HypertensionNCT05883605
Clinical Validation of The Combei BP 880 W
A study testing Measuring blood pressure in people with Hypertension, run by International Society for Vascular Health at 1 site in France.
Trial summary
Clinical validation of the wrist blood pressure measuring device Combei BP 880 W for home use according to "the Universal Standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for… Read more about this study
- Who is it for?
- Ages 18 to 80 yearsHealthy volunteers can take part
- Conditions the study is for
- Hypertension
- What is being tested?
- Measuring blood pressure
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 85 participants
- Where
- 1 siteFrance
- Study dates
- 10 Jan 2023to 27 May 2023
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 80 years
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- Patient older than 12 years;
- Patient who signed the informed consent form;
- Patient followed-up at site (in-patient or out-patient);
- Patient with wrist circumference between 12.5 and 21.5 cm*. *in accordance with the Indications For Use of the Combei BP 880 W device.
Who cannot join
- Patient unable to give a consent or understand properly protocol information;
- Patient suffering from arrhythmia;
- Patient with poor quality of Korotkov sounds;
- Patient for whom K5 sounds are absent.
Trial locations and contacts
1 site in 1 country.
About this study
Official title: Clinical Validation of the Wrist Blood Pressure Measuring Device Combei BP 880 W for Home Use According to "the Universal" Standard (AAMI/ESH/ISO)" in the General Population
Clinical validation of the wrist blood pressure measuring device Combei BP 880 W for home use according to "the Universal Standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO)" in the general population.
Read the detailed description
Members of the AAMI, ESH and ISO committees achieved a consensus on an optimal validation standard and agreed on a single protocol for the validation of BP measuring devices that will replace all previous ones.
Therefore, the Association for the Advancement of Medical Instrumentation / European Society of Hypertension / International Organization for Standardization Universal Standard (AAMI/ESH/ISO 81060-2:2018) is now considered as the standard protocol for the validation of non-invasive blood pressure measuring device.
This study aims to demonstrate that the Combei BP 880 W wrist device fulfills the requirements of the new protocol of clinical validation and its Amendment 1, 2020-01 for limb circumference distribution criteria.
this is a monocentric, prospective, non-randomized, non-interventional, cross-study.
Combei BP 880 W wrist device (Sphygmomanometer Under Test, SUT). The reference device is the mercury sphygmomanometer The primary objective of the study is to assess the accuracy of the automatic oscillometric BP measuring device at the wrist level, the Combei BP 880 W, in the general population, following AAMI/ESH/ISO 81060-2:2018 protocol and its Amendment 1, 2020-01 for limb circumference distribution criteria.
The Combei BP 880 W device will pass the test if:
- The differences of at least 255 individual paired determinations of the SUT and of the observers' readings with the reference sphygmomanometer for all subjects, calculated separately for systolic blood pressure and diastolic blood pressure, shall have the mean value of the differences, be within or equal to +/- 5.0mmHg and a standard deviation no greater than 8.0 mmHg.
- For the systolic blood pressure and diastolic blood pressure of at least 85 subjects, the standard deviation of the averaged paired determinations per subject of the SUT and of the observers' readings with the reference sphygmomanometer shall meet the criteria listed in Table 1 of AAMI/ESH/ISO 81060-2:2018.
The secondary objective is to confirm the safety of use of this device, intended for home use.
Inclusion criteria
- Patient older than 12 years;
- Patient who signed the informed consent form;
- Patient followed-up at site (in-patient or out-patient);
- Patient with wrist circumference between 12.5 and 21.5 cm*. *in accordance with the Indications For Use of the Combei BP 880 W device.
Non-inclusion criteria
- Patient unable to give a consent or understand properly protocol information;
- Patient suffering from arrhythmia;
- Patient with poor quality of Korotkov sounds;
- Patient for whom K5 sounds are absent. The validation team will consist of three persons: two observers trained in accurate BP measurement and a supervisor.
- The reference sphygmomanometer is calibrated just before the beginning of the study.
- The observers are trained to the use of the SUT.
- In-patients or out-patients are included in the study after reading and signing the ICF.
- After inclusion, the patient is screened to verify that he/she satisfies the inclusion/non-inclusion criteria of the protocol. If not, the patient will be excluded from the study and no blood pressure measure will be performed on this patient.
- The patient is placed comfortably, and the reference sphygmomanometer is placed on his/her left upper arm, except if left upper arm is wounded. In this case, the reference sphygmomanometer is placed on his/her right upper arm.
- Measurements start with alternance between measurements with the reference sphygmomanometer and the SUT according to the same arm sequential BP measurements method.
- For each measurement with the reference sphygmomanometer, the two observers using mercury sphygmomanometers and a double-headed stethoscope, blinded from each other's result, and then by the supervisor using the tested device according to the manufacturer recommendations. The diastolic blood pressure should be measured at Korotkoff K5 sound. The agreement between the 2 observers will be checked all over the evaluation period by the supervisor to make sure that the difference between the two observers is no more than 4 mmHg for systolic and diastolic BP values.Otherwise, the measurement should be repeated. The intra individual BP variability for the reference BP measurements must be < 12 mmHg for SBP and < 8 mmHg for DBP.
- The results are written down on a source document and then reported in the CRF. Any AE, ADE or device malfunction occurring during the protocol will be also documented and reported in the CRF. For this study, all source documents should be written in English language.
- At the end of the measurements, the patient participation to the study is over.
Study procedures:
- Validation team and pre-study training The validation team will consist of three persons: two observers trained in accurate BP measurement and a supervisor. The reference sphygmomanometer is calibrated just before the beginning of the study. The observers are trained to the use of the SUT.
- Inclusion of the patients In-patients or out-patients of the investigational site are included in the study after reading and signing the ICF. After inclusion, the patient is screened to verify that he/she satisfies the inclusion/non-inclusion criteria of the protocol. If not, the patient will be excluded from the study and no blood pressure measure will be performed on this patient.
- Cuffs The cuffs used for reference auscultatory BP measurement will have an inflatable bladder length that covers 75-100% of the arm circumference of each patient and width that covers 37-50% of the arm circumference. Detailed description of cuffs used for reference BP measurement will be reported for each patient.
- Blood pressure measurements
The validation will be performed according to the (AAMI/ESH/ISO) Universal Standard validation protocol described in detail elsewhere10. The patients will have their reference BP measured by the two observers using a mercury sphygmomanometer and a double-headed stethoscope, blinded from each other's result, and then by the supervisor using the tested devices. The agreement between the 2 observers will be checked all over the evaluation period by the supervisor to make sure that the difference between the two observers is no more than 4 mmHg for systolic and diastolic BP values. Otherwise, the reference measurement should be repeated. The intra individual variability of the reference BP measurements must be < 12 mm Hg for SBP and < 8 mm Hg for DBP.
The circumference of the patient's upper arm will be measured to ensure that the cuff-size being used is adequate for the subject. The upper arm midpoint is first determined by marking the arm posteriorly at a point halfway between the acromion and olecranon, measured while the arm is flexed 90 degrees at the elbow with the palm facing up. The subject's upper arm circumference shall be determined by measuring at the midpoint of the upper arm while the elbow is relaxed, and the arm is dangling freely to the side.
The circumference of the patient's wrist will be measured to ensure that the circumference is within the acceptable range 12.5 - 21.5 cm for the Combei device, otherwise the subject will be excluded.
Measurements made by the mercury sphygmomanometer, will be performed according to the "same arm, consecutive measurements" (Figure 1), on the left arm supported at heart level.
Measurements by the Combei BP 880 W device will be performed at the wrist level of the same arm supported at the heart level, as recommended by the manufacturer with the following procedure.
- Ask the subject to relax for 5-10 min.
- make sure that the subject :
- is comfortably seated with legs uncrossed and feet flat on the floor,
- has the back, elbow and forearm supported,
- has the measurement site at the level of the left ventricle of the heart,
- and other conditions as recommended by the validation protocol and the device manual user.
In total, nine consecutive BP measurements will be performed in each patient using the mercury sphygmomanometers (5 times) and the tested devices (4 times).
All nine-sequential same-arm measurements will be recorded as shown in Table 2, starting with the standard mercury sphygmomanometer, followed by the device. A resting period of a minimum of 60 seconds shall be observed between the end of one measurement and the beginning of the following measurement.
Table 2: Sequential measurements of BP recorded by the standard mercury sphygmomanometer and the tested device:
R0 Entry BP, observers 1 and 2 each with the mercury standard T0 Device detection BP, supervisor R1 Observers 1 and 2 with mercury standard T1 Supervisor with the test instrument R2 Observers 1 and 2 with mercury standard T2 Supervisor with the test instrument R3 Observers 1 and 2 with mercury standard T3 Supervisor with the test instrument R4 Observers 1 and 2 with mercury standard
The first auscultatory and the first device measurement represent the recruitment pressures (R0 and T0) are not used in the evaluation of the test device accuracy. Each of the reference BP measurements (R) is the average of the simultaneous readings of the two observers. For each subject, the device measurements T1, T2 and T3 will be compared to observer measurements. Each of the test device measurements is compared to the average of the previous and succeeding reference BP readings (T1 vs average of R1 and R2, T2 vs average of R2 and R3, T3 vs average of R3 and R4).
At the end of the measurements, the patient participation to the study is over.
The Statistical Analysis: Differences are calculated by subtracting the reference BP measurement from the test device measurement. The mean BP difference (SUT versus reference sphygmomanometer) and its standard deviation (SD) will be calculated for the analysis. The same criteria will be used for systolic blood pressure and diastolic blood pressure measurements. The statistical analysis will be performed according to the requirements of the AAMI/ESH/ISO standard 2018 and its Amendment 1, 2020-01 for limb circumference distribution criteria.
- Conditions
- Hypertension
- Being tested
- Measuring blood pressure
- Keywords
- validation, Blood pressure., wrist device, electronic BP device, home Blood pressure
- Medical subject headings
- Hypertension
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Measuring blood pressureDevice
Measuring blood pressure using electronic device and using a mercury sphygmomanometer
| Study group | Role | What they receive |
|---|---|---|
| one population of normotensive and hypertensive | Other | treated and untreated normotensive and hypertensive patients Receives: Device: Measuring blood pressure |
What is being measured
The primary outcome is the main question the study is designed to answer.
Accuracy of blood pressure measurement
differences < than 5 +/- 8 mmHg according to the protocol by comparison to the gold standard (mercury sphygmomanometer)
Measured: Subject will participate to only one session of measurement. The duration of one session is about 45 minutes.
Secondary outcomes (1)
differences between the tested device and the gold standard (mercury sphygmomanometer according to the baseline blood pressure and to the wrist circumference
differences < than 5 +/- 8 mmHg according to the protocol by comparison to the gold standard (mercury sphygmomanometer)
Measured: Subject will participate only one time to the study. one session of blood pressure measurement will be performed. the duration of one session is about 45 minutes
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted1 Jun 2023
- Study start10 Jan 2023
- Primary completion23 May 2023
- Study completion27 May 2023
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for HypertensionStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Other
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 85 participants (actual)
Sponsor and investigators
- Lead sponsor
- International Society for Vascular Health
- Sponsor type
- University, hospital or other
- Responsible party
- Principal investigator, Catherine DOGNON HADDAD, International Society for Vascular Health
- Investigators
- ROLAND ASMAR, Study chair, International Society of Vascular Health
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT05883605
- Sponsor's protocol ID
- VALCOMBW22
- Status last verified
- May 2023
- First submitted
- 31 May 2022
- Last updated
- 1 Jun 2023
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
Study at a glance
From the registry record, last updated by the sponsor on 1 Jun 2023. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Clinical Validation of the Wrist Blood Pressure Measuring Device Combei BP 880 W for Home Use According to "the Universal" Standard (AAMI/ESH/ISO)" in the General Population
- ClinicalTrials.gov ID
- NCT05883605
- Acronym / sponsor ID
- VALCOMBW22
- Conditions
- Hypertension
- Keywords
- validation, Blood pressure., wrist device, electronic BP device, home Blood pressure
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 85 participants (actual)
- Lead sponsor
- International Society for Vascular HealthUniversity, hospital or other
- Study start
- 10 Jan 2023 (actual)
- Primary completion
- 23 May 2023 (actual)
- Study completion
- 27 May 2023 (actual)
- First posted
- 1 Jun 2023 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Clinical validation of the wrist blood pressure measuring device Combei BP 880 W for home use according to "the Universal Standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO)" in the general population.
Detailed description
Members of the AAMI, ESH and ISO committees achieved a consensus on an optimal validation standard and agreed on a single protocol for the validation of BP measuring devices that will replace all previous ones.
Therefore, the Association for the Advancement of Medical Instrumentation / European Society of Hypertension / International Organization for Standardization Universal Standard (AAMI/ESH/ISO 81060-2:2018) is now considered as the standard protocol for the validation of non-invasive blood pressure measuring device.
This study aims to demonstrate that the Combei BP 880 W wrist device fulfills the requirements of the new protocol of clinical validation and its Amendment 1, 2020-01 for limb circumference distribution criteria.
this is a monocentric, prospective, non-randomized, non-interventional, cross-study.
Combei BP 880 W wrist device (Sphygmomanometer Under Test, SUT). The reference device is the mercury sphygmomanometer The primary objective of the study is to assess the accuracy of the automatic oscillometric BP measuring device at the wrist level, the Combei BP 880 W, in the general population, following AAMI/ESH/ISO 81060-2:2018 protocol and its Amendment 1, 2020-01 for limb circumference distribution criteria.
The Combei BP 880 W device will pass the test if:
- The differences of at least 255 individual paired determinations of the SUT and of the observers' readings with the reference sphygmomanometer for all subjects, calculated separately for systolic blood pressure and diastolic blood pressure, shall have the mean value of the differences, be within or equal to +/- 5.0mmHg and a standard deviation no greater than 8.0 mmHg.
- For the systolic blood pressure and diastolic blood pressure of at least 85 subjects, the standard deviation of the averaged paired determinations per subject of the SUT and of the observers' readings with the reference sphygmomanometer shall meet the criteria listed in Table 1 of AAMI/ESH/ISO 81060-2:2018.
The secondary objective is to confirm the safety of use of this device, intended for home use.
Inclusion criteria
- Patient older than 12 years;
- Patient who signed the informed consent form;
- Patient followed-up at site (in-patient or out-patient);
- Patient with wrist circumference between 12.5 and 21.5 cm*. *in accordance with the Indications For Use of the Combei BP 880 W device.
Non-inclusion criteria
- Patient unable to give a consent or understand properly protocol information;
- Patient suffering from arrhythmia;
- Patient with poor quality of Korotkov sounds;
- Patient for whom K5 sounds are absent. The validation team will consist of three persons: two observers trained in accurate BP measurement and a supervisor.
- The reference sphygmomanometer is calibrated just before the beginning of the study.
- The observers are trained to the use of the SUT.
- In-patients or out-patients are included in the study after reading and signing the ICF.
- After inclusion, the patient is screened to verify that he/she satisfies the inclusion/non-inclusion criteria of the protocol. If not, the patient will be excluded from the study and no blood pressure measure will be performed on this patient.
- The patient is placed comfortably, and the reference sphygmomanometer is placed on his/her left upper arm, except if left upper arm is wounded. In this case, the reference sphygmomanometer is placed on his/her right upper arm.
- Measurements start with alternance between measurements with the reference sphygmomanometer and the SUT according to the same arm sequential BP measurements method.
- For each measurement with the reference sphygmomanometer, the two observers using mercury sphygmomanometers and a double-headed stethoscope, blinded from each other's result, and then by the supervisor using the tested device according to the manufacturer recommendations. The diastolic blood pressure should be measured at Korotkoff K5 sound. The agreement between the 2 observers will be checked all over the evaluation period by the supervisor to make sure that the difference between the two observers is no more than 4 mmHg for systolic and diastolic BP values.Otherwise, the measurement should be repeated. The intra individual BP variability for the reference BP measurements must be < 12 mmHg for SBP and < 8 mmHg for DBP.
- The results are written down on a source document and then reported in the CRF. Any AE, ADE or device malfunction occurring during the protocol will be also documented and reported in the CRF. For this study, all source documents should be written in English language.
- At the end of the measurements, the patient participation to the study is over.
Study procedures:
- Validation team and pre-study training The validation team will consist of three persons: two observers trained in accurate BP measurement and a supervisor. The reference sphygmomanometer is calibrated just before the beginning of the study. The observers are trained to the use of the SUT.
- Inclusion of the patients In-patients or out-patients of the investigational site are included in the study after reading and signing the ICF. After inclusion, the patient is screened to verify that he/she satisfies the inclusion/non-inclusion criteria of the protocol. If not, the patient will be excluded from the study and no blood pressure measure will be performed on this patient.
- Cuffs The cuffs used for reference auscultatory BP measurement will have an inflatable bladder length that covers 75-100% of the arm circumference of each patient and width that covers 37-50% of the arm circumference. Detailed description of cuffs used for reference BP measurement will be reported for each patient.
- Blood pressure measurements
The validation will be performed according to the (AAMI/ESH/ISO) Universal Standard validation protocol described in detail elsewhere10. The patients will have their reference BP measured by the two observers using a mercury sphygmomanometer and a double-headed stethoscope, blinded from each other's result, and then by the supervisor using the tested devices. The agreement between the 2 observers will be checked all over the evaluation period by the supervisor to make sure that the difference between the two observers is no more than 4 mmHg for systolic and diastolic BP values. Otherwise, the reference measurement should be repeated. The intra individual variability of the reference BP measurements must be < 12 mm Hg for SBP and < 8 mm Hg for DBP.
The circumference of the patient's upper arm will be measured to ensure that the cuff-size being used is adequate for the subject. The upper arm midpoint is first determined by marking the arm posteriorly at a point halfway between the acromion and olecranon, measured while the arm is flexed 90 degrees at the elbow with the palm facing up. The subject's upper arm circumference shall be determined by measuring at the midpoint of the upper arm while the elbow is relaxed, and the arm is dangling freely to the side.
The circumference of the patient's wrist will be measured to ensure that the circumference is within the acceptable range 12.5 - 21.5 cm for the Combei device, otherwise the subject will be excluded.
Measurements made by the mercury sphygmomanometer, will be performed according to the "same arm, consecutive measurements" (Figure 1), on the left arm supported at heart level.
Measurements by the Combei BP 880 W device will be performed at the wrist level of the same arm supported at the heart level, as recommended by the manufacturer with the following procedure.
- Ask the subject to relax for 5-10 min.
- make sure that the subject :
- is comfortably seated with legs uncrossed and feet flat on the floor,
- has the back, elbow and forearm supported,
- has the measurement site at the level of the left ventricle of the heart,
- and other conditions as recommended by the validation protocol and the device manual user.
In total, nine consecutive BP measurements will be performed in each patient using the mercury sphygmomanometers (5 times) and the tested devices (4 times).
All nine-sequential same-arm measurements will be recorded as shown in Table 2, starting with the standard mercury sphygmomanometer, followed by the device. A resting period of a minimum of 60 seconds shall be observed between the end of one measurement and the beginning of the following measurement.
Table 2: Sequential measurements of BP recorded by the standard mercury sphygmomanometer and the tested device:
R0 Entry BP, observers 1 and 2 each with the mercury standard T0 Device detection BP, supervisor R1 Observers 1 and 2 with mercury standard T1 Supervisor with the test instrument R2 Observers 1 and 2 with mercury standard T2 Supervisor with the test instrument R3 Observers 1 and 2 with mercury standard T3 Supervisor with the test instrument R4 Observers 1 and 2 with mercury standard
The first auscultatory and the first device measurement represent the recruitment pressures (R0 and T0) are not used in the evaluation of the test device accuracy. Each of the reference BP measurements (R) is the average of the simultaneous readings of the two observers. For each subject, the device measurements T1, T2 and T3 will be compared to observer measurements. Each of the test device measurements is compared to the average of the previous and succeeding reference BP readings (T1 vs average of R1 and R2, T2 vs average of R2 and R3, T3 vs average of R3 and R4).
At the end of the measurements, the patient participation to the study is over.
The Statistical Analysis: Differences are calculated by subtracting the reference BP measurement from the test device measurement. The mean BP difference (SUT versus reference sphygmomanometer) and its standard deviation (SD) will be calculated for the analysis. The same criteria will be used for systolic blood pressure and diastolic blood pressure measurements. The statistical analysis will be performed according to the requirements of the AAMI/ESH/ISO standard 2018 and its Amendment 1, 2020-01 for limb circumference distribution criteria.
Study design
- Primary purpose
- Other
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| one population of normotensive and hypertensive | Other | Device: Measuring blood pressuretreated and untreated normotensive and hypertensive patients |
Who may qualify
- Ages
- 18 to 80 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (4 listed)
- Patient older than 12 years;
- Patient who signed the informed consent form;
- Patient followed-up at site (in-patient or out-patient);
- Patient with wrist circumference between 12.5 and 21.5 cm*. *in accordance with the Indications For Use of the Combei BP 880 W device.
Key exclusion criteria (4 listed)
- Patient unable to give a consent or understand properly protocol information;
- Patient suffering from arrhythmia;
- Patient with poor quality of Korotkov sounds;
- Patient for whom K5 sounds are absent.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Measuring blood pressureMeasuring blood pressure using electronic device and using a mercury sphygmomanometer | Device | one population of normotensive and hypertensive |
Primary outcome
Accuracy of blood pressure measurement
Time frame: Subject will participate to only one session of measurement. The duration of one session is about 45 minutes.
differences < than 5 +/- 8 mmHg according to the protocol by comparison to the gold standard (mercury sphygmomanometer)
1 secondary outcome
differences between the tested device and the gold standard (mercury sphygmomanometer according to the baseline blood pressure and to the wrist circumference
Time frame: Subject will participate only one time to the study. one session of blood pressure measurement will be performed. the duration of one session is about 45 minutes
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 1 Jun 2023.
Related trials for Hypertension
Active, not recruiting
Digital Reinforcement of Motivational Interviewing Micro-Skills Among Primary Care Healthcare Professionals
Hypertension (Without Type 2 Diabetes Mellitus)
Recruiting
Investigating Vascular Properties of HEMI and SPG Signals in Individuals With or at Risk for Chronic Kidney Disease
Chronic Kidney Disease, Hypertension (HTN)
Recruiting
Trial Phase: Syndemic Adapted Medly Uganda
Mental Health
Not yet recruitingPhase 2/3
An Open-Label Extension Study to Evaluate Inhaled Treprostinil SMI in Participants With Pulmonary Hypertension Associated With Chronic Obstructive Pulmonary Disease (PH-COPD)
Pulmonary Hypertension Associated With Chronic Obstructive Pulmonary Disease
Enrolling by invitation
Testing Implementation Strategies to Support Clinic Fidelity to an Outpatient Hypertension Bundle
Pregnancy Related, Hypertension
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
ROLAND ASMAR study chair
International Society of Vascular Health
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Measuring blood pressure in people with Hypertension, run by International Society for Vascular Health at 1 site in France.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 4 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 1 Jun 2023. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Accuracy of blood pressure measurement (time frame: Subject will participate to only one session of measurement. The duration of one session is about 45 minutes.). It is what the study was designed to measure, not a finding.