Foot Ulcer, DiabeticPASNCT06026813

Pressure Alternating Shoes

A study testing test pressure alternating shoes in people with Foot Ulcer, Diabetic, run by University of Texas Southwestern Medical Center at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The project is designed to develop and test Pressure Alternating Shoes (PAS), which will periodically off-load certain regions of the foot in order to prevent foot ulcers. An automated dual layer insole compromised of an active pressurized actuator array in combination with a passive compliant layer on top of each… Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
Conditions the study is for
Foot Ulcer, Diabetic
What is being tested?
test pressure alternating shoes
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
21 participants
Where
1 siteUnited States
Study dates
7 Mar 2024to 11 Jul 2024
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Age 18 or greater
  • Ability to wear insoles in shoes provided
  • Ability to walk unaided
  • Age 18 or greater
  • Ability to wear insoles in shoes provided
  • Ability to walk unaided
Show all 13Show fewer
  • Diagnosis of diabetic neuropathy
  • Excluision:
  • Charcot foot
  • Knee Pain
  • Previous amputations
  • Inflammatory diseases such as rheumatoid arthritis
  • Open wounds, ulcers, sores or blistesr on the feet; signs of infection in the feet

Who cannot join

  • Diabetic Neuropathy
  • Charcot foot
  • Knee pain
  • Previous amputations
  • Inflammatory diseases such as rheumatoid arthritis
  • Open wounds, ulcers, sores or blisters on the feet; signs of infection in the feet
  • Diabetic population:

Trial locations and contacts

1 site in 1 country.

United States1 site
  • University of Texas Southwestern Medical Center

    Dallas, Texas, 75390-8560

Other trials in United States

About this study

Official title: Pressure Alternating Shoes (PAS) for Prevention of Diabetic Foot Ulcers

The project is designed to develop and test Pressure Alternating Shoes (PAS), which will periodically off-load certain regions of the foot in order to prevent foot ulcers. An automated dual layer insole compromised of an active pressurized actuator array in combination with a passive compliant layer on top of each actuator to modulate and distribute the plantar surface pressure as desired will be tested. This device will allow us to simultaneously load and offload select areas of the foot using the active layer by inflating and deflating individual actuators using pressurized air. After offloading, the remaining load will be distributed to other areas with inflated actuators. Automatic modulation will be provided through programmable control hardware which will cyclically relieve mechanical loading based on a prescribed duration and frequency.

Read the detailed description

The two-tier human subjects study will be conducted to assess the biomechanical characteristics of PAS. In the first tier, we will test PAS in healthy subjects and in the second tier, we will test PAS in Diabetic Neuropathy patients (DN).

Subjects will walk on a treadmill in their usual daily shoes for 5 minutes, then with standard diabetic shoes with the PAS insoles for 5 minutes. Subjects will wear body worn sensors that assess position of the body. Subjects will have the temperature of their feet measured via a special camera after walking. Subjects will rest for 30 minute washout period and have the blood flow in the soles of the feet measured via a special camera. 1 healthy subject will undergo MRI of the foot and ankle.

Being tested
test pressure alternating shoes
Medical subject headings
Diabetic Foot

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • test pressure alternating shoesDevice

    test pressure alternating shoes

Study groups in NCT06026813
Study groupRoleWhat they receive
Healthy controlsExperimentalPatients without diabetes who do not have foot wounds or history of amputation
Receives: Device: test pressure alternating shoes
Patients with Diabetic NeuropathyExperimentalPatients with diagnosed diabetes and neuropathy who do not have foot wounds or history of amputation.
Receives: Device: test pressure alternating shoes

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Maximum Average Pressure at Baseline With Pressure Alternating Shoes, Right Foot (Before Any Cells Were Offloaded)

    Maximum Average pressure values in the plantar regions before offloading during walking. The insole consists of seven air cells that align with specific regions: the big toe (cell 1), the area spanning from the second toe to the fifth toe (cell 2), metatarsal heads (cells 3 and 4), the midfoot (cells 5 and 6), and the heel (cell 7). Measurements were only obtained from the right foot.

    Measured: Before any cells were offloaded (First 40-seconds of the walk)

  2. Maximum Average Pressure at After Offloading - Diabetic Footwear Equipped With PAS Device

    Maximum Average pressure values in the plantar regions after offloading cells during walking. The insole consists of seven air cells that align with specific regions: the big toe (cell 1), the area spanning from the second toe to the fifth toe (cell 2), metatarsal heads (cells 3 and 4), the midfoot (cells 5 and 6), and the heel (cell 7). Measurements were only obtained from the right foot. The results reported represent results for footwear with PAS device.

    Measured: After offloading (approx. 41- 200 seconds walk)

  3. Peak Interface Pressure at Baseline (Before Offloading) - Diabetic Footwear Equipped With PAS Device

    Peak Interface Pressure values in the plantar regions before offloading during walking. The insole consists of seven air cells that align with specific regions: the big toe (cell 1), the area spanning from the second toe to the fifth toe (cell 2), metatarsal heads (cells 3 and 4), the midfoot (cells 5 and 6), and the heel (cell 7). Measurements were only obtained from the right foot. The results reported represent results for footwear with PAS device.

    Measured: Baseline (First 40-seconds of the walk)

  4. Peak Interface Pressure at After Offloading - Diabetic Footwear Equipped With PAS Device

    Peak Interface Pressure values in the plantar regions after offloading during walking. The insole consists of seven air cells that align with specific regions: the big toe (cell 1), the area spanning from the second toe to the fifth toe (cell 2), metatarsal heads (cells 3 and 4), the midfoot (cells 5 and 6), and the heel (cell 7). Measurements were only obtained from the right foot. The results reported represent results for footwear with PAS device.

    Measured: After offloading (approx. 41- 200 seconds walk)

  5. Max Average Pressure Change Among All Participants - Diabetic Footwear Equipped With PAS Device

    Percentage change due to offloading for each participant is measured by dividing the pressure difference between the after offloading and before offloading values divided by the before offloading value. Measurements were only obtained from the right foot. The results reported represent results for footwear with PAS device.

    Measured: Baseline (first 40-second walk), 201 seconds after the intervention ended

  6. Peak Pressure Change Among All Participants - Diabetic Footwear Equipped With PAS Device

    Peak pressure change due to offloading for each participant is measured by dividing the pressure difference between the peak after offloading and peak before offloading values divided by the peak before offloading value. Measurements were only obtained from the right foot. The results reported represent results for footwear with PAS device.

    Measured: Baseline (first 40-second walk), 201 seconds after the intervention ended

Secondary outcomes (11)
  1. Plantar Skin Temperature at Baseline, Right Foot

    Plantar skin temperature measured by a noncontact thermal imaging camera (*Celcius) before the intervention and without footwear

    Measured: Baseline before the intervention (30 minutes after start of visit 1)

  2. Plantar Skin Temperature Post Intervention, Right Foot

    Plantar skin temperature measured by a noncontact thermal imaging camera (*Celcius) post intervention, without footwear

    Measured: Post intervention (215 seconds)

  3. Tissue Oxygenation (Oxygen Saturation) - for Plantar Lateral Foot

    Plantar foot tissue oxygenation (Oxygen saturation) is measured by noncontact hyperspectral imaging camera (StO2). Only the right foot was analyzed and without footwear.

    Measured: Baseline (30 minutes after start of visit 1)

  4. Tissue Oxygenation (Oxyhemoglobin and Deoxyhemoglobin ) - for Plantar Lateral Foot

    Plantar foot tissue oxygenation (Oxyhemoglobin and Deoxyhemoglobin) is measured by noncontact hyperspectral imaging (HSI) camera. Only the right foot was analyzed and without footwear. The accepted measure of oxyhemoglobin and deoxyhemoglobin is arbitrary units (AU). NIR spectrum light passes through skin and is reflected off the blood supplying the tissue. Wavelength dependent light absorption of hemoglobin differs if it is carrying oxygen or not, therefore detecting oxygenated and deoxygenated blood. This was derived from Beer-Lambert Law but would need to be verified by the manufacturer. There is no researcher drawn calculation as the device provides the arbitrary unit based on the reflectance detected back to the imaging device. No reference values or any standardization available but one would infer higher oxyhemoglobin level indicates better oxygenation and thus a higher deoxyhemoglobin would be indicative of poor oxygenation.

    Measured: Baseline (30 minutes after start of visit 1)

  5. Tissue Oxygenation (Oxygen Saturation) - for Plantar Medial Foot

    Plantar Medial foot tissue oxygenation (Oxygen saturation) is measured by noncontact hyperspectral imaging camera (StO2). Only the right foot was analyzed and without footwear.

    Measured: Baseline (30 minutes after start of visit 1)

  6. Tissue Oxygenation (Oxyhemoglobin and Deoxyhemoglobin) - for Plantar Medial Foot

    Plantar medial foot tissue oxygenation (Oxyhemoglobin and Deoxyhemoglobin) is measured by noncontact hyperspectral imaging (HSI) camera. Only the right foot was analyzed. Only the right foot was analyzed and without footwear. The accepted measure of oxyhemoglobin and deoxyhemoglobin is arbitrary units (AU). NIR spectrum light passes through skin and is reflected off the blood supplying the tissue. Wavelength dependent light absorption of hemoglobin differs if it is carrying oxygen or not, therefore detecting oxygenated and deoxygenated blood. This was derived from Beer-Lambert Law but would need to be verified by the manufacturer. There is no researcher drawn calculation as the device provides the arbitrary unit based on the reflectance detected back to the imaging device. No reference values or any standardization available but one would infer higher oxyhemoglobin level indicates better oxygenation and thus a higher deoxyhemoglobin would be indicative of poor oxygenation.

    Measured: Baseline (30 minutes after start of visit 1)

  7. Tissue Oxygenation (Oxygen Saturation) - for Plantar Lateral Foot

    Plantar foot tissue oxygenation (Oxygen saturation) is measured by noncontact hyperspectral imaging camera (StO2). Only the right foot was analyzed and without footwear.

    Measured: Post Gait (approx. 2015 seconds post intervention)

  8. Tissue Oxygenation (Oxyhemoglobin and Deoxyhemoglobin ) - for Plantar Lateral Foot

    Plantar foot tissue oxygenation (Oxyhemoglobin and Deoxyhemoglobin) is measured by noncontact hyperspectral imaging (HSI) camera. Only the right foot was analyzed and without footwear. The accepted measure of oxyhemoglobin and deoxyhemoglobin is arbitrary units (AU). NIR spectrum light passes through skin and is reflected off the blood supplying the tissue. Wavelength dependent light absorption of hemoglobin differs if it is carrying oxygen or not, therefore detecting oxygenated and deoxygenated blood. This was derived from Beer-Lambert Law but would need to be verified by the manufacturer. There is no researcher drawn calculation as the device provides the arbitrary unit based on the reflectance detected back to the imaging device. No reference values or any standardization available but one would infer higher oxyhemoglobin level indicates better oxygenation and thus a higher deoxyhemoglobin would be indicative of poor oxygenation.

    Measured: Post Gait (approx. 2015 seconds post intervention)

  9. Tissue Oxygenation (Oxygen Saturation) - for Plantar Medial Foot

    Plantar Medial foot tissue oxygenation (Oxygen saturation) is measured by noncontact hyperspectral imaging camera (StO2). Only the right foot was analyzed and without footwear.

    Measured: Post Gait (approx. 2015 seconds post intervention)

  10. Tissue Oxygenation (Oxyhemoglobin and Deoxyhemoglobin) - for Plantar Medial Foot

    Plantar medial foot tissue oxygenation (Oxyhemoglobin and Deoxyhemoglobin) is measured by noncontact hyperspectral imaging (HSI) camera. Only the right foot was analyzed and without footwear. The accepted measure of oxyhemoglobin and deoxyhemoglobin is arbitrary units (AU). NIR spectrum light passes through skin and is reflected off the blood supplying the tissue. Wavelength dependent light absorption of hemoglobin differs if it is carrying oxygen or not, therefore detecting oxygenated and deoxygenated blood. This was derived from Beer-Lambert Law but would need to be verified by the manufacturer. There is no researcher drawn calculation as the device provides the arbitrary unit based on the reflectance detected back to the imaging device. No reference values or any standardization available but one would infer higher oxyhemoglobin level indicates better oxygenation and thus a higher deoxyhemoglobin would be indicative of poor oxygenation.

    Measured: Post Gait (approx. 2015 seconds post intervention)

  11. Balance

    Positional sense measured by body-worn sensors (cm)

    Measured: Baseline

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted7 Sep 2023
  2. Study start7 Mar 2024
  3. Primary completion11 Jul 2024
  4. Study completion11 Jul 2024

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Prevention
Allocation
Non-randomised
Intervention model
Single group
Masking
None (open label)
Enrolment
21 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06026813
Sponsor's protocol ID
STU-2022-1038
Other IDs
7R21AG061471-02
Acronym
PAS
Status last verified
Jul 2025
First submitted
25 Aug 2023
Last updated
12 Aug 2025
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing test pressure alternating shoes in people with Foot Ulcer, Diabetic, run by University of Texas Southwestern Medical Center at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 13 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 12 Aug 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 12 Aug 2025). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Maximum Average Pressure at Baseline With Pressure Alternating Shoes, Right Foot (Before Any Cells Were Offloaded) (time frame: Before any cells were offloaded (First 40-seconds of the walk)). The study lists 6 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 12 Aug 2025.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in