PUMCH Study Into Individualized Scanning for Coronary Artery Disease

An observational study of Coronary Artery Disease and Coronary Computed Tomographic Angiography, run by Peking Union Medical College Hospital.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

Cardiac computed tomographic angiography (CCTA) is a non-invasive diagnostic imaging technique for visualization of the coronary arteries and thus, frequently used in the evaluation of coronary artery disease (CAD). Read more about this study

Who is it for?
Ages 18 years and older
How many people
300 participants
Where
No sites listed yet
Study dates
1 Jan 2023to 30 Jun 2024
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: The study aims to consecutively include 300 patients scheduled for coronary CTA examination to rule out CAD in PUMCH.

Who may be able to join

  • Age: Patient aged older than 18;
  • Deny previous cardiovascular history;
  • Atypical or typical complaints of angina and referred for CCTA to rule out CAD with Iopromide 370 mgI/ml by clinicians
  • Possibility of 'flash' mode scan: patient heart rate below 70 beats per minute and regular rhythm;
  • Patients or families are able to understand the study protocol and willing to participate in the study and provide written informed consent.

Who cannot join

  • Any contraindications for prospective ECG gate-control CT coronary angiography, including:
  • Inability to perform a breath hold for at least the expected scan time;
  • Unstable angina;
  • Hemodynamic instability;
  • Pregnant or lactating women;
  • Renal insufficiency (defined as eGFR<30 ml/min/1.73m2);
Show all 13Show fewer
  • Previous history of adverse reactions to iodinated contrast agents;
  • Confirmed or suspected hyperthyroidism or pheochromocytoma;
  • Atrial fibrillation or arrhythmia; severe congestive heart failure (class IV according to NYHA classification of cardiac function);
  • History of CAD
  • Patients undergoing PCI before coronary CTA examination, or patients with a previous history of heart bypass surgery
  • Insufficient cannula venous access (preferred 18G, minimal 20G cannula);
  • Patients who are participating other clinical studies。

Trial locations and contacts

No sites are listed yet.

About this study

Cardiac computed tomographic angiography (CCTA) is a non-invasive diagnostic imaging technique for visualization of the coronary arteries and thus, frequently used in the evaluation of coronary artery disease (CAD). CT technology is improving continuously, and various technological advances not only increase diagnostic accuracy, but also provide a substantial reduction in radiation dose and scan acquisition time. These modifications challenge optimal synchronization and timing of scan protocols in CM administration due to a shorter data acquisition window, hereby creating opportunities for injection strategies with a decrease in total amount of contrast media (CM). [4-6] Reducing the radiation dose of CT angiography and injected CM dose have become a routine need and trend in clinical practice. In addition to this, physical factors such as cardiac output and body weight are considered important factors with regard to variability in vascular enhancement. A standard 'one size fits all' protocol with a standard injected CM volume, independent of weight and length of the patient has proven to be outdated and precision medicine in the future should be based on individually tailored scan and CM injection protocols that are more scientific and involve various parameters such as individual tube voltage, patients weight and heart rate to benefit patients by reducing radiation exposure and CM dose while fulfilling the diagnostic purpose.

Prospective studies focused on modifying both scan and injection parameters were completed in the Dutch (representing European) patient population with body weight varying between 40 and 130kg by Maastricht University Medical Center (MUMC) and have shown very promising results in Dutch population. In theory, these scan and injection protocols should be applicable to both average Dutch population (e.g. European population) and other heterogeneous patient populations, i.e. world-wide patient populations with any BMI category. However, so far, the effectiveness of this approach has not been deliberately discussed in the (on average) heavier North American population or the lighter Asian population. So, prior to promotion of the individually tailored CT scan protocols for global use, we need to obtain enough evidence in terms of the diagnostic confidence from those protocols in Chinese patients who could represent Asian population.

Read the detailed description

Primary objective: To prospectively evaluate the diagnostic quality (including the intravascular attenuation of the coronary arteries and image quality) in a previously validated individualized scan and CM injection CCTA protocol whereas both scan and injection parameters are tailored to the individual patient in Chinese population.

Secondary Objective(s):

  1. To evaluate the radiation dose and contrast media dose of this individualized approach for CCTA in a Chinese patient population.
  2. To evaluate the injection parameters required for the average patient population.

Subjects will be screened according to clinical data and enrolled in the study after providing informed consent. Patients will be scanned by Siemens Force CT scanner, prospective ECG gate-control and narrow window scanning technology, as well as intelligent KV (CARE-KV Tech) scanning technology. The CARE-kV software will determine optimal scan parameters for each patient, without any human intervention. By this means, each patient will be included in a specific tube-voltage arm based on in the condition of individual patient.

Care kV is a software that collects information about the patient's own attenuation values/ x-ray absorption during a so-called scout scan. Based on these parameters and taking into account the required scan quality (e.g. parenchymal, bone or vascular), the software determines the optimal scan settings for the patient. This includes both a variable tube current (in mAs) along the z-axis of the patient and an individual tube voltage (in kV). The tube voltage can be selected between 70 and 150kV in steps of 10kV. For this study, the scan quality slider will be set to 'vascular' or '11', which is commonly used in clinical studies and clinical practice for CCTA.

Iopromide injection (iodine concentration 370 mgI/ml) (Ultravist, Bayer Healthcare, Berlin, Germany) is administered as a bolus through the cubital vein using dual-syringe high-pressure injector (CT high-pressure injector: Stellant, MEDRAD, Pittsburg, PA, USA). The CM injection parameters are computed according to patient and examination-related parameters (patient weight, CM concentration, scan duration and maximum flow rate) using the Personalized Patients Protocol Technology (P3T) software supplied in the high-pressure injector.

P3T (Personalized Patient Protocol Technology) software computes individualized contrast injection protocols. P3T Cardiac/Pulmonary arteries are indicated for use with CAT of the cardiac structures, coronary arteries, chambers of the heart, pulmonary vasculature, thoracic and abdominal aorta. During initial setup, options such as minimum and maximum iodine loads, minimum injection duration, and other settings are configured according to site preferences and guidelines. These options are stored as a "preset", which can then be selected to calculate an injection protocol. At that time, patient and procedure specific parameters (patient weight, CM concentration, scan duration, and maximum flow rate) are entered by the radiographer. The software calculates the iodine load by multiplying the patient weight by a pre-determined dosing factor, applying the configured minimums and maximums. Then, the injection duration is determined based on the duration of the scan and minimum injection duration. Using the iodine load, injection duration, and CM concentration, the individualized flow rate and CM volumes are computed. The radiographer may edit the injection protocol if necessary. Dosing factors were derived based on pharmacokinetic model to simulate vascular enhancement in the regions of interest, targeting 350-400 HU across a wide range of patients.

An 18G or 20G venous access in the antecubital vein is preferred for higher flow rates. If low flow rates are required, venous access through an IV needle as small as 22G (max flow: 6.5 mL/sec) may be used.

  • Test bolus + saline flush, next CM bolus and saline flush 40ml. Prewarmed (37 º C [99˚F]) CM (Iopromide 370mgI/ml, Bayer).
  • Injection using a programmable dual-head CT injector (Stellant, Bayer), via an 18-20G intravenous injection catheter in the left or right antecubital vein. We propose to use the appropriate gauge for the expected flow rate, according to local safety regulations.
  • P3T Cardiac Certegra/ Use patient body weight adjusted protocol: P3T Cardiac, Certegra. (This will show flow rates if one sets the injection time and contrast media concentration).
  • Adjust for selected kV setting by adapting the IDR: -10% for 110kV, -20% for 100kV, -30% for 90kV, -40% for 80kV and -50% for 70kV.
  • In all patients, a test bolus injection at the level of the ascending aorta is performed to assess optimal start delay; 20ml of CM at a flow rate according to P3T protocol. Test bolus is followed by 40ml of saline flush at the same flow rate.

Study archiving: patient protocol (radiation dose information), ECG information, contrast media information, scout views calcium score scan, heart scan (Bv40, 0.6mm slices).

For this study we chose to include 300 patients to ensure a representation of normal patient variance. Given the assumed non-diagnostic rate to be around 5%, the 95% CI of non-diagnostic rate with 300 sample size is [2.8 %, 8.1%]. Although there is no hypothesis test planned for this study, the result from a recent published study taking similar design on individualized scan and injection protocol of liver CT provides support on adequate sample size of the current study. Based on the study, 256 patients in total could enable the stratification performed based on age (<60 and ≥60 years) and weight (<75 and ≥75 kg) in the population weighed below 115 kg.

If 300 patients are not included after 18 months, the inclusion period will be over none the less.

Medical subject headings
Coronary Artery Disease

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

Study groups in NCT05497466
Study groupRoleWhat they receive
Subjects who are male or female outpatients aged 18 years and scheduled for CCTA to rule out CAD300 patients scheduled for coronary CTA examination to rule out CAD in PUMCH. Each subject will be selected and assigned in a specific tube-voltage arm by the scanner based on the patient's body habitus for individualized CTA scans, without any intervention

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. The diagnostic quality produced by individualized scan and injection protocol.

    When using an individualized scan and injection protocol, definition of diagnostic quality of the scan is the imaging result could satisfy the following two criteria:

    1. Intravascular attenuation of the target coronary arteries should be over 325 Hounsfield Units (HU).
    2. Subjective overall image quality should be diagnostic (image quality > 2 of a 5-point Likert scale).

    Measured: 48 hours post examination

Secondary outcomes (11)
  1. Objective evaluation on image quality

    The images are transferred to the Siemens image processing workstation for post-processing. The post-vascular enhancement CT value of chambers of the heart, aorta and all branch vessels of coronary arteries is measured with standard deviation (SD) as image noise; the CT value of the inferior wall myocardium is measured with SD as background noise. Signal-To-Noise Ratio (SNR) is calculated as follows: SNR=enhanced CT value/SD.

    Measured: 48 hours post examination

  2. Subjective evaluation on image quality

    According to the 17-segment coronary artery segmentation standards of the American Heart Association, the images are scored by two radiologists who are blinded to protocol and have more than 5 years of work experience in CCTA with the 5-point Likert scale for scoring of coronary artery image quality: 5 points (excellent), images have fine anatomical details, sharp edges, and no obvious noise; 4 points (good), images have fine anatomical details and slightly increased noise; 3 points (fair), the visualization of most of anatomical results on the image satisfies diagnostic needs, with obvious but acceptable noise; 2 points (poor), images have blurred anatomical details that are difficult to distinguish and obvious noise; 1 point (very poor), images have extremely obvious noise that results in failed diagnosis.

    Measured: 48 hours post examination

  3. contrast media dose in ml

    injection parameters will be continuously monitored and automatically measured by a data acquisition program, Certegra™ Informatics Platform (Bayer Healthcare, Berlin, Germany). The contrast media dose could be recorded and read out directly in the system.

    Measured: During Injection procedure

  4. contrast media flow rate in ml/s

    injection parameters will be continuously monitored and automatically measured by a data acquisition program, Certegra™ Informatics Platform (Bayer Healthcare, Berlin, Germany). Contrast media flow rate could be recorded and read out from the system.

    Measured: During Injection procedure

  5. contrast media peak flow rate [ml/s]

    injection parameters will be continuously monitored and automatically measured by a data acquisition program, Certegra™ Informatics Platform (Bayer Healthcare, Berlin, Germany). The peak flow rate could be recorded and read out from the system

    Measured: During Injection procedure

  6. Time-to-peak of contrast media [s]

    injection parameters will be continuously monitored and automatically measured by a data acquisition program, Certegra™ Informatics Platform (Bayer Healthcare, Berlin, Germany). Time to peak of contrast media could be recorded and read out from the system

    Measured: During Injection procedure

  7. iodine delivery rate (IDR) in mgI/s

    IDR= concentration of contrast media (mgI/ml) administrated X flow rate(ml/s).injection parameters will be continuously monitored and automatically measured by a data acquisition program, Certegra™ Informatics Platform (Bayer Healthcare, Berlin, Germany). IDR of contrast media could be read out from the system.

    Measured: During Injection procedure

  8. Patient radiation dose:volume CT dose index (CTDIvol) in mGy

    Radiation parameters will be continuously monitored and automatically measured by a data acquisition programs Radimetrics Enterpise Platforms (Bayer Healthineers, Berlin, Germany). CTDIvol could be read out directly from the system.

    Measured: 48 hours post examination

  9. Patient radiation dose: size-specific dose estimates (SSDEs) of chest in mGy

    Radiation parameters will be continuously monitored and automatically measured by a data acquisition programs Radimetrics Enterpise Platforms (Bayer Healthineers, Berlin, Germany). SSDEs of chest could be read out directly from the system

    Measured: 48 hours post examination

  10. Patient radiation dose: dose length product (DLP) in mGy*cm

    Radiation parameters will be continuously monitored and automatically measured by a data acquisition programs Radimetrics Enterpise Platforms (Bayer Healthineers, Berlin, Germany). DLP could be read out from the system.

    Measured: 48 hours post examination

  11. Patient radiation dose: effective dose (ED) in mSv

    Radiation parameters will be continuously monitored and automatically measured by a data acquisition programs Radimetrics Enterpise Platforms (Bayer Healthineers, Berlin, Germany). ED could be read out from the system.

    Measured: 48 hours post examination

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted11 Aug 2022
  2. Study start1 Jan 2023 (estimated)
  3. Primary completion30 Jun 2024 (estimated)
  4. Study completion30 Jun 2024 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
300 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05497466
Sponsor's protocol ID
HS-2806 COE V1.3
Acronym
COE
Status last verified
Jul 2022
First submitted
31 Jul 2022
Last updated
11 Aug 2022
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided: There is a rule in China, for sharing any individual patient information, the study responsible must get authority's approval

Questions about this trial

What is this clinical trial studying?

An observational study of Coronary Artery Disease and Coronary Computed Tomographic Angiography, run by Peking Union Medical College Hospital.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 13 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 11 Aug 2022. Status can change between updates, so confirm with the study contact before making plans.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Zhengyu Jin (+86 10 69159553, jinzy@pumch.cn), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

The diagnostic quality produced by individualized scan and injection protocol. (time frame: 48 hours post examination). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 11 Aug 2022; AzTrials last refreshed its copy on 6 Nov 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in