Congenital Heart Disease in ChildrenNCT05617547
Evaluating Pulse Oximetry Bias in Children With Darker Skin Pigmentation
An observational study of Congenital Heart Disease in Children and Cardiomyopathies, run by Stanford University at 1 site in United States.
Trial summary
In this prospective study, the investigators will enroll 154 children with arterial lines to determine the accuracy of pulse oximeters in children with darker skin pigmentation. Studies in adults suggest pulse oximeters may overestimate the true level of oxygenation in the blood as measured directly by co-oximetry. Read more about this study
- Who is it for?
- Ages Up to 21 years
- Conditions the study is for
- Congenital Heart Disease in Children, Cardiomyopathies
- What is being tested?
- Pulse Oximeter (Massimo)
- How many people
- 154 participants
- Where
- 1 siteUnited States
- Study dates
- 31 Aug 2022to 1 Oct 2024
- Phase
- Observational
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages Up to 21 years
- All sexes
- Child, Adult
Study population: All children ≤21 years of age receiving care at Lucile Packard Children's Hospital and who require an arterial vascular access catheter as part of their clinical care will be eligible for inclusion in the study.
Who may be able to join
- Age ≤21 years of age
- Requires arterial vascular access as part of routine clinical care
- Patient or legally authorized representative (LAR) willing to provide written informed consent
Who cannot join
- Anemia defined as a Hgb <8 g/dL
- Methemoglobinemia or carbon monoxide poisoning where the SpO2 is known to be inaccurate
- Skin condition, such as epidermolysis bullosa, where pulse oximetry not expected to be accurate, or application of skin probes is medically inadvisable
- Non-pulsatile patients (e.g., left ventricular assist device (LVAD), extracorporeal membrane oxygenation (ECMO)
- Lack of informed consent
Trial locations and contacts
1 site in 1 country.
United States1 site, 1 recruiting
- Recruiting
Lucile Packard Children's Hospital
Palo Alto, California, 94304
Christopher Almond, MD, 650-724-2439, calmond@stanford.edu
Rohan Taneja, MBA, 6507242439, rohant@stanford.edu
About this study
Official title: Prospective Clinical Study to Evaluate the Accuracy of Pulse Oximeters in Children With Darker Skin Pigmentation
In this prospective study, the investigators will enroll 154 children with arterial lines to determine the accuracy of pulse oximeters in children with darker skin pigmentation. Studies in adults suggest pulse oximeters may overestimate the true level of oxygenation in the blood as measured directly by co-oximetry. However, pediatric data are relatively limited. This study, which is funded by the FDA through the Stanford-UCSF (University of California San Francisco) Clinical Excellence in Regulatory Science and Innovation (CERSI) Program, will determine if the error/bias is associated with skin pigmentation and whether the error falls outside FDA standards. The broader purpose of the study is to work toward eliminating health disparities.
Read the detailed description
As a conventional patient vital sign, pulse oximetry is used widely to determine whether a patient is adequately oxygenated. However, studies suggest that peripheral pulse oximetry (Sp02) systematically overestimates the true arterial oxygen saturation (Sa02) in patients with darker skin pigmentation. This error or bias places patients with darker skin pigmentation at considerable risk by failing to detect important levels of hypoxemia that drive critical treatment decisions like medication usage for Severe Acute Respiratory Syndrome (SARS)-COV-2 (COVID) infection, hospital admission, ICU transfer and intubation. Except for the original description of the problem published by Bickler and colleagues in 2005, most studies (including one pediatric study) have been limited to retrospective studies where the racial/ethnic category is used as a proxy for skin pigmentation, so-called "paired" oximetry measurements may be separated by several minutes where the oxygen saturation can readily change, and important technical factors such a perfusion quality, motion artifact, and light transmittance are not reliably documented. This information is critical because emerging studies suggest that perfusion effects, as measured by the perfusion index (PI), may account for much of the error observed in patients with darker pigmentation. The limitations of prior studies can be readily overcome with a prospective study where skin pigmentation is measured objectively and at the location of the oximeter sensor, SpO2/SaO2 (arterial oxygen saturation) measurements are collected nearly simultaneously while at steady state, and technical factors such as perfusion status, transmittance, and temperature can be accurately recorded. This study will enroll 154 children at Stanford's Lucile Packard Children's Hospital to determine whether the mean bias (error) in FDA-cleared pulse oximeters increases with darker skin pigmentation, and whether this error falls outside of the FDA standard of 3%. The primary study hypothesis is that in children ≤21 years of age managed in a real-world hospital setting, the mean bias (error) in FDA-cleared pulse oximeters increases with increased skin pigmentation as measured by colorimetry and standard pigmentation scales (Fitzpatrick and von Luschan pigmentation scales). The secondary hypothesis is that the mean bias is mitigated when technical factors like perfusion index are accounted for in adjusted models. Data generated from this study will help to identify the impact of darker skin pigmentation on medical device performance in the pediatric population and ultimately help eliminate health disparities.
- Being tested
- Pulse Oximeter (Massimo)
- Medical subject headings
- Cardiomyopathies
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Pulse Oximeter (Massimo)Diagnostic test
The Massimo LNCS (low noise cabled sensors) pulse oximeter will be used to estimate the oxygen saturation and compared to the gold standard, the arterial oxygen saturation in the blood as measured by co-oximetry. The pulse oximeters is FDA cleared for this purpose so patients will not be exposed to any interventions that are not standard of care.
| Study group | Role | What they receive |
|---|---|---|
| Pulse Oximetry Observational Cohort | Not described Receives: Diagnostic Test: Pulse Oximeter (Massimo) |
What is being measured
The primary outcome is the main question the study is designed to answer.
Saturation by pulse oximetry (Sp02)
Oxygen saturation as measured by pulse oximetry
Measured: Up to 2 weeks after consent and arterial line placement (where Sp02 will be compared to simultaneous measurements of the Sa02, or oxygen saturation using blood from the artery)
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted15 Nov 2022
- Study start31 Aug 2022
- Primary completion1 Oct 2024 (estimated)
- Study completion1 Oct 2024 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Congenital Heart Disease in ChildrenStudies of the same condition, open and under way.
- More trials run by Stanford UniversityThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Cohort
- Time perspective
- Cross-sectional
- Enrolment
- 154 participants (estimated)
Sponsor and investigators
- Lead sponsor
- Stanford University
- Sponsor type
- University, hospital or other
- Collaborators
- Food and Drug Administration (FDA), University of California, San Francisco
- Responsible party
- Principal investigator, Christopher Almond, Stanford University
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT05617547
- Sponsor's protocol ID
- SU_IRB_66947
- Status last verified
- Mar 2024
- First submitted
- 1 Sep 2022
- Last updated
- 1 Apr 2024
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated device
- Yes
- Individual data sharing
- Undecided
Study at a glance
From the registry record, last updated by the sponsor on 1 Apr 2024. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Prospective Clinical Study to Evaluate the Accuracy of Pulse Oximeters in Children With Darker Skin Pigmentation
- ClinicalTrials.gov ID
- NCT05617547
- Acronym / sponsor ID
- SU_IRB_66947
- Study type
- Observational
- Recruitment status
- Unknown status
- Enrolment
- 154 participants (estimated)
- Lead sponsor
- Stanford UniversityUniversity, hospital or other
- Collaborators
- Food and Drug Administration (FDA), University of California, San Francisco
- Study start
- 31 Aug 2022 (actual)
- Primary completion
- 1 Oct 2024 (estimated)
- Study completion
- 1 Oct 2024 (estimated)
- First posted
- 15 Nov 2022 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
In this prospective study, the investigators will enroll 154 children with arterial lines to determine the accuracy of pulse oximeters in children with darker skin pigmentation. Studies in adults suggest pulse oximeters may overestimate the true level of oxygenation in the blood as measured directly by co-oximetry. However, pediatric data are relatively limited. This study, which is funded by the FDA through the Stanford-UCSF (University of California San Francisco) Clinical Excellence in Regulatory Science and Innovation (CERSI) Program, will determine if the error/bias is associated with skin pigmentation and whether the error falls outside FDA standards. The broader purpose of the study is to work toward eliminating health disparities.
Detailed description
As a conventional patient vital sign, pulse oximetry is used widely to determine whether a patient is adequately oxygenated. However, studies suggest that peripheral pulse oximetry (Sp02) systematically overestimates the true arterial oxygen saturation (Sa02) in patients with darker skin pigmentation. This error or bias places patients with darker skin pigmentation at considerable risk by failing to detect important levels of hypoxemia that drive critical treatment decisions like medication usage for Severe Acute Respiratory Syndrome (SARS)-COV-2 (COVID) infection, hospital admission, ICU transfer and intubation. Except for the original description of the problem published by Bickler and colleagues in 2005, most studies (including one pediatric study) have been limited to retrospective studies where the racial/ethnic category is used as a proxy for skin pigmentation, so-called "paired" oximetry measurements may be separated by several minutes where the oxygen saturation can readily change, and important technical factors such a perfusion quality, motion artifact, and light transmittance are not reliably documented. This information is critical because emerging studies suggest that perfusion effects, as measured by the perfusion index (PI), may account for much of the error observed in patients with darker pigmentation. The limitations of prior studies can be readily overcome with a prospective study where skin pigmentation is measured objectively and at the location of the oximeter sensor, SpO2/SaO2 (arterial oxygen saturation) measurements are collected nearly simultaneously while at steady state, and technical factors such as perfusion status, transmittance, and temperature can be accurately recorded. This study will enroll 154 children at Stanford's Lucile Packard Children's Hospital to determine whether the mean bias (error) in FDA-cleared pulse oximeters increases with darker skin pigmentation, and whether this error falls outside of the FDA standard of 3%. The primary study hypothesis is that in children ≤21 years of age managed in a real-world hospital setting, the mean bias (error) in FDA-cleared pulse oximeters increases with increased skin pigmentation as measured by colorimetry and standard pigmentation scales (Fitzpatrick and von Luschan pigmentation scales). The secondary hypothesis is that the mean bias is mitigated when technical factors like perfusion index are accounted for in adjusted models. Data generated from this study will help to identify the impact of darker skin pigmentation on medical device performance in the pediatric population and ultimately help eliminate health disparities.
Study design
- Observational model
- Cohort
- Time perspective
- Cross-sectional
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Pulse Oximetry Observational Cohort | – | Diagnostic Test: Pulse Oximeter (Massimo) |
Who may qualify
- Ages
- Up to 21 years
- Sex
- All
- Age groups
- Child, Adult
- Study population
- All children ≤21 years of age receiving care at Lucile Packard Children's Hospital and who require an arterial vascular access catheter as part of their clinical care will be eligible for inclusion in the study.
Key inclusion criteria (3 listed)
- Age ≤21 years of age
- Requires arterial vascular access as part of routine clinical care
- Patient or legally authorized representative (LAR) willing to provide written informed consent
Key exclusion criteria (5 listed)
- Anemia defined as a Hgb <8 g/dL
- Methemoglobinemia or carbon monoxide poisoning where the SpO2 is known to be inaccurate
- Skin condition, such as epidermolysis bullosa, where pulse oximetry not expected to be accurate, or application of skin probes is medically inadvisable
- Non-pulsatile patients (e.g., left ventricular assist device (LVAD), extracorporeal membrane oxygenation (ECMO)
- Lack of informed consent
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Pulse Oximeter (Massimo)The Massimo LNCS (low noise cabled sensors) pulse oximeter will be used to estimate the oxygen saturation and compared to the gold standard, the arterial oxygen saturation in the blood as measured by co-oximetry. The pulse oximeters is FDA cleared for this purpose so patients will not be exposed to any interventions that are not standard of care. | Diagnostic test | Pulse Oximetry Observational Cohort |
Primary outcome
Saturation by pulse oximetry (Sp02)
Time frame: Up to 2 weeks after consent and arterial line placement (where Sp02 will be compared to simultaneous measurements of the Sa02, or oxygen saturation using blood from the artery)
Oxygen saturation as measured by pulse oximetry
Results and safety
Results are not available in the source record.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 1 Apr 2024.
- Individual participant dataSharing planned: Undecided
Related trials for Congenital Heart Disease in Children
Not yet recruiting
Autologous Thymus Transplant in Children With Congenital Heart Disease: A Pilot Study
Congenital Heart Disease (CHD)
RecruitingPhase 3
Trial Investigating the Efficacy and Safety of Weekly Lonapegsomatropin Compared to Daily Somatropin in Children and Adolescents With Short Stature or Growth Failure Due to Growth Hormone Sufficient Disorders
Turner Syndrome, Short Stature Homeobox Gene Mutation
RecruitingPhase 1/2
Butyric Acid Supplementation for Gut Improvement After Cardiac Surgery in Kids
Congenital Heart Disease (CHD)
Not yet recruiting
Starry-Sky Projection and Maternal Voice for Sleep After Pediatric Cardiac Surgery
Congenital Heart Disease (CHD), Postoperative Sleep
RecruitingObservational study
Defining the Brain Phenotype of Children With Williams Syndrome
Williams Syndrome, Duplication
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Congenital Heart Disease in Children and Cardiomyopathies, run by Stanford University at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: status not confirmed, as of the sponsor's last update on 1 Apr 2024. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Results are not available in the source record.
How can a doctor refer a patient or contact the research team?
Contact Christopher S Almond, MD, MPH (650-7237913, calmond@stanford.edu), or a site near the patient. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Saturation by pulse oximetry (Sp02) (time frame: Up to 2 weeks after consent and arterial line placement (where Sp02 will be compared to simultaneous measurements of the Sa02, or oxygen saturation using blood from the artery)). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 1 Apr 2024.