Braindex Cerebral Tissue Oxygen Saturation (SctO2) Measurement Algorithm Assessment and Improvement

A study testing Venus Jugular O2 Saturation optical fiber in people with Post-Operative Confusion, run by Braindex at 1 site in France.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

The objective of this clinical trial is to optimize Braindex brain tissue oxygen saturation (SctO2) measurement algorithm compared to a mixed reference value of jugular venous and systemic arterial saturation, during proven reproducible and standardized loss of brain autoregulation induced during transcatheter aortic… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Post-Operative Confusion
What is being tested?
Venus Jugular O2 Saturation optical fiber
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
40 participants
Where
1 siteFrance
Study dates
11 May 2023to 30 Oct 2023
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patient, male or female, over 18 years of age
  • Patient to undergo percutaneous femoral aortic valve replacement with aortic valve deployment by balloon (Edwards Sapien valve) under simple sedation with remifentanil in TIVA, with spontaneous ventilation
  • Patient affiliated or entitled to a social security plan
  • Patient having received written and informed information about the study and having signed a free and free and informed consent to participate in the study.

Who cannot join

  • Emergency surgery
  • Patient weighing less than 40 kg
  • Patient with a history of ischemic stroke
  • Patient with a contraindication to jugular and arterial catheterization
  • Patient with a contraindication to the Braindex device
  • Protected patient: adult under guardianship, curatorship or other legal protection, deprived of liberty by judicial or administrative decision
  • Pregnant, parturient or breastfeeding woman
  • Patient hospitalized without consent

Trial locations and contacts

1 site in 1 country.

Pierre CHARLIER contact

pierre.charlier@braindex.fr

France1 site, 1 recruiting
  • Jacques CARTIER Private Hospital

    Massy, Île-de-France Region, 91300

    Julien AMOUR, profjulien.amour@gmail.com

    Recruiting

Other trials in France

About this study

Official title: In Vivo Optimization Study of the Braindex Cerebral Tissue Oxygen Saturation (SctO2) Measurement Algorithm Compared to a Mixed Reference Value of Jugular Venous and Systemic Arterial Saturation During Proven Cerebral Autoregulation Loss

The objective of this clinical trial is to optimize Braindex brain tissue oxygen saturation (SctO2) measurement algorithm compared to a mixed reference value of jugular venous and systemic arterial saturation, during proven reproducible and standardized loss of brain autoregulation induced during transcatheter aortic valve replacement procedures. The main question it aims to answer is: assessing and improving the sensitivity/specificity of Braindex SctO2 measurements.

Study type: single-center, prospective, interventional, open-label, exploratory clinical trial.

Read more

Participant population/health status:

  • Patient, male or female, over 18 years of age
  • Patient scheduled for percutaneous femoral aortic valve replacement with aortic valve balloon deployment (Edwards Sapien valve) under simple sedation with Remifentanil (TIVA), with spontaneous ventilation.

The expected outcome of this research is to optimize in vivo the algorithm for calculating the SctO2 of the BRAINDEX sensor in surgical conditions, in a context of proven low cerebral flow, on a target population. It is expected, in the long run, a better quality of detection of the loss of cerebral auto-regulation leading to a faster and optimal correction, thus potentially limiting the risk of POCD, subcortical stroke and postoperative cognitive decline.

Read the detailed description

SPONSOR BRAINDEX sas Parc Eurasanté 12 avenue Pierre Mauroy 59120 Loos

INVESTIGATOR COORDINATOR Professor Julien AMOUR Institut de Perfusion, de Réanimation et d'Anesthésie de Chirurgie Cardiaque Paris Sud (IPRA) Hôpital Privé Jacques Cartier 6 avenue du Noyer Lambert 91300 Massy, France

TITLE In vivo optimization study of the algorithm for measuring cerebral tissue oxygen saturation (SctO2) by Braindex compared with a mixed reference value of jugular venous and systemic arterial saturation during proven loss of cerebral auto-regulation.

ID-RCB NUMBER 2022-A02770-43

JUSTIFICATION / BACKGROUND The incidence of postoperative cognitive impairment (POCD) in cardiac surgery is of the order of 30-50%. Loss of cerebral autoregulation is a factor in POCD and cognitive decline. Cerebral near-infrared oximetry (NIRS), a commercially available non-invasive monitoring system, aims to detect and limit the loss of cerebral autoregulation during intraoperative hemodynamic instability.

NIRS currently measures cerebral tissue oxygen saturation (SctO2) over a patchy tissue volume, and corresponds to blood sampling comprising arterial, venous and capillary components.

The performance of cerebral oximetry devices currently on the market has been clinically validated to a limited extent, often in healthy volunteers with no comorbidity, or in cardio-vascular situations (hypoxemia tests) outside any surgical procedure. For a number of devices, test results have been extrapolated and marketed without any real per-procedural clinical validation, and without being able to fully assert the correlation between a real loss of cerebral autoregulation, which is difficult to affirm, and the fall in SctO2. Beyond clinical validation, the various devices present a number of limitations and technical imperfections, notably a restricted sampling zone, a low number of transmitters and receivers, and variable signal processing algorithm quality. All these factors may explain the mixed results of recent clinical meta-analyses concerning the benefits of this technology.

The BRAINDEX sensor offers the advantage of an optical imager equipped with 60 LEDs emitting 6 wavelengths (660, 730, 780, 810, 850 and 900nm) and 28 photodiodes (optical sensors). This set of optodes translates into a measurement of 420 optical vectors per cerebral hemisphere, enhancing the resolution of cerebral oxygenation measurement and the number of analysis zones compared with current devices on the market.

Initial efficiency tests have already been carried out in silico using simulation models and on a test bench. We now wish to optimize and calibrate the signal analysis algorithm for measuring cerebral tissue oxygen saturation (SctO2) in vivo, in a context of proven loss of cerebral autoregulation, and compare it with a mixed value of jugular venous and systemic arterial saturation, a commonly accepted reference.

In this sense, percutaneous aortic valve replacement or TAVI (Sapiens valve, Edwards) offers the decisive advantage of generating a standardized and perfectly reproducible circulatory arrest, leading to a planned loss of cerebral autoregulation, throughout the duration of aortic valve occlusion by the valve deployment balloon.

The aim of the research is to optimize the Braindex algorithm in vivo during the TAVI procedure, and in particular during the loss of cerebral autoregulation linked to circulatory arrest systematically induced by valve deployment. This study will also rely on the measurement of jugular venous oxygen saturation (SvjO2) and arterial oxygen saturation (SaO2), which are the reference measurements for the previous devices.

OBJECTIVES The primary objective of this research is to optimize an algorithm for estimating SctO2 (SctO2Exp.), in vivo, in patients scheduled for a femoral TAVI procedure and systematically exposed to a period of loss of cerebral autoregulation during deployment of the percutaneous aortic valve.

Secondary objectives are:

  • To evaluate the quality of EEG (electroencephalogram) signals from electrodes in the BRAINDEX sensor,
  • Evaluate the possibility of calculating the ANI (Analgesia Nociception Index) from PPG (photoplethysmography) signals obtained with the BRAINDEX sensor
  • Study the technical performance of the BRAINDEX sensor.

RESEARCH SCHEME This is an exploratory, prospective, interventional, open clinical investigation aimed at developing and optimizing the algorithm for estimating SctO2Exp.

INCLUSION CRITERIA

  • Patient, male or female, over 18 years of age
  • Patient undergoing percutaneous aortic valve replacement with aortic valve balloon deployment (Edwards Sapien valve) under simple sedation with remifentanil in AIVOC, with spontaneous ventilation
  • Patient affiliated to or entitled under a social security scheme
  • Patient having received written and informed information about the study and having signed a free and informed consent to participate in the study.

NON-INCLUSION CRITERIA

  • Emergency surgery
  • Patient weighing less than 40kg
  • Patient with history of ischemic stroke
  • Patient with a contraindication to catheterization arterial catheters
  • Patient with a contraindication to the Braindex device
  • Protected patient: adult under guardianship, curatorship or other legal protection deprived of liberty by judicial or administrative decision
  • Pregnant, parturient or breast-feeding women
  • Patient hospitalized without consent

RESEARCH PROCEDURE The research procedure consists in optimizing the Braindex algorithm for measuring in vivo SctO2Exp before, during and after a proven loss of cerebral autoregulation linked to circulatory arrest programmed by percutaneous aortic valve deployment via transvalvular balloon, on a vigorous patient sedated with remifentanil in IVCA.

Continuous measurement of SvjO2, the usual reference measurement for commercially available devices, will be performed by implanting an optical fiber (CEVOX,PULSIONMedicalSystemsSEHans-Riedl-Str.17 85622 Feldkirchen Germany) through an internal jugular catheter, in addition to the conventional external electro systolic drive catheter. The following parameters will be recorded every 2 seconds: invasive blood pressure (systolic, mean and diastolic), heart rate, SaO2 on the pulse wave. SvjO2 will be recorded continuously throughout the implantation procedure. The times at which the surgical incision begins, the onset of sinus tachycardia (overdriving) induced by external electro-systolic stimulation, the start of inflation of the intra-avalvular aortic balloon, maximum balloon occlusion, the end of complete balloon deflation, the end of stimulation and the end of skin closure will be recorded.

Jugular venous and arterial gasometry (1 ml per sample) will be taken to perform initial calibration of the CEVOX fiber at the start of the procedure (skin incision), then before overdriving, then when overdriving is stopped and finally at the end of skin closure, to ensure that there is no derivation of the estimated continuous value of the CEVOX fiber.

ADDED ACT Placement of the CEVOX probe for continuous SvjO2 monitoring will be performed under local anaesthetic, in addition to remifentanil sedation, under ultrasound guidance.

Installation of the BRAINDEX non-invasive forehead sensor.

EVALUATION CRITERIA

Primary endpoints :

  • NIRS (Near-infrared spectroscopy) signals recorded via the BRAINDEX sensor to optimize the algorithm
  • SctO2Exp value obtained from jugular measurement to optimize algorithm.

Secondary endpoints :

  • EEG signals recorded via BRAINDEX sensor and three-axis acceleration information via accelerometer
  • PPG (photoplethysmocardiography) signals via BRAINDEX sensor and ANI signals via ANIv2 monitor
  • Sensor operation, optode detachment, electrode detachment

STUDY SIZE As this is an exploratory study, no statistical hypothesis could be defined. As the main objective of the study is to optimize an in vivo algorithm for estimating SctO2Exp, it is important to obtain a sample representative of the diversity of hemodynamic situations observed during TAVI procedures, which offer the advantage of being highly standardized and reproducible. The inclusion of 40 patients should enable us to achieve this diversity. The inclusion of 40 patients should enable us to achieve this diversity.

INTENDED NUMBER OF CENTERS One center: Institut de Perfusion, de Réanimation et d'Anesthésie de Chirurgie Cardiaque Paris Sud (IPRA) at Hôpital Privé Jacques Cartier (Massy).

DURATION OF RESEARCH

  • Length of inclusion period: 10 months (projected start date of IC: 01/03/2023)
  • Duration of each participant's participation: duration of the surgical procedure with follow-up of adverse events during the 24 postoperative hours.
  • Total research duration: 10 months and 24 hours (projected end of CI: 05/01/2024).

STATISTICAL ANALYSIS OF DATA Analyses will be performed on all included patients. Main analysis NIRS signals collected over the entire operative period of the TAVI procedure will be used to develop the algorithm for calculating SctO2Exp.

The raw data will be denoised using a butter Worth 0.05Hz order 6 low-pass filter and subtracting potential artifacts, external light interference and high-frequency interference (physiological and caused by the electromagnetic environment). The regional oxygen concentration SctO2Exp will then be calculated using the modified Beer-Lambert law, expressed by the equation below, on all optical vectors, In order to optimize the algorithm obtained, the concentrations thus estimated, SctO2Exp, will be compared with the reference measurement, SctO2Ref (estimated as being equal to 25% SaO2 and 75% SvjO2), and will be used to group together the distances and vectors required for a more accurate estimate of oxygen saturation.

Secondary analysis

Evaluation of EEG signal quality:

  • EEG signal quality in the face of physiological and environmental artefacts environmental artifacts associated with the TAVI procedure (patient handling, electromagnetic environment, etc.), will be assessed by calculating a signal-to-noise ratio defined by the power density in the spectral band of interest (0.1 - 4hz) divided by the remaining spectral content and signal saturation indicators (considered saturated if greater than ±200mV). The proportion of saturated ratio will be described, - Signal variations in relation to voluntary or involuntary movements of the patient's head measured along 3 axes will be calculated. A variation of more than 0.5g in any axis will be considered as a head movement (voluntary or involuntary). The proportion of variation over 0.5g will be described.

Assessment of the possibility of calculating ANI :

  • The PPG signals recorded by the BRAINDEX sensor will be used to calculate ANI. The agreement between the ANI obtained from the PPG signals via the BRAINDEX sensor and the ANI obtained from the ANIv2 monitor will be assessed using the intraclass correlation coefficient (ICC) and the graphical method of Bland and Altman. The ICC and its 95% confidence interval will be determined using a 2-factor random analysis of variance, patient and method.

Study the technical performance of the BRAINDEX sensor:

  • The operational reliability of the BRAINDEX sensor will be assessed using a log file indicating potential abnormal operation of the system
  • The identification of optode detachment during measurement will be evaluated by estimating the variance of ambient noise with diodes off
  • The indication of electrode detachment during measurement will be evaluated by estimating the impedance of the electrodes by the acquisition component

EXPECTED BENEFITS The expected spin-offs of this research are to optimize the BRAINDEX sensor's SctO2Exp calculation algorithm in vivo, under actual patient management conditions, in proven situations of loss of cerebral auto regulation.

Being tested
Venus Jugular O2 Saturation optical fiber
Keywords
Delirium, NIRS, cognitive decline, elderly, confusion
Medical subject headings
Delirium, Cognitive Dysfunction, Confusion

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Venus Jugular O2 Saturation optical fiberDevice

    Also called CEVOX,PULSION Medical Systems

    A continuous measurement of VjO2S, the usual reference measurement of the marketed devices, will be performed by implanting an optical fiber (CEVOX, PULSION Medical Systems) through an internal jugular catheter in addition to the classical catheter of the external electro-systolic drive probe. The following parameters will be recorded every 2 seconds: invasive blood pressure (systolic, mean and diastolic), heart rate, SaO2 on the pulse wave. VjO2S will be recorded continuously during the entire implantation procedure. The time of the beginning of the surgical incision, of the beginning of sinus tachycardia (overdriving) induced by external electro-systolic stimulation, of the beginning of inflation of the intravalvular aortic balloon, of the maximum occlusion of the balloon, of the end of complete deflation of the balloon, of the end of stimulation and of the end of skin closure will be recorded.

Study groups in NCT05897060
Study groupRoleWhat they receive
Patients undergoing TAVI procedureExperimentalThe study is interventional and consists of a single group of patients undergoing a Transcatheter Aortic Valve Implantation (TAVI), receiving in addition to the usual treatment, an optical fiber placed retrograde in the jugular bulb, in order to record the values of SJVO2 every second.
Receives: Device: Venus Jugular O2 Saturation optical fiber

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Optimization of SctO2 Braindex algorithm

    SctO2 which is Tissular Cerebral Saturation in Oxygène expressed in percentage

    Measured: Duration of the patient's Trans-catheter Aortic Valve Implantation procedure

Secondary outcomes (2)
  1. Validation of Braindex EEG measures

    EEG raw signals recorded via the BRAINDEX sensor and transformed in spectral edge frequencies expressed in Hertz

    Measured: Duration of the patient's Trans-catheter Aortic Valve Implantation procedure

  2. Validation of Analgesia Nociception Index (MDoloris) acquired via frontal Braindex frontal PPG

    ANI Braindex measurements acquired via frontal PPG compared to ANI MDoloris measurements acquired via MDoloris ECG sensor. Trending values without any unit on a range from 0 to 100.

    Measured: Duration of the patient's Trans-catheter Aortic Valve Implantation procedure

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted9 Jun 2023
  2. Study start11 May 2023
  3. Primary completion30 Oct 2023 (estimated)
  4. Study completion30 Oct 2023 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
40 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05897060
Sponsor's protocol ID
2022-A02770-43
Acronym
POC-SctO2
Status last verified
May 2023
First submitted
16 May 2023
Last updated
9 Jun 2023
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study testing Venus Jugular O2 Saturation optical fiber in people with Post-Operative Confusion, run by Braindex at 1 site in France.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 9 Jun 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Thierry CUSSAC (+33(7)82230369, thierry.cussac@braindex.fr), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Optimization of SctO2 Braindex algorithm (time frame: Duration of the patient's Trans-catheter Aortic Valve Implantation procedure). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.