Infant, PrematureMODERATION NeoNCT05030012

Maintaining Optimal HVNI Delivery Using Automatic Titration of Oxygen in Preterm Infants

A study of Infant, Premature and Neonatal Respiratory Distress, run by Vapotherm, Inc. at 3 sites in United States.

Terminated

Stopped early. The study ended before it was planned to and will not resume.

Official record
Search all trials

Trial summary

Oxygen treatment is common in management of preterm babies requiring intensive care. Delivery of too much or too little oxygen increase the risk of damage to eyes and lungs, and contributes to death and disability. Read more about this study

Who is it for?
Any age
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
15 participants
Where
3 sitesUnited States
Study dates
2 Sep 2021to 19 Oct 2023
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Any age
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Infants delivered at a gestational age of less than or equal to 35 6/7 weeks (Preterm) being treated with high velocity nasal insufflation therapy for the management of respiratory distress syndrome
  • Patients that clinically require SpO2 maintenance within the target range of 90-95%
  • A need for supplemental oxygen as demonstrated by a required FiO2 > 0.25 at enrollment
  • Requiring a flow rate of greater than 2 L•min-1 such that the assumed inspired oxygen fraction matched delivered oxygen fraction (definition of HVNI).
  • A minimum of 6 manual FiO2 adjustments in the 24hr period prior to trial enrollment.
  • Willing/Able to complete informed consent.

Who cannot join

  • Current patient weight of <1000g or >3500g at time of study
  • Major congenital abnormalities
  • Hemodynamic instability, defined as being outside of a normotensive range based on an infant's individual characteristics by clinician
  • Persistent unresolved apnea defined as: requiring 6 stimulations or more per 6 hours
  • Seizures
  • Ongoing sepsis
  • Meningitis
  • Clinician's concern regarding stability of the infant

Trial locations and contacts

3 sites in 1 country.

United States3 sites
  • University of Utah Hospital

    Salt Lake City, Utah, 84132

  • Seattle Children's Hospital

    Seattle, Washington, 98195

  • Children's National Hospital & Research Institute

    Washington D.C., District of Columbia, 20010

Other trials in United States

About this study

Official title: Maintaining Optimal Delivery Using Automatic Titration of Oxygen in Preterm Infants Receiving High Velocity Nasal Insufflation Therapy

Oxygen treatment is common in management of preterm babies requiring intensive care. Delivery of too much or too little oxygen increase the risk of damage to eyes and lungs, and contributes to death and disability. Oxygen control in preterm infants requires frequent adjustments in the amount of oxygen delivered to the baby. This is generally performed manually by a clinician attending the baby, and generally directed to maintaining a specific range of blood oxygen saturation. The manual control often results in only half of the time in the specified range, with the baby experiencing high and low blood oxygen saturations.

The technology being studied is designed to assist the clinician in maintaining blood oxygen saturation within target range by measuring oxygen saturation and automatically adjusting the amount of oxygen delivered for babies receiving high velocity nasal insufflation (an advanced form of high flow oxygen therapy). The proposed study will evaluate the efficacy and safety of the automatic control of oxygen by the new technology, as compared to manual control, among babies receiving high velocity therapy in a neonatal intensive care unit.

Read the detailed description

Detailed Description: Supplemental oxygen is commonly administered to babies in neonatal intensive care units. The goal of oxygen therapy is to maintain normal oxygenation while minimizing hypoxemia and hypoxemia. Preterm infants are particularly vulnerable to oxygen toxicity and oxidative stress leading to retinopathy of prematurity (ROP), bronchopulmonary dysplasia (BPD), and periventricular leukomalacia (PVL). It's also well known that preterm infants experience hypoxic events exposing the baby to low oxygen levels. These hypoxic events vary as the infant matures, but exposure to prolonged and frequent episodes of hypoxemia is associated with increased morbidity and mortality. The delivery of oxygen is generally controlled by a clinician, and the control decisions are generally made using a non-invasive measure of blood oxygen saturation called pulse oximetry (SpO2) and is a standard of care in the neonatal intensive care unit. The most frequent item adjusted by clinicians to maintain SpO2 within specific target ranges is the fraction of the inspired oxygen (FiO2).

In a recent study by Reynolds, et al., caregiver manual control of oxygen delivered to NICU babies receiving high velocity therapy resulted in only 49% of the total 24-hour study period with the babies within the target SpO2 range (90-95%). by the caregiver based on the monitored oxygen saturation. Similar to the Reynolds findings, Hagadorn et al., conducted a study in 14 centers and showed that preterm infants under 28 weeks' gestation receiving oxygen spent on average only 48% of the time with SpO2 within the prescribed target range, about 36% of the time above and 16% of the time with SpO2 below the target range.

Preterm infants have frequent fluctuations in SpO2 due to their cardio-respiratory instability requiring frequent adjustments of FiO2 . Consequently, these particularly vulnerable infants spend significant time with SpO2 outside the optimal target range and are often exposed to extremes of hypoxemia and hyperoxaemia. The automatic oxygen control system continuously monitors the oxygen saturation and adjusts the oxygen delivery to maintain oxygen saturation within the target range. The efficacy of this mode of oxygen control was demonstrated by Reynolds, et al. in 2018 from two centers in the United Kingdom. Automated control of FiO2 can significantly improve compliance of oxygen saturation targeting and may significantly reduces exposure to hypoxemia as well as hyperoxaemia. The high velocity nasal insufflation therapy is a common mode of non-invasive respiratory support in preterm infants.

Unlike prior studies, this study will include a set of hypothesis-driven safety endpoints (proportion of time above or below target range), stratification by body mass at enrolment, and skin pigmentation phenotype. The objective of this randomized control trial is to evaluate the efficacy of the controller (Vapotherm Oxygen Assist Module [OAM]) in maintaining the SpO2 within target range for premature infants receiving high velocity therapy and presenting with a labile FiO2 requirement.

Being tested
Automated Control, Manual Control
Keywords
High Velocity Nasal Insufflation (HVNI), Preterm Infants, Noninvasive Ventilatory Support, Automatic Oxygen Titration, Closed Loop Oxygen Control, High Flow Nasal Cannula
Medical subject headings
Premature Birth, Pulmonary Atelectasis, Bronchopulmonary Dysplasia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Automated ControlDevice

    Also called Oxygen Assist Module (OAM), Vapotherm OAM + Vapotherm Precision Flow

    The purpose of this intervention is to evaluate that the efficacy and safety during the automated performance of the Vapotherm OAM are not inferior to standard care practice (manual control) in maintaining SpO2 levels of 90-95% in preterm infants requiring oxygen adjustments while being treated with high velocity nasal insufflation (HVNI) therapy.

  • Manual ControlDevice

    Also called Vapotherm Precision Flow, Treatment as Usual (TAU), Standard of Care (SOC)

    The purpose of this intervention is to establish an active comparator via standard care practice against which to evaluate the efficacy and safety of the automated arm of the study, for determination of non-inferiority of that automated control arm to manual control in maintaining SpO2 levels of 90-95% in preterm infants being treated with high velocity nasal insufflation (HVNI) therapy.

Study groups in NCT05030012
Study groupRoleWhat they receive
Automated Control (OAM)ExperimentalIn this arm, FiO2 levels delivered via high-velocity nasal insufflation therapy (Vapotherm Precision Flow) will be adjusted by the Oxygen Assist Module (OAM) to keep the infants pulse oxygen saturation within a target range (90-95%). Clinical staff will have the ability to override FiO2 levels when required, and instructed to do so.
Receives: Device: Automated Control
Manual Control (Manual)Active comparatorIn this arm, FiO2 levels delivered via high-velocity nasal insufflation therapy (Vapotherm Precision Flow) will be manually adjusted by clinical staff to keep infants' oxygen saturation between 90-95%.
Receives: Device: Manual Control

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Primary Safety Objective - Percentage of Time Outside of SpO2 Target Range

    Percentage of time spent outside target oxygen saturation range, measured by pulse oximetry (SpO2). A lower value indicates a better outcome.

    Measured: Through study completion, two consecutive 24-hour periods

  2. Primary Performance Objective - Percentage of Time Within SpO2 Target Range

    Percentage of time spent within target oxygen saturation range, measured by pulse oximetry (SpO2). A higher value indicates a better outcome.

    Measured: Through study completion, two consecutive 24-hour periods

Secondary outcomes (2)
  1. Performance Endpoint, Percentage of Time IN Range (SpO2 in 90-95% With FiO2 Special at .21) Across the Two Weight Groups.

    Percentage of time spent within target oxygen saturation range across two weight groups (1000g-2500g, 2501g-3500g), measured by pulse oximetry (SpO2). A higher value indicates a better outcome.

    Measured: Through study completion, two consecutive 24-hour periods

  2. Secondary Performance Objective 2 - Percentage of Time Within SpO2 Target Range (Skin Pigmentation)

    Percentage of time spent within target oxygen saturation range across two skin pigmentation groups (light, dark), measured by pulse oximetry (SpO2). A higher value indicates a better outcome.

    Measured: Through study completion, two consecutive 24-hour periods

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted1 Sep 2021
  2. Study start2 Sep 2021
  3. Primary completion19 Oct 2023
  4. Study completion19 Oct 2023

Why it stopped: No path to obtain FDA approval for the legacy Precision Flow OAM.

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Crossover
Model description
This study will be a prospective, multi-center, randomized, crossover trial evaluating automated oxygen titration versus manual titration in maintaining oxygen saturation levels in preterm infants requiring noninvasive ventilatory support via high velocity nasal insufflation (HVNI).
Masking
None (open label)
Enrolment
15 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05030012
Sponsor's protocol ID
RP-NIOA2019001Reg
Acronym
MODERATION Neo
Status last verified
Jun 2024
First submitted
25 Aug 2021
Last updated
26 Jun 2024
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Publications

  • Reynolds PR, Miller TL, Volakis LI, Holland N, Dungan GC, Roehr CC, Ives K. Randomised cross-over study of automated oxygen control for preterm infants receiving nasal high flow. Arch Dis Child Fetal Neonatal Ed. 2019 Jul;104(4):F366-F371. doi: 10.1136/archdischild-2018-315342. Epub 2018 Nov 21.

    PubMed 30464005
  • Hagadorn JI, Furey AM, Nghiem TH, Schmid CH, Phelps DL, Pillers DA, Cole CH; AVIOx Study Group. Achieved versus intended pulse oximeter saturation in infants born less than 28 weeks' gestation: the AVIOx study. Pediatrics. 2006 Oct;118(4):1574-82. doi: 10.1542/peds.2005-0413.

    PubMed 17015549

Questions about this trial

What is this clinical trial studying?

A study of Infant, Premature and Neonatal Respiratory Distress, run by Vapotherm, Inc. at 3 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: stopped early, as of the sponsor's last update on 26 Jun 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 3 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 26 Jun 2024). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Primary Safety Objective - Percentage of Time Outside of SpO2 Target Range (time frame: Through study completion, two consecutive 24-hour periods). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT05030012, which also lists 2 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.