Normal SubjectNCT06496763

Assessment of an Automated Optical Coherence Tomography and Camera : Effectiveness of Vision-700

An observational study of Normal Subject and Glaucoma, run by Crystalvue Medical Coporation at 1 site in Taiwan.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The objective of study is to assess the repeatability and agreement of the retina and cornea measurement by Vision-700 in comparison with those by 3D OCT-1. Read more about this study

Who is it for?
Ages 22 years and olderHealthy volunteers can take part
Conditions the study is for
Normal Subject, Glaucoma, Retinal Disease
What is being tested?
Vision-700
Type of treatment
Medical device
How many people
159 participants
Where
1 siteTaiwan
Study dates
1 Sep 2022to 31 Jul 2023
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 22 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Study population: Healthy subjects, Glaucoma subjects, Retinal diseasse subjects, Cornea condition subjects

Who may be able to join

  • Subject 22 years old or older on the date of informed consent; Subjects able to understand the written informed consent and willing to participate as evidenced by signing the informed consent.
  • Normal Group:
  • (1)Subjects presenting at the site with normal eyes bilaterally (cataracts are acceptable). (2)IOP ≤ 21 mmHg bilaterally; BCVA 20/40 or better bilaterally. (3) No history of corneal, glaucoma, or retinal diseases. (4)No history of ocular surgeries, including laser refractive surgery, cataract surgery, glaucoma surgery and retinal surgery. (5)Subjects without any current ocular pathology other than cataract in the normal eye(s), as determined by self-report and/or investigator assessment at the study visit.
  • Glaucoma Group:
  • With a current BCVA of 20/40 or better in the glaucoma study eyes. (2) History of visual field defects within the previous six months from the study visit or measured on the day of the study visit that is consistent with glaucomatous optic nerve damage. (3) Subjects who have been diagnosed with glaucoma in the glaucoma study eyes based on the presence of visual field defects, regardless of the optic nerve head and RNFL appearance. (4) Subjects without retinal disease or cornea condition in the glaucoma study eyes, as determined by self-report and/or investigator assessment at the study visit.
  • Retinal Disease Group:
Show all 9Show fewer
  • (1)Subjects with a current BCVA of 20/400 or better in the retinal disease study eye(s) at the study visit. (2)Subjects diagnosed with retinal pathology including but not limited to any one of following criteria: a. Non Exudative Macular Degeneration (dry AMD), Diabetic Macular Edema, Macular Hole, Epiretinal Membrane, Cystoid Macular Edema, or other retinal disease in the study eye(s) as confirmed within the past six months, who exhibit structural lesions in the study eye; b. Neovascular Macular Degeneration (wet AMD), Diabetic Retinopathy, Retinal Artery Occlusion, Retinal Vein Occlusion or other retinal disease in the study eye(s) as confirmed within the past six months, who exhibit vascular and/or ischemic lesions in the study eye. (3) Subjects without glaucoma or cornea condition in the retinal disease study eye(s), as determined by self-report and/or investigator assessment at the study visit.
  • Cornea Condition Group:
  • History of refractive surgery, contact lens (wear contact lens more than 6 months), dry eye or keratoconus.

Who cannot join

  • Subjects unable to tolerate ophthalmic imaging;
  • Subjects with ocular media not sufficiently clear to obtain acceptable OCT images;
  • Subjects with a history of leukemia, dementia or multiple sclerosis.
  • Subjects with history of hydroxychloroquine or chloroquine use
  • Subjects with photosensitivity
  • Subject having Photodynamic therapy (PDT) within 6 months
  • Subjects taking photosensitivity drug currently
  • Subject has a condition or is in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study.

Trial locations and contacts

1 site in 1 country.

Taiwan1 site
  • Tri-Services General Hospital

    Taipei County

Other trials in Taiwan

About this study

Official title: Assessment of an Automated Optical Coherence Tomography and Camera : Effectiveness o

The objective of study is to assess the repeatability and agreement of the retina and cornea measurement by Vision-700 in comparison with those by 3D OCT-1.

Read the detailed description

The first objective of study is to assess the repeatability and agreement of the retina and cornea measurement by Vision-700 in comparison with those by 3D OCT-1, and Cornea Measurements by RTVue XR + CAM. The second objective is to compare the agreement of fundus photography quality between Vision-700 with 3D OCT-1.

Being tested
Vision-700
Medical subject headings
Glaucoma, Retinal Diseases, Corneal Diseases

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Vision-700Device

    Also called 3D OCT-1 Maestro

    OCT machines used for diagnostic purposes

Study groups in NCT06496763
Study groupRoleWhat they receive
Normal Group:(1)Subjects presenting at the site with normal eyes bilaterally (cataracts are acceptable). (2)IOP ≤ 21 mmHg bilaterally; BCVA 20/40 or better bilaterally. (3) No history of corneal, glaucoma, or retinal diseases. (4)No history of ocular surgeries, including laser refractive surgery, cataract surgery, glaucoma surgery and retinal surgery. (5)Subjects without any current ocular pathology other than cataract in the normal eye(s), as determined by self-report and/or investigator assessment at the study visit.
Receives: Device: Vision-700
Glaucoma Group(1) With a current BCVA of 20/40 or better in the glaucoma study eyes. (2) History of visual field defects within the previous six months from the study visit or measured on the day of the study visit that is consistent with glaucomatous optic nerve damage. (3) Subjects who have been diagnosed with glaucoma in the glaucoma study eyes based on the presence of visual field defects, regardless of the optic nerve head and RNFL appearance. (4) Subjects without retinal disease or cornea condition in the glaucoma study eyes, as determined by self-report and/or investigator assessment at the study visit.
Receives: Device: Vision-700
Retinal Disease Group(1)Subjects with a current BCVA of 20/400 or better in the retinal disease study eye(s) at the study visit. (2)Subjects diagnosed with retinal pathology including but not limited to any one of following criteria: a. Non Exudative Macular Degeneration (dry AMD), Diabetic Macular Edema, Macular Hole, Epiretinal Membrane, Cystoid Macular Edema, or other retinal disease in the study eye(s) as confirmed within the past six months, who exhibit structural lesions in the study eye; b. Neovascular Macular Degeneration (wet AMD), Diabetic Retinopathy, Retinal Artery Occlusion, Retinal Vein Occlusion or other retinal disease in the study eye(s) as confirmed within the past six months, who exhibit vascular and/or ischemic lesions in the study eye. (3) Subjects without glaucoma or cornea condition in the retinal disease study eye(s), as determined by self-report and/or investigator assessment at the study visit.
Receives: Device: Vision-700
Cornea Condition GroupHistory of refractive surgery, contact lens (wear contact lens more than 6 months), dry eye or keratoconus.
Receives: Device: Vision-700

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. The Precision of Vision-700 in Retinal and Cornea Measurement

    The Precision included the repeatability and reproducibility.

    Measured: 2022/9/1

  2. The Precision of 3D OCT-1 in Retinal and Cornea Measurements

    The Precision included the repeatability and reproducibility.

    Measured: 2022/9/1

  3. The Agreement of Vision-700 and 3D OCT-1 in Retinal and Cornea Measurements

    the agreement study

    Measured: 2022/9/1

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted11 Jul 2024
  2. Study start1 Sep 2022
  3. Primary completion1 Sep 2022
  4. Study completion31 Jul 2023

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
159 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06496763
Sponsor's protocol ID
Effectiveness of Vision-700
Status last verified
Jul 2024
First submitted
4 Jul 2024
Last updated
11 Jul 2024
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

An observational study of Normal Subject and Glaucoma, run by Crystalvue Medical Coporation at 1 site in Taiwan.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 9 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 11 Jul 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Taiwan. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

The Precision of Vision-700 in Retinal and Cornea Measurement (time frame: 2022/9/1). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 11 Jul 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in