Pulmonary Disease, Chronic ObstructiveNCT05655832

A Study to Investigate the Association of Real-world Sensor-derived Biometric Data With Clinical Parameters and Patient-reported Outcomes for Monitoring Disease Activity in Patients With COPD

A study testing Vivalink wearable device in people with Pulmonary Disease, Chronic Obstructive, run by Merck Healthcare KGaA, Darmstadt, Germany, an affiliate of Merck KGaA, Darmstadt, Germany at 11 sites in Germany.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this multicenter, prospective cohort study is to investigate the correlation of real-world sensor-derived biometric data obtained via a wearable device with clinical parameters and patient-reported outcomes (PROs) for monitoring disease activity and predicting exacerbations for participants with Chronic… Read more about this study

Who is it for?
Ages 40 to 80 years
Conditions the study is for
Pulmonary Disease, Chronic Obstructive
What is being tested?
Vivalink wearable device
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
77 participants
Where
11 sitesGermany
Study dates
5 Dec 2022to 31 Oct 2023
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 40 to 80 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • For participants with COPD:
  • Participants ≥40 and ≤80 years at baseline
  • Diagnosis of COPD stage II to IV
  • History of moderate or severe exacerbations (≥2 moderate exacerbations or ≥1 severe exacerbations in any 12-month time window during last 3 years prior to inclusion and ≥1 moderate or severe exacerbations in the last 12 months prior to inclusion, considering that the last 12 months may reflect lower exacerbation rate due to Covid-19 measures)
  • For participants in the calibration cohort:
  • Participants ≥40 and ≤80 years at baseline

Who cannot join

  • For participants with COPD:
  • Clinically relevant and/or serious concurrent medical conditions including, but not limited to visual problems, severe mental illness or cognitive impairment, musculoskeletal or movement disorders, cardiac disease (e.g., heart failure, arrythmia [esp. atrial fibrillation and conduction blocks]), lung cancer (currently treated) that in the opinion of the Investigator, would interfere with participant's ability to participate in the study or draw meaningful conclusions from the study
  • Participants with a cardiac pacemaker, defibrillators, or other implanted electronic devices
  • Participants with known allergies or sensitivity to silicon or hydrogel
  • Less than 6 weeks since previous moderate/severe exacerbation
  • For participants in the calibration cohort:
Show all 9Show fewer
  • Participants with a cardiac pacemaker, defibrillators, or other implanted electronic devices
  • Participants with known allergies or sensitivity to silicon or hydrogel
  • Diagnosis of pulmonary disease including, but not limited to COPD, asthma, pulmonary fibrosis, with impact on the lung function and exercise capacity

Trial locations and contacts

11 sites in 1 country.

11 sites in 1 country

Germany11 sites
  • Praxis an der Oper

    Berlin

  • Lungenzentrum Darmstadt GmbH

    Darmstadt

  • Städtische Kliniken Darmstadt

    Darmstadt

  • Lungenzentrum Frankfurt

    Frankfurt

  • Thoraxklinik Heidelberg gGmbH

    Heidelberg

  • ZERO Praxen

    Mannheim

  • Pneumologisches Studienzentrum München-West

    München

  • Pneumologische Gemeinschaftspraxis Saarbrücken

    Saarbrücken

  • RespiRatio / Lungenpraxis

    Schleswig

  • Pneumologische Praxis Wiesbaden

    Wiesbaden

  • Lungenpraxis Dr. Franz / Dr. Weber

    Witten

Other trials in Germany

About this study

Official title: A Study to Investigate the Association of Real-world Sensor-derived Biometric Data With Clinical Parameters and Patient-reported Outcomes for Monitoring Disease Activity in Patients With Chronic Obstructive Pulmonary Disease (COPD)

The purpose of this multicenter, prospective cohort study is to investigate the correlation of real-world sensor-derived biometric data obtained via a wearable device with clinical parameters and patient-reported outcomes (PROs) for monitoring disease activity and predicting exacerbations for participants with Chronic Obstructive Pulmonary Disease (COPD). The cohort of participants with COPD will be followed for 3 months. A calibration cohort with non-COPD participants will be included and followed for 2 weeks.

Being tested
Vivalink wearable device
Medical subject headings
Pulmonary Disease, Chronic Obstructive

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Vivalink wearable deviceDevice

    a CE marked device modified to add a temperature measurement algorithm in addition to ECG and respiratory rate measurements

Study groups in NCT05655832
Study groupRoleWhat they receive
COPD cohortExperimentalNot described
Receives: Device: Vivalink wearable device
Calibration participants cohortExperimentalNot described
Receives: Device: Vivalink wearable device

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Physical Activity

    An activity flag is extracted from the accelerometer by Vivalink, by using a predefined threshold for adult movement. For stair climbing, first periodic movement was determined, by using frequency analysis on specific time windows, and generating a ratio to the total spectrum indicating periodic activity over a certain threshold.

    Measured: Day 0(Baseline) and Day 8 to Day 14

  2. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Heart Rate

    Heart rate is provided by Vivalink.

    Measured: Day 0(Baseline) and Day 8 to Day 14

  3. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Heart Rate Variability (SDRR, SDNN, SDNNI, RMSSD, ln(RMSSD))

    Heart rate variability reflecting differences in time intervals between 2 R-waves in the ECG (milliseconds) SDRR (Standard Deviation of Intervals between Heartbeats), SDNN (Standard Deviation of Intervals between Heartbeats, after removing abnormal Beats), SDNNI (Mean of the Standard Deviations of all the NN intervals for each 5 min Segment of a 24-h HRV Recording), and RMSSD (Mean of the Standard Deviations of all the NN intervals for each 5 min Segment of a 24-hour HRV Recording) and In(RMSDD)

    Measured: Day 0(Baseline) and Day 8 to Day 14

  4. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Heart Rate Variability (pNN50)

    pNN50 is the percentage of adjacent NN intervals that differ from each other by more than 50 milliseconds.

    Measured: Day 0(Baseline) and Day 8 to Day 14

  5. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Heart Rate Variability (Stress Index)

    Baevsky's Stress Index is a heart rate variability (HRV) measure used to assess autonomic nervous system activity and physiological stress, especially in monitoring chronic obstructive pulmonary disease (COPD) exacerbations. It is calculated as: amplitude of the mode (AMo) divided by two times the mode (Mo) multiplied by the difference between the maximum and minimum RR intervals (MxDMn). AMo is the percentage of RR intervals at the most frequent value, Mo is the most common RR interval, and MxDMn is the range of RR intervals. The index typically ranges from 50 to over 900. Lower values (50-150) indicate low stress and better autonomic balance, while higher values (above 500) reflect increased stress and sympathetic activity. Values above 900 are considered very high stress. This is a single composite score with no subscales; higher scores represent worse outcomes.

    Measured: Day 0(Baseline) and Day 8 to Day 14

  6. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Heart Rate Variability (LF and HF)

    Applying a Fast Fourier Transformation (FFT) or autoregressive (AR) modeling one can separate Heart rate variability (HRV) into its component ultra-low-frequency (ULF), very low frequency (VLF), Low-Frequency power (LF), and High-Frequency power (HF) rhythms that operate within different frequency ranges. Given in absolute values of power (milliseconds squared). LF power, low frequency power (0.04-0.15 Hz). HF power, high frequency power (0.15-0.40 Hz). LF/HF Ratio, spectral HRV index computed as (LF/HF).

    Measured: Day 0(Baseline) and Day 8 to Day 14

  7. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Heart Rate Variability (LF/HF)

    Applying a Fast Fourier Transformation (FFT) or autoregressive (AR) modeling one can separate Heart rate variability (HRV) into its component ultra-low-frequency (ULF), very low frequency (VLF), Low-Frequency power (LF), and High-Frequency power (HF) rhythms that operate within different frequency ranges. Given in absolute values of power (milliseconds squared). LF power, low frequency power (0.04-0.15 Hz). HF power, high frequency power (0.15-0.40 Hz). LF/HF Ratio, spectral HRV index computed as (LF/HF).

    Measured: Day 0(Baseline) and Day 8 to Day 14

  8. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Temperature

    Temperature is provided by Vivalink. The value for temperature is derived by Vivalink from the display temperature and then calibrated using initial calibration values, in an IP protected process. The sensor temperature is considered only as a relative value to evaluate changes in the temperature, and not as an objective human body temperature value, meaning no thresholds relative to normal human body temperature are considered, and it will not be used as a marker for fever or hypothermia.

    Measured: Day 0(Baseline) and Day 8 to Day 14

  9. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Respiratory Rate

    Respiration rate is provided by Vivalink.

    Measured: Day 0(Baseline) and Day 8 to Day 14

  10. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Cough Frequency

    Cough Frequency was provided by vivalink.

    Measured: Day 0(Baseline) and Day 8 to Day 14

  11. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Sleep Patterns

    The basis of the sleep pattern calculations is the self-reported bedtimes. With the same technique as the cough frequency prediction, inactivity signals can be predicted from the labeled data to improve the bedtime accuracy, and the changes in accelerometer (step detection algorithms) can be used to quantify the number of clear breaks in the sleep (standing up, strong cough, etc.).

    Measured: Day 0(Baseline) and Day 8 to Day 14

  12. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Resting Heart Rate

    Resting Heart Rate is provided by Vivalink.

    Measured: Day 0(Baseline) and Day 8 to Day 14

  13. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Inspiration vs Expiration Time Ratio

    Using the breathing signal one can determine the inspiration and expiration peaks. The difference between said peaks in milliseconds can be used to determine the ratio of inspiration (distance from lower point to next peak) vs expiration (distance from peak to next lower point).

    Measured: Day 0(Baseline) and Day 8 to Day 14

  14. Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Frequency of Additional Medication

    Count of the number of times the use of additional medication as a log activity is reported per day.

    Measured: Day 0(Basseline) and Day 8 to Day 14

  15. Prediction of Moderate or Severe COPD Exacerbations by Building a Statistical Model Employing Sensor-Derived Data and Demographic and Medical Covariates - Accuracy

    Accuracy was calculated as (True Positives + True Negatives) / Total Population. True Positives (TP) are events correctly predicted as exacerbations. True Negatives (TN) are events correctly predicted as non-exacerbations. Total Population refers to the total number of events evaluated. Accuracy scores reflect XGBoost algorithm performance using random and time-based 70/30 data splits. The values were calculated in form of percentage where 100% is the ideal scenario for perfect predictability.

    Measured: Up to 3 months

  16. Prediction of Moderate or Severe COPD Exacerbations by Building a Statistical Model Employing Sensor-Derived Data and Demographic and Medical Covariates - Precision

    Precision was calculated as True Positives / (True Positives + False Positives). True Positives (TP) are events correctly predicted as exacerbations. False Positives (FP) are events incorrectly predicted as exacerbations. Total Population refers to the total number of events evaluated. Precision scores reflect XGBoost algorithm performance using random and time-based 70/30 data splits. The values were calculated in form of percentage where 100% is the ideal scenario for perfect predictability.

    Measured: Up to 3 months

  17. Prediction of Moderate or Severe COPD Exacerbations by Building a Statistical Model Employing Sensor-Derived Data and Demographic and Medical Covariates - Recall

    Recall was calculated as True Positives / (True Positives + False Negatives). True Positives (TP) are events correctly predicted as exacerbations. False Negatives (FN) are events incorrectly predicted as non-exacerbations. Total Population refers to the total number of events evaluated. Recall scores reflect XGBoost algorithm performance using random and time-based 70/30 data splits. The values were calculated in form of percentage where 100% is the ideal scenario for perfect predictability.

    Measured: Up to 3 months

  18. Prediction of Moderate or Severe COPD Exacerbations by Building a Statistical Model Employing Sensor-Derived Data and Demographic and Medical Covariates - Specificity

    Specificity was calculated as True Negatives / (True Negatives + False Positives). True Negatives (TN) are events correctly predicted as non-exacerbations. False Positives (FP) are events incorrectly predicted as exacerbations. Total Population refers to the total number of events evaluated. Specificity scores reflect XGBoost algorithm performance using random and time-based 70/30 data splits. The values were calculated in form of percentage where 100% is the ideal scenario for perfect predictability.

    Measured: Up to 3 months

Secondary outcomes (19)
  1. Correlation of Sensor-Collected Data With COPD Assessment Test (CAT) Questionnaire: Participants Health Status and Symptoms at Baseline and Study End

    Participants health status and symptoms at baseline (Day 0) and study end will be measured as the summary score across items of the CAT questionnaire that consists of 8-items in which participants can choose a score from 0 to 5, for each visit.

    Measured: Baseline (Day 0) and at 3 months

  2. Correlation of Sensor-Collected Data With Lung Function (FEV1) at Baseline and Study End

    Lung function was assessed using plethysmography.

    Measured: Baseline (Day 0) and at 3 months

  3. Correlation of Sensor-Collected Data With Lung Function (FVC) at Baseline and Study End

    Lung function was assessed using plethysmography.

    Measured: Baseline (Day 0) and at 3 months

  4. Correlation of Sensor-Collected Data With Lung Function (FEV1/FVC) at Baseline and Study End

    Lung function was assessed using plethysmography.

    Measured: Baseline (Day 0) and at 3 months

  5. Correlation of Sensor-Collected Data With COPD Assessment Test (CAT) Questionnaire: Lung Function and Lab Values at Baseline and Study End (White Blood Cells Count)

    Lung function was assessed using plethysmography and lab values including Complete Blood Count with differential, Blood Gas Analysis, procalcitonin and CRP will be assessed as per standard practice at baseline (Day 0) and at study end

    Measured: Baseline (Day 0) and at 3 months

  6. Correlation of Sensor-Collected Data With COPD Assessment Test (CAT) Questionnaire: Lung Function and Lab Values at Baseline and Study End (Erythrocytes Count)

    Lung function was assessed using plethysmography and lab values including Complete Blood Count with differential, Blood Gas Analysis, procalcitonin and CRP will be assessed as per standard practice at baseline (Day 0) and at study end

    Measured: Baseline (Day 0) and at 3 months

  7. Correlation of Sensor-Collected Data With COPD Assessment Test (CAT) Questionnaire: Lung Function and Lab Values at Baseline and Study End (Partial Pressure of Oxygen (pO2))

    Lung function was assessed using plethysmography and lab values including Complete Blood Count with differential, Blood Gas Analysis, procalcitonin and CRP will be assessed as per standard practice at baseline (Day 0) and at study end

    Measured: Baseline (Day 0) and at 3 months

  8. Correlation of Sensor-Collected Data With COPD Assessment Test (CAT) Questionnaire: Lung Function and Lab Values at Baseline and Study End (Partial Pressure of Carbon Dioxide (pCO2))

    Lung function was assessed using plethysmography and lab values including Complete Blood Count with differential, Blood Gas Analysis, procalcitonin and CRP will be assessed as per standard practice at baseline (Day 0) and at study end

    Measured: Baseline (Day 0) and at 3 months

  9. Correlation of Sensor-Collected Data With COPD Assessment Test (CAT) Questionnaire: Lung Function and Lab Values at Baseline and Study End (O2 Saturation)

    Lung function was assessed using plethysmography and lab values including Complete Blood Count with differential, Blood Gas Analysis, procalcitonin and CRP will be assessed as per standard practice at baseline (Day 0) and at study end

    Measured: Baseline (Day 0) and at 3 months

  10. Correlation of Sensor-Collected Data With Number, Date of Onset, and Duration of Mild, Moderate, and Severe Exacerbations

    Exacerbations are classified as mild if they are treated with short-acting bronchodilators only, moderate if they are treated additionally with antibiotics or oral corticosteroids, or severe if the patient visits the emergency room or requires hospitalization because of an exacerbation.

    Measured: Up to 3 months

  11. Association Between Sensor Parameters (Heart Rate and Resting Heart Rate) and CAT Score

    During the observation period the CAT score was obtained via a digital application daily. The daily CAT questionnaire summary score was computed. The fixed effect estimate represents the change in CAT score per unit change in the corresponding parameter. Linear mixed models were performed to assess the association between the CAT score and each sensor parameter (Heart Rate and Resting Heart Rate). Data was calculated through linear mixed model; reported as "fixed effect estimate" with measure type as "number" and measure dispersion as "95% CI."

    Measured: 7 days before Severe/Moderate Excarbations(S/M E) (7-day window period)

  12. Association Between Sensor Parameters (Respiration Rate) and CAT Score

    During the observation period the CAT score was obtained via a digital application daily. The daily CAT questionnaire summary score was computed. The fixed effect estimate represents the change in CAT score per unit change in the corresponding parameter. Linear mixed models were performed to assess the association between the CAT score and each sensor parameter (Respiration Rate). Data was calculated through linear mixed model; reported as "fixed effect estimate" with measure type as "number" and measure dispersion as "95% CI."

    Measured: 7 days before Severe/Moderate Excarbations(S/M E) (7-day window period) and 14 days before S/M E (1 day window period)

  13. Association Between Sensor Parameters (SDRR, SDNN, SDNNI, RMSSD, In(RMSSD)) and CAT Score

    During the observation period the CAT score was obtained via a digital application daily. The daily CAT questionnaire summary score was computed. The fixed effect estimate represents the change in CAT score per unit change in the corresponding parameter. Linear mixed models were performed to assess the association between the CAT score and each sensor parameter (SDRR, SDNN, SDNNI, RMSSD, In(RMSSD)). Data was calculated through linear mixed model; reported as "fixed effect estimate" with measure type as "number" and measure dispersion as "95% CI."

    Measured: 7 days before Severe/Moderate Excarbations(S/M E) (7-day window period) and 14 days before S/M E (1 day window period)

  14. Association Between Sensor Parameters (Stress Index, LF/HF) and CAT Score

    During the observation period, participants completed the CAT questionnaire daily via a digital app. Stress Index, based on heart rate variability (HRV), assessed autonomic activity and physiological stress. It was calculated as AMo/(2 * Mo * MxDMn), where AMo is the % of RR intervals at the most frequent value, Mo is the most common RR interval, and MxDMn is the RR interval range. Stress Index values range from 50-900; lower values (50-150) indicate low stress and better autonomic balance, while higher values (>500) reflect increased stress and sympathetic activity. LF power reflects sympathetic activity; HF power reflects parasympathetic activity. Linear mixed models assessed associations between CAT score and each sensor parameters. Fixed effect estimates represent change in CAT score per unit change in each parameter. Data was calculated through linear mixed model; reported as "fixed effect estimate" with measure type as "number" and measure dispersion as "95% CI".

    Measured: 7 days before Severe/Moderate Excarbations(S/M E) (7-day window period) and 14 days before S/M E (1 day window period)

  15. Association Between Sensor Parameters (pNN50) and CAT Score

    During the observation period the CAT score was obtained via a digital application daily. The daily CAT questionnaire summary score was computed. The fixed effect estimate represents the change in CAT score per unit change in the corresponding parameter. Linear mixed models were performed to assess the association between the CAT score and each sensor parameter (pNN50). Data was calculated through linear mixed model; reported as "fixed effect estimate" with measure type as "number" and measure dispersion as "95% CI."

    Measured: 7 days before Severe/Moderate Excarbations(S/M E) (7-day window period) and 14 days before S/M E (1 day window period)

  16. Association Between Sensor Parameters (Temperature) and CAT Score

    During the observation period the CAT score was obtained via a digital application daily. The daily CAT questionnaire summary score was computed. The fixed effect estimate represents the change in CAT score per unit change in the corresponding parameter. Linear mixed models were performed to assess the association between the CAT score and each sensor parameter (Temperature). Data was calculated through linear mixed model; reported as "fixed effect estimate" with measure type as "number" and measure dispersion as "95% CI."

    Measured: 14 days before S/M E (1 day window period)

  17. Association Between Sensor Parameters (Physical Activity) and CAT Score

    During the observation period the CAT score was obtained via a digital application daily. The daily CAT questionnaire summary score was computed. The fixed effect estimate represents the change in CAT score per unit change in the corresponding parameter. Linear mixed models were performed to assess the association between the CAT score and each sensor parameter (Physical activity). Data was calculated through linear mixed model; reported as "fixed effect estimate" with measure type as "number" and measure dispersion as "95% CI."

    Measured: 14 days before S/M E (1 day window period)

  18. Association Between Sensor Parameters (Sleep Pattern) and CAT Score

    During the observation period the CAT score was obtained via a digital application daily. The daily CAT questionnaire summary score was computed. The fixed effect estimate represents the change in CAT score per unit change in the corresponding parameter. Linear mixed models were performed to assess the association between the CAT score and each sensor parameter (Sleep pattern). Data was calculated through linear mixed model; reported as "fixed effect estimate" with measure type as "number" and measure dispersion as "95% CI."

    Measured: 14 days before S/M E (1 day window period)

  19. Predicting the CAT Score by Building a Statistical Model Employing Sensor-Derived Data and Demographic and Medical Covariates

    Patients' health status and symptoms at baseline (Day 0) were measured using the CAT questionnaire, an 8-item tool with scores ranging from 0 to 5 per item. CAT scores were collected daily via a digital application during the observation period. Various machine learning algorithms were evaluated for predictive performance using metrics including accuracy, specificity, sensitivity, precision, positive predictive value (PPV), negative predictive value (NPV), and area under the ROC curve. R² (coefficient of determination) was computed for CAT score prediction models, defined as R² = 1 - (SS_res / SS_tot), where SS_res is the residual sum of squares and SS_tot is the total sum of squares. R² values range from 0 to 1, with higher values indicating better model fit.

    Measured: Up to 3 months

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted19 Dec 2022
  2. Study start5 Dec 2022
  3. Primary completion31 Oct 2023
  4. Study completion31 Oct 2023

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Non-randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
77 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05655832
Sponsor's protocol ID
MS202559_0001
Status last verified
Dec 2025
First submitted
18 Nov 2022
Last updated
15 Jan 2026
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No: We are committed to enhancing public health through responsible sharing of clinical trial data. Following approval of a new product or a new indication for an approved product in both the US and the European Union, the study sponsor and/or its affiliated companies will share study protocols, anonymized patient data and study level data, and redacted clinical study reports with qualified scientific and medical researchers, upon request, as necessary for conducting legitimate research. Further information on how to request data can be found on our website http://bit.ly/IPD21

Questions about this trial

What is this clinical trial studying?

A study testing Vivalink wearable device in people with Pulmonary Disease, Chronic Obstructive, run by Merck Healthcare KGaA, Darmstadt, Germany, an affiliate of Merck KGaA, Darmstadt, Germany at 11 sites in Germany.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 9 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 15 Jan 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 11 sites in Germany. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 15 Jan 2026). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Chronic Obstructive Pulmonary Disease (COPD) Exacerbations of Sensor-collected Parameters During Observation Period - Physical Activity (time frame: Day 0(Baseline) and Day 8 to Day 14). The study lists 18 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 15 Jan 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in