Multifocal Intraocular LensesNCT04865016
Low Add Multifocal Versus Extended Depth of Focus Intraocular Lens
A study testing Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00) in people with Multifocal Intraocular Lenses and Extended Depth of Focus, run by Marmara University at 1 site in Turkey (Türkiye).
Trial summary
The innovations in phacoemulsification surgery and advances in surgical equipment ensure a safe extraction of the crystalline lens and implantation of intraocular lenses (IOL). As a result of developing intraocular lens technologies and changing patient demands, many multifocal, accommodative, toric, toric-multifocal… Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Multifocal Intraocular Lenses, Extended Depth of Focus
- What is being tested?
- Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00)
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 29 participants
- Where
- 1 siteTurkey (Türkiye)
- Study dates
- 1 Oct 2016to 1 Mar 2019
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Inclusion criteria were patients with bilateral visually significant cataract seeking spectacle independence, especially at far and intermediate distances, age of ≥18 years, preoperative corneal astigmatism less than 1.00 D, axial length (AL) between 21 - 26.5 mm, and availability to attend all follow-up visits.
Who cannot join
- Patients were excluded from the study if any of the following conditions were present: high visual demand for near vision, amblyopia, glaucoma, corneal opacity, severe/moderate to severe dry eye, pupillary abnormalities and dysfunction, corneal astigmatism ≥1.00 D, AL <21 mm or ≥26.5 mm, diabetes mellitus with retinal changes, active uveitis or history of uveitis, previous ocular surgery and lack of attendance to any follow-up visit.
Trial locations and contacts
1 site in 1 country.
Turkey (Türkiye)1 site
Marmara University School of Medicine, Department of Ophthalmology
Istanbul, 34854
About this study
Official title: Clinical Outcomes With a Low Add Multifocal and an Extended Depth of Focus Intraocular Lenses Both Implanted With Mini-monovision
The innovations in phacoemulsification surgery and advances in surgical equipment ensure a safe extraction of the crystalline lens and implantation of intraocular lenses (IOL). As a result of developing intraocular lens technologies and changing patient demands, many multifocal, accommodative, toric, toric-multifocal, and extended depth of focus (EDOF) IOLs have been introduced and the original aim of restoration of far vision with implanted monocular IOLs has been evolved to an objective of improved vision at different distances with total spectacle independence. This study aimed to compare the visual acuity at different distances, contrast sensitivity, spectacle needs, photic phenomena, and quality of life parameters of patients bilaterally implanted with a low add bifocal (Tecnis ZKB00) or an EDOF IOL (Tecnis Symfony), both with planned mini-monovision to improve near vision.
Read the detailed description
The innovations in phacoemulsification surgery and advances in surgical equipment ensure a safe extraction of the crystalline lens and implantation of intraocular lenses (IOL). As a result of developing intraocular lens technologies and changing patient demands, many multifocal, accommodative, toric, toric-multifocal, and extended depth of focus (EDOF) IOLs have been introduced and the original aim of restoration of far vision with implanted monocular IOLs has been evolved to an objective of improved vision at different distances with total spectacle independence. This study aimed to compare the visual acuity at different distances, contrast sensitivity, spectacle needs, photic phenomena, and quality of life parameters of patients bilaterally implanted with a low add bifocal (Tecnis ZKB00) or an EDOF IOL (Tecnis Symfony), both with planned mini-monovision to improve near vision. In this regard, binocular logMAR uncorrected visual acuities (UVA), monocular defocus curves, CS with CSV 1000-E and Pelli-Robson Test (PRT), spectacle needs and quality of life parameters with NEI RQL-42 questionnaire were evaluated at postoperative 1, 3, and 6 months.
- Being tested
- Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00)
- Keywords
- low add multifocal intraocular lens, MIOL, ZKB00, extended depth of focus, EDOF, ZXR00, mini-monovision
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00)Device
This study included 30 eyes of 15 cataract patients undergoing cataract surgery with implantation of a low add (+2.75 Diopters [D]) bifocal (Tecnis ZKB00; Johnson and Johnson Surgical Vision Inc., Jacksonville, Florida, USA) IOL (MIOL Group) and 30 eyes of 15 patients with implantation of an EDOF (Tecnis Symfony [ZXR00]; Johnson and Johnson Surgical Vision Inc., Jacksonville, Florida, USA) IOL (EDOF Group).
| Study group | Role | What they receive |
|---|---|---|
| MIOL Group | Active comparator | 30 eyes of 15 cataract patients undergoing cataract surgery were implanted with a low add (+2.75 Diopters [D]) bifocal (Tecnis ZKB00; Johnson and Johnson Surgical Vision Inc., Jacksonville, Florida, USA) IOL (MIOL Group). Receives: Device: Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00) |
| EDOF Group | Active comparator | 30 eyes of 15 cataract patients undergoing cataract surgery were implanted with an EDOF (Tecnis Symfony [ZXR00]; Johnson and Johnson Surgical Vision Inc., Jacksonville, Florida, USA) IOL (EDOF Group). Receives: Device: Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00) |
What is being measured
The primary outcome is the main question the study is designed to answer.
The visual acuity
monocular and binocular uncorrected and corrected distance (UDVA, CDVA; at 6 m), intermediate (UIVA, CIVA; at 60 cm), and near (UNVA, CNVA; at 40 cm) logMAR visual acuity were assessed with electronic shelled chart.Distance visual acuity was measured with LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Snellen Charts under photopic conditions (85 cd/m2). Intermediate and near visual acuities were measured and recorded with a Turkish reading chart prepared with reference to Bailey-Lovie and ETDRS (Early Treatment Diabetic Retinopathy Study) reading charts covering international standards.13 The logMAR values of the chart designed for 35 cm distance were corrected with the formula "log10(standard distance/new distance [cm])" according to the distance used (40 and 60 cm).
Measured: Preoperative
The visual acuity
monocular and binocular uncorrected and corrected distance (UDVA, CDVA; at 6 m), intermediate (UIVA, CIVA; at 60 cm), and near (UNVA, CNVA; at 40 cm) logMAR visual acuity were assessed with electronic shelled chart.Distance visual acuity was measured with LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Snellen Charts under photopic conditions (85 cd/m2). Intermediate and near visual acuities were measured and recorded with a Turkish reading chart prepared with reference to Bailey-Lovie and ETDRS (Early Treatment Diabetic Retinopathy Study) reading charts covering international standards.13 The logMAR values of the chart designed for 35 cm distance were corrected with the formula "log10(standard distance/new distance [cm])" according to the distance used (40 and 60 cm).
Measured: Postoperative 1st month
The visual acuity
monocular and binocular uncorrected and corrected distance (UDVA, CDVA; at 6 m), intermediate (UIVA, CIVA; at 60 cm), and near (UNVA, CNVA; at 40 cm) logMAR visual acuity were assessed with electronic shelled chart.Distance visual acuity was measured with LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Snellen Charts under photopic conditions (85 cd/m2). Intermediate and near visual acuities were measured and recorded with a Turkish reading chart prepared with reference to Bailey-Lovie and ETDRS (Early Treatment Diabetic Retinopathy Study) reading charts covering international standards.13 The logMAR values of the chart designed for 35 cm distance were corrected with the formula "log10(standard distance/new distance [cm])" according to the distance used (40 and 60 cm).
Measured: Postoperative 3rd month
The visual acuity
monocular and binocular uncorrected and corrected distance (UDVA, CDVA; at 6 m), intermediate (UIVA, CIVA; at 60 cm), and near (UNVA, CNVA; at 40 cm) logMAR visual acuity were assessed with electronic shelled chart.Distance visual acuity was measured with LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Snellen Charts under photopic conditions (85 cd/m2). Intermediate and near visual acuities were measured and recorded with a Turkish reading chart prepared with reference to Bailey-Lovie and ETDRS (Early Treatment Diabetic Retinopathy Study) reading charts covering international standards.13 The logMAR values of the chart designed for 35 cm distance were corrected with the formula "log10(standard distance/new distance [cm])" according to the distance used (40 and 60 cm).
Measured: Postoperative 6th month
The contrast sensitivity (CS),
Contrast sensitivity (CS) was evaluated at 3, 6, 12 and 18 cycle per degree (cpd) spatial frequencies with CSV-1000 E test (VectorVision, Ohio, USA) under mesopic (3.5 candela [cd]/m2) conditions at 2.5 m with distance refractive correction. The test was conducted with and without glare, and as a glare source, two halogen lamps located either side of the test chart and producing 2.5 cd/m2 luminance at eye plane were used. CS at different spatial frequencies was compared with the physiologic CS range for normal subjects of similar age. CS was also evaluated with the LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Pelli-Robson Test (PRT) under photopic (85 cd/m2) and mesopic (3.5 cd/m2) conditions at 3 m monocularly with distance refractive correction. To the best of our knowledge, PRT CS normal values, performed under similar conditions, are not available in the literature, so the data was used only for intergroup comparisons.
Measured: Preoperative
The contrast sensitivity (CS),
Contrast sensitivity (CS) was evaluated at 3, 6, 12 and 18 cycle per degree (cpd) spatial frequencies with CSV-1000 E test (VectorVision, Ohio, USA) under mesopic (3.5 candela [cd]/m2) conditions at 2.5 m with distance refractive correction. The test was conducted with and without glare, and as a glare source, two halogen lamps located either side of the test chart and producing 2.5 cd/m2 luminance at eye plane were used. CS at different spatial frequencies was compared with the physiologic CS range for normal subjects of similar age. CS was also evaluated with the LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Pelli-Robson Test (PRT) under photopic (85 cd/m2) and mesopic (3.5 cd/m2) conditions at 3 m monocularly with distance refractive correction. To the best of our knowledge, PRT CS normal values, performed under similar conditions, are not available in the literature, so the data was used only for intergroup comparisons.
Measured: Postoperative 1st month
The contrast sensitivity (CS),
Contrast sensitivity (CS) was evaluated at 3, 6, 12 and 18 cycle per degree (cpd) spatial frequencies with CSV-1000 E test (VectorVision, Ohio, USA) under mesopic (3.5 candela [cd]/m2) conditions at 2.5 m with distance refractive correction. The test was conducted with and without glare, and as a glare source, two halogen lamps located either side of the test chart and producing 2.5 cd/m2 luminance at eye plane were used. CS at different spatial frequencies was compared with the physiologic CS range for normal subjects of similar age. CS was also evaluated with the LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Pelli-Robson Test (PRT) under photopic (85 cd/m2) and mesopic (3.5 cd/m2) conditions at 3 m monocularly with distance refractive correction. To the best of our knowledge, PRT CS normal values, performed under similar conditions, are not available in the literature, so the data was used only for intergroup comparisons.
Measured: Postoperative 3rd month
The contrast sensitivity (CS),
Contrast sensitivity (CS) was evaluated at 3, 6, 12 and 18 cycle per degree (cpd) spatial frequencies with CSV-1000 E test (VectorVision, Ohio, USA) under mesopic (3.5 candela [cd]/m2) conditions at 2.5 m with distance refractive correction. The test was conducted with and without glare, and as a glare source, two halogen lamps located either side of the test chart and producing 2.5 cd/m2 luminance at eye plane were used. CS at different spatial frequencies was compared with the physiologic CS range for normal subjects of similar age. CS was also evaluated with the LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Pelli-Robson Test (PRT) under photopic (85 cd/m2) and mesopic (3.5 cd/m2) conditions at 3 m monocularly with distance refractive correction. To the best of our knowledge, PRT CS normal values, performed under similar conditions, are not available in the literature, so the data was used only for intergroup comparisons.
Measured: Postoperative 6th month
Quality of life measurement
Quality of life parameters were evaluated by the validated Turkish version of the National Eye Institute Refractive Error Quality of Life Instrument 42 (NEI RQL-42) questionnaire. It includes 13 subscales of quality of life parameters calculated according to a scoring key, and a "total score" can be obtained by calculating the mean value of all 42 questions.
Measured: Preoperative
Quality of life measurement
Quality of life parameters were evaluated by the validated Turkish version of the National Eye Institute Refractive Error Quality of Life Instrument 42 (NEI RQL-42) questionnaire. It includes 13 subscales of quality of life parameters calculated according to a scoring key, and a "total score" can be obtained by calculating the mean value of all 42 questions.
Measured: Postoperative 1st month
Quality of life measurement
Quality of life parameters were evaluated by the validated Turkish version of the National Eye Institute Refractive Error Quality of Life Instrument 42 (NEI RQL-42) questionnaire. It includes 13 subscales of quality of life parameters calculated according to a scoring key, and a "total score" can be obtained by calculating the mean value of all 42 questions.
Measured: Postoperative 3rd month
Quality of life measurement
Quality of life parameters were evaluated by the validated Turkish version of the National Eye Institute Refractive Error Quality of Life Instrument 42 (NEI RQL-42) questionnaire. It includes 13 subscales of quality of life parameters calculated according to a scoring key, and a "total score" can be obtained by calculating the mean value of all 42 questions.
Measured: Postoperative 6th month
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted29 Apr 2021
- Study start1 Oct 2016
- Primary completion1 Sep 2018
- Study completion1 Mar 2019
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (4)Papers the sponsor has linked to this study.
- Other trials for Multifocal Intraocular LensesStudies of the same condition, open and under way.
- More trials run by Marmara UniversityThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Enrolment
- 29 participants (actual)
Sponsor and investigators
- Lead sponsor
- Marmara University
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
- Investigators
- Ayse Ebru Toker, Professor, Study director, Marmara University
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT04865016
- Sponsor's protocol ID
- 09.2016.309
- Status last verified
- Apr 2021
- First submitted
- 27 Apr 2021
- Last updated
- 29 Apr 2021
- Results posted
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
Publications
Rosen E, Alio JL, Dick HB, Dell S, Slade S. Efficacy and safety of multifocal intraocular lenses following cataract and refractive lens exchange: Metaanalysis of peer-reviewed publications. J Cataract Refract Surg. 2016 Feb;42(2):310-28. doi: 10.1016/j.jcrs.2016.01.014.
PubMed 27026457Kim JS, Jung JW, Lee JM, Seo KY, Kim EK, Kim TI. Clinical Outcomes Following Implantation of Diffractive Multifocal Intraocular Lenses With Varying Add Powers. Am J Ophthalmol. 2015 Oct;160(4):702-9.e1. doi: 10.1016/j.ajo.2015.07.021. Epub 2015 Jul 21.
PubMed 26209232Kretz FT, Gerl M, Gerl R, Muller M, Auffarth GU; ZKB00 Study Group. Clinical evaluation of a new pupil independent diffractive multifocal intraocular lens with a +2.75 D near addition: a European multicentre study. Br J Ophthalmol. 2015 Dec;99(12):1655-9. doi: 10.1136/bjophthalmol-2015-306811. Epub 2015 May 18.
PubMed 25987651Pedrotti E, Bruni E, Bonacci E, Badalamenti R, Mastropasqua R, Marchini G. Comparative Analysis of the Clinical Outcomes With a Monofocal and an Extended Range of Vision Intraocular Lens. J Refract Surg. 2016 Jul 1;32(7):436-42. doi: 10.3928/1081597X-20160428-06.
PubMed 27400074
Study at a glance
From the registry record, last updated by the sponsor on 29 Apr 2021. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Clinical Outcomes With a Low Add Multifocal and an Extended Depth of Focus Intraocular Lenses Both Implanted With Mini-monovision
- ClinicalTrials.gov ID
- NCT04865016
- Acronym / sponsor ID
- 09.2016.309
- Keywords
- low add multifocal intraocular lens, MIOL, ZKB00, extended depth of focus, EDOF, ZXR00, mini-monovision
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 29 participants (actual)
- Lead sponsor
- Marmara UniversityUniversity, hospital or other
- Study start
- 1 Oct 2016 (actual)
- Primary completion
- 1 Sep 2018 (actual)
- Study completion
- 1 Mar 2019 (actual)
- First posted
- 29 Apr 2021 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The innovations in phacoemulsification surgery and advances in surgical equipment ensure a safe extraction of the crystalline lens and implantation of intraocular lenses (IOL). As a result of developing intraocular lens technologies and changing patient demands, many multifocal, accommodative, toric, toric-multifocal, and extended depth of focus (EDOF) IOLs have been introduced and the original aim of restoration of far vision with implanted monocular IOLs has been evolved to an objective of improved vision at different distances with total spectacle independence. This study aimed to compare the visual acuity at different distances, contrast sensitivity, spectacle needs, photic phenomena, and quality of life parameters of patients bilaterally implanted with a low add bifocal (Tecnis ZKB00) or an EDOF IOL (Tecnis Symfony), both with planned mini-monovision to improve near vision.
Detailed description
The innovations in phacoemulsification surgery and advances in surgical equipment ensure a safe extraction of the crystalline lens and implantation of intraocular lenses (IOL). As a result of developing intraocular lens technologies and changing patient demands, many multifocal, accommodative, toric, toric-multifocal, and extended depth of focus (EDOF) IOLs have been introduced and the original aim of restoration of far vision with implanted monocular IOLs has been evolved to an objective of improved vision at different distances with total spectacle independence. This study aimed to compare the visual acuity at different distances, contrast sensitivity, spectacle needs, photic phenomena, and quality of life parameters of patients bilaterally implanted with a low add bifocal (Tecnis ZKB00) or an EDOF IOL (Tecnis Symfony), both with planned mini-monovision to improve near vision. In this regard, binocular logMAR uncorrected visual acuities (UVA), monocular defocus curves, CS with CSV 1000-E and Pelli-Robson Test (PRT), spectacle needs and quality of life parameters with NEI RQL-42 questionnaire were evaluated at postoperative 1, 3, and 6 months.
Study design
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| MIOL Group | Active comparator | Device: Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00)30 eyes of 15 cataract patients undergoing cataract surgery were implanted with a low add (+2.75 Diopters [D]) bifocal (Tecnis ZKB00; Johnson and Johnson Surgical Vision Inc., Jacksonville, Florida, USA) IOL (MIOL Group). |
| EDOF Group | Active comparator | Device: Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00)30 eyes of 15 cataract patients undergoing cataract surgery were implanted with an EDOF (Tecnis Symfony [ZXR00]; Johnson and Johnson Surgical Vision Inc., Jacksonville, Florida, USA) IOL (EDOF Group). |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (1 listed)
- Inclusion criteria were patients with bilateral visually significant cataract seeking spectacle independence, especially at far and intermediate distances, age of ≥18 years, preoperative corneal astigmatism less than 1.00 D, axial length (AL) between 21 - 26.5 mm, and availability to attend all follow-up visits.
Key exclusion criteria (1 listed)
- Patients were excluded from the study if any of the following conditions were present: high visual demand for near vision, amblyopia, glaucoma, corneal opacity, severe/moderate to severe dry eye, pupillary abnormalities and dysfunction, corneal astigmatism ≥1.00 D, AL <21 mm or ≥26.5 mm, diabetes mellitus with retinal changes, active uveitis or history of uveitis, previous ocular surgery and lack of attendance to any follow-up visit.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00)This study included 30 eyes of 15 cataract patients undergoing cataract surgery with implantation of a low add (+2.75 Diopters [D]) bifocal (Tecnis ZKB00; Johnson and Johnson Surgical Vision Inc., Jacksonville, Florida, USA) IOL (MIOL Group) and 30 eyes of 15 patients with implantation of an EDOF (Tecnis Symfony [ZXR00]; Johnson and Johnson Surgical Vision Inc., Jacksonville, Florida, USA) IOL (EDOF Group). | Device | EDOF Group, MIOL Group |
Primary outcome
The visual acuity
Time frame: Preoperative
monocular and binocular uncorrected and corrected distance (UDVA, CDVA; at 6 m), intermediate (UIVA, CIVA; at 60 cm), and near (UNVA, CNVA; at 40 cm) logMAR visual acuity were assessed with electronic shelled chart.Distance visual acuity was measured with LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Snellen Charts under photopic conditions (85 cd/m2). Intermediate and near visual acuities were measured and recorded with a Turkish reading chart prepared with reference to Bailey-Lovie and ETDRS (Early Treatment Diabetic Retinopathy Study) reading charts covering international standards.13 The logMAR values of the chart designed for 35 cm distance were corrected with the formula "log10(standard distance/new distance [cm])" according to the distance used (40 and 60 cm).
The visual acuity
Time frame: Postoperative 1st month
monocular and binocular uncorrected and corrected distance (UDVA, CDVA; at 6 m), intermediate (UIVA, CIVA; at 60 cm), and near (UNVA, CNVA; at 40 cm) logMAR visual acuity were assessed with electronic shelled chart.Distance visual acuity was measured with LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Snellen Charts under photopic conditions (85 cd/m2). Intermediate and near visual acuities were measured and recorded with a Turkish reading chart prepared with reference to Bailey-Lovie and ETDRS (Early Treatment Diabetic Retinopathy Study) reading charts covering international standards.13 The logMAR values of the chart designed for 35 cm distance were corrected with the formula "log10(standard distance/new distance [cm])" according to the distance used (40 and 60 cm).
The visual acuity
Time frame: Postoperative 3rd month
monocular and binocular uncorrected and corrected distance (UDVA, CDVA; at 6 m), intermediate (UIVA, CIVA; at 60 cm), and near (UNVA, CNVA; at 40 cm) logMAR visual acuity were assessed with electronic shelled chart.Distance visual acuity was measured with LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Snellen Charts under photopic conditions (85 cd/m2). Intermediate and near visual acuities were measured and recorded with a Turkish reading chart prepared with reference to Bailey-Lovie and ETDRS (Early Treatment Diabetic Retinopathy Study) reading charts covering international standards.13 The logMAR values of the chart designed for 35 cm distance were corrected with the formula "log10(standard distance/new distance [cm])" according to the distance used (40 and 60 cm).
The visual acuity
Time frame: Postoperative 6th month
monocular and binocular uncorrected and corrected distance (UDVA, CDVA; at 6 m), intermediate (UIVA, CIVA; at 60 cm), and near (UNVA, CNVA; at 40 cm) logMAR visual acuity were assessed with electronic shelled chart.Distance visual acuity was measured with LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Snellen Charts under photopic conditions (85 cd/m2). Intermediate and near visual acuities were measured and recorded with a Turkish reading chart prepared with reference to Bailey-Lovie and ETDRS (Early Treatment Diabetic Retinopathy Study) reading charts covering international standards.13 The logMAR values of the chart designed for 35 cm distance were corrected with the formula "log10(standard distance/new distance [cm])" according to the distance used (40 and 60 cm).
The contrast sensitivity (CS),
Time frame: Preoperative
Contrast sensitivity (CS) was evaluated at 3, 6, 12 and 18 cycle per degree (cpd) spatial frequencies with CSV-1000 E test (VectorVision, Ohio, USA) under mesopic (3.5 candela [cd]/m2) conditions at 2.5 m with distance refractive correction. The test was conducted with and without glare, and as a glare source, two halogen lamps located either side of the test chart and producing 2.5 cd/m2 luminance at eye plane were used. CS at different spatial frequencies was compared with the physiologic CS range for normal subjects of similar age. CS was also evaluated with the LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Pelli-Robson Test (PRT) under photopic (85 cd/m2) and mesopic (3.5 cd/m2) conditions at 3 m monocularly with distance refractive correction. To the best of our knowledge, PRT CS normal values, performed under similar conditions, are not available in the literature, so the data was used only for intergroup comparisons.
The contrast sensitivity (CS),
Time frame: Postoperative 1st month
Contrast sensitivity (CS) was evaluated at 3, 6, 12 and 18 cycle per degree (cpd) spatial frequencies with CSV-1000 E test (VectorVision, Ohio, USA) under mesopic (3.5 candela [cd]/m2) conditions at 2.5 m with distance refractive correction. The test was conducted with and without glare, and as a glare source, two halogen lamps located either side of the test chart and producing 2.5 cd/m2 luminance at eye plane were used. CS at different spatial frequencies was compared with the physiologic CS range for normal subjects of similar age. CS was also evaluated with the LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Pelli-Robson Test (PRT) under photopic (85 cd/m2) and mesopic (3.5 cd/m2) conditions at 3 m monocularly with distance refractive correction. To the best of our knowledge, PRT CS normal values, performed under similar conditions, are not available in the literature, so the data was used only for intergroup comparisons.
The contrast sensitivity (CS),
Time frame: Postoperative 3rd month
Contrast sensitivity (CS) was evaluated at 3, 6, 12 and 18 cycle per degree (cpd) spatial frequencies with CSV-1000 E test (VectorVision, Ohio, USA) under mesopic (3.5 candela [cd]/m2) conditions at 2.5 m with distance refractive correction. The test was conducted with and without glare, and as a glare source, two halogen lamps located either side of the test chart and producing 2.5 cd/m2 luminance at eye plane were used. CS at different spatial frequencies was compared with the physiologic CS range for normal subjects of similar age. CS was also evaluated with the LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Pelli-Robson Test (PRT) under photopic (85 cd/m2) and mesopic (3.5 cd/m2) conditions at 3 m monocularly with distance refractive correction. To the best of our knowledge, PRT CS normal values, performed under similar conditions, are not available in the literature, so the data was used only for intergroup comparisons.
The contrast sensitivity (CS),
Time frame: Postoperative 6th month
Contrast sensitivity (CS) was evaluated at 3, 6, 12 and 18 cycle per degree (cpd) spatial frequencies with CSV-1000 E test (VectorVision, Ohio, USA) under mesopic (3.5 candela [cd]/m2) conditions at 2.5 m with distance refractive correction. The test was conducted with and without glare, and as a glare source, two halogen lamps located either side of the test chart and producing 2.5 cd/m2 luminance at eye plane were used. CS at different spatial frequencies was compared with the physiologic CS range for normal subjects of similar age. CS was also evaluated with the LCD screen version (CSO Vision Chart, Mod CVC02, version 1.3.0, Florence, Italy) of Pelli-Robson Test (PRT) under photopic (85 cd/m2) and mesopic (3.5 cd/m2) conditions at 3 m monocularly with distance refractive correction. To the best of our knowledge, PRT CS normal values, performed under similar conditions, are not available in the literature, so the data was used only for intergroup comparisons.
Quality of life measurement
Time frame: Preoperative
Quality of life parameters were evaluated by the validated Turkish version of the National Eye Institute Refractive Error Quality of Life Instrument 42 (NEI RQL-42) questionnaire. It includes 13 subscales of quality of life parameters calculated according to a scoring key, and a "total score" can be obtained by calculating the mean value of all 42 questions.
Quality of life measurement
Time frame: Postoperative 1st month
Quality of life parameters were evaluated by the validated Turkish version of the National Eye Institute Refractive Error Quality of Life Instrument 42 (NEI RQL-42) questionnaire. It includes 13 subscales of quality of life parameters calculated according to a scoring key, and a "total score" can be obtained by calculating the mean value of all 42 questions.
Quality of life measurement
Time frame: Postoperative 3rd month
Quality of life parameters were evaluated by the validated Turkish version of the National Eye Institute Refractive Error Quality of Life Instrument 42 (NEI RQL-42) questionnaire. It includes 13 subscales of quality of life parameters calculated according to a scoring key, and a "total score" can be obtained by calculating the mean value of all 42 questions.
Quality of life measurement
Time frame: Postoperative 6th month
Quality of life parameters were evaluated by the validated Turkish version of the National Eye Institute Refractive Error Quality of Life Instrument 42 (NEI RQL-42) questionnaire. It includes 13 subscales of quality of life parameters calculated according to a scoring key, and a "total score" can be obtained by calculating the mean value of all 42 questions.
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 29 Apr 2021.
- Background: PubMed 27026457Rosen E, Alio JL, Dick HB, Dell S, Slade S. Efficacy and safety of multifocal intraocular lenses following cataract and refractive lens exchange: Metaanalysis of peer-reviewed publications. J Cataract Refract Surg. 2016 Feb;42(2):310-28. doi: 10.1016/j.jcrs.2016.01.014.
- Background: PubMed 26209232Kim JS, Jung JW, Lee JM, Seo KY, Kim EK, Kim TI. Clinical Outcomes Following Implantation of Diffractive Multifocal Intraocular Lenses With Varying Add Powers. Am J Ophthalmol. 2015 Oct;160(4):702-9.e1. doi: 10.1016/j.ajo.2015.07.021. Epub 2015 Jul 21.
- Background: PubMed 25987651Kretz FT, Gerl M, Gerl R, Muller M, Auffarth GU; ZKB00 Study Group. Clinical evaluation of a new pupil independent diffractive multifocal intraocular lens with a +2.75 D near addition: a European multicentre study. Br J Ophthalmol. 2015 Dec;99(12):1655-9. doi: 10.1136/bjophthalmol-2015-306811. Epub 2015 May 18.
- Background: PubMed 27400074Pedrotti E, Bruni E, Bonacci E, Badalamenti R, Mastropasqua R, Marchini G. Comparative Analysis of the Clinical Outcomes With a Monofocal and an Extended Range of Vision Intraocular Lens. J Refract Surg. 2016 Jul 1;32(7):436-42. doi: 10.3928/1081597X-20160428-06.
Related trials for Multifocal Intraocular Lenses
Active, not recruiting
Clinical Trial With Artiflex Presbyopic (Artiplus)
Presbyopia, Ametropia
RecruitingObservational study
Visual Adaptation and Quality of Life After Different IOL Implantation Strategies
Cataract, Aphakia - Lens Capsule Present
Recruiting
Multicentric Clinical Investigation to Assess Safety and Performance of Lux HighAdd IOL
Cataract
Enrolling by invitation
Visual Outcomes in Patients Contralaterally Implanted With PanOptix Pro and Clareon Vivity Compared to Bilateral Implantation of PanOptix Pro
Cataract, Multifocal Intraocular Lens
Recruiting
Visual Performance of Clareon PanOptix vs. a Comparable Multifocal Intraocular Lens
Age Related Cataracts
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Ayse Ebru Toker, Professor study director
Marmara University
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Tecnis +2.75 D (ZKB00) and Tecnis Symfony (ZXR00) in people with Multifocal Intraocular Lenses and Extended Depth of Focus, run by Marmara University at 1 site in Turkey (Türkiye).
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 1 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 29 Apr 2021. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Turkey (Türkiye). Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
The visual acuity (time frame: Preoperative). The study lists 12 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT04865016, which also lists 4 linked publications.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 29 Apr 2021.