Atrial Fibrillation (AF)SVtagNCT06721221
Impact of Catheter Stability on the Outcomes With Very High Power Short Duration Ablation
A study testing Point-by-point PVI with the QDot Micro catheter in QMODE+ setting with the Svtag software in people with Atrial Fibrillation (AF) and Atrial Fibrillation Ablation, run by Semmelweis University at 1 site in Hungary.
Trial summary
Background: Atrial fibrillation (AF), the most common cardiac rhythm disorder can be treated with pulmonary vein isolation (PVI). One technique for PVI is point-by-point radiofrequency ablation. Very high power short duration ablation is one of the most advanced technologies for radiofrequency ablation. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Atrial Fibrillation (AF), Atrial Fibrillation Ablation, Pulmonary Vein Isolation
- What is being tested?
- Point-by-point PVI with the QDot Micro catheter in QMODE+ setting with the Svtag software
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 100 participants
- Where
- 1 siteHungary
- Study dates
- 1 Aug 2024to 31 Dec 2026
- Phase
- Interventional
- Are there results?
- Not yet: the study is under way
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Symptomatic paroxysmal or persistent AF
- Age >18 years
- Willingness to sign the informed consent form
Who cannot join
- Contraindication to ablation
- Contraindication of long-term anticoagulation
- Long-standing persistent AF
- History of PVI
- History of cardiac surgery
- Pregnancy
Show all 10Show fewer
- Active malignancy
- Life expectancy \<1 year
- Valvular AF
- Reversible cause of AF (e.g. hyperthyroidism).
Trial locations and contacts
1 site in 1 country.
About this study
Background: Atrial fibrillation (AF), the most common cardiac rhythm disorder can be treated with pulmonary vein isolation (PVI). One technique for PVI is point-by-point radiofrequency ablation. Very high power short duration ablation is one of the most advanced technologies for radiofrequency ablation. However, acute efficacy results with this technology vary in a wide range. Improvements in automated tagging modules, incorporating tracking of cardiac and respiratory motion and enhanced stability algorithms, may allow for a better assessment of lesion quality and location and may improve the so called first-pass isolation rate (an indicator for acute procedural efficacy).
Objective: To assess the acute procedural outcomes of very high power short duration PVI with the new enhanced stability software.
Read more
Methods: Investigator-initiated, prospective, single-arm study on one hundred symptomatic patients with paroxysmal AF will undergo PVI with the QDOT catheter using a power setting of 90W(QMODE+). The inter-tag distance will be 6 mm posteriorly and 4 mm anteriorly, and the enhanced stability algorithm will be used in all cases. After creating the isolation circle, the presence or absence of first-pass isolation will be assessed on each side by the presence of entrance block. Primary endpoint will be first-pass isolation rate. Secondary outcomes are as follows: procedure time, left atrial dwell time, RF time, number of RF tags, use of a steerable sheath, occurrence of serious adverse events.
Hypothesis: Very high power short duration PVI using the new stability software results in a higher rate of first-pass isolation than previously published.
Read the detailed description
Background Atrial fibrillation (AF) is the most common sustained cardiac arrhythmia in adults. Its prevalence increases with age, affecting 3/100 people, and its incidence is also correlated with age, affecting 30/100.000 people per year. Its clinical significance stems from common complications (stroke, heart failure) and concomitant higher mortality. The most effective rhythm control treatment method for atrial fibrillation is catheter ablation, which seeks to maintain sinus rhythm. Isolation of the triggering pulmonary veins (PVI) from the left atrium is the cornerstone of AF ablation. PVI has undergone significant technical development over the past decade; however, the chances of arrhythmia recurrence and the occurrence of procedural complications are not negligible. Experiments have shown that lesions with a slightly smaller surface diameter and less depth develop during high power short duration (HPSD) ablation. The shorter ablation time might result in more consistent lesion creation if the catheter stability is good during the radiofrequency (RF) application. On the other hand, if the catheter position is unstable during the RF application, a huge variety might occur between the created lesion sizes. Regarding the safety aspects, the shallower lesions might cause less damage to the surrounding tissues. The latest application of HPSD technology is the so-called "very high power short duration" (vHPSD) ablation with application of 90W power for 3-4 seconds. The ablation index does not work for such a short ablation time; thus, other lesion-predicting parameters should be used, such as impedance drop and catheter stability. At the moment, there are two software products for ablation point visualization. The commercially available legacy software was originally built to assess catheter stability over a long ablation period (e.g. with the ablation index), but the system is unable to differentiate between poor and good catheter stability during a 4 seconds time frame. Conversely, the new SVTAG software is adapted to assess catheter stability over a 4-second period; thus, the operator might get better feedback on the actual lesion creation. This might result in a more optimized lesion set and a more effective PVI, but this hypothesis has not been tested yet.
Objective: To evaluate the acute efficacy of radiofrequency PVI performed with 90W, using the new stability software version.
Primary hypothesis: Pulmonary vein isolation is effective with the Svtag software that gives proper feedback on catheter stability, resulting in short procedures, high first-pass isolation rate, and few acute reconnections.
Study Design: Multicenter, prospective, single-arm study.
Research period: The planned research period starts in the first quarter of 2023 until December 2025.
Study subjects: Planned number of included subjects: 150. Inclusion criteria: Symptomatic paroxysmal or persistent AF, Age >18 years, Willingness to sign the informed consent form.
Exclusion criteria: Contraindication to ablation, Contraindication of long-term anticoagulation, Long-standing persistent AF, History of PVI, History of cardiac surgery, Pregnancy, Active malignancy, Life expectancy <1 year, Valvular AF, Reversible cause of AF (e.g. hyperthyroidism).
Study procedure: PVI will be performed via femoral access after transseptal puncture, guided by fluoroscopy pressure monitoring, intracardiac echocardiography (ICE) or transesophageal echocardiography (TEE). A fast anatomical left atrial map will be created; then, point-by- point PVI will be performed with QDot Micro catheter in QMODE plus with the Svtag software. Inter-tag distance should be ≤ 6 mm on the posterior wall and ≤ 4 mm on the anterior wall. All applications' duration should be 4 seconds. After creating the isolation circle, the presence or absence of first-pass isolation will be assessed on each side by the presence of entrance block. If PVI is complete, this is defined as first-pass isolation. If PVI is not complete at this point, touch-up lesions will be delivered to reach the isolation of all pulmonary veins. After that, acute PV reconnections will be evaluated during a 20 minutes waiting period. In case of acute reconnection, additional applications will be delivered to complete the PVI. No further waiting is necessary after that.
Primary endpoint: First-pass isolation rate.Secondary endpoints: Occurrence of any adverse events related to the procedure during the hospital stay and 1 month after the procedure. Acute reconnection: pulmonary vein reconnection during the 20 minutes waiting period. Impact of scar burden in the on the FPI rate assessed by EA mapping (during the waiting period). 12-month AF-free survival
Nature of the study: Investigator-initiated clinical trial, there is no commissioning company. The study is not for commercial gain. Not a drug test. Subjects will not undergo any additional procedures in connection to the study. The study protocol is in compliance with routine clinical practice, no additional radiation is applied.
Adverse events: PVI is a routinely used, safe procedure these days. During the current clinical trial, for the participating patients, there is no additional intervention that is different from the everyday routine, except for the 20-minute waiting period after the PVI. During the procedure, patients are exposed to the following potential complications, which in total occur in 1% of cases: Complications involving the vascular access site: femoral pseudoaneurysm / arterio- venous fistula / hematoma with bleeding requiring blood transfusion, Pericardial fluid, Pericardial tamponade, Stroke / TIA, Esophageal injury / perforation / fistula, Myocardial infarction, Phrenic nerve paresis, Pulmonary vein stenosis, Conduction system damage requiring pacemaker implantation, Death. The participating centers' Electrophysiology Laboratories are prepared and capable of providing fast and adequate care for these complications.
Statistical analysis: Continuous variables are expressed as means and standard deviations, or medians and interquartile ranges, depending on the normal or non-normal distribution. Categorical variables are expressed in numbers and percentages. Continuous variables are compared by parametric or non-parametric tests, depending on the distribution, while categorical variables are compared by Chi-square test or Fisher's exact test. A two-tailed p-value of <0.05 is considered statistically significant. Statistical analyses will be performed using IBM SPSS 25 (Apache Software Foundation, USA) and GraphPad Prism 7.1 (GraphPad Softwares Inc., USA) software, or other statistical software products.
Data management: All study patient data is entered into an electronic database (REDCap) in a coded and unique manner, with a unique identifier, to which project staff has a password-protected, defined level of access. Each person involved in the study is given a unique identification code, and the data stored in the database is linked to that unique identifier. Thus, a data set will be incomprehensible and unusable for an external (unauthorized) user. The data belonging to the unique identification code, with which the patient's identity can be clearly indicated (name, place and date of birth, clinical reference number, identification number, identity card number, etc.) are not available from the database and are stored separately from it. Access to personal data will be limited to authorized staff (both clinical and research) at the local hospital site.
Funding: A research grant is provided by the Johnson and Johnson Company.
Ethics approval: An ethics approval application is submitted separately in case of all participating centers to their national or local ethics committee.
- Being tested
- Point-by-point PVI with the QDot Micro catheter in QMODE+ setting with the Svtag software
- Keywords
- pulmonary vein isolation, catheter stability, first-pass isolation
- Medical subject headings
- Atrial Fibrillation
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Point-by-point PVI with the QDot Micro catheter in QMODE+ setting with the Svtag softwareDevice
PVI will be performed via femoral access after transseptal puncture, guided by fluoroscopy pressure monitoring, intracardiac echocardiography (ICE) or transesophageal echocardiography (TEE). A fast anatomical left atrial map will be created; then, point-by-point PVI will be performed with QDot Micro catheter in QMODE plus with the Svtag software. Inter-tag distance should be ≤ 6 mm on the posterior wall and ≤ 4 mm on the anterior wall. All applications' duration should be 4 seconds. After creating the isolation circle, the presence or absence of first-pass isolation will be assessed on each side by the presence of entrance block. If PVI is complete, this is defined as first-pass isolation. If PVI is not complete at this point, touch-up lesions will be delivered to reach the isolation of all pulmonary veins. After that, acute PV reconnections will be evaluated during a 20 minutes waiting period.
| Study group | Role | What they receive |
|---|---|---|
| SVtag arm. Point-by-point PVI with the Svtag software | Experimental | PVI will be performed with QDot Micro catheter in QMODE plus with the Svtag software. Receives: Device: Point-by-point PVI with the QDot Micro catheter in QMODE+ setting with the Svtag software |
What is being measured
The primary outcome is the main question the study is designed to answer.
First-pass isolation rate
After creating the isolation circle, the presence or absence of first-pass isolation will be assessed on each side by the presence of entrance block. If PVI is complete, this is defined as first-pass isolation. If PVI is not complete at this point, touch-up lesions will be delivered to reach the isolation of all pulmonary veins.
Measured: Endpoint is assessed intra-procedurally
Secondary outcomes (3)
Acute reconnection
Pulmonary vein reconnection during the 20 minutes waiting period
Measured: Assessed 20 minutes after PVI is reached during the procedure
12-months AF freedom
Freedom from AF during 12-month follow-up
Measured: 12-month from procedure
Complications or adverse events
Occurrence of any adverse events related to the procedure during the hospital stay and 1 month after the procedure
Measured: 1 month
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted6 Dec 2024
- Study start1 Aug 2024
- Primary completion31 Dec 2026 (estimated)
- Study completion31 Dec 2026 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (9)Papers the sponsor has linked to this study.
- Other trials for Atrial Fibrillation (AF)Studies of the same condition, open and under way.
- More trials run by Semmelweis UniversityThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 100 participants (estimated)
Sponsor and investigators
- Lead sponsor
- Semmelweis University
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT06721221
- Sponsor's protocol ID
- BM/8942-1/2023
- Other IDs
- BWI-IIS-688
- Acronym
- SVtag
- Status last verified
- Dec 2024
- First submitted
- 3 Dec 2024
- Last updated
- 11 Dec 2024
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- No: All study patient data is entered into an electronic database (REDCap) in a coded and unique manner, with a unique identifier, to which project staff has a password-protected, defined level of access. Each person involved in the study is given a unique identification code, and the data stored in the database is linked to that unique identifier. Thus, a data set will be incomprehensible and unusable for an external (unauthorized) user. The data belonging to the unique identification code, with which the patient's identity can be clearly indicated (name, place and date of birth, clinical reference number, identification number, identity card number, etc.) are not available from the database and are stored separately from it. Access to personal data will be limited to authorized staff (both clinical and research) at the local hospital site.
Publications
Reddy VY, Grimaldi M, De Potter T, Vijgen JM, Bulava A, Duytschaever MF, Martinek M, Natale A, Knecht S, Neuzil P, Purerfellner H. Pulmonary Vein Isolation With Very High Power, Short Duration, Temperature-Controlled Lesions: The QDOT-FAST Trial. JACC Clin Electrophysiol. 2019 Jul;5(7):778-786. doi: 10.1016/j.jacep.2019.04.009. Epub 2019 May 8.
PubMed 31320006Kewcharoen J, Techorueangwiwat C, Kanitsoraphan C, Leesutipornchai T, Akoum N, Bunch TJ, Navaravong L. High-power short duration and low-power long duration in atrial fibrillation ablation: A meta-analysis. J Cardiovasc Electrophysiol. 2021 Jan;32(1):71-82. doi: 10.1111/jce.14806. Epub 2020 Nov 18.
PubMed 33155303Nakagawa H, Ikeda A, Sharma T, Govari A, Ashton J, Maffre J, Lifshitz A, Fuimaono K, Yokoyama K, Wittkampf FHM, Jackman WM. Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With High Power-Short Duration and Moderate Power-Moderate Duration: Effects of Thermal Latency and Contact Force on Lesion Formation. Circ Arrhythm Electrophysiol. 2021 Jul;14(7):e009899. doi: 10.1161/CIRCEP.121.009899. Epub 2021 Jun 17.
PubMed 34138641Hussein A, Das M, Riva S, Morgan M, Ronayne C, Sahni A, Shaw M, Todd D, Hall M, Modi S, Natale A, Dello Russo A, Snowdon R, Gupta D. Use of Ablation Index-Guided Ablation Results in High Rates of Durable Pulmonary Vein Isolation and Freedom From Arrhythmia in Persistent Atrial Fibrillation Patients: The PRAISE Study Results. Circ Arrhythm Electrophysiol. 2018 Sep;11(9):e006576. doi: 10.1161/CIRCEP.118.006576.
PubMed 30354288Taghji P, El Haddad M, Phlips T, Wolf M, Knecht S, Vandekerckhove Y, Tavernier R, Nakagawa H, Duytschaever M. Evaluation of a Strategy Aiming to Enclose the Pulmonary Veins With Contiguous and Optimized Radiofrequency Lesions in Paroxysmal Atrial Fibrillation: A Pilot Study. JACC Clin Electrophysiol. 2018 Jan;4(1):99-108. doi: 10.1016/j.jacep.2017.06.023. Epub 2017 Sep 27.
PubMed 29600792Solimene F, Schillaci V, Shopova G, Urraro F, Arestia A, Iuliano A, Maresca F, Agresta A, La Rocca V, De Simone A, Stabile G. Safety and efficacy of atrial fibrillation ablation guided by Ablation Index module. J Interv Card Electrophysiol. 2019 Jan;54(1):9-15. doi: 10.1007/s10840-018-0420-5. Epub 2018 Jul 30.
PubMed 30058055Szegedi N, Sallo Z, Perge P, Piros K, Nagy VK, Osztheimer I, Merkely B, Geller L. The role of local impedance drop in the acute lesion efficacy during pulmonary vein isolation performed with a new contact force sensing catheter-A pilot study. PLoS One. 2021 Sep 16;16(9):e0257050. doi: 10.1371/journal.pone.0257050. eCollection 2021.
PubMed 34529678Haissaguerre M, Jais P, Shah DC, Takahashi A, Hocini M, Quiniou G, Garrigue S, Le Mouroux A, Le Metayer P, Clementy J. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med. 1998 Sep 3;339(10):659-66. doi: 10.1056/NEJM199809033391003.
PubMed 9725923Hindricks G, Potpara T, Dagres N, Arbelo E, Bax JJ, Blomstrom-Lundqvist C, Boriani G, Castella M, Dan GA, Dilaveris PE, Fauchier L, Filippatos G, Kalman JM, La Meir M, Lane DA, Lebeau JP, Lettino M, Lip GYH, Pinto FJ, Thomas GN, Valgimigli M, Van Gelder IC, Van Putte BP, Watkins CL; ESC Scientific Document Group. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J. 2021 Feb 1;42(5):373-498. doi: 10.1093/eurheartj/ehaa612. No abstract available.
PubMed 32860505
Study at a glance
From the registry record, last updated by the sponsor on 11 Dec 2024. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Impact of Catheter Stability on the Outcomes With Very High Power Short Duration Ablation
- ClinicalTrials.gov ID
- NCT06721221
- Acronym / sponsor ID
- SVtag / BM/8942-1/2023
- Keywords
- pulmonary vein isolation, catheter stability, first-pass isolation
- Study type
- Interventional
- Recruitment status
- Active, not recruiting
- Enrolment
- 100 participants (estimated)
- Lead sponsor
- Semmelweis UniversityUniversity, hospital or other
- Study start
- 1 Aug 2024 (actual)
- Primary completion
- 31 Dec 2026 (estimated)
- Study completion
- 31 Dec 2026 (estimated)
- First posted
- 6 Dec 2024 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Background: Atrial fibrillation (AF), the most common cardiac rhythm disorder can be treated with pulmonary vein isolation (PVI). One technique for PVI is point-by-point radiofrequency ablation. Very high power short duration ablation is one of the most advanced technologies for radiofrequency ablation. However, acute efficacy results with this technology vary in a wide range. Improvements in automated tagging modules, incorporating tracking of cardiac and respiratory motion and enhanced stability algorithms, may allow for a better assessment of lesion quality and location and may improve the so called first-pass isolation rate (an indicator for acute procedural efficacy).
Objective: To assess the acute procedural outcomes of very high power short duration PVI with the new enhanced stability software.
Methods: Investigator-initiated, prospective, single-arm study on one hundred symptomatic patients with paroxysmal AF will undergo PVI with the QDOT catheter using a power setting of 90W(QMODE+). The inter-tag distance will be 6 mm posteriorly and 4 mm anteriorly, and the enhanced stability algorithm will be used in all cases. After creating the isolation circle, the presence or absence of first-pass isolation will be assessed on each side by the presence of entrance block. Primary endpoint will be first-pass isolation rate. Secondary outcomes are as follows: procedure time, left atrial dwell time, RF time, number of RF tags, use of a steerable sheath, occurrence of serious adverse events.
Hypothesis: Very high power short duration PVI using the new stability software results in a higher rate of first-pass isolation than previously published.
Detailed description
Background Atrial fibrillation (AF) is the most common sustained cardiac arrhythmia in adults. Its prevalence increases with age, affecting 3/100 people, and its incidence is also correlated with age, affecting 30/100.000 people per year. Its clinical significance stems from common complications (stroke, heart failure) and concomitant higher mortality. The most effective rhythm control treatment method for atrial fibrillation is catheter ablation, which seeks to maintain sinus rhythm. Isolation of the triggering pulmonary veins (PVI) from the left atrium is the cornerstone of AF ablation. PVI has undergone significant technical development over the past decade; however, the chances of arrhythmia recurrence and the occurrence of procedural complications are not negligible. Experiments have shown that lesions with a slightly smaller surface diameter and less depth develop during high power short duration (HPSD) ablation. The shorter ablation time might result in more consistent lesion creation if the catheter stability is good during the radiofrequency (RF) application. On the other hand, if the catheter position is unstable during the RF application, a huge variety might occur between the created lesion sizes. Regarding the safety aspects, the shallower lesions might cause less damage to the surrounding tissues. The latest application of HPSD technology is the so-called "very high power short duration" (vHPSD) ablation with application of 90W power for 3-4 seconds. The ablation index does not work for such a short ablation time; thus, other lesion-predicting parameters should be used, such as impedance drop and catheter stability. At the moment, there are two software products for ablation point visualization. The commercially available legacy software was originally built to assess catheter stability over a long ablation period (e.g. with the ablation index), but the system is unable to differentiate between poor and good catheter stability during a 4 seconds time frame. Conversely, the new SVTAG software is adapted to assess catheter stability over a 4-second period; thus, the operator might get better feedback on the actual lesion creation. This might result in a more optimized lesion set and a more effective PVI, but this hypothesis has not been tested yet.
Objective: To evaluate the acute efficacy of radiofrequency PVI performed with 90W, using the new stability software version.
Primary hypothesis: Pulmonary vein isolation is effective with the Svtag software that gives proper feedback on catheter stability, resulting in short procedures, high first-pass isolation rate, and few acute reconnections.
Study Design: Multicenter, prospective, single-arm study.
Research period: The planned research period starts in the first quarter of 2023 until December 2025.
Study subjects: Planned number of included subjects: 150. Inclusion criteria: Symptomatic paroxysmal or persistent AF, Age >18 years, Willingness to sign the informed consent form.
Exclusion criteria: Contraindication to ablation, Contraindication of long-term anticoagulation, Long-standing persistent AF, History of PVI, History of cardiac surgery, Pregnancy, Active malignancy, Life expectancy <1 year, Valvular AF, Reversible cause of AF (e.g. hyperthyroidism).
Study procedure: PVI will be performed via femoral access after transseptal puncture, guided by fluoroscopy pressure monitoring, intracardiac echocardiography (ICE) or transesophageal echocardiography (TEE). A fast anatomical left atrial map will be created; then, point-by- point PVI will be performed with QDot Micro catheter in QMODE plus with the Svtag software. Inter-tag distance should be ≤ 6 mm on the posterior wall and ≤ 4 mm on the anterior wall. All applications' duration should be 4 seconds. After creating the isolation circle, the presence or absence of first-pass isolation will be assessed on each side by the presence of entrance block. If PVI is complete, this is defined as first-pass isolation. If PVI is not complete at this point, touch-up lesions will be delivered to reach the isolation of all pulmonary veins. After that, acute PV reconnections will be evaluated during a 20 minutes waiting period. In case of acute reconnection, additional applications will be delivered to complete the PVI. No further waiting is necessary after that.
Primary endpoint: First-pass isolation rate.Secondary endpoints: Occurrence of any adverse events related to the procedure during the hospital stay and 1 month after the procedure. Acute reconnection: pulmonary vein reconnection during the 20 minutes waiting period. Impact of scar burden in the on the FPI rate assessed by EA mapping (during the waiting period). 12-month AF-free survival
Nature of the study: Investigator-initiated clinical trial, there is no commissioning company. The study is not for commercial gain. Not a drug test. Subjects will not undergo any additional procedures in connection to the study. The study protocol is in compliance with routine clinical practice, no additional radiation is applied.
Adverse events: PVI is a routinely used, safe procedure these days. During the current clinical trial, for the participating patients, there is no additional intervention that is different from the everyday routine, except for the 20-minute waiting period after the PVI. During the procedure, patients are exposed to the following potential complications, which in total occur in 1% of cases: Complications involving the vascular access site: femoral pseudoaneurysm / arterio- venous fistula / hematoma with bleeding requiring blood transfusion, Pericardial fluid, Pericardial tamponade, Stroke / TIA, Esophageal injury / perforation / fistula, Myocardial infarction, Phrenic nerve paresis, Pulmonary vein stenosis, Conduction system damage requiring pacemaker implantation, Death. The participating centers' Electrophysiology Laboratories are prepared and capable of providing fast and adequate care for these complications.
Statistical analysis: Continuous variables are expressed as means and standard deviations, or medians and interquartile ranges, depending on the normal or non-normal distribution. Categorical variables are expressed in numbers and percentages. Continuous variables are compared by parametric or non-parametric tests, depending on the distribution, while categorical variables are compared by Chi-square test or Fisher's exact test. A two-tailed p-value of <0.05 is considered statistically significant. Statistical analyses will be performed using IBM SPSS 25 (Apache Software Foundation, USA) and GraphPad Prism 7.1 (GraphPad Softwares Inc., USA) software, or other statistical software products.
Data management: All study patient data is entered into an electronic database (REDCap) in a coded and unique manner, with a unique identifier, to which project staff has a password-protected, defined level of access. Each person involved in the study is given a unique identification code, and the data stored in the database is linked to that unique identifier. Thus, a data set will be incomprehensible and unusable for an external (unauthorized) user. The data belonging to the unique identification code, with which the patient's identity can be clearly indicated (name, place and date of birth, clinical reference number, identification number, identity card number, etc.) are not available from the database and are stored separately from it. Access to personal data will be limited to authorized staff (both clinical and research) at the local hospital site.
Funding: A research grant is provided by the Johnson and Johnson Company.
Ethics approval: An ethics approval application is submitted separately in case of all participating centers to their national or local ethics committee.
Study design
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| SVtag arm. Point-by-point PVI with the Svtag software | Experimental | Device: Point-by-point PVI with the QDot Micro catheter in QMODE+ setting with the Svtag softwarePVI will be performed with QDot Micro catheter in QMODE plus with the Svtag software. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (3 listed)
- Symptomatic paroxysmal or persistent AF
- Age >18 years
- Willingness to sign the informed consent form
Key exclusion criteria (10 listed)
- Contraindication to ablation
- Contraindication of long-term anticoagulation
- Long-standing persistent AF
- History of PVI
- History of cardiac surgery
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Point-by-point PVI with the QDot Micro catheter in QMODE+ setting with the Svtag softwarePVI will be performed via femoral access after transseptal puncture, guided by fluoroscopy pressure monitoring, intracardiac echocardiography (ICE) or transesophageal echocardiography (TEE). A fast anatomical left atrial map will be created; then, point-by-point PVI will be performed with QDot Micro catheter in QMODE plus with the Svtag software. Inter-tag distance should be ≤ 6 mm on the posterior wall and ≤ 4 mm on the anterior wall. All applications' duration should be 4 seconds. After creating the isolation circle, the presence or absence of first-pass isolation will be assessed on each side by the presence of entrance block. If PVI is complete, this is defined as first-pass isolation. If PVI is not complete at this point, touch-up lesions will be delivered to reach the isolation of all pulmonary veins. After that, acute PV reconnections will be evaluated during a 20 minutes waiting period. | Device | SVtag arm. Point-by-point PVI with the Svtag software |
Primary outcome
First-pass isolation rate
Time frame: Endpoint is assessed intra-procedurally
After creating the isolation circle, the presence or absence of first-pass isolation will be assessed on each side by the presence of entrance block. If PVI is complete, this is defined as first-pass isolation. If PVI is not complete at this point, touch-up lesions will be delivered to reach the isolation of all pulmonary veins.
3 secondary outcomes
Acute reconnection
Time frame: Assessed 20 minutes after PVI is reached during the procedure
12-months AF freedom
Time frame: 12-month from procedure
Complications or adverse events
Time frame: 1 month
Results and safety
This study is still under way. Results are not available in the source record yet.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 11 Dec 2024.
- Background: PubMed 31320006Reddy VY, Grimaldi M, De Potter T, Vijgen JM, Bulava A, Duytschaever MF, Martinek M, Natale A, Knecht S, Neuzil P, Purerfellner H. Pulmonary Vein Isolation With Very High Power, Short Duration, Temperature-Controlled Lesions: The QDOT-FAST Trial. JACC Clin Electrophysiol. 2019 Jul;5(7):778-786. doi: 10.1016/j.jacep.2019.04.009. Epub 2019 May 8.
- Background: PubMed 33155303Kewcharoen J, Techorueangwiwat C, Kanitsoraphan C, Leesutipornchai T, Akoum N, Bunch TJ, Navaravong L. High-power short duration and low-power long duration in atrial fibrillation ablation: A meta-analysis. J Cardiovasc Electrophysiol. 2021 Jan;32(1):71-82. doi: 10.1111/jce.14806. Epub 2020 Nov 18.
- Background: PubMed 34138641Nakagawa H, Ikeda A, Sharma T, Govari A, Ashton J, Maffre J, Lifshitz A, Fuimaono K, Yokoyama K, Wittkampf FHM, Jackman WM. Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With High Power-Short Duration and Moderate Power-Moderate Duration: Effects of Thermal Latency and Contact Force on Lesion Formation. Circ Arrhythm Electrophysiol. 2021 Jul;14(7):e009899. doi: 10.1161/CIRCEP.121.009899. Epub 2021 Jun 17.
- Background: PubMed 30354288Hussein A, Das M, Riva S, Morgan M, Ronayne C, Sahni A, Shaw M, Todd D, Hall M, Modi S, Natale A, Dello Russo A, Snowdon R, Gupta D. Use of Ablation Index-Guided Ablation Results in High Rates of Durable Pulmonary Vein Isolation and Freedom From Arrhythmia in Persistent Atrial Fibrillation Patients: The PRAISE Study Results. Circ Arrhythm Electrophysiol. 2018 Sep;11(9):e006576. doi: 10.1161/CIRCEP.118.006576.
- Background: PubMed 29600792Taghji P, El Haddad M, Phlips T, Wolf M, Knecht S, Vandekerckhove Y, Tavernier R, Nakagawa H, Duytschaever M. Evaluation of a Strategy Aiming to Enclose the Pulmonary Veins With Contiguous and Optimized Radiofrequency Lesions in Paroxysmal Atrial Fibrillation: A Pilot Study. JACC Clin Electrophysiol. 2018 Jan;4(1):99-108. doi: 10.1016/j.jacep.2017.06.023. Epub 2017 Sep 27.
- Background: PubMed 30058055Solimene F, Schillaci V, Shopova G, Urraro F, Arestia A, Iuliano A, Maresca F, Agresta A, La Rocca V, De Simone A, Stabile G. Safety and efficacy of atrial fibrillation ablation guided by Ablation Index module. J Interv Card Electrophysiol. 2019 Jan;54(1):9-15. doi: 10.1007/s10840-018-0420-5. Epub 2018 Jul 30.
- Background: PubMed 34529678Szegedi N, Sallo Z, Perge P, Piros K, Nagy VK, Osztheimer I, Merkely B, Geller L. The role of local impedance drop in the acute lesion efficacy during pulmonary vein isolation performed with a new contact force sensing catheter-A pilot study. PLoS One. 2021 Sep 16;16(9):e0257050. doi: 10.1371/journal.pone.0257050. eCollection 2021.
- Background: PubMed 9725923Haissaguerre M, Jais P, Shah DC, Takahashi A, Hocini M, Quiniou G, Garrigue S, Le Mouroux A, Le Metayer P, Clementy J. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med. 1998 Sep 3;339(10):659-66. doi: 10.1056/NEJM199809033391003.
- Background: PubMed 32860505Hindricks G, Potpara T, Dagres N, Arbelo E, Bax JJ, Blomstrom-Lundqvist C, Boriani G, Castella M, Dan GA, Dilaveris PE, Fauchier L, Filippatos G, Kalman JM, La Meir M, Lane DA, Lebeau JP, Lettino M, Lip GYH, Pinto FJ, Thomas GN, Valgimigli M, Van Gelder IC, Van Putte BP, Watkins CL; ESC Scientific Document Group. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J. 2021 Feb 1;42(5):373-498. doi: 10.1093/eurheartj/ehaa612. No abstract available.
- Individual participant dataSharing planned: No. All study patient data is entered into an electronic database (REDCap) in a coded and unique manner, with a unique identifier, to which project staff has a password-protected, defined level of access. Each person involved in the study is given a unique identification code, and the data stored in the database is linked to that unique identifier. Thus, a data set will be incomprehensible and unusable for an external (unauthorized) user. The data belonging to the unique identification code, with which the patient's identity can be clearly indicated (name, place and date of birth, clinical reference number, identification number, identity card number, etc.) are not available from the database and are stored separately from it. Access to personal data will be limited to authorized staff (both clinical and research) at the local hospital site.
Related trials for Atrial Fibrillation (AF)
Active, not recruiting
Horizon 360 Protocol for the Treatment of Paroxysmal Atrial Fibrillation With the Sphere-360™ Catheter and Affera™ Mapping and Ablation System (Horizon 360)
Paroxysmal AF
Active, not recruiting
CONVERGE Post-Approval Study (PAS)
Chronic Atrial Fibrillation
RecruitingPhase 2
Safety in Adult Participants With Atrial Fibrillation Who Are Treated With Anticoagulation
Atrial Fibrillation (AF)
RecruitingPhase 3
Stroke and Systemic Embolism Prevention in Adult Participants With Atrial Fibrillation
Atrial Fibrillation (AF)
Not yet recruitingObservational study
Hummingbird ICE Study
ATRIAL APPENDAGE CLOSURE for ATRIAL FIBRILLATION, Atrial Fibrillation (AF)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Point-by-point PVI with the QDot Micro catheter in QMODE+ setting with the Svtag software in people with Atrial Fibrillation (AF) and Atrial Fibrillation Ablation, run by Semmelweis University at 1 site in Hungary.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 10 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: under way, not looking for participants, as of the sponsor's last update on 11 Dec 2024. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Hungary. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
This study is still under way. Results are not available in the source record yet.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
First-pass isolation rate (time frame: Endpoint is assessed intra-procedurally). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT06721221, which also lists 9 linked publications.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 11 Dec 2024.