Sensorineural Hearing Loss, BilateralNCT05230498

Study of Sound and Speech Perception in New Cochlear Implanted Subjects Using or Not an Anatomy-based Fitting

A study testing anatomy-based fitting then default fitting and default fitting then anatomy-based fitting in people with Sensorineural Hearing Loss, Bilateral, run by MED-EL Elektromedizinische Geräte GmbH at 1 site in France.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Main objective: Compare the recognition of environmental sounds with an anatomy-based fitting and with a default fitting adult patients newly implanted with a MED-EL cochlear implant. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Sensorineural Hearing Loss, Bilateral
What is being tested?
anatomy-based fitting then default fitting and default fitting then anatomy-based fitting
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
17 participants
Where
1 siteFrance
Study dates
15 Feb 2022to 15 Nov 2024
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Adult patient (>= 18 years old) speaking French
  • Patient who fulfils the criteria for cochlear implantation

Who cannot join

  • retro-cochlear pathology: auditory neuropathy, vestibular schwannoma

Trial locations and contacts

1 site in 1 country.

France1 site
  • CHU

    Bordeaux, 33076

Other trials in France

About this study

Official title: Comparison of an Anatomy-based Fitting and a Conventional Fitting in Newly Implanted Cochlear Patients. Prospective Monocentric Randomized Double-blind Crossover Study.

Main objective:

Compare the recognition of environmental sounds with an anatomy-based fitting and with a default fitting adult patients newly implanted with a MED-EL cochlear implant.

Secondary objectives:

Compare speech recognition in quiet with an anatomy-based fitting and with a default fitting in adult patients newly implanted with a MED-EL cochlear implant.

Compare speech recognition in noise with an anatomy-based fitting and with a default fitting in adult patients newly implanted with a MED-EL cochlear implant.

Read the detailed description

Introduction: Cochlear implantation allows the rehabilitation of profound bilateral deafness, restoring speech perception and verbal communication when the traditional hearing aid no longer provides satisfactory hearing gain. A cochlear implant includes an electrode array and its functioning is based on the principle of cochlear tonotopy: each electrode encodes a frequency spectrum according to its position in the cochlea (high frequencies are assigned to the basal electrodes and low frequencies to the apical electrodes). The cochlear implant thus breaks down the frequency spectrum into a number of frequency bands via bandpass filters corresponding to the number of electrodes in the implant. During the fitting these bands can be modified by the audiologist. The fitting software developed by the manufacturers proposed a default fitting with a lower limit between 100 and 250 Hz according to the brands and an upper limit of about 8500 Hz. The frequency bands assigned to each electrode follow a logarithmic scale with the high frequencies for the basal electrodes and the low frequencies for the apical electrodes. This distribution takes into account the number of active electrodes but does not take into account the anatomy and the natural cochlear tonotopy specific to each patient. Several studies have analyzed the anatomical variations of the cochlear dimensions: size of the cochlea and the ratio between the contact surfaces of the electrodes with the cochlea are variable from one patient to another. The insertion depth during surgery is also variable due to parameters related to the patients as well as to the operator, which seems to impact the understanding of speech in noise. Mathematical algorithms have recently been developed to estimate the cochlear tonotopy of each patient from a CT scan assessment. CT imaging of the implanted ear combined with 3D reconstruction software, provides cochlear length measurements Using this approach it is possible to measure the position of each electrode relative to the cochlear apex. Recently, MED-EL (Austria) has developed a new approach based on CT-scan and tuning of the frequencies associated with each electrode using anatomical information of position of the electrodes in the cochlea: this fitting is called anatomy-based fitting.

Main objective:

Compare the recognition of environmental sounds with an anatomy-based fitting and with a default fitting adult patients newly implanted with a MED-EL cochlear implant.

Secondary objectives:

Compare speech recognition in quiet with an anatomy-based fitting and with a default fitting in adult patients newly implanted with a MED-EL cochlear implant.

Compare speech recognition in noise with an anatomy-based fitting and with a default fitting in adult patients newly implanted with a MED-EL cochlear implant.

Plan of the study:

It is a prospective open monocentric randomized crossover study: measures will be done on the patient at 6 weeks and 12 weeks post-activation.

Being tested
anatomy-based fitting then default fitting, default fitting then anatomy-based fitting
Keywords
cochlear implant strategy, anatomy-based fitting
Medical subject headings
Hearing Loss, Sensorineural

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • anatomy-based fitting then default fittingDevice

    Cochlear implant with anatomy-based fitting then default fitting

  • default fitting then anatomy-based fittingDevice

    Cochlear implant with anatomy-based fitting then default fitting

Study groups in NCT05230498
Study groupRoleWhat they receive
Cochlear Implant (CI) with default fitting then anatomy-based fittingActive comparatorCochlear Implant with default fitting first during 6 weeks then with anatomy-based fitting during 6 weeks
Receives: Device: default fitting then anatomy-based fitting
Cochlear Implant (CI) with anatomy-based fitting then default fittingActive comparatorCochlear Implant with anatomy-based fitting during 6 weeks then with default fitting during 6 weeks
Receives: Device: anatomy-based fitting then default fitting

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Recognition of Environmental sounds

    The environmental sound recognition is evaluated with the Environmental Sound Identification Test (TISE, Treville-Protain et al. 2019). The patient has to recognize 24 environmental sounds. Each good answer is scored 1 yielding a total between 0 and 1 (or 0% and 100%).

    Measured: at 6 weeks post-activation

  2. Recognition of Environmental sounds

    The environmental sound recognition is evaluated with the Environmental Sound Identification Test (TISE, Treville-Protain et al. 2019). The patient has to recognize 24 environmental sounds. Each good answer is scored 1 yielding a total between 0 and 1 (or 0% and 100%).

    Measured: at 12 weeks post-activation

Secondary outcomes (4)
  1. Speech recognition in quiet

    The speech recognition in quiet is evaluated with 3 lists of 10 disyllabic words. The patient has to recognize 30 words. Each good answer is scored 1 yielding a total between 0 and 1 (or 0% and 100%).

    Measured: at 6 weeks post-activation

  2. Speech recognition in quiet

    The speech recognition in quiet is evaluated with 3 lists of 10 disyllabic words. The patient has to recognize 30 words. Each good answer is scored 1 yielding a total between 0 and 1 (or 0% and 100%).

    Measured: at 12 weeks post-activation

  3. Speech recognition in noise

    The speech recognition in noise is evaluated with the French-language "Rapid speech in noise" (VRB) test (Leclerc et al. 2018). The speech level is at 65 dB SPL. The patient has to recognize 3 target words by sentence. The 8 sentences are played with signal-to-noise ratios between +18 dB and -3 dB by steps of 3 dB. The SRT50 (threshold for 50% intelligibility in noise) is obtained by SRT50 = 19,5 - R, with R = number of correct answers (on 24).

    Measured: at 6 weeks post-activation

  4. Speech recognition in noise

    The speech recognition in noise is evaluated with the French-language VRB test (Leclerc et al. 2018). The speech level is at 65 dB SPL. The patient has to recognize 3 target words by sentence. The 8 sentences are played with signal-to-noise ratios between +18 dB and -3 dB by steps of 3 dB. The SRT50 (threshold for 50% intelligibility in noise) is obtained by SRT50 = 19,5 - R, with R=number of correct answers (on 24).

    Measured: at 12 weeks post-activation

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted9 Feb 2022
  2. Study start15 Feb 2022
  3. Primary completion15 Nov 2024
  4. Study completion15 Nov 2024

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Randomised
Intervention model
Crossover
Model description
Two arms A and B: Arm A: patient's fitting with default fitting --\> 6 weeks use --\> tests and patient's fitting with anatomy-based fitting --\> 6 weeks use --\> tests Arm B: patient's fitting with anatomy-based fitting --\> 6 weeks use --\> tests and patient's fitting with default fitting --\> 6 weeks use --\> tests
Masking
Double blind
Who is masked
Participant, Investigator
Enrolment
17 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05230498
Sponsor's protocol ID
MEDEL_anatfit_Bordeaux_study
Status last verified
Dec 2024
First submitted
28 Jan 2022
Last updated
5 Dec 2024
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing anatomy-based fitting then default fitting and default fitting then anatomy-based fitting in people with Sensorineural Hearing Loss, Bilateral, run by MED-EL Elektromedizinische Geräte GmbH at 1 site in France.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 5 Dec 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Recognition of Environmental sounds (time frame: at 6 weeks post-activation). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.