Clinical Performance Testing of Philips FAST SpO2 With Masimo Pulse Oximetry Sensors Across Skin Pigmentation
A study testing Philips FAST SpO2 with Masimo Pulse Oximetry Sensors in people with Healthy, run by Masimo Corporation at 1 site in United States.
Trial summary
Validate the performance of Philips FAST SpO2 with Masimo Pulse Oximetry Sensors in determining functional arterial oxygen saturation (SpO2) using arterial saturation (SaO2) as a reference in the range of 70-100% in subjects of varying skin pigmentation. Read more about this study
- Who is it for?
- Ages 18 to 50 yearsHealthy volunteers can take part
- Conditions the study is for
- Healthy
- What is being tested?
- Philips FAST SpO2 with Masimo Pulse Oximetry Sensors
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 70 participants
- Where
- 1 siteUnited States
- Study dates
- 13 Nov 2023to 14 Jan 2025
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 50 years
- All sexes
- Healthy volunteers accepted
- Adult
Who may be able to join
- Subject is 18 to 50 years of age.
- Subject weighs a minimum of 110 lbs.
- Baseline heart rate ≥ 45 bpm and ≤ 90 bpm.
- Blood Pressure: Systolic BP ≤ 140 mmHg and ≥ 90 mmHg, Diastolic BP ≤ 90 mmHg and ≥ 50 mmHg, and if systolic BP is lower than 100 mmHg and/or diastolic BP is lower than 60 mmHg, subject passes an orthostatic blood pressure test.
- Hemoglobin value ≥ 11 g/dL.
- CO value ≤ 3.0% FCOHb.
Show all 38Show fewer
- Subject can read and communicate in English and understands the study and the risks involved.
- Exclusion Criteria (Ineligible Subjects) ( = per clinician discretion):
- Subject is pregnant or breastfeeding.
- Subject has a BMI > 35.
- Subject has a history of fainting (vasovagal syncope), blacking out or losing consciousness during or after a blood draw, or has a fear of blood draws.
- Subject has open wounds, inflamed tattoos, or piercings, and/or has any visible healing wounds that a medical professional determines may place them at an increased risk for participation. *
- Subject has finger deformities, nail deformities, nail polish, and/or gel/acrylic that can interfere with study device placement. *
- Subject has known drug or alcohol abuse.
- Subject uses recreational drugs. *
- Subject experiences frequent or severe headaches and/or migraine headaches, migraine auras, altitude sickness, and/or headaches accompanied by visual changes or sensitivity to light or sound.
- Subject has experienced a concussion or head injury with loss of consciousness within the past 12 months.
- Subject has any history of a stroke, myocardial infarction (heart attack), and/or seizures.
- Subject has any chronic bleeding disorder (e.g., hemophilia).
- Subject has taken anticoagulant medication within the past 30 days (excluding nonsteroidal anti-inflammatory drugs (NSAIDs)).
- Subject has donated blood within the past 4 weeks.
- Subject has any cardiac dysrhythmia (e.g., atrial fibrillation) and has not received clearance from their physician to participate.
- Subject has a known neurological and/or psychiatric disorder (e.g., schizophrenia, bipolar disorder, Multiple Sclerosis, Huntington's disease) that interferes with the subject's level of consciousness. *
- Subject has taken opioid pain medication 24 hours before the study.
- Subject has any infectious disease (e.g., Hepatitis, HIV, Tuberculosis, Flu, Malaria, Measles). *
- Subject is taking medications known to treat any type of infectious disease.
- Subject has either signs or history of peripheral ischemia or carpal tunnel syndrome.
- Subject has had invasive surgery within the past year, including but not limited to major dental surgery, appendectomy, plastic surgery, jaw surgery, major ENT surgery, major abdominal and/or pelvic surgery, heart surgery, or thoracic surgery. *
- Subject has symptoms of congestion, head cold, or other illnesses.
- Subject has been in a severe car accident(s) or a similar type of accident(s) requiring hospitalization within the past 12 months.
- Subject has any cancer or history of cancer (not including skin cancer). *
- Subject has chronic unresolved asthma, lung disease (including COPD) and/or respiratory disease.
- Subject is allergic to lidocaine, chlorhexidine, latex, adhesives, or plastic.
- Subject has a heart condition, insulin-dependent diabetes, uncontrolled hypertension, or hypercholesterolemia.
- Subject delivered vaginally, had a pregnancy terminated, had a miscarriage with hospitalization, or had a C-section within the past 6 months.
- Subject intends on participating in any heavy lifting, repetitive movement of their wrist (including riding a motorcycle, tennis), exercise (e.g., working out, riding a bike, riding a skateboard), or any activity that will put additional stress on the wrist within 24 hours following a study that involves an arterial line.
- Difficulty inserting an intravenous line in the subject's hand or arm and/or an arterial line in the radial artery of the subject's wrist.
- Subject has any medical condition which in the judgment of the investigator and/or medical staff, renders them ineligible for participation in this study or subject is deemed ineligible by the discretion of the investigator/study staff.
Who cannot join
None listed.
Trial locations and contacts
1 site in 1 country.
About this study
Validate the performance of Philips FAST SpO2 with Masimo Pulse Oximetry Sensors in determining functional arterial oxygen saturation (SpO2) using arterial saturation (SaO2) as a reference in the range of 70-100% in subjects of varying skin pigmentation.
- Conditions
- Healthy
- Being tested
- Philips FAST SpO2 with Masimo Pulse Oximetry Sensors
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Philips FAST SpO2 with Masimo Pulse Oximetry SensorsDevice
Noninvasive pulse oximeter
| Study group | Role | What they receive |
|---|---|---|
| Test Subject | Experimental | All subjects who are enrolled into the test group and participate in data collection have the Philips FAST SpO2 with Masimo Pulse Oximetry Sensors Receives: Device: Philips FAST SpO2 with Masimo Pulse Oximetry Sensors |
What is being measured
The primary outcome is the main question the study is designed to answer.
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® Adt)
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® Adt) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device.
Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n.
Participants in this arm could receive one or more sensor types used for data collection.
Measured: 1 hour
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® Adt CS)
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® Adt CS) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device.
Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n.
Participants in this arm could receive one or more sensor types used for data collection.
Measured: 1 hour
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® YI)
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® YI) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device.
Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n.
Participants in this arm could receive one or more sensor types used for data collection.
Measured: 1 hour
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® DBI)
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® DBI) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device.
Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n.
Participants in this arm could receive one or more sensor types used for data collection.
Measured: 1 hour
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (LNCS® Adtx)
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (LNCS® Adtx) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device.
Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n.
Participants in this arm could receive one or more sensor types used for data collection.
Measured: 1 hour
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® TC-I)
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® TC-I) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device.
Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n.
Participants in this arm could receive one or more sensor types used for data collection.
Measured: 1 hour
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® DCI)
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® DCI) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device.
Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n.
Participants in this arm could receive one or more sensor types used for data collection.
Measured: 1 hour
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted28 Nov 2023
- Study start13 Nov 2023
- Primary completion14 Jan 2025
- Study completion14 Jan 2025
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for HealthyStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Other
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 70 participants (actual)
Sponsor and investigators
- Lead sponsor
- Masimo Corporation
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT06148623
- Sponsor's protocol ID
- CIP-1088
- Status last verified
- Mar 2026
- First submitted
- 20 Nov 2023
- Last updated
- 2 Apr 2026
- Results posted
- Yes, on ClinicalTrials.gov
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
Study at a glance
From the registry record, last updated by the sponsor on 2 Apr 2026. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Clinical Performance Testing of Philips FAST SpO2 With Masimo Pulse Oximetry Sensors Across Skin Pigmentation
- ClinicalTrials.gov ID
- NCT06148623
- Acronym / sponsor ID
- CIP-1088
- Conditions
- Healthy
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 70 participants (actual)
- Lead sponsor
- Masimo CorporationIndustry (pharma, biotech, devices)
- Study start
- 13 Nov 2023 (actual)
- Primary completion
- 14 Jan 2025 (actual)
- Study completion
- 14 Jan 2025 (actual)
- First posted
- 28 Nov 2023 (actual)
- Results
- Posted 11 Mar 2026
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Validate the performance of Philips FAST SpO2 with Masimo Pulse Oximetry Sensors in determining functional arterial oxygen saturation (SpO2) using arterial saturation (SaO2) as a reference in the range of 70-100% in subjects of varying skin pigmentation.
Study design
- Primary purpose
- Other
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Test Subject | Experimental | Device: Philips FAST SpO2 with Masimo Pulse Oximetry SensorsAll subjects who are enrolled into the test group and participate in data collection have the Philips FAST SpO2 with Masimo Pulse Oximetry Sensors |
Who may qualify
- Ages
- 18 to 50 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult
Key inclusion criteria (38 listed)
- Subject is 18 to 50 years of age.
- Subject weighs a minimum of 110 lbs.
- Baseline heart rate ≥ 45 bpm and ≤ 90 bpm.
- Blood Pressure: Systolic BP ≤ 140 mmHg and ≥ 90 mmHg, Diastolic BP ≤ 90 mmHg and ≥ 50 mmHg, and if systolic BP is lower than 100 mmHg and/or diastolic BP is lower than 60 mmHg, subject passes an orthostatic blood pressure test.
- Hemoglobin value ≥ 11 g/dL.
Key exclusion criteria (0 listed)
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Philips FAST SpO2 with Masimo Pulse Oximetry SensorsNoninvasive pulse oximeter | Device | Test Subject |
Primary outcome
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® Adt)
Time frame: 1 hour
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® Adt) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection.
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® Adt CS)
Time frame: 1 hour
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® Adt CS) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection.
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® YI)
Time frame: 1 hour
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® YI) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection.
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® DBI)
Time frame: 1 hour
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® DBI) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection.
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (LNCS® Adtx)
Time frame: 1 hour
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (LNCS® Adtx) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection.
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® TC-I)
Time frame: 1 hour
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® TC-I) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection.
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® DCI)
Time frame: 1 hour
The SpO₂ accuracy of Philips FAST SpO₂ with Masimo Pulse Oximeter Sensor (RD SET® DCI) was determined by calculating the accuracy root mean-squared (ARMS) difference between the measured values (SpO₂) to the reference values (SaO₂) in accordance with ISO-80601-2-61. ARMS is a standard statistical method used in pulse oximetry performance evaluations to quantify the overall accuracy of a device. Arms= √(∑(i=1 to n) ((SpO₂_i-SaO₂_i )^2 ))/n. Participants in this arm could receive one or more sensor types used for data collection.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 2 Apr 2026.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Healthy
Recruiting
Trained vs. Untrained: Lower Extremity Pressure Changes
Trained Athletes, Untrained
Not yet recruitingPhase 1
Comparison of Pharmacokinetic/Pharmacodynamic Characteristics and Safety of GZR101-80 Injection With Separately Injected, Simultaneous Doses of GZR33 Injection and Insulin Aspart
Healthy Participants
Active, not recruitingPhase 2
Effect of Equol Supplementation on Arterial Stiffness and Cognition in Healthy Volunteers
Arterial Stiffness, White Matter Lesions
Enrolling by invitationObservational study
Screening for Hematology Branch Protocols
Hematologic Disease and Disorders, Donors
RecruitingPhase 1
A Study of Single and Multiple Subcutaneous or Intravenous Doses of ADX-001 in Healthy Adults to Assess Safety, Effects, and Immune Response
Healthy
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Philips FAST SpO2 with Masimo Pulse Oximetry Sensors in people with Healthy, run by Masimo Corporation at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 38 inclusion and 0 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 2 Apr 2026. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 11 Mar 2026). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Oxygen Saturation Performance of Philips FAST SpO₂ With Masimo Pulse Oximeter Sensor (RD SET® Adt) (time frame: 1 hour). The study lists 7 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 2 Apr 2026.