation, grade can be 1-4, grade 1 correlates with a view of entirety of vocal cords while grade 4 correlates with no view at all of any part of the vocal cords

    Measured: during intubation

  5. Pre-operative Airway Assessment (Thyromental Distance)

    thyromental distance is the distance measured between the chin and the thyroid cartilage (both landmarks are readily palpable on physical exam), this distance is typically measured in centimeters or as a measure of the number of finger breadths (as reported here). Finger breadths is defined as the number of finger widths that can fit within the defined distance.

    Measured: pre-operative assessment

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted28 Apr 2021
  2. Study start29 Jun 2021
  3. Primary completion1 Sep 2021
  4. Study completion1 Sep 2021

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
79 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04864236
Sponsor's protocol ID
21-000022
Status last verified
Sep 2022
First submitted
19 Apr 2021
Last updated
21 Oct 2022
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Publications

  • Seger CD, Wang L, Dong X, Tebon P, Kwon S, Liew EC, Marijic J, Umar S, Hoftman NN. A Novel Negative Pressure Isolation Device for Aerosol Transmissible COVID-19. Anesth Analg. 2020 Sep;131(3):664-668. doi: 10.1213/ANE.0000000000005052.

    PubMed 32541251

Questions about this trial

What is this clinical trial studying?

A study testing Novel isolation device to contain aerosol in people with Aerosol Containment and Aerosol Generating Procedure, run by John Shin at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 21 Oct 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 21 Oct 2022). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Particle Count Measurements During Intubation (time frame: particle counts will be measured continuously from the time of patient's entry into the operating room until the patient has been intubated for surgery, approximately 15 minutes total). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT04864236, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 21 Oct 2022.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in