AzTrials
AzTrials

HealthyNCT05264116

SpO2 Accuracy of Noninvasive Disposable Pulse Oximeter Sensor With Motion Conditions

A study testing STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERS in people with Healthy and Hypoxia, run by Stryker Sustainability Solutions at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this clinical study is to validate the SpO2 accuracy of the Stryker Sustainability Solutions pulse oximetry sensors during motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. Read more about this study

Who is it for?
Ages 18 to 55 yearsHealthy volunteers can take part
Conditions the study is for
Healthy, Hypoxia
What is being tested?
STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERS
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
12 participants
Where
1 siteUnited States
Study dates
2 Feb 2022to 3 Feb 2022
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 55 years
  • All sexes
  • Healthy volunteers accepted
  • Adult

Who may be able to join

  • Good general health with no evidence of pre-existing medical problems
  • Fluent in both written and spoken English
  • Must be able to review informed consent and is willing to comply with study procedures

Who cannot join

  • Obesity (BMI >30)
  • Known history of heart disease, lung disease, kidney or liver disease
  • Prior diagnosis of asthma, sleep apnea, or use of CPAP
  • Diabetes
  • Clotting disorder
  • Hemoglobinopathy or history of anemia that in the opinion of the investigators would make them unsuitable for study participation
Show all 16Show fewer
  • Any serious systemic illness
  • Current smoker
  • Any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigator would interfere with the sensors working correctly
  • History of fainting or vasovagal response
  • History of sensitivity to local anesthesia
  • Prior diagnosis of Raynaud's disease
  • Unacceptable collateral circulation based on exam by investigators (Allen's test)
  • Pregnant, lactating, or trying to get pregnant
  • Unable or unwilling to provide informed consent, or unable or unwilling to comply with study procedures
  • Any other condition which in the opinion of the investigators would make them unsuitable for the study

Trial locations and contacts

1 site in 1 country.

United States1 site
  • UCSF Hypoxia Research Laboratory

    San Francisco, California, 94133

Other trials in United States

About this study

The purpose of this clinical study is to validate the SpO2 accuracy of the Stryker Sustainability Solutions pulse oximetry sensors during motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of the reprocessed sensors.

Read the detailed description

In this study, the level of oxygen within the blood will be reduced in a controlled manner by reducing the concentration of oxygen the participant breathes. The accuracy of a noninvasive pulse oximeter sensor will be assessed by comparison to the oxygen saturation measurements from a laboratory blood gas analyzer.

It is required that the Accuracy Root Mean Square (ARMS) performance of the Stryker pulse oximetry sensors will meet a specification of +/-3% or better in motion conditions for the range of 70 - 100% SaO2 (typically, saturation is determined once with air breathing and then at one of 6 levels, e.g. 94%, 90%, 85%, 80%, 75% and 70% saturation for about 30-60 seconds at each level), thereby demonstrating an acceptable SpO2 accuracy performance specification. Motion is defined as controlled rubbing or tapping of the subject's fingers in a 2-4 Hz or random sine motion with 2-3 cm amplitude.

Conditions
Healthy, Hypoxia
Being tested
STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERS
Medical subject headings
Hypoxia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERSDevice

    Reprocessed noninvasive pulse oximeter sensors.

Study groups in NCT05264116
Study groupRoleWhat they receive
Healthy adult participantsExperimentalAll participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors.
Receives: Device: STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERS

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Accuracy of Sensor by Arms Calculation Under Motion Conditions

    Percentage of SpO2 (oxygen saturation by pulse oximetry) measured by the Reprocessed Oximeter pulse oximetry sensors during motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. Accuracy root mean square (ARMS) between measured SpO2 and reference SaO2 (arterial oxygen saturation) must meet the 3% specification for each reprocessed pulse oximetry sensor style. Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square error (Arms) value. Standard deviation of the differences is computed as the precision. Square root of the sum of the squares of bias and precision is computed as the Arms Error value.

    Measured: 1-2 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted3 Mar 2022
  2. Study start2 Feb 2022
  3. Primary completion3 Feb 2022
  4. Study completion3 Feb 2022

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
12 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05264116
Sponsor's protocol ID
CRD10318
Status last verified
Feb 2023
First submitted
22 Feb 2022
Last updated
20 May 2025
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERS in people with Healthy and Hypoxia, run by Stryker Sustainability Solutions at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 16 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 20 May 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 10 Mar 2023). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Accuracy of Sensor by Arms Calculation Under Motion Conditions (time frame: 1-2 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 20 May 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in