Coronary StenosisNCT05250193

DK Score Coronary Scoring Balloon: a RCT Trial to Evaluate Acute Lumen Gain (SCORE CHINA)

A study testing DK ScoreTM Coronary Scoring Balloon and Non-Slip Element Coronary Dilatation Catheter in people with Coronary Stenosis, run by DK Medical Technology (Suzhou) Co., Ltd. at 4 sites in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

A prospective, Multi-center, Randomized Study designed to evaluate the acute lumen gain and device procedural success of the DK ScoreTM Scoring balloon versus NSE Scoring Balloon for patients with coronary arteries stenosis. Read more about this study

Who is it for?
Ages 18 to 75 years
Conditions the study is for
Coronary Stenosis
What is being tested?
DK ScoreTM Coronary Scoring Balloon and Non-Slip Element Coronary Dilatation Catheter
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
200 participants
Where
4 sitesChina
Study dates
24 Mar 2021to 30 Dec 2021
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 75 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patients related
  • Age ≥ 18 years and ≤ 75 years.
  • Patients with symptomatic coronary artery disease with stable and unstable angina pectoralis remote myocardial infarction , asymptomatic subjects with coronary atherosclerotic stenosis causing myocardial ischemia.
  • Patients with indications for coronary artery bypass grafting.
  • Patients are able to follow protocol requirements and data collection procedures
  • Patients understand the purpose of the study, will voluntarily participate in the study and sign informed consent. Patients are willing to undergo clinical follow-up as required by this study.
Show all 11Show fewer
  • Lesions Related:
  • Reference vessel diameter (RVD) between 2.0 and 4.0 mm by visual estimation.. Target lesion length ≤ 20 mm.
  • More than 70% stenosis (or more than 50% stenosis combined with symptoms of ischemia). TIMI≥1.
  • One target lesion needs to be Treated. Treatment of non-target lesion, if any, must be completed prior to treatment of target lesion and must be deemed a clinical angiographic success as visually assessed by the physician. The non-target lesion need be located at another coronary artery different from the target lesion.
  • Investigator estimates that the target lesion need Cutting/Scoring balloon treatment. Target lesion is fully pre-dilated that scoring balloon is able to pass.

Who cannot join

  • Patients Related :
  • AMI within 1 week.
  • Patients has congestive heart failure or NYHA IV.
  • Renal dysfunctional (CRE>443umol/L).
  • Patient underwent heat transplant.
  • Pregnant or nursing.
Show all 18Show fewer
  • Lifespan expectancy is less than 12 months
  • Patients with history of peptic ulcer or gastrointestinal bleeding, or the patients have a bleeding tendency evaluated by investigator. Patients are forbidden to use anticoagulation agents or anti-platelet drugs.
  • Patients are allergic to paclitaxel or heparin
  • Difficult for evaluation or treatment.
  • Patents are currently participating in another investigational drug or device study that has not yet completed its primary endpoint.
  • Patients participated in any other clinical trials less than 1 month.
  • Interventional surgery is expected to be extremely risky unsuitable for study evaluated by investigator.
  • Patient had poor compliance record and could not follow the study.
  • Lesion Related:
  • Target Lesion located in Left Main Artery lesions and Ostial lesion within 5mm to Left Main Artery.
  • Angiography shows thrombus in Target Lesion
  • Chronic Total Occlusion Lesion

Trial locations and contacts

4 sites in 1 country.

China4 sites
  • Beijing Anzhen Hospital affiliated to Capital University of Medical

    Beijing, Beijing Municipality

  • Taiyuan Central Hospital

    Taiyuan, Shanxi

  • Tianjin Chest Hospita

    Tianjin, Tianjin Municipality

  • Xuzhou Third People Hospital

    Xuzhou, Jiangsu

Other trials in China

About this study

Official title: A Prospective, Multi-center, Randomized Study to Evaluate the Safety and Efficacy of DK ScoreTM Scoring Balloon Vesus Non-Slip Element Coronary Dilatation Catheter for Coronary Artery Stenosis

A prospective, Multi-center, Randomized Study designed to evaluate the acute lumen gain and device procedural success of the DK ScoreTM Scoring balloon versus NSE Scoring Balloon for patients with coronary arteries stenosis.

Read the detailed description

This is a prospective, multi-center, randomized controlled, open-label, noninferior study to evaluate Acute Lumen Gain after dilatation. A total of 200 patients will be enrolled from 4 sites in China. All patients enrolled will be randomly assigned to the test group (DK ScoreTM group, n=100) and the control group (NSE group, n=100) with randomized allocation ratio of 1:1. Patients in the test group and the control group will be treated with DK ScoreTM Scoring balloon catheters or NSE Coronary Dilatation Catheter. A 30-day after procedure follow-up will be conducted for all 200 patients.

Being tested
DK ScoreTM Coronary Scoring Balloon, Non-Slip Element Coronary Dilatation Catheter
Keywords
Scoring balloon; DK Score; NSE
Medical subject headings
Coronary Stenosis

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • DK ScoreTM Coronary Scoring BalloonDevice

    Also called DK Score

    After pre-dilation balloon is used (if any), DK Score is used as final dilatation balloon before Drug Eluting Stent implantation or Drug Coated Balloon deployment.

  • Non-Slip Element Coronary Dilatation CatheterDevice

    Also called NSE

    After pre-dilation balloon is used (if any), NSE is used as final dilatation balloon before Drug Eluting Stent implantation or Drug Coated Balloon deployment.

Study groups in NCT05250193
Study groupRoleWhat they receive
DK ScoreTM Coronary Scoring Balloon (DK Score)ExperimentalDK ScoreTM Coronary Scoring Balloon Manufactured by DK Medical Technolgy CO.,LTD
Receives: Device: DK ScoreTM Coronary Scoring Balloon
Non-Slip Element Coronary Dilatation Catheter (NSE)Active comparatorNon-Slip Element Coronary Dilatation Catheter Manufactured by GOODMAN CO.,LTD
Receives: Device: Non-Slip Element Coronary Dilatation Catheter

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Numerical Acute Lumen Gain in mm (QCA)

    Acute Lumen Gain for both arms will be measure with QCA, as accessed by independent core-lab.

    Angiographic endpoint will be in-lesion acute lumen gain defined as minimal lumen diameter (MLD) in lumen after dilatation by either arm device minus baseline MLD.

    Measured: During the procedure

Secondary outcomes (5)
  1. Rate of Device success in percent

    Successful delivery to the target lesion, deployment, achieving less than 50% residual stenosis, and retrieval after procedure.

    Measured: During the procedure

  2. Rate of Target Lesion Failure in percent

    Defined as any Target Lesion Failure after procedure

    Measured: 30±7 Days

  3. Rate of Patient-oriented Composite Endpoint (POCE) in percent

    Patient-oriented Composite Endpoint after procedure

    Measured: 30±7 Days

  4. Rate of other Complications in percent

    Incidence of complications associated with the use of scoring balloon catheters

    Measured: During the procedure

  5. Rate of Clinical Success in percent

    Defined as <20% diameter stenosis after stenting for DCB, without in-hospital major adverse cardiac events/MI/TLR After PCI

    Measured: During the procedure

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted22 Feb 2022
  2. Study start24 Mar 2021
  3. Primary completion2 Sep 2021
  4. Study completion30 Dec 2021

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
Single blind
Who is masked
Participant
Enrolment
200 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05250193
Sponsor's protocol ID
VP-P-2019-006
Status last verified
Feb 2022
First submitted
5 Feb 2022
Last updated
10 Mar 2022
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing DK ScoreTM Coronary Scoring Balloon and Non-Slip Element Coronary Dilatation Catheter in people with Coronary Stenosis, run by DK Medical Technology (Suzhou) Co., Ltd. at 4 sites in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 11 inclusion and 18 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 10 Mar 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 4 sites in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Numerical Acute Lumen Gain in mm (QCA) (time frame: During the procedure). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 10 Mar 2022.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in