10-year Follow-up of a Wide Diameter Bone Anchored Hearing Implant

A study testing Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm abutment, Baha® BIA300 Implant (diameter 4.5mm; length 4mm) with 6mm abutment and Early loading in people with Hearing Loss, run by Radboud University Medical Center at 1 site in Netherlands.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

A single centre, open, comparative, parallel group, prospective clinical investigation with a single 10 year follow up visit. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Hearing Loss
What is being tested?
Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm abutment, Baha® BIA300 Implant (diameter 4.5mm; length 4mm) with 6mm abutment and Early loading
Could I receive a placebo?
No, everyone receives an active treatment
How many people
51 participants
Where
1 siteNetherlands
Study dates
24 Jun 2019to 21 Sep 2020
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • To be included in the original trials, patients had to be at least 18 years old.

Who cannot join

  • Exclusion criteria were an inability to follow investigational procedures, any factor at the discretion of the investigator that was considered to contraindicate participation, and any disease or treatment known to compromise the bone quality at the implant site. For patients who had lost or removed the implant, the time to implant loss was recorded. Patients who could not attend the 5 or 10 year follow-up visit for other reasons were included in the implant survival analysis as well; the last available information regarding implant survival was obtained verbally from the patient, from medical records, or from information captured from the original investigations. Furthermore, patients were excluded if they had a bone thickness less than 4 mm at the implant site during implantation.

Trial locations and contacts

1 site in 1 country.

Netherlands1 site
  • Radboudumc

    Nijmegen, Gelderland, 6525EX

Other trials in Netherlands

About this study

Official title: Long-term Follow-up of a Wide Diameter Bone Anchored Hearing Implant: the 10-year Experience on Stability, Survival and Tolerability of the BIA300®

A single centre, open, comparative, parallel group, prospective clinical investigation with a single 10 year follow up visit.

Read the detailed description

This study is a continuation of two previous conducted trials:

  1. Stability, long term survival and tolerability of a novel Baha® implant system, a multi-centre investigation new BIA300 (test) implant vs previous generation (control) implant:
  2. CAG5173: 6-month(1) and 3-year(2) data.
  3. CBAS5562: 5-year data(3).
  4. 3 weeks loading: stability, long term survival and tolerability of a novel Baha® implant system, a multi-centre investigation 3 week loading vs 6+ week loading (= same group as control group trial 1). 6- month(4) and 3-year(5) data.

In this study the (5-year data for trial 2 and) 10-year data regarding stability, survival and soft tissue tolerability from the above mentioned trials will be evaluated and compared. The previous trials were multicentre, but the current study will be single-centred and therefore only include the participants of the Radboud site. Data will be collected at a visit at (5 years, for trial 2 and) 10 years (+/- 6 months) after implantation.

Up to now, no RCT of these implants (with the loading of 3 weeks after implantation) has assessed stability, survival and tolerability over a 10-year follow-up period.

Conditions
Hearing Loss
Being tested
Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm abutment, Baha® BIA300 Implant (diameter 4.5mm; length 4mm) with 6mm abutment, Early loading
Medical subject headings
Hearing Loss

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm abutmentDevice

    The control implant included the previous generation as-machined titanium Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm conically shaped abutment.

  • Baha® BIA300 Implant (diameter 4.5mm; length 4mm) with 6mm abutmentDevice

    The test implant and the implant used for the 3-week loading trial included the wide diameter titanium implant (diameter 4.5mm; length 4mm) with 6mm rounded, apically converging titanium abutment developed by Cochlear Bone Anchored Solutions AB (Mölnlyncke, Sweden). This system with an additional minor change to the internal abutment connection design was later commercialized under the name Cochlear™ Baha® BIA300 Implant with abutment. In addition to the difference in abutment shape and the wider diameter, the test implant incorporates small-sized threads at the implant neck and the moderately rough TiOblast™ (Dentsply, Mölndal, Sweden) surface on the intraosseous part of the implant6-8.

  • Early loadingProcedure or surgery

    Loading the implant at 3 weeks after implantation (instead of 6 weeks after implantation)

Study groups in NCT05058066
Study groupRoleWhat they receive
Control implant loaded at 6 weeks post-surgeryActive comparatorThe control implant included the previous generation as-machined titanium Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm conically shaped abutment.
Receives: Device: Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm abutment
Test implant loaded at 6 weeks post-surgeryExperimentalThe test implant and the implant used for the 3-week loading trial included the wide diameter titanium implant (diameter 4.5mm; length 4mm) with 6mm rounded, apically converging titanium abutment developed by Cochlear Bone Anchored Solutions AB (Mölnlyncke, Sweden). This system with an additional minor change to the internal abutment connection design was later commercialized under the name Cochlear™ Baha® BIA300 Implant with abutment.
Receives: Device: Baha® BIA300 Implant (diameter 4.5mm; length 4mm) with 6mm abutment
Test implant loaded at 3 weeks post-surgeryExperimentalThe test implant and the implant used for the 3-week loading trial included the wide diameter titanium implant (diameter 4.5mm; length 4mm) with 6mm rounded, apically converging titanium abutment developed by Cochlear Bone Anchored Solutions AB (Mölnlyncke, Sweden). This system with an additional minor change to the internal abutment connection design was later commercialized under the name Cochlear™ Baha® BIA300 Implant with abutment.
Receives: Device: Baha® BIA300 Implant (diameter 4.5mm; length 4mm) with 6mm abutment; Procedure: Early loading

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. implant stability

    Implant Stability Quotiënt (ISQ) will be measured at the 10y-visit. The resonance frequency analysis renders an ISQ value ranging from 1 to 100. Measurements shall be performed at the abutment level, as is done in the previous investigations. The highest and lowest ISQ value out of two perpendicular measurements shall be recorded as ISQ High and ISQ low. The measurement should preferably be done by the same Osstell instrument (Osstell, Gothenburg, Sweden) that was used in the previous completed study. SmartPeg Type 43 and 55 shall be used during the measurement (Osstell, Gothenburg, Sweden). The procedures for obtaining ISQ values is described in the manufacturer's instruction that will be handed out to the investigators.

    Measured: 10 years after implantation

Secondary outcomes (3)
  1. implant survival

    All patients will be asked if they have experienced any implant osseointegration problems. The time from implantation until implant loss or removal will be noted. In case of implant removal, reason for removal shall be recorded.

    Measured: 10 years after implantation

  2. soft tissue reaction

    Soft tissue reactions will be measured using the Holgers' classification \& IPS-score.

    The Holgers' classification is designed to capture signs and symptoms of inflammation or infection at the site of implantation (scale 0-4).

    The IPS-score is a complete standardized assessment scale for skin complications for both percutaneous and transcutaneous implants for bone conduction devices (BCDs) with a proposed treatment advice based on the outcome. The IPS score has three subscales: inflammation (0-4); pain (0-2), and skin height (0-2).

    Measured: 10 years after implantation

  3. device use

    To evaluate the long-term device use of a Baha®.

    Measured: 10 years after surgery

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted27 Sep 2021
  2. Study start24 Jun 2019
  3. Primary completion21 Sep 2020
  4. Study completion21 Sep 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Non-randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
51 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05058066
Sponsor's protocol ID
CAG5173/CBAS5562 10y
Status last verified
Sep 2021
First submitted
7 Sep 2021
Last updated
18 Nov 2025
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Publications

  • Teunissen EM, Caspers CJI, Vijverberg MA, Mylanus EAM, Hol MKS. Long-Term Follow-up of a Wide-Diameter Bone-Anchored Hearing Implant: 10-Year Experience on Stability, Survival, and Tolerability of an Implant-Abutment Combination. Otol Neurotol. 2023 Jan 1;44(1):40-46. doi: 10.1097/MAO.0000000000003763. Epub 2022 Nov 23.

    PubMed 36417764

Questions about this trial

What is this clinical trial studying?

A study testing Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm abutment, Baha® BIA300 Implant (diameter 4.5mm; length 4mm) with 6mm abutment and Early loading in people with Hearing Loss, run by Radboud University Medical Center at 1 site in Netherlands.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 18 Nov 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Netherlands. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

implant stability (time frame: 10 years after implantation). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT05058066, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 18 Nov 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in