Clinical Trial on a Novel Blood Pumping System in 4008A Hemodialysis Machine

A study testing Convergence Dialyzer in people with Renal Dialysis, run by D.med Consulting GmbH at 1 site in North Macedonia.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The aim of this clinical trial will be to verify the feasibility of using an impeller system instead of a roller pump for hemodialysis, using the established 4008 hemodialysis system and a Convergence Dialyzer with design changes made to the FX-Coral P 600 dialyzer. Read more about this study

Who is it for?
Ages 18 to 85 years
Conditions the study is for
Renal Dialysis
What is being tested?
Convergence Dialyzer
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
8 participants
Where
1 siteNorth Macedonia
Study dates
19 Jan 2022to 22 Feb 2022
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 85 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patients must be up to 18-85 years
  • Hemodialysis Schedule: 3 times per week; app. 3- 5 hour treatment
  • Stable clinical conditions: no recent (last 2 weeks) hospitalization events, no recent main surgery (last 3 weeks), no acute cardiological problems, hemoglobine levels in the DOQI Guideline ranges.
  • Hemodynamic stability during dialysis i.c. number of sessions complicated by acute hypotension events or arrhythmia <25% in the last 3 Months
  • Well-functioning AVF vascular access: access recirculation ≤10%
  • No coagulation disorders and anticoagulant therapy
  • Patients must be using a similar size dialyzer as Convergence dialyzer
  • Signed Inform consent form.

Who cannot join

  • Unstable clinical condition: recent hospitalization, recent surgery, anemia, active neoplastic diseases,shock or unstable cardiac function and intolerable to extracorporeal treatment
  • Hemodynamic instability during the dialysis sessions; more than 25% of the sessions complicated by acute hypotension or arrhythmic events.Presence of clinically significant abnormality on a 12-lead ECG at Screening that, at the investigator's clinical judgment, may compromise the safety of the patient or affect the outcome of the study.
  • History or presence of congestive heart failure with symptoms consistent with New York Heart Association Class III or IV functional status within 6 months prior to the study Convergence system treatment.
  • Abnormal liver function values including but not limited to the aspartate transaminase, alanine transaminase or alkaline phosphatase ≥5 × upper limit normal.
  • Any uncontrolled clinically significant respiratory disease in the investigator's opinion (e.g., chronic obstructive pulmonary disease, cystic fibrosis, bronchiectasis, asthma).
  • Temporary excluded will be patients with increased body temperature due to some infective disease, dehydration, diarrhea, hyperglycemia,
Show all 14Show fewer
  • Severe bleeding, or history or evidence of bleeding diathesis or coagulopathy with the risk of bleeding or have to put on dialysis without anticoagulant.
  • Using temporary or permanent dialysis catheter
  • Uncontrolled diabetes mellitus or hypertension, at the discretion of investigator.
  • Known pregnancy and lactating without pregnancy test
  • Patient has known allergy or hypersensitivity reaction to dialyzer membrane or other disposable
  • Any conditions significantly affecting the nervous system (i.e., neuropathic conditions or nervous system damage
  • Patient shows evidence of a condition (psychological, emotional problems, any disorders or resultant therapy) that is likely to invalidate health information, consent, or limit the ability of the patient to comply with the protocol requirements in the opinion of the investigator
  • Any other conditions at the discretion of investigator not suitable for patients to participate in the study

Trial locations and contacts

1 site in 1 country.

North Macedonia1 site
  • Special Hospital for Nephrology and Dialysis Diamed

    Skopje, 1000

Other trials in North Macedonia

About this study

The aim of this clinical trial will be to verify the feasibility of using an impeller system instead of a roller pump for hemodialysis, using the established 4008 hemodialysis system and a Convergence Dialyzer with design changes made to the FX-Coral P 600 dialyzer.

Conditions
Renal Dialysis
Being tested
Convergence Dialyzer

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Convergence DialyzerDevice

    The Convergence dialyzer with design changes made to FX Coral 600 dialyzer will be tested together with Sidecar. In the dialyzer is integrated centrifugal-flow blood pump.The centrifugal-flow blood pump replaces the peristaltic roller pump of the extracorporeal circuit which pumps blood from patient to the dialyzer and from dialyzer to the patient. The centrifugal-flow blood pump is designed to transport blood from the patient to the dialyzer, and after purification, return the blood to patient.

Study groups in NCT05217095
Study groupRoleWhat they receive
Interventional GroupOtherEach patient will be treated for one session of dialysis after enrolled. Each of them will have only one treatment in the middle of the week (Wednesday or Thursday). The patients will be treated using 4008A dialysis machines in combination with a sidecar and the convergence dialyzer as investigational devices.Each patient will be treated according their regular treatment and laboratory analysis will be taken. The blood level of free hemoglobin ( fHb ) will be taken on 30 min by use of HemoCue device. ACT time will be measured on 15 min and for that 0.5 ml blood from venous line will be taken.
Receives: Device: Convergence Dialyzer

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Incidence of Alarms

    Incidence of alarms leading to end of treatment, cause identification

    Measured: 1 study day

  2. Product defects

    Product defects leading to end of treatment, cause identification

    Measured: 1 study day

  3. Serious adverse event / adverse event/ Serious adverse device effect/ Adverse device effect/Unanticipated serious adverse device effect

    Incidence of Serious adverse event / adverse event/ Serious adverse device effect/ Adverse device effect/Unanticipated serious adverse device effect

    Measured: 1 study day

  4. Incidence of "recoverable" alarms

    Incidence of "recoverable" alarms, cause of alarms i.e.arterial/venous pressure outside of set range/limits

    Measured: 1 study day

Secondary outcomes (17)
  1. Completion Rate

    The completion rate of a dialysis session according to prescription. The study is considered to be successful, if at least 8 patients reach the effective dialysis time.

    Measured: 1 study day

  2. Average Blood Flow Rate

    The average blood flow rate is measured during the dialysis treatment. Maintenance of preset blood flow rates throughout the dialysis in at least 8 out of 10 patients

    Measured: 1 study day

  3. Arterial pressure

    The arterial pressure is measured in mmHg.

    Measured: 1 study day

  4. Venous pressure

    The venous pressure is measured in mmHg.

    Measured: 1 study day

  5. Impeller RPM

    The rotation of the impeller pump is measured in rpm.

    Measured: 1 study day

  6. Dialysis Dose as single pool Kt/V

    The delivered dialysis dose as single-pool Kt/V. Achievement of a spKt/V > 1.4 in at least 8 out of 10 patients.

    Measured: 1 study day

  7. Dialysis Dose as Urea Reduction Ratio (URR)

    The delivered dialysis dose as URR.

    Measured: 1 study day

  8. Ultrafiltration Volume

    Total ultrafiltration volume (ml). Achievement of preset total ultrafiltration volume (ml) to reach dry weight target in at least 8 out of 10 patients.

    Measured: 1 study day

  9. Body weight change

    Patient body weight change (before treatment vs. after treatment).

    Measured: 1 study day

  10. pH

    The blood level of pH is measured pre/post HD. Achievement of results in laboratory acceptable ranges,

    Measured: 1 study day

  11. Bicarbonate

    The blood level of Bicarbonate is measured pre/post HD. Achievement of results in laboratory acceptable ranges.

    Measured: 1 study day

  12. Potassium

    The blood level of Potassium is measured pre/post HD. Achievement of results in laboratory acceptable ranges,

    Measured: 1 study day

  13. Sodium

    The blood level of Sodium is measured pre/post HD. Achievement of results in laboratory acceptable ranges.

    Measured: 1 study day

  14. Chloride

    The blood level of Chloride is measured pre/post HD. Achievement of results in laboratory acceptable ranges.

    Measured: 1 study day

  15. Free hemoglobin

    The blood level of free hemoglobin (fHb measured on 30 min) to be in acceptable ranges for hemolysis

    Measured: 1 study day

  16. Anticoagulation

    The total dose of anticoagulation consumption not to exceed patient treatment dose as I.U per kg

    Measured: 1 study day

  17. Clotting Score

    The clotting score of dialyzer(visual check, using common scale for dialyzer study)

    1. No visible thrombosed fibers
    2. A few visible thrombosed fibers
    3. Moderately visible thrombosed fibers
    4. Many visible thrombosed fibers
    5. Completely visible thrombosed fibers

    So the score goes from 1 (good) to 5 (bad).

    Measured: 1 study day

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted1 Feb 2022
  2. Study start19 Jan 2022
  3. Primary completion22 Jan 2022
  4. Study completion22 Feb 2022

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
8 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05217095
Sponsor's protocol ID
D.Med 20-01 Convergence
Status last verified
Feb 2022
First submitted
13 Dec 2021
Last updated
2 Mar 2022
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Convergence Dialyzer in people with Renal Dialysis, run by D.med Consulting GmbH at 1 site in North Macedonia.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 8 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 2 Mar 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in North Macedonia. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Incidence of Alarms (time frame: 1 study day). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 2 Mar 2022.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in