Atrial FibrillationHPSD-RemapNCT05459831
Efficacy Comparison of High and Very High Power Short Duration Pulmonary Vein Isolation
A study testing QDot Micro Catheter in people with Atrial Fibrillation, run by Semmelweis University at 1 site in Hungary.
Trial summary
Objective: To evaluate the long-term durability of PVI performed with 90W and 50W (QDot catheter). Primary hypothesis: Pulmonary vein isolation performed with 90W (QMODE+) and inter-tag distance <5 mm is non-inferior to pulmonary vein isolation performed with 50W guided by ablation index (QMODE) and inter-tag distance… Read more about this study
- Who is it for?
- Ages 18 to 100 years
- Conditions the study is for
- Atrial Fibrillation
- What is being tested?
- QDot Micro Catheter
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 46 participants
- Where
- 1 siteHungary
- Study dates
- 15 Feb 2022to 1 Oct 2023
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 100 years
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Symptomatic paroxysmal/persistent AF
- Age >18 years
- Willingness to sign informed consent form.
Who cannot join
- Contraindication to ablation
- Contraindication of long-term anticoagulation
- Long-standing persistent AF
- History of PVI
- History of cardiac surgery
- Pregnancy
Show all 10Show fewer
- Active malignancy
- Life expectancy <1 year
- Valvular AF
- Reversible cause of AF (e.g. hyperthyroidism).
Trial locations and contacts
1 site in 1 country.
About this study
Objective: To evaluate the long-term durability of PVI performed with 90W and 50W (QDot catheter).
Primary hypothesis: Pulmonary vein isolation performed with 90W (QMODE+) and inter-tag distance <5 mm is non-inferior to pulmonary vein isolation performed with 50W guided by ablation index (QMODE) and inter-tag distance <5 mm.
Read more
Study Design: 40 patients will be randomized in a 1:1 ratio to 50W and 90W PVI.
Study procedures: Initial PVI will be performed via femoral access and double transseptal puncture guided by fluoroscopy and pressure monitoring. If necessary intracardiac echocardiography (ICE) will be used for the transseptal puncture. A fast anatomical left atrial map will be created with a multipolar mapping catheter; then, point-by-point PVI will be performed with QDot catheter with 50W or 90W according to randomization. Three months after the initial procedure, patients will undergo a repeated high density left atrial mapping with PentaRay or OctaRay catheter to evaluate the durability of the PVI. Exact gap numbers and locations will be registered. If necessary, reablation of the pulmonary veins (rePVI) will be performed.
Inclusion criteria: Symptomatic paroxysmal/persistent AF, >18 years, willingness to sign informed consent form.
Exclusion criteria: Long-standing persistent AF, active malignancy, contraindication for anticoagulant therapy, life expectancy <1 year, valvular AF, hyperthyroidism.
Endpoints: At repeat procedure, a high density left atrial map will be created with PentaRay or if available OctaRay catheter to evaluate the durability of the PVI. Primary endpoint will be pulmonary vein reconnection.
Read the detailed description
- Background Atrial fibrillation (AF) is the most common sustained cardiac arrhythmia in adults. Its prevalence increases with age, affecting 3/100 people, and its incidence is also correlated with age, affecting 30/100,000 people per year. Its clinical significance stems from common complications (stroke, heart failure) and concomitant higher mortality. The most effective treatment method of atrial fibrillation is catheter ablation, which seeks to maintain sinus rhythm. Isolation of the triggering pulmonary veins (PVI) from the left atrium is the cornerstone of AF ablation. PVI has undergone significant technical development over the past decade; however, the chances of arrhythmia recurrence and the occurrence of procedural complications are not negligible. Experiments have shown that lesions with a larger surface diameter but less depth develop during HPSD ablation; the higher the energy is, the shallower and wider the lesions are. These lesion properties make it easier to create a solid line from the ablation points, while the shallower lesion causes less damage to the surrounding tissues. The latest application of HPSD technology is the so-called "very high power short duration" (vHPSD) ablation when we apply 90W power. No data are available on long-term efficacy or safety comparing ablation with HPSD vs. vHPSD.
- Objective: To evaluate the durability of radiofrequency PVI performed with 90W (QMODE plus) and 50W (QMODE) with the QDot catheter.
- Primary hypothesis: Pulmonary vein isolation performed with 90W (QMODE+) and inter-tag distance <5 mm is non-inferior to pulmonary vein isolation performed with 50W guided by ablation index (QMODE) and inter-tag distance <5 mm.
- Study Design: Single-center, prospective, randomized study. Patients will be randomly assigned in a 1:1 ratio to HPSD (50W) and vHPSD (90W) PVI.
- Research period Planned research period from the second quarter of 2022 for a maximum of 1 year.
- Study subjects
- Planned number of included subjects: 40.
- Study procedures:
- Initial PVI will be performed via femoral access after double transseptal puncture guided by fluoroscopy and pressure monitoring. If necessary ICE will be used for the transseptal puncture. A fast anatomical left atrial map will be created with a multipolar mapping catheter; then, point-by-point PVI will be performed with QDot catheter with 50W or 90W according to randomization. After creating the isolation circle, the presence or absence of first-pass isolation will be assessed by multipolar catheters. If PVI is not complete at this point, it will be completed to reach complete isolation of all veins. After that, a 20 minutes waiting period will be the next step to evaluate presence or absence of acute reconnection of the PV-s. In case of acute reconnection, touch-up applications will be delivered to complete the PVI.
- Three months after the initial procedure, patients will undergo a repeated high density left atrial mapping with PentaRay or, if available OctaRay catheter to evaluate the durability of the PVI. Exact gap numbers and locations will be registered (based on the 16 segment PV model). If necessary, rePVI will be performed.
- Statistical analysis Continuous variables are expressed as mean and standard deviation, or median and interquartile ranges, depending on the normal of non-normal distribution. Categorical variables are expressed in numbers and percentages. Continuous variables are compared by parametric or non-parametric tests, depending on the distribution, while categorical variables are compared by Chi-square test or Fisher's exact test. The predictive value of different ablation techniques in outcome is planned to be analyzed by logistic regression. A two-tailed p-value of <0.05 will be considered statistically significant. Statistical analyses will be performed using IBM SPSS 25 (Apache Software Foundation, USA) and GraphPad Prism 7.1 (GraphPad Softwares Inc., USA) software products.
- Data management All study patient data is entered into Semmelweis University's electronic database (accessible only from Semmelweis University's network) in a coded and unique manner, with a unique identifier, to which project staff has a password-protected, defined level of access. Each person involved in the study is given a unique identification code, and the data stored in the database is linked to that unique identifier. Thus, a data set will be incomprehensible and unusable for an external (unauthorized) user. The data belonging to the unique identification code, with which the patient's identity can be clearly indicated (name, place and date of birth, clinical reference number, identification number, identity card number, etc.) are not available from the database and are stored separately from it. Access to personal data is restricted to institutional leaders and program professional leaders.
- Personal and material requirements The personal and material requirements of the research are available at the Heart and Vascular Center of Semmelweis University. The personal requirements are given by the staff of the Electrophysiological Laboratory, and the material conditions are provided by the equipment of the same laboratory.
- Ethics approval
An ethics approval was obtained from the Hungarian National Public Health Center (No. 8119-2/2022/EÜIG).
- Conditions
- Atrial Fibrillation
- Being tested
- QDot Micro Catheter
- Keywords
- atrial fibrillation, catheter ablation, high-power, short-duration, pulmonary vein isolation, durability
- Medical subject headings
- Atrial Fibrillation
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
QDot Micro CatheterDevice
Atrial fibrillation radiofrequency ablation with 90W or 50W energy setting.
| Study group | Role | What they receive |
|---|---|---|
| Pulmonary vein isolation with 50W energy setting | Active comparator | In this group of subjects, the initial pulmonary vein isolation procedure will be performed using 50W radiofrequency energy. This power setting will be used for all the ablation points. Receives: Device: QDot Micro Catheter |
| Pulmonary vein isolation with 90W energy setting | Active comparator | In this group of subjects, the initial pulmonary vein isolation procedure will be performed using 90W radiofrequency energy. This power setting will be used for all the ablation points. Receives: Device: QDot Micro Catheter |
What is being measured
The primary outcome is the main question the study is designed to answer.
Pulmonary vein reconnection
Presence or absence of pulmonary vein reconnection at 3 month remapping procedure
Measured: 3 month
Secondary outcomes (8)
Procedure time
Procedure time of the initial pulmonary vein isolation procedure
Measured: Immediately after the procedure
Fluoroscopy time
Fluoroscopy time of the initial pulmonary vein isolation procedure
Measured: Immediately after the procedure
Fluoroscopy dose
Fluoroscopy dose of the initial pulmonary vein isolation procedure
Measured: Immediately after the procedure
Major complication
Major complication of the initial pulmonary vein isolation procedure
Measured: 3 month
First pass isolation
First pass isolation during the initial pulmonary vein isolation procedure
Measured: Immediately after the procedure
Acute reconnection
Acute reconnection during the initial pulmonary vein isolation procedure
Measured: Immediately after the procedure
Major complication of the second (remapping) procedure
Major complication of the second (remapping) procedure
Measured: 1 week
12-month atrial fibrillation freedom
freedom from atrial fibrillation during the 12-month follow-up
Measured: 12 months
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted15 Jul 2022
- Study start15 Feb 2022
- Primary completion1 Dec 2022
- Study completion1 Oct 2023
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for Atrial FibrillationStudies of the same condition, open and under way.
- More trials run by Semmelweis UniversityThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Enrolment
- 46 participants (actual)
Sponsor and investigators
- Lead sponsor
- Semmelweis University
- Sponsor type
- University, hospital or other
- Responsible party
- Principal investigator, Nandor Szegedi, Semmelweis University
- Investigators
- László Gellér, Ph.D., DSc, Study chair, Semmelweis University
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT05459831
- Sponsor's protocol ID
- BWI- IIS-653
- Acronym
- HPSD-Remap
- Status last verified
- Nov 2023
- First submitted
- 4 Jul 2022
- Last updated
- 15 Nov 2023
- Results posted
- No
- Data monitoring committee
- Yes
- FDA-regulated drug
- No
- FDA-regulated device
- No
Publications
Szegedi N, Sallo Z, Nagy VK, Osztheimer I, Hizoh I, Lakatos B, Boussoussou M, Orban G, Boga M, Ferencz AB, Komlosi F, Toth P, Perge P, Kovacs A, Merkely B, Geller L. Long-Term Durability of High- and Very High-Power Short-Duration PVI by Invasive Remapping: The HPSD Remap Study. Circ Arrhythm Electrophysiol. 2024 Feb;17(2):e012402. doi: 10.1161/CIRCEP.123.012402. Epub 2024 Jan 29.
PubMed 38284286
Study at a glance
From the registry record, last updated by the sponsor on 15 Nov 2023. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Efficacy Comparison of High and Very High Power Short Duration Pulmonary Vein Isolation
- ClinicalTrials.gov ID
- NCT05459831
- Acronym / sponsor ID
- HPSD-Remap / BWI- IIS-653
- Conditions
- Atrial Fibrillation
- Keywords
- atrial fibrillation, catheter ablation, high-power, short-duration, pulmonary vein isolation, durability
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 46 participants (actual)
- Lead sponsor
- Semmelweis UniversityUniversity, hospital or other
- Study start
- 15 Feb 2022 (actual)
- Primary completion
- 1 Dec 2022 (actual)
- Study completion
- 1 Oct 2023 (actual)
- First posted
- 15 Jul 2022 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Objective: To evaluate the long-term durability of PVI performed with 90W and 50W (QDot catheter).
Primary hypothesis: Pulmonary vein isolation performed with 90W (QMODE+) and inter-tag distance <5 mm is non-inferior to pulmonary vein isolation performed with 50W guided by ablation index (QMODE) and inter-tag distance <5 mm.
Study Design: 40 patients will be randomized in a 1:1 ratio to 50W and 90W PVI.
Study procedures: Initial PVI will be performed via femoral access and double transseptal puncture guided by fluoroscopy and pressure monitoring. If necessary intracardiac echocardiography (ICE) will be used for the transseptal puncture. A fast anatomical left atrial map will be created with a multipolar mapping catheter; then, point-by-point PVI will be performed with QDot catheter with 50W or 90W according to randomization. Three months after the initial procedure, patients will undergo a repeated high density left atrial mapping with PentaRay or OctaRay catheter to evaluate the durability of the PVI. Exact gap numbers and locations will be registered. If necessary, reablation of the pulmonary veins (rePVI) will be performed.
Inclusion criteria: Symptomatic paroxysmal/persistent AF, >18 years, willingness to sign informed consent form.
Exclusion criteria: Long-standing persistent AF, active malignancy, contraindication for anticoagulant therapy, life expectancy <1 year, valvular AF, hyperthyroidism.
Endpoints: At repeat procedure, a high density left atrial map will be created with PentaRay or if available OctaRay catheter to evaluate the durability of the PVI. Primary endpoint will be pulmonary vein reconnection.
Detailed description
- Background Atrial fibrillation (AF) is the most common sustained cardiac arrhythmia in adults. Its prevalence increases with age, affecting 3/100 people, and its incidence is also correlated with age, affecting 30/100,000 people per year. Its clinical significance stems from common complications (stroke, heart failure) and concomitant higher mortality. The most effective treatment method of atrial fibrillation is catheter ablation, which seeks to maintain sinus rhythm. Isolation of the triggering pulmonary veins (PVI) from the left atrium is the cornerstone of AF ablation. PVI has undergone significant technical development over the past decade; however, the chances of arrhythmia recurrence and the occurrence of procedural complications are not negligible. Experiments have shown that lesions with a larger surface diameter but less depth develop during HPSD ablation; the higher the energy is, the shallower and wider the lesions are. These lesion properties make it easier to create a solid line from the ablation points, while the shallower lesion causes less damage to the surrounding tissues. The latest application of HPSD technology is the so-called "very high power short duration" (vHPSD) ablation when we apply 90W power. No data are available on long-term efficacy or safety comparing ablation with HPSD vs. vHPSD.
- Objective: To evaluate the durability of radiofrequency PVI performed with 90W (QMODE plus) and 50W (QMODE) with the QDot catheter.
- Primary hypothesis: Pulmonary vein isolation performed with 90W (QMODE+) and inter-tag distance <5 mm is non-inferior to pulmonary vein isolation performed with 50W guided by ablation index (QMODE) and inter-tag distance <5 mm.
- Study Design: Single-center, prospective, randomized study. Patients will be randomly assigned in a 1:1 ratio to HPSD (50W) and vHPSD (90W) PVI.
- Research period Planned research period from the second quarter of 2022 for a maximum of 1 year.
- Study subjects
- Planned number of included subjects: 40.
- Study procedures:
- Initial PVI will be performed via femoral access after double transseptal puncture guided by fluoroscopy and pressure monitoring. If necessary ICE will be used for the transseptal puncture. A fast anatomical left atrial map will be created with a multipolar mapping catheter; then, point-by-point PVI will be performed with QDot catheter with 50W or 90W according to randomization. After creating the isolation circle, the presence or absence of first-pass isolation will be assessed by multipolar catheters. If PVI is not complete at this point, it will be completed to reach complete isolation of all veins. After that, a 20 minutes waiting period will be the next step to evaluate presence or absence of acute reconnection of the PV-s. In case of acute reconnection, touch-up applications will be delivered to complete the PVI.
- Three months after the initial procedure, patients will undergo a repeated high density left atrial mapping with PentaRay or, if available OctaRay catheter to evaluate the durability of the PVI. Exact gap numbers and locations will be registered (based on the 16 segment PV model). If necessary, rePVI will be performed.
- Statistical analysis Continuous variables are expressed as mean and standard deviation, or median and interquartile ranges, depending on the normal of non-normal distribution. Categorical variables are expressed in numbers and percentages. Continuous variables are compared by parametric or non-parametric tests, depending on the distribution, while categorical variables are compared by Chi-square test or Fisher's exact test. The predictive value of different ablation techniques in outcome is planned to be analyzed by logistic regression. A two-tailed p-value of <0.05 will be considered statistically significant. Statistical analyses will be performed using IBM SPSS 25 (Apache Software Foundation, USA) and GraphPad Prism 7.1 (GraphPad Softwares Inc., USA) software products.
- Data management All study patient data is entered into Semmelweis University's electronic database (accessible only from Semmelweis University's network) in a coded and unique manner, with a unique identifier, to which project staff has a password-protected, defined level of access. Each person involved in the study is given a unique identification code, and the data stored in the database is linked to that unique identifier. Thus, a data set will be incomprehensible and unusable for an external (unauthorized) user. The data belonging to the unique identification code, with which the patient's identity can be clearly indicated (name, place and date of birth, clinical reference number, identification number, identity card number, etc.) are not available from the database and are stored separately from it. Access to personal data is restricted to institutional leaders and program professional leaders.
- Personal and material requirements The personal and material requirements of the research are available at the Heart and Vascular Center of Semmelweis University. The personal requirements are given by the staff of the Electrophysiological Laboratory, and the material conditions are provided by the equipment of the same laboratory.
- Ethics approval
An ethics approval was obtained from the Hungarian National Public Health Center (No. 8119-2/2022/EÜIG).
Study design
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Pulmonary vein isolation with 50W energy setting | Active comparator | Device: QDot Micro CatheterIn this group of subjects, the initial pulmonary vein isolation procedure will be performed using 50W radiofrequency energy. This power setting will be used for all the ablation points. |
| Pulmonary vein isolation with 90W energy setting | Active comparator | Device: QDot Micro CatheterIn this group of subjects, the initial pulmonary vein isolation procedure will be performed using 90W radiofrequency energy. This power setting will be used for all the ablation points. |
Who may qualify
- Ages
- 18 to 100 years
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (3 listed)
- Symptomatic paroxysmal/persistent AF
- Age >18 years
- Willingness to sign informed consent form.
Key exclusion criteria (10 listed)
- Contraindication to ablation
- Contraindication of long-term anticoagulation
- Long-standing persistent AF
- History of PVI
- History of cardiac surgery
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| QDot Micro CatheterAtrial fibrillation radiofrequency ablation with 90W or 50W energy setting. | Device | Pulmonary vein isolation with 50W energy setting, Pulmonary vein isolation with 90W energy setting |
Primary outcome
Pulmonary vein reconnection
Time frame: 3 month
Presence or absence of pulmonary vein reconnection at 3 month remapping procedure
8 secondary outcomes
Procedure time
Time frame: Immediately after the procedure
Fluoroscopy time
Time frame: Immediately after the procedure
Fluoroscopy dose
Time frame: Immediately after the procedure
Major complication
Time frame: 3 month
First pass isolation
Time frame: Immediately after the procedure
Acute reconnection
Time frame: Immediately after the procedure
Major complication of the second (remapping) procedure
Time frame: 1 week
12-month atrial fibrillation freedom
Time frame: 12 months
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 15 Nov 2023.
- Derived from this study: PubMed 38284286Szegedi N, Sallo Z, Nagy VK, Osztheimer I, Hizoh I, Lakatos B, Boussoussou M, Orban G, Boga M, Ferencz AB, Komlosi F, Toth P, Perge P, Kovacs A, Merkely B, Geller L. Long-Term Durability of High- and Very High-Power Short-Duration PVI by Invasive Remapping: The HPSD Remap Study. Circ Arrhythm Electrophysiol. 2024 Feb;17(2):e012402. doi: 10.1161/CIRCEP.123.012402. Epub 2024 Jan 29.
Related trials for Atrial Fibrillation
Active, not recruiting
CONVERGE Post-Approval Study (PAS)
Chronic Atrial Fibrillation
Recruiting
Early Closure of Left Atrial Appendage for Patients With Atrial Fibrillation and Ischemic Stroke Despite Anticoagulation Therapy
Ischemic Stroke, Atrial Fibrillation
Active, not recruitingObservational study
Concomitant Left Atrial Appendage Closure and Pulsed Field Ablation-Asia
Concomitant Procedures, Atrial Fibrillation
Active, not recruiting
Horizon 360 Protocol for the Treatment of Paroxysmal Atrial Fibrillation With the Sphere-360™ Catheter and Affera™ Mapping and Ablation System (Horizon 360)
Paroxysmal AF
RecruitingPhase 2
Safety in Adult Participants With Atrial Fibrillation Who Are Treated With Anticoagulation
Atrial Fibrillation (AF)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
László Gellér, Ph.D., DSc study chair
Semmelweis University
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing QDot Micro Catheter in people with Atrial Fibrillation, run by Semmelweis University at 1 site in Hungary.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 10 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 15 Nov 2023. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Hungary. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Pulmonary vein reconnection (time frame: 3 month). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT05459831, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 15 Nov 2023; AzTrials last refreshed its copy on 7 Nov 2026.