Healthy VolunteersNCT05968079
Performance Comparison Between Masimo W1™ and Apple Watch Series 8
A study testing Masimo W1™ and Apple Watch Series 8 in people with Healthy Volunteers, run by Masimo Corporation at 1 site in United States.
Trial summary
This study is also designed to compare the performance of Masimo W1™ and Apple Watch Series 8 against FDA cleared hospital grade pulse oximetry technology under desaturation conditions. Read more about this study
- Who is it for?
- Ages 18 to 55 yearsHealthy volunteers can take part
- Conditions the study is for
- Healthy Volunteers
- What is being tested?
- Masimo W1™ and Apple Watch Series 8
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 20 participants
- Where
- 1 siteUnited States
- Study dates
- 6 Sep 2023to 11 Oct 2023
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 55 years
- All sexes
- Healthy volunteers accepted
- Adult
Who may be able to join
- Subject is 18 to 55 years of age.
- Subject is American Society Anesthesiologist status 1 (ASA I)
- Subject has a BMI between 18 and 35.
- Subject is able to read and communicate in English and understands the study and the risks involved.
Who cannot join
- Subject is currently taking any medications which in the opinion of the principal investigator would not be suitable for participation in the study.
- Subject is wearing nail polish that cannot be removed, gel nails, and/or acrylic nails that can interfere with study device's placement.
- Participants with conditions or skin abnormalities at or around site of sensor placement that could affect the placement on the sensor or prevent monitoring of physiological parameters during the study, such as psoriasis, eczema, angioma, scar tissue, burn, fungal infection, substantial skin breakdown.
- Subject has participated in an investigational drug study within one month prior to the start of the study.
- Subject has failed the Allen's Test to confirm patency of the collateral artery.
- Subject is female with a positive pregnancy test, or is female and is unwilling to use effective birth control between the time of screening and study termination.
Show all 16Show fewer
- Subject has a positive urine cotinine or drug test.
- Subject has a reported allergy to Lidocaine.
- Subject has clinically significant anemia or other hemoglobinopathy.
- Subject has a room air saturation of less than 95% by pulse oximetry.
- Subject has a clinically significant abnormal EKG.
- Subject has a clinically significant abnormal pulmonary function test via spirometry.
- Subject is intolerant to a breathing mask apparatus.
- Subject has a COHb greater than 3%, or MetHb greater than 2% verified by laboratory co-oximeter.
- Subject has another condition, which in the opinion of the principal investigator would not be suitable for participation in the study.
- Subject is unwilling or unable to provide informed consent or comply with the study procedures.
Trial locations and contacts
1 site in 1 country.
United States1 site
Human Pharmacology and Physiology Laboratory (HPPL) - Duke University Medical Center
Durham, North Carolina, 27705
About this study
This study is also designed to compare the performance of Masimo W1™ and Apple Watch Series 8 against FDA cleared hospital grade pulse oximetry technology under desaturation conditions.
- Conditions
- Healthy Volunteers
- Being tested
- Masimo W1™ and Apple Watch Series 8
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Masimo W1™ and Apple Watch Series 8Device
Noninvasive wearable health monitoring device
| Study group | Role | What they receive |
|---|---|---|
| Masimo W1™ and Apple Watch Series 8 | Experimental | All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo W1™ and Apple Watch Series 8. Receives: Device: Masimo W1™ and Apple Watch Series 8 |
What is being measured
The primary outcome is the main question the study is designed to answer.
Accuracy of Masimo W1™ and Apple Watch Series 8 in Determining SpO2 (Oxygen Saturation) Values in Subjects During Desaturation
Accuracy will be determined by comparing Masimo W1™ SpO2 measurements and Apple Watch Series 8 SpO2 measurements with FDA-cleared pulse oximetry device SpO2 measurements and calculating the Arithmetic root mean square (ARMS) value. All values presented below for SpO2 (Bias, Precision and ARMS) are in units of %.
The pairwise samples consist of paired SpO2 timing measurements from both the reference sensor, Masimo W1, and the Apple Watch.
Measured: 1-6 hours
Successful Read Rate for Motion Conditions Under Slow Desaturation for Masimo W1™ and Apple Watch Series 8
This analysis examines the successful read rates for motion conditions during slow desaturation events for the Masimo W1™ and Apple Watch Series 8. The motion condition consists of a combined value representing three types of simulated motion: drinking, scratching the nose, and checking the time.
Successful Read Rate = 1 - No Read Rate (Nf / Nspotcheck) x 100 (%) Nf = Number of Reads without Valid SpO2 Nspotcheck = Number of All Spot Checks
Measured: 1-6 hours
Detection Rate of Fast Desaturation Events Under Various Conditions for Masimo W1™ and Apple Watch Series 8
Comparison of the detection rate of fast desaturation events under different conditions for the Masimo W1™ and Apple Watch Series 8. The first condition involves no motion, with the palm facing down and the Apple Watch in sleep mode, while the second condition involves no motion, with the palm facing sideways and manual spot checks performed on the Apple Watch. The goal is to evaluate the performance of both devices in detecting desaturation events under these varying circumstances.
Detection Rate = (Nt / Ndesat) x 100 (%) Nt = Number of Detected Event by Test Device
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo W1™ and Apple Watch Series 8.
Ndesat = Number of All Valid Fast Desaturation Events by Reference SpO2
Measured: 1-6 hours
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted1 Aug 2023
- Study start6 Sep 2023
- Primary completion11 Oct 2023
- Study completion11 Oct 2023
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Healthy VolunteersStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Primary purpose
- Other
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 20 participants (actual)
Sponsor and investigators
- Lead sponsor
- Masimo Corporation
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT05968079
- Sponsor's protocol ID
- MACL0003
- Status last verified
- Apr 2025
- First submitted
- 13 Jul 2023
- Last updated
- 9 Apr 2025
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Study at a glance
From the registry record, last updated by the sponsor on 9 Apr 2025. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Performance Comparison Between Masimo W1™ and Apple Watch Series 8
- ClinicalTrials.gov ID
- NCT05968079
- Acronym / sponsor ID
- MACL0003
- Conditions
- Healthy Volunteers
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 20 participants (actual)
- Lead sponsor
- Masimo CorporationIndustry (pharma, biotech, devices)
- Study start
- 6 Sep 2023 (actual)
- Primary completion
- 11 Oct 2023 (actual)
- Study completion
- 11 Oct 2023 (actual)
- First posted
- 1 Aug 2023 (actual)
- Results
- Posted 9 Apr 2025
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
This study is also designed to compare the performance of Masimo W1™ and Apple Watch Series 8 against FDA cleared hospital grade pulse oximetry technology under desaturation conditions.
Study design
- Primary purpose
- Other
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Masimo W1™ and Apple Watch Series 8 | Experimental | Device: Masimo W1™ and Apple Watch Series 8All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo W1™ and Apple Watch Series 8. |
Who may qualify
- Ages
- 18 to 55 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult
Key inclusion criteria (4 listed)
- Subject is 18 to 55 years of age.
- Subject is American Society Anesthesiologist status 1 (ASA I)
- Subject has a BMI between 18 and 35.
- Subject is able to read and communicate in English and understands the study and the risks involved.
Key exclusion criteria (16 listed)
- Subject is currently taking any medications which in the opinion of the principal investigator would not be suitable for participation in the study.
- Subject is wearing nail polish that cannot be removed, gel nails, and/or acrylic nails that can interfere with study device's placement.
- Participants with conditions or skin abnormalities at or around site of sensor placement that could affect the placement on the sensor or prevent monitoring of physiological parameters during the study, such as psoriasis, eczema, angioma, scar tissue, burn, fungal infection, substantial skin breakdown.
- Subject has participated in an investigational drug study within one month prior to the start of the study.
- Subject has failed the Allen's Test to confirm patency of the collateral artery.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Masimo W1™ and Apple Watch Series 8Noninvasive wearable health monitoring device | Device | Masimo W1™ and Apple Watch Series 8 |
Primary outcome
Accuracy of Masimo W1™ and Apple Watch Series 8 in Determining SpO2 (Oxygen Saturation) Values in Subjects During Desaturation
Time frame: 1-6 hours
Accuracy will be determined by comparing Masimo W1™ SpO2 measurements and Apple Watch Series 8 SpO2 measurements with FDA-cleared pulse oximetry device SpO2 measurements and calculating the Arithmetic root mean square (ARMS) value. All values presented below for SpO2 (Bias, Precision and ARMS) are in units of %. The pairwise samples consist of paired SpO2 timing measurements from both the reference sensor, Masimo W1, and the Apple Watch.
Successful Read Rate for Motion Conditions Under Slow Desaturation for Masimo W1™ and Apple Watch Series 8
Time frame: 1-6 hours
This analysis examines the successful read rates for motion conditions during slow desaturation events for the Masimo W1™ and Apple Watch Series 8. The motion condition consists of a combined value representing three types of simulated motion: drinking, scratching the nose, and checking the time. Successful Read Rate = 1 - No Read Rate (Nf / Nspotcheck) x 100 (%) Nf = Number of Reads without Valid SpO2 Nspotcheck = Number of All Spot Checks
Detection Rate of Fast Desaturation Events Under Various Conditions for Masimo W1™ and Apple Watch Series 8
Time frame: 1-6 hours
Comparison of the detection rate of fast desaturation events under different conditions for the Masimo W1™ and Apple Watch Series 8. The first condition involves no motion, with the palm facing down and the Apple Watch in sleep mode, while the second condition involves no motion, with the palm facing sideways and manual spot checks performed on the Apple Watch. The goal is to evaluate the performance of both devices in detecting desaturation events under these varying circumstances. Detection Rate = (Nt / Ndesat) x 100 (%) Nt = Number of Detected Event by Test Device All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo W1™ and Apple Watch Series 8. Ndesat = Number of All Valid Fast Desaturation Events by Reference SpO2
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 9 Apr 2025.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Healthy Volunteers
Not yet recruitingPhase 1
Dynamic Magnetic Resonance Imaging of Human Subarachnoid Space Using Intrathecal Gadolinium Contrast
Healthy Volunteers
RecruitingObservational study
Survey to Determine Incarcerated Persons' Views on Surrogate Decision Making
Healthy Volunteers
Active, not recruitingPhase 1
A Study to Learn How Different Amounts of the Study Medicine Called PF-07985631 Are Tolerated and Act in the Body of Healthy Adults
Healthy
Not yet recruitingObservational study
Collection and Distribution of Blood Components From Healthy Donors for In Vitro Research Use
Blood Donors, Apheresis
RecruitingEarly Phase 1
Study Brain Mechanisms of Frustration With Magnetoencephalography in Healthy Volunteers
Irritability
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Masimo W1™ and Apple Watch Series 8 in people with Healthy Volunteers, run by Masimo Corporation at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 4 inclusion and 16 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 9 Apr 2025. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 9 Apr 2025). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Accuracy of Masimo W1™ and Apple Watch Series 8 in Determining SpO2 (Oxygen Saturation) Values in Subjects During Desaturation (time frame: 1-6 hours). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 9 Apr 2025.