Premature BirthNCT05250102
Neonatal Pulse Oximetry Sensor Study
A study testing Pulse oximeter sensors in people with Premature Birth, run by Medline Industries at 2 sites in United States.
Trial summary
To validate the SpO2 accuracy, bias, and precision of Medline's reprocessed pulse oximetry sensors as compared to SaO2 in arterial blood samples as assessed by CO-oximetry in neonates. Read more about this study
- Who is it for?
- Ages 1 day to 28 days
- Conditions the study is for
- Premature Birth
- What is being tested?
- Pulse oximeter sensors
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 23 participants
- Where
- 2 sitesUnited States
- Study dates
- 13 Jun 2022to 10 Oct 2023
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 1 day to 28 days
- All sexes
- No healthy volunteers
- Child
Who may be able to join
- Subjects who are postnatal 28 days or younger (neonates)
- Subjects who are likely to receive one or more ABG measurements as part of their clinical SOC
- Subjects who weigh less than 5 Kg (weight range for thesensors)
Who cannot join
- Subjects with current signs and symptoms of a clinically significant Patent Ductus Arteriosus (PDA) combined with a current or planned arterial line placement which will affect the validity of the co-oximetry measurement
- Subjects with physical malformation of hands, fingers, feet, or toes that would limit the ability to place sensors for this study
- Subjects judged by the Principal Investigator (PI) to be inappropriate for participation in this study
- Subjects for whom placing a pulse oximeter will cause dermatological issues (e.g. allergic to foam rubber or adhesive tape)
Trial locations and contacts
2 sites in 1 country.
United States2 sites
Eastern Carolina University
Greenville, North Carolina, 27858
Children's Hospital Los Angeles
Los Angeles, California, 90027
About this study
Official title: Accuracy and Precision of Peripheral Capillary Oxygen Saturation of Reprocessed Pulse Oximetry Sensors Compared to Oxygen Saturation in Arterial Blood Samples Assessed by CO-oximetry in Neonates
To validate the SpO2 accuracy, bias, and precision of Medline's reprocessed pulse oximetry sensors as compared to SaO2 in arterial blood samples as assessed by CO-oximetry in neonates.
Read the detailed description
Pulse oximetry helps in measuring peripheral capillary oxygen saturation (SpO2) continuously and non-invasively, and provides an indirect measurement of arterial oxygenation (SaO2) based on the red and infrared light-absorption characteristics of oxygenated and deoxygenated hemoglobin. Uses of pulse oximetry include detection of hypoxia, avoidance of hyperoxia, titration of fractional inspired oxygen, and enabling weaning from mechanical ventilation. Arterial blood gas (ABG) analysis, such as by the use of CO-oximeter, provides a direct measurement of SaO2. However, ABG analysis requires time, expense, and arterial access. Therefore, this study aims to validate the SpO2 accuracy of pulse oximetry sensors (manufactured by Nellcor and Masimo, and reprocessed by Medline Industries, LP) in neonates as compared to ABG measurements as part of their clinical standard of care (SOC), as assessed by CO-oximetry.
- Conditions
- Premature Birth
- Being tested
- Pulse oximeter sensors
- Medical subject headings
- Premature Birth
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Pulse oximeter sensorsDevice
Pulse oximeter reading will be compared to CO-oximeter reading.
| Study group | Role | What they receive |
|---|---|---|
| Pulse oximeter sensor readings | Other | Subjects who satisfy all of the inclusion criteria will be eligible to participate in the study, will receive intervention (the Masimo and Nellcor sensors, and their arterial blood sample will be taken in the normal course of care). This is a single arm study. Those who don't meet the inclusion/ exclusion criteria won't be eligible to participate. Receives: Device: Pulse oximeter sensors |
What is being measured
The primary outcome is the main question the study is designed to answer.
Bland-Altman Analysis for Nellcor Pulse Oximeter
BIAS Measurement and 95% confidence interval calculated using a Bland-Altman plot (mean vs. difference of SaO2 and SpO2) and an error plot (SaO2 vs. difference of SaO2 and SpO2) were generated. The B\&A plot analysis is a simple way to evaluate a bias between the mean differences, and to estimate an agreement interval, within which 95% of the differences of the second method, compared to the first one, fall. Data can be analyzed both as unit differences plot and as percentage differences plot. We ran a percentage mean difference plot.
Measured: 15 episodic minutes
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted22 Feb 2022
- Study start13 Jun 2022
- Primary completion10 Oct 2023
- Study completion10 Oct 2023
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (3)Papers the sponsor has linked to this study.
- Other trials for Premature BirthStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 23 participants (actual)
Sponsor and investigators
- Lead sponsor
- Medline Industries
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT05250102
- Sponsor's protocol ID
- MED-2020-DIV65-001
- Status last verified
- Oct 2024
- First submitted
- 10 Feb 2022
- Last updated
- 30 Oct 2024
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- No
Publications
Ascha M, Bhattacharyya A, Ramos JA, Tonelli AR. Pulse Oximetry and Arterial Oxygen Saturation during Cardiopulmonary Exercise Testing. Med Sci Sports Exerc. 2018 Oct;50(10):1992-1997. doi: 10.1249/MSS.0000000000001658.
PubMed 29771822Dawson JA, Davis PG, O'Donnell CP, Kamlin CO, Morley CJ. Pulse oximetry for monitoring infants in the delivery room: a review. Arch Dis Child Fetal Neonatal Ed. 2007 Jan;92(1):F4-7. doi: 10.1136/adc.2006.102749. No abstract available.
PubMed 17185428Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571-82. doi: 10.1080/10543400701329422.
PubMed 17613642
Study at a glance
From the registry record, last updated by the sponsor on 30 Oct 2024. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Accuracy and Precision of Peripheral Capillary Oxygen Saturation of Reprocessed Pulse Oximetry Sensors Compared to Oxygen Saturation in Arterial Blood Samples Assessed by CO-oximetry in Neonates
- ClinicalTrials.gov ID
- NCT05250102
- Acronym / sponsor ID
- MED-2020-DIV65-001
- Conditions
- Premature Birth
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 23 participants (actual)
- Lead sponsor
- Medline IndustriesIndustry (pharma, biotech, devices)
- Study start
- 13 Jun 2022 (actual)
- Primary completion
- 10 Oct 2023 (actual)
- Study completion
- 10 Oct 2023 (actual)
- First posted
- 22 Feb 2022 (actual)
- Results
- Posted 30 Oct 2024
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
To validate the SpO2 accuracy, bias, and precision of Medline's reprocessed pulse oximetry sensors as compared to SaO2 in arterial blood samples as assessed by CO-oximetry in neonates.
Detailed description
Pulse oximetry helps in measuring peripheral capillary oxygen saturation (SpO2) continuously and non-invasively, and provides an indirect measurement of arterial oxygenation (SaO2) based on the red and infrared light-absorption characteristics of oxygenated and deoxygenated hemoglobin. Uses of pulse oximetry include detection of hypoxia, avoidance of hyperoxia, titration of fractional inspired oxygen, and enabling weaning from mechanical ventilation. Arterial blood gas (ABG) analysis, such as by the use of CO-oximeter, provides a direct measurement of SaO2. However, ABG analysis requires time, expense, and arterial access. Therefore, this study aims to validate the SpO2 accuracy of pulse oximetry sensors (manufactured by Nellcor and Masimo, and reprocessed by Medline Industries, LP) in neonates as compared to ABG measurements as part of their clinical standard of care (SOC), as assessed by CO-oximetry.
Study design
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Pulse oximeter sensor readings | Other | Device: Pulse oximeter sensorsSubjects who satisfy all of the inclusion criteria will be eligible to participate in the study, will receive intervention (the Masimo and Nellcor sensors, and their arterial blood sample will be taken in the normal course of care). This is a single arm study. Those who don't meet the inclusion/ exclusion criteria won't be eligible to participate. |
Who may qualify
- Ages
- 1 day to 28 days
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Child
Key inclusion criteria (3 listed)
- Subjects who are postnatal 28 days or younger (neonates)
- Subjects who are likely to receive one or more ABG measurements as part of their clinical SOC
- Subjects who weigh less than 5 Kg (weight range for thesensors)
Key exclusion criteria (4 listed)
- Subjects with current signs and symptoms of a clinically significant Patent Ductus Arteriosus (PDA) combined with a current or planned arterial line placement which will affect the validity of the co-oximetry measurement
- Subjects with physical malformation of hands, fingers, feet, or toes that would limit the ability to place sensors for this study
- Subjects judged by the Principal Investigator (PI) to be inappropriate for participation in this study
- Subjects for whom placing a pulse oximeter will cause dermatological issues (e.g. allergic to foam rubber or adhesive tape)
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Pulse oximeter sensorsPulse oximeter reading will be compared to CO-oximeter reading. | Device | Pulse oximeter sensor readings |
Primary outcome
Bland-Altman Analysis for Nellcor Pulse Oximeter
Time frame: 15 episodic minutes
BIAS Measurement and 95% confidence interval calculated using a Bland-Altman plot (mean vs. difference of SaO2 and SpO2) and an error plot (SaO2 vs. difference of SaO2 and SpO2) were generated. The B\&A plot analysis is a simple way to evaluate a bias between the mean differences, and to estimate an agreement interval, within which 95% of the differences of the second method, compared to the first one, fall. Data can be analyzed both as unit differences plot and as percentage differences plot. We ran a percentage mean difference plot.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 30 Oct 2024.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Background: PubMed 29771822Ascha M, Bhattacharyya A, Ramos JA, Tonelli AR. Pulse Oximetry and Arterial Oxygen Saturation during Cardiopulmonary Exercise Testing. Med Sci Sports Exerc. 2018 Oct;50(10):1992-1997. doi: 10.1249/MSS.0000000000001658.
- Background: PubMed 17185428Dawson JA, Davis PG, O'Donnell CP, Kamlin CO, Morley CJ. Pulse oximetry for monitoring infants in the delivery room: a review. Arch Dis Child Fetal Neonatal Ed. 2007 Jan;92(1):F4-7. doi: 10.1136/adc.2006.102749. No abstract available.
- Background: PubMed 17613642Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571-82. doi: 10.1080/10543400701329422.
- Individual participant dataSharing planned: No
Related trials for Premature Birth
RecruitingPhase 3
Selective Antibiotics When Symptoms Develop Versus Universal Antibiotics for Preterm Neonates
Sepsis, PROM, Preterm (Pregnancy)
Recruiting
The Effect of Breast Milk, 10% and 20% Dextrose on Pain and Physiological Variables During Heel Puncture in Infants
Newborn Care, Pain Management
Not yet recruiting
The Hemodynamic Impact of High-frequency Oscillatory Ventilation and Periodic Sighs in Preterm Infants
Preterm, Respiratory Distress Syndrome in Premature Infant
Active, not recruiting
Prone Positioning During Delayed Cord Clamping
Delayed Cord Clamping
Not yet recruiting
Strap To Reduce Apnea in Preterm Infants: A Randomized Crossover Trial
Apnea of Prematurity, Intermittent Hypoxemia
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
2 sites in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Pulse oximeter sensors in people with Premature Birth, run by Medline Industries at 2 sites in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 30 Oct 2024. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 2 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 30 Oct 2024). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Bland-Altman Analysis for Nellcor Pulse Oximeter (time frame: 15 episodic minutes). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT05250102, which also lists 3 linked publications.