Sleep Apnea SyndromesEventLAB FFNCT06317077

Investigation of the Accordance Between Event Detection of prismaLINE Devices and Polysomnography Using Full Face Masks

A study testing PAP therapy setting and exclusive use of Löwenstein Medical full face masks in people with Sleep Apnea Syndromes, run by Löwenstein Medical Technology GmbH & Co. KG at 1 site in Germany.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Sleep-related breathing disorders are highly prevalent and are usually treated with positive airway pressure (PAP) therapy. To determine the quality of therapy, PAP therapy devices measure the frequency of residual breathing events. Read more about this study

Who is it for?
Ages 18 to 80 years
Conditions the study is for
Sleep Apnea Syndromes
What is being tested?
PAP therapy setting and exclusive use of Löwenstein Medical full face masks
Could I receive a placebo?
No, everyone receives an active treatment
How many people
49 participants
Where
1 siteGermany
Study dates
2 Dec 2024to 17 Jun 2026
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 80 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Polysomnographic diagnosis performed at the investigator site
  • Indication for CPAP therapy (initial setting)
  • Age of the participants: ≥ 18 ≤ 80 years
  • AHI diagnosis night ≥ 15/h
  • Capacity to consent
  • Agreement to use a full face mask of Löwenstein Medical during the titration night
Show all 15Show fewer
  • Existence of a signed informed consent form for participation in the clinical investigation and a signed declaration of consent for the use of data within the scope of the DSGVO
  • AHI PSG (diagnosis and therapy) and AHI of the device (therapy) are completely available
  • Evaluation of the mask/device functionalities is available
  • Clear documentation of (predictable) adverse events and (predictable) adverse device effects
  • Existence of a signed informed consent form for participation in the clinical investigation and a signed declaration of consent for the use of data within the scope of the DSGVO
  • Reliable evaluation of the polysomnography (titration night) is possible: no missing signals, good signal quality, no deviation of the applied CPAP pressure course from the course specified for the clinical investigation
  • Total sleep time during the titration night ≥ 4 hrs
  • Proportion of the therapy time with high leakage (more than 50 l/min) < 10%
  • The quality assurance process did not identify any deviations from the source data that could not be corrected

Who cannot join

  • Presence of a contraindication to CPAP therapy
  • Participation in another clinical investigation that influences the adjustment to CPAP therapy by stipulate device settings or titration
  • pregnant or breastfeeding woman

Trial locations and contacts

1 site in 1 country.

Germany1 site
  • Krankenhaus Bethanien

    Solingen, North Rhine-Westphalia, 42699

Other trials in Germany

About this study

Sleep-related breathing disorders are highly prevalent and are usually treated with positive airway pressure (PAP) therapy. To determine the quality of therapy, PAP therapy devices measure the frequency of residual breathing events. For this purpose, breathing events of different classes are detected, counted and identified.

This clinical investigation aims to evaluate the performance of prismaLINE devices based on the accuracy of the apnea-hypopnea index (AHI) recorded in the devices. During the PAP titration night, the AHI is similarly detected via polysomnography (PSG) device and a subsequent manual scoring of the recorded PSG data. The validation of the device AHI accuracy refers to the alignment with the manually scored AHI.

Read more

Further goals of this clinical investigation are the confirmation of safety of the prismaLINE devices, the confirmation of performance and safety of LM full face masks (full face masks of the manufacturer Löwenstein Medical Technology), as well as the investigation of the influence of LM full face masks on therapy quality.

The most relevant inclusion criteria are diagnosis of a sleep-related breathing disorder and indication for therapy with continuous positive airway pressure (CPAP). Treatment takes place with prismaLINE PAP therapy devices in combination with LM full face masks.

Read the detailed description

The duration of the clinical investigation per participant is two or three nights: diagnosis night with the following CPAP titration night. A further titration night (CPAP or APAP) might be necessary to find optimal therapy settings. But only the CPAP titration night with the best therapy result will be analysed.

The titration night is performed according to a predefined titration scheme. Therapy pressure will be gradually increased until breathing events are treated sufficiently. Due to the need of a certain number of remaining breathing events for sufficient performance evaluation, therapy pressure will be gradually reduced after reaching the ideal pressure level to provoke additional breathing events. The titration night is additionally monitored and recorded via polysomnography (PSG). No significant negative impact on patients' condition based on subtherapeutic phase is expected.

After the end of the clinical investigation participants will be treated according to clinical routine.

The performance of the prismaLINE therapy device will be evaluated by the accuracy of the alignment between the device AHI and the manually scored PSG AHI (using AASM hypopnea scoring rule 1A). The alignment between device AHI versus PSG AHI will be defined for coincident titration night recordings of both, therapy and PSG device.

The accuracy of the device AHI is defined by the absolute or relative deviation from the manually scored PSG AHI (ΔAHI_rel or ΔAHI_abs) and classified in acceptable, good, very good or unacceptable. The maximum deviation (ΔAHI_rel or ΔAHI_abs) is used throughout to determine the corresponding device AHI values for each category.

To achieve the primary endpoint, the proportion of the investigation population with an at least acceptable device AHI must be ≥ 80%.

Being tested
PAP therapy setting, exclusive use of Löwenstein Medical full face masks
Keywords
sleep-related breathing disorders, obstructive sleep apnea, event classification, device, full face mask, apnea-hypopnea-index, polysomnography, accuracy
Medical subject headings
Sleep Apnea Syndromes, Sleep Apnea, Obstructive

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • PAP therapy settingOther

    Subtherapeutic phases during titration to provoke breathing events after reaching the pressure level with which the breathing events are sufficiently treated

  • exclusive use of Löwenstein Medical full face masksOther

    Use of a Löwenstein Medical full face mask as a requirement for participation in the clinical investigation

Study groups in NCT06317077
Study groupRoleWhat they receive
Treatment groupOtherAll participants of the clinical investigation undergo the standard stages of the clinical routine within an initial PAP therapy setting: a diagnostic night followed by a titration night (or if necessary two titration nights) with prescribed titration scheme. All participants use a full face mask of Löwenstein Medical.
Receives: Other: PAP therapy setting; Other: exclusive use of Löwenstein Medical full face masks

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Delta AHI [n/h]= Difference of device AHI (prismaLINE devices of the type WM100TD) and the manual scored AHI of polysomnographic data identified in the Matlab-based SOMNOmat tool

    comparison of PSG AHI (according to AASM hypopnea scoring rule 1A and 1B) and device AHI for the period in which the records of both devices are available

    Measured: 1 night

Secondary outcomes (14)
  1. N_AE device [n]= number and severity of occurred (predictable) adverse events and (predictable) adverse device effects, counted during study duration

    Confirmation of safety of the prisma therapy devices (type WM100TD)

    Measured: 1 night

  2. Mask fit score= Leakage, pressure points, stable pressure level, comfort determined via questionnaire

    Confirmation of performance of the full face masks of Löwenstein Medical

    Measured: 1 night

  3. N_AE mask [n]= number and severity of occurred (predictable) adverse events and (predictable) adverse device effects, counted during study duration

    Confirmation of safety of the full face masks of Löwenstein Medical

    Measured: 1 night

  4. Delta AI [n/h]= difference of apnea-indices (obstructive/central/mixed) between the therapy device and the manually evaluated PSG data identified in the Matlab-based tool SOMNOmat

    Investigation of the accordance of the individual event indices of the synchronized device and PSG files

    Measured: 1 night

  5. Delta HI [n/h]= difference of hypopnea-indices (obstructive/central) between the therapy device and the manually evaluated PSG data identified in the Matlab-based tool SOMNOmat

    Investigation of the accordance of the individual event indices of the synchronized device and PSG files

    Measured: 1 night

  6. Delta Event_individual [n]= difference of the number (with time, duration, type) of residual breathing events (apneas/hypopneas) device vs. PSG identified in the Matlab-based tool SOMNOmat

    Comparison of all AHI-relevant individual events of the synchronized device and PSG files

    Measured: 1 night

  7. N_influenced phases [n]= number (with time, duration, type) of phases with significant deviations between the synchronized device and PSG files identified in the Matlab-based tool SOMNOmat

    investigation of influencing factors for accordance between therapy device and the polysomnography (e.g. phases with increased pressure or increased leakage)

    Measured: 1 night

  8. Delta AHI_1A [n/h]= difference between apnea indices using scoring rule 1A from AASM manual version 3.0

    investigation of impact of AASM hypopnea scoring rules 1A from AASM manual, version 3.0

    Measured: 1 night

  9. Delta AHI_1B [n/h]= difference between apnea indices using scoring rule 1B from AASM manual version 3.0

    investigation of impact of AASM hypopnea scoring rules 1B from AASM manual, version 3.0

    Measured: 1 night

  10. Delta AHI_mean [n/h]= difference between apnea indices using mean of scoring rule 1A and 1B from AASM manual version 3.0 (AHI_1A + AHI_1B)/2

    investigation of impact of AASM hypopnea mean of scoring rule 1A and 1B from AASM manual, version 3.0

    Measured: 1 night

  11. Delta AHI_Routine [n/h]= difference between unmachined apnea indices of device and PSG

    comparison of original device AHI and PSG AHI to investigate the alignment of routine device AHI with routine PSG AHI (1A/1B)

    Measured: 1 night

  12. N_impact factors [n]= number and type of factors that can increase the risk for sleep related breathing disorders like comorbidities and smoking

    investigation of influencing factors and identification of confounders

    Measured: 1 night

  13. corr_flow-sleep= correlation between flow signal and sleep state

    identification of correlation between flow signal and sleep state

    Measured: 1 night

  14. calc_marker = possible correlations of calculated sleep apnea syndrome markers (e. g. hypoxic burden) with medical history data, diagnosis biomarker, therapy pressure and sleep stage plus analysis of deviations between different calculation methods

    identification of correlation of modern versus classic diagnostic sleep apnea syndrome markers

    Measured: 1 night

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted19 Mar 2024
  2. Study start2 Dec 2024
  3. Primary completion17 Jun 2026
  4. Study completion17 Jun 2026

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
49 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06317077
Sponsor's protocol ID
I5-R1-C-FA
Acronym
EventLAB FF
Status last verified
Nov 2024
First submitted
2 Mar 2024
Last updated
12 Aug 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study testing PAP therapy setting and exclusive use of Löwenstein Medical full face masks in people with Sleep Apnea Syndromes, run by Löwenstein Medical Technology GmbH & Co. KG at 1 site in Germany.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 15 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 12 Aug 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Germany. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Delta AHI [n/h]= Difference of device AHI (prismaLINE devices of the type WM100TD) and the manual scored AHI of polysomnographic data identified in the Matlab-based SOMNOmat tool (time frame: 1 night). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.