Lung CancerINSPECTNCT06380361

InvestigatioN of a Smart Probe for Lung lEsion Characterization Using Impedance Technology

A study testing BioSpy System in people with Lung Cancer, run by Sensome at 2 sites in 2 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The objective of the study is to evaluate the feasibility of the BioSpy System sensor to differentiate tissues that are encountered during bronchoscopic biopsy of endobronchial tumors and peripheral lung nodules and masses. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Lung Cancer
What is being tested?
BioSpy System
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
30 participants
Where
2 sitesin 2 countries
Study dates
28 Mar 2024to 23 Apr 2025
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Age > 18 years
  • Subjects with lesions eligible for lung biopsy under general anesthesia.
  • Lesion localization:
  • Central or proximal lesions ≥ 10 mm in diameter confirmed by imaging (CT scan and/or PET scan) and/or endobronchial visual control; or
  • Peripheral lesions ≥ 20 mm in diameter confirmed by imaging (CT scan and/or PET scan) and/or ultrasound analysis (RP EBUS with central localization of the ultrasound probe) during the procedure.
  • Written Informed Consent to participate in the study.

Who cannot join

  • Target lesion <10 mm for central and <20 mm for peripheral lesions (as determined on previous imaging)
  • Contra-indication to bronchoscopy procedures
  • Contra-indication to general anesthesia
  • Any subject that is, according to the discretion of the investigator, not eligible for study participation
  • Known lactating or confirmation of positive pregnancy test according to site specific standard of care (e.g. test, verbal communication)

Trial locations and contacts

2 sites in 2 countries.

2 sites in 2 countries

Australia1 site
  • Royal Brisbane and Women's Hospital

    Brisbane, Queensland, 4019

Other trials in Australia

France1 site
  • Marie-Lannelongue Hospital

    Le Plessis-Robinson, 92350

Other trials in France

About this study

The objective of the study is to evaluate the feasibility of the BioSpy System sensor to differentiate tissues that are encountered during bronchoscopic biopsy of endobronchial tumors and peripheral lung nodules and masses.

Read the detailed description

Participants in this study require a lung biopsy to confirm if they have a lung tumour. A biopsy is a procedure where a piece of lung tissue is removed for testing. Current biopsy techniques cannot guarantee the exact location of lung tumours, so biopsies can be taken from healthy tissue.

The BioSpy System (BSS) is a device with a probe, which is placed inside a normal biopsy needle. The tip of the probe contains sensors, that will scan the composition of the tissue which is in contact with the sensors.

The INSPECT study aims at collecting the sensor data, with the goal of being able to differentiate the different tissues.

All study participants will have a biopsy with the BSS. Participants will have a hospital visit for the procedure, and will have to stay up to 16 hours after their procedure as per normal hospital practice.

Conditions
Lung Cancer
Being tested
BioSpy System
Keywords
Lung cancer, Lung masses, Lung Nodes, Bronchoscopic biopsy
Medical subject headings
Lung Neoplasms

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • BioSpy SystemDevice

    One arm only - in all eligible study patients, the study device will be used. In patients with suspicion of lung cancer a bronchoscope will be inserted to navigate to the lesion where a biopsy sample will be taken for histology analysis. In the current trial, the device will be used to perform electrophysiological measurements in the lesion during bronchoscopic biopsy procedure.

Study groups in NCT06380361
Study groupRoleWhat they receive
All patients will be allocated to the same study arm.ExperimentalDuring the bronchoscopic biopsy procedure, the study device will be entered through the bronchoscope to the level of the lesion. At lesion level, the BioSpy System sensor will make several measurements of the tumors and peripheral lung nodules and masses.
Receives: Device: BioSpy System

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. The Ability of BioSpy System to Acquire Electrophysiological Measurements in the Relevant Tissues During Bronchoscopic Biopsy.

    The primary endpoint was defined as the proportion of patients in which at least one non-anomalous biophysical measurement was obtained by the BSS in the lesion. This endpoint represents the procedural success rate being defined as the BioSpy System obtaining at least one non-anomalous impedance measurement in the lesion during the procedure, divided by the total number of patients in whom BSS was used.

    Measured: During the bronchoscopy procedure, up to 79 minutes

Secondary outcomes (2)
  1. The Ability of BioSpy System to Differentiate the Lesion From Healthy Tissue

    Electrical impedance measurements were continuously acquired during bronchoscopy using the BioSpy System. The physician annotated time intervals corresponding to sensor contact with suspected lesion tissue and with healthy tissue, using bronchoscopic visualization and imaging guidance.

    Impedance data collected within these annotated windows were extracted for analysis. Each window was matched to the corresponding histopathological diagnosis from biopsy samples obtained at the same location.

    Machine learning model predictions (lesion vs. healthy) based on the impedance data were compared with the biopsy results to assess diagnostic performance.

    Models were evaluated with leave-one-patient-out training: one patient was held for testing while the model trained on the others. This cycle repeated until all patients were tested, and the mean performance across iterations was computed. Mean accurary, mean sensitivity and mean specificity of detecting lesion are reported.

    Measured: During the bronchoscopy procedure, up to 79 minutes

  2. The Ability of BioSpy System to Differentiate Various Lesion Types

    The BioSpy System's ability to differentiate tumoral, inflamed, necrotic and fibrotic tissues was assessed using continuous electrical impedance measurements recorded during bronchoscopy.

    Physicians annotated intervals where the sensor contacted suspected lesion or healthy tissue guided by bronchoscopic visualization and imaging. Impedance data from these annotated windows were extracted and paired with histopathological diagnoses from biopsies taken at the same location.

    Due to limitations in available labels, the secondary endpoint was evaluated by training a machine-learning model to distinguish cancer from all other tissue types.

    Model performance was assessed with a leave-one-patient-out procedure: one patient was used for testing while the others were used for training, repeating the process until all had been tested.

    Mean accuracy, mean sensitivity and mean specificity for cancer detection are reported.

    Measured: During the bronchoscopy procedure, up to 79 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted23 Apr 2024
  2. Study start28 Mar 2024
  3. Primary completion23 Apr 2025
  4. Study completion23 Apr 2025

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Model description
Patient with suspected lung cancer coming to the hospital for bronchoscopic biopsy will be selected for this trial.
Masking
None (open label)
Enrolment
30 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06380361
Sponsor's protocol ID
SEN-ONCO-1
Acronym
INSPECT
Status last verified
Mar 2026
First submitted
27 Mar 2024
Last updated
21 Apr 2026
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing BioSpy System in people with Lung Cancer, run by Sensome at 2 sites in 2 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 21 Apr 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in Australia, France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 21 Apr 2026). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

The Ability of BioSpy System to Acquire Electrophysiological Measurements in the Relevant Tissues During Bronchoscopic Biopsy. (time frame: During the bronchoscopy procedure, up to 79 minutes). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 21 Apr 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in