BGM and HbA1c POC Device Evaluation

A study testing CareSens N-Eco (i-Sens), • CareSens N-Premier (i-Sens), CodeFree (SD Biosensor), GlucoNavii GDH (SD Biosensor) and Affinion (Abbott), HbA1c 501 (HemoCue), A1Care (iSens), DCA Vantage (Siemens) in people with Diabetes, run by Foundation for Innovative New Diagnostics, Switzerland at 3 sites in 3 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record

Search all trials

Trial summary

This study aims to assess analytical performance of four BGMs and four HbA1c POC devices in comparison to suitable comparator or reference methods, with the view to evaluate performance in the hands of trained users, as well as intended users who are unfamiliar with the system (lay users, for BGMs only). Read more about this study

Who is it for?
Ages 13 years and older
Conditions the study is for
Diabetes
What is being tested?
CareSens N-Eco (i-Sens), • CareSens N-Premier (i-Sens), CodeFree (SD Biosensor), GlucoNavii GDH (SD Biosensor) and Affinion (Abbott), HbA1c 501 (HemoCue), A1Care (iSens), DCA Vantage (Siemens)
Could I receive a placebo?
No, everyone receives an active treatment
How many people
910 participants
Where
3 sitesin 3 countries
Study dates
9 Jun 2023to 24 Jul 2024
Phase
Interventional

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 13 years and older
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Arm 1:
  • 13-17 years, diagnosed with diabetes and able to manage condition independently (ideally at least 10% of total Arm 1 population)
  • •≥18 years, diagnosed with diabetes
  • Able and willing to provide informed consent/assent
  • Naïve to the SMBG use of the investigational devices
  • 10% naïve to any SMBG (defined as not having performed any SMBG in the last 24 months)
Show all 14Show fewer
  • Haematocrit within 20-60%
  • Pre-screen BGM result qualifies for inclusion in any of the five glucose concentration ranges based on % of subjects required to meet the target number per concentration range
  • Arm 2:
  • •≥18 years
  • Able and willing to provide informed consent
  • Diagnosed or not diagnosed with diabetes
  • Haemoglobin values ≥ 8g/dL
  • Pre-screen HbA1c result qualifies for inclusion in the target HbA1c range and % of subjects required for the mid-range values

Who cannot join

  • Arm 1:
  • <13 years
  • 13-17 years and not diagnosed with diabetes or unable to manage condition independently
  • Previous use of investigational BGM for SMBG
  • Haematocrit outside normal range
  • Pre-screen BGM result falling in a glucose concentration range that already has sufficient participants.
Show all 10Show fewer
  • Arm 2:
  • <18 years
  • Haemoglobin value < 8g/dL
  • Pre-screen HbA1c result falling in a concentration range that already has sufficient participants.

Trial locations and contacts

3 sites in 3 countries.

3 sites in 3 countries

Cambodia1 site
  • Sihanouk Hospital Centre of Hope

    Phnom Penh, Phnom Penh, 12253

Other trials in Cambodia

Colombia1 site
  • Asociación Colombiana de Diabetes

    Bogotá, Bogota D.C., 111311

Other trials in Colombia

Nigeria1 site
  • Lagos University Teaching Hospital

    Lagos, Lagos

Other trials in Nigeria

About this study

Official title: Performance Evaluation of Blood Glucose Meters and HbA1c Point-of-care Devices

This study aims to assess analytical performance of four BGMs and four HbA1c POC devices in comparison to suitable comparator or reference methods, with the view to evaluate performance in the hands of trained users, as well as intended users who are unfamiliar with the system (lay users, for BGMs only). System usability will also be assessed for all BGMs (including lancing devices) and HbA1c devices.

Conditions
Diabetes
Being tested
CareSens N-Eco (i-Sens), • CareSens N-Premier (i-Sens), CodeFree (SD Biosensor), GlucoNavii GDH (SD Biosensor), Affinion (Abbott), HbA1c 501 (HemoCue), A1Care (iSens), DCA Vantage (Siemens)
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • CareSens N-Eco (i-Sens), • CareSens N-Premier (i-Sens), CodeFree (SD Biosensor), GlucoNavii GDH (SD Biosensor)Diagnostic test

    BGM Test

  • Affinion (Abbott), HbA1c 501 (HemoCue), A1Care (iSens), DCA Vantage (Siemens)Diagnostic test

    HbA1c Test

Study groups in NCT06170515
Study groupRoleWhat they receive
BGM EvaluationOtherNot described
Receives: Diagnostic Test: CareSens N-Eco (i-Sens), • CareSens N-Premier (i-Sens), CodeFree (SD Biosensor), GlucoNavii GDH (SD Biosensor)
HbA1c evaluationOtherNot described
Receives: Diagnostic Test: Affinion (Abbott), HbA1c 501 (HemoCue), A1Care (iSens), DCA Vantage (Siemens)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. BGM Performace

    For BGMs: Performance accuracy is based on ISO 15197 criteria, mean relative bias, 95% limits of agreement, and regression slopes of each BGM with respect to two independently validated comparator BGMs (each investigational BGM will be compared to each comparator BGMs independently).

    Measured: 1 year

  2. HbA1c Performance

    For HbA1c: Performance accuracy is based on regression slopes, mean relative bias, 95% limits of agreement, and of each HbA1c device versus the reference method and the concordance between capillary and venous blood results.

    Measured: 1 year

Secondary outcomes (8)
  1. Comprehension

    Proportion of label comprehension, system usability and result/symbol interpretation scale rating per question across all lay users stratified by BGM and lancing device; Total system usability score scale per BGM and lancing device.

    Measured: 1 year

  2. Usability

    Proportion of system usability and result/symbol interpretation scale rating per question across all lay users stratified by BGM and lancing device; Total system usability score scale per BGM and lancing device.

    Measured: 1 year

  3. Interpretation

    Proportion of result/symbol interpretation scale rating per question across all lay users stratified by BGM and lancing device; Total system usability score scale per BGM and lancing device

    Measured: 1 year

  4. Number of errors as observed

    Number and type of errors observed; proportion of error type observed per BGM and lancing device .

    Measured: 1 year

  5. Type of errors as observed

    Number and type of errors observed; proportion of error type observed per BGM and lancing device .

    Measured: 1 year

  6. Proportion of label comprehension

    Proportion of system usability scale rating per question across all healthcare professionals, stratified by HbA1c POC device; Total system usability score scale per HbA1c POC device

    Measured: 1 year

  7. Number of errors as displayed.

    Number and type of errors as displayed on the BGM or HbA1c POC devices, stratified by type of user (BGMs only) and device

    Measured: 1 year

  8. Type of errors as displayed

    Number and type of errors as displayed on the BGM or HbA1c POC devices, stratified by type of user (BGMs only) and device

    Measured: 1 year

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted14 Dec 2023
  2. Study start9 Jun 2023
  3. Primary completion24 Oct 2023
  4. Study completion24 Jul 2024

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Non-randomised
Intervention model
Parallel groups
Model description
Arm 1: Evaluation of BGM's (listed below) * CareSens N-Eco (i-Sens) * CareSens N-Premier (i-Sens) * CodeFree (SD Biosensor) * GlucoNavii GDH (SD Biosensor) Comparator BGMs: Ascensia Contour Next One and Nova Statstrip Xpress (BG); using strip lots meeting CLSI POCT12 acceptance criteria as validated at an experienced laboratory using traceable analysers Arm 2: HbA1c evaluation (listed below) * Affinion (Abbott) * HbA1c 501 (HemoCue) * A1Care (iSens) * DCA Vantage (Siemens) Reference: NGSP-certified HbA1c laboratory method, validated with ERL reference samples
Masking
None (open label)
Enrolment
910 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06170515
Sponsor's protocol ID
NC002
Status last verified
Sep 2026
First submitted
27 Oct 2023
Last updated
17 Sep 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Singh P, Lenters-Westra E, Choun K, Fasanmade O, Erkosar B, Watson M, Haldane C, Vetter B. Analytical performance of four HbA1c point-of-care devices: a cross-sectional evaluation in low- and middle-income countries. BMJ Open. 2026 May 12;16(5):e113491. doi: 10.1136/bmjopen-2025-113491.

    PubMed 42129995

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in