Non-interventional Analytical Performance Study on Left-over SamplesNCT06518967
Performance Study with the SiMoT System for in Vitro Diagnostic Tests
An observational study of Non-interventional Analytical Performance Study on Left-over Samples, run by Centro di Innovazione Regionale in Single Molecule Digital Assay at 1 site in Italy.
Trial summary
Advanced and performing technologies able to analyze specific biomarkers for the diagnosis, prognosis and prediction as well as for the monitoring, the personalized care and the evaluation of the therapeutic response to pathologies, is currently considered extremely relevant. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Non-interventional Analytical Performance Study on Left-over Samples
- What is being tested?
- analytical performance study
- How many people
- 50 participants
- Where
- 1 siteItaly
- Study dates
- 16 Jul 2024to 31 May 2025
- Phase
- Observational
- Are there results?
- Not yet: the study is under way
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Study population: The study involves around 50 patients with solid tumors who will come to the Gynecological Oncologic Clinic at IRCCS Giovanni Paolo II Cancer Institute and who are prescribed for pre-hospitalization procedures as per standard clinical practice at the investigational site.
Who may be able to join
- Informed Consent and personal data handling consent signed by the subject;
- > 18 years of age
- presence of neoplastic or suspected neoplastic pathology at the time of sample collection
Who cannot join
- < 18 years of age
- Participation in a clinical trial with an investigational drug or a medical device within the 30 days preceding and during the present investigation.
Trial locations and contacts
1 site in 1 country.
Italy1 site
IRCCS Istituto Tumori Giovanni Paolo II, Gynecological Oncologic Clinic
Bari, Bari, 70124
About this study
Official title: Analytical Performance Study of the Handheld Device Single-Molecule with a Large Transistor (SiMoT) for the Near-patient Testing of Biological Samples
Advanced and performing technologies able to analyze specific biomarkers for the diagnosis, prognosis and prediction as well as for the monitoring, the personalized care and the evaluation of the therapeutic response to pathologies, is currently considered extremely relevant. Such technologies work only if molecular markers specific for a given pathology have been already validated. This approach is considered of strategic importance for the health and well-being of citizens as well as for the sustainability of public spending. Often, however, the analysis of the markers is carried out on tissues or fluids of the organs involved, being hence quite invasive procedures. In this scenario, it is strategic to develop devices and platforms that allow the analysis of markers in a minimally invasive or non-invasive way, through the analysis of peripheral biological fluids, such as blood or urine. The goal is to provide clinicians with the attacker's advantage against a plethora of progressive diseases by mass-screening individuals who are affected by a disease long before symptoms appear or to allow the identification of those patients who have a high probability of recurring or responding to target-oriented pharmacological approaches. This study pertains to the definition of the analytical performance of the Single-Sensor Single-Molecule with a large Transistor (SiMoT) device. In the context of the present performance study, the IVD SiMoT system will be equipped with an accessory-cartridge developed and produced without any diagnostic purpose but only to verify the analytical performance of the SiMoT system, in the detection of exogenous biochemical molecules suitably selected to act as markers of interest.
Read the detailed description
Advanced and performing technologies able to analyze specific biomarkers for the diagnosis, prognosis and prediction as well as for the monitoring, the personalized care and the evaluation of the therapeutic response to pathologies, is currently considered extremely relevant. Such technologies work only if molecular markers specific for a given pathology have been already validated. This approach is considered of strategic importance for the health and well-being of citizens as well as for the sustainability of public spending. Among the pathologies that will increasingly be monitored and treated thanks to enabling technologies for precision medicine, tumors are certainly of particular importance. Often, however, the analysis of the markers is carried out on tissues or fluids of the organs involved, being hence quite invasive procedures. In this scenario, it is strategic to develop devices and platforms that allow the analysis of markers in a minimally invasive or non-invasive way, through the analysis of peripheral biological fluids, such as blood or urine. The goal is to provide clinicians with the attacker's advantage against a plethora of progressive diseases by mass-screening individuals who are affected by a disease long before symptoms appear or to allow the identification of those patients who have a high probability of recurring or responding to target-oriented pharmacological approaches.
This study pertains to the definition of the analytical performance of the Single-Sensor Single-Molecule with a large Transistor (SiMoT) device. The SiMoT device is an in vitro diagnostic - IVD system (REGULATION (EU) 2017/746, IVDR Art. 2, point 2) that performs decentralized analyses (near-patient testing, such as, for example, patient's home, emergency unit, ambulance) composed of a palm reader completed by machine-learning / artificial-intelligence software. The system works through an accessory (IVDR Art. 2, point 4) represented by a disposable cartridge, which determines the specific diagnostic application based on the molecular markers detected by the accessory. In the context of the present performance study, the IVD SiMoT system will be equipped with an accessory-cartridge developed and produced without any diagnostic purpose but only to verify the analytical performance of the SiMoT system, in the detection of exogenous biochemical molecules suitably selected to act as markers of interest.
The present performance study aims to verify the qualitative detection (on/off) of target molecules (both proteins and nucleic acids) with a single molecule threshold in 0.1 mL (10^-20 M). The threshold is established by the limit of identification (LOI).
Analytical performance will be evaluated according to standard statistical approaches such as: analytical sensitivity, analytical specificity, trueness (bias), precision (repeatability and reproducibility), accuracy (resulting from trueness and precision), limits of detection and identification, measurement range, possible linearity, threshold values, including the determination of appropriate criteria for the collection and handling of samples and the control of known endogenous and exogenous interferences, cross reactions.
Risk evaluation (Risk-to-Benefits rationale) There are no identifiable physical, psychic and/or social risks for the patients enrolled in the study directly deriving from the procedures that will be followed for the enrollment and collection of the related data.
- Being tested
- analytical performance study
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
analytical performance studyDiagnostic test
analytical performance study on left-over samples with the SiMoT IVD system for in vitro diagnostic tests
| Study group | Role | What they receive |
|---|---|---|
| Open | Test analysis perfomed in open to train the artificial intelligence algorithm. Receives: Diagnostic Test: analytical performance study | |
| Blind | Test analysis performed in blind to minimize biases when evaluating the predictive performance of the IVD under investigation. Receives: Diagnostic Test: analytical performance study |
What is being measured
The primary outcome is the main question the study is designed to answer.
To evaluate the analytical performance in plasma, serum, and urine samples supplemented with non-endogenous molecules (three proteins and two nucleotide sequences) of the SiMoT handheld device
Qualitative detection of specific marker at single molecule level
Measured: December 2024
Secondary outcomes (1)
Duration of analytical analysis performance
To define the time to obtain the test result (time-to-result) for a complete analysis.
Measured: December 2024
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted25 Jul 2024
- Study start16 Jul 2024
- Primary completion30 Nov 2024
- Study completion31 May 2025 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Non-interventional Analytical Performance Study on Left-over SamplesStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Other
- Time perspective
- Prospective
- Enrolment
- 50 participants (estimated)
Sponsor and investigators
- Lead sponsor
- Centro di Innovazione Regionale in Single Molecule Digital Assay
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT06518967
- Sponsor's protocol ID
- SiMoT-01
- Status last verified
- Feb 2025
- First submitted
- 15 Jul 2024
- Last updated
- 5 Feb 2025
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- No
Study at a glance
From the registry record, last updated by the sponsor on 5 Feb 2025. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Analytical Performance Study of the Handheld Device Single-Molecule with a Large Transistor (SiMoT) for the Near-patient Testing of Biological Samples
- ClinicalTrials.gov ID
- NCT06518967
- Acronym / sponsor ID
- SiMoT-01
- Study type
- Observational
- Recruitment status
- Active, not recruiting
- Enrolment
- 50 participants (estimated)
- Lead sponsor
- Centro di Innovazione Regionale in Single Molecule Digital AssayUniversity, hospital or other
- Study start
- 16 Jul 2024 (actual)
- Primary completion
- 30 Nov 2024 (actual)
- Study completion
- 31 May 2025 (estimated)
- First posted
- 25 Jul 2024 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Advanced and performing technologies able to analyze specific biomarkers for the diagnosis, prognosis and prediction as well as for the monitoring, the personalized care and the evaluation of the therapeutic response to pathologies, is currently considered extremely relevant. Such technologies work only if molecular markers specific for a given pathology have been already validated. This approach is considered of strategic importance for the health and well-being of citizens as well as for the sustainability of public spending. Often, however, the analysis of the markers is carried out on tissues or fluids of the organs involved, being hence quite invasive procedures. In this scenario, it is strategic to develop devices and platforms that allow the analysis of markers in a minimally invasive or non-invasive way, through the analysis of peripheral biological fluids, such as blood or urine. The goal is to provide clinicians with the attacker's advantage against a plethora of progressive diseases by mass-screening individuals who are affected by a disease long before symptoms appear or to allow the identification of those patients who have a high probability of recurring or responding to target-oriented pharmacological approaches. This study pertains to the definition of the analytical performance of the Single-Sensor Single-Molecule with a large Transistor (SiMoT) device. In the context of the present performance study, the IVD SiMoT system will be equipped with an accessory-cartridge developed and produced without any diagnostic purpose but only to verify the analytical performance of the SiMoT system, in the detection of exogenous biochemical molecules suitably selected to act as markers of interest.
Detailed description
Advanced and performing technologies able to analyze specific biomarkers for the diagnosis, prognosis and prediction as well as for the monitoring, the personalized care and the evaluation of the therapeutic response to pathologies, is currently considered extremely relevant. Such technologies work only if molecular markers specific for a given pathology have been already validated. This approach is considered of strategic importance for the health and well-being of citizens as well as for the sustainability of public spending. Among the pathologies that will increasingly be monitored and treated thanks to enabling technologies for precision medicine, tumors are certainly of particular importance. Often, however, the analysis of the markers is carried out on tissues or fluids of the organs involved, being hence quite invasive procedures. In this scenario, it is strategic to develop devices and platforms that allow the analysis of markers in a minimally invasive or non-invasive way, through the analysis of peripheral biological fluids, such as blood or urine. The goal is to provide clinicians with the attacker's advantage against a plethora of progressive diseases by mass-screening individuals who are affected by a disease long before symptoms appear or to allow the identification of those patients who have a high probability of recurring or responding to target-oriented pharmacological approaches.
This study pertains to the definition of the analytical performance of the Single-Sensor Single-Molecule with a large Transistor (SiMoT) device. The SiMoT device is an in vitro diagnostic - IVD system (REGULATION (EU) 2017/746, IVDR Art. 2, point 2) that performs decentralized analyses (near-patient testing, such as, for example, patient's home, emergency unit, ambulance) composed of a palm reader completed by machine-learning / artificial-intelligence software. The system works through an accessory (IVDR Art. 2, point 4) represented by a disposable cartridge, which determines the specific diagnostic application based on the molecular markers detected by the accessory. In the context of the present performance study, the IVD SiMoT system will be equipped with an accessory-cartridge developed and produced without any diagnostic purpose but only to verify the analytical performance of the SiMoT system, in the detection of exogenous biochemical molecules suitably selected to act as markers of interest.
The present performance study aims to verify the qualitative detection (on/off) of target molecules (both proteins and nucleic acids) with a single molecule threshold in 0.1 mL (10^-20 M). The threshold is established by the limit of identification (LOI).
Analytical performance will be evaluated according to standard statistical approaches such as: analytical sensitivity, analytical specificity, trueness (bias), precision (repeatability and reproducibility), accuracy (resulting from trueness and precision), limits of detection and identification, measurement range, possible linearity, threshold values, including the determination of appropriate criteria for the collection and handling of samples and the control of known endogenous and exogenous interferences, cross reactions.
Risk evaluation (Risk-to-Benefits rationale) There are no identifiable physical, psychic and/or social risks for the patients enrolled in the study directly deriving from the procedures that will be followed for the enrollment and collection of the related data.
Study design
- Observational model
- Other
- Time perspective
- Prospective
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Open | – | Diagnostic Test: analytical performance studyTest analysis perfomed in open to train the artificial intelligence algorithm. |
| Blind | – | Diagnostic Test: analytical performance studyTest analysis performed in blind to minimize biases when evaluating the predictive performance of the IVD under investigation. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
- Study population
- The study involves around 50 patients with solid tumors who will come to the Gynecological Oncologic Clinic at IRCCS Giovanni Paolo II Cancer Institute and who are prescribed for pre-hospitalization procedures as per standard clinical practice at the investigational site.
Key inclusion criteria (3 listed)
- Informed Consent and personal data handling consent signed by the subject;
- > 18 years of age
- presence of neoplastic or suspected neoplastic pathology at the time of sample collection
Key exclusion criteria (2 listed)
- < 18 years of age
- Participation in a clinical trial with an investigational drug or a medical device within the 30 days preceding and during the present investigation.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| analytical performance studyanalytical performance study on left-over samples with the SiMoT IVD system for in vitro diagnostic tests | Diagnostic test | Blind, Open |
Primary outcome
To evaluate the analytical performance in plasma, serum, and urine samples supplemented with non-endogenous molecules (three proteins and two nucleotide sequences) of the SiMoT handheld device
Time frame: December 2024
Qualitative detection of specific marker at single molecule level
1 secondary outcome
Duration of analytical analysis performance
Time frame: December 2024
Results and safety
This study is still under way. Results are not available in the source record yet.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 5 Feb 2025.
- Individual participant dataSharing planned: No
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Non-interventional Analytical Performance Study on Left-over Samples, run by Centro di Innovazione Regionale in Single Molecule Digital Assay at 1 site in Italy.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: under way, not looking for participants, as of the sponsor's last update on 5 Feb 2025. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Italy. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
This study is still under way. Results are not available in the source record yet.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
To evaluate the analytical performance in plasma, serum, and urine samples supplemented with non-endogenous molecules (three proteins and two nucleotide sequences) of the SiMoT handheld device (time frame: December 2024). It is what the study was designed to measure, not a finding.