Electrolyte DisorderNCT06726460
Validation of the Accuracy of a Novel POCT Dry Electrolyte Analysis System in the Acute Care Setting
An observational study of Electrolyte Disorder and Arrhythmia, run by Zhujiang Hospital at 1 site in China.
Trial summary
The goal of this study is to compare the performance of a novel Point-of-Care Testing (POCT) dry electrolyte analyzer (P1) with the Nova Stat Profile pHOx Ultra Analyzer System (Nova) in an acute care setting. The main questions it aims to answer are: 1. Read more about this study
- Who is it for?
- Ages 28 days and older
- Conditions the study is for
- Electrolyte Disorder, Arrhythmia, Stroke
- What is being tested?
- Nova Stat Profile pHOx Ultra Analyzer System(Nova) and Novel POCT dry electrolyte analysis system(P1)
- How many people
- 200 participants
- Where
- 1 siteChina
- Study dates
- 17 Dec 2024to 31 Oct 2025
- Phase
- Observational
- Are there results?
- Not yet: the study is under way
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 28 days and older
- All sexes
- No healthy volunteers
- Child, Adult, Older adult
Study population: The study population consists of patients presenting to the emergency department (ED) or admitted to the emergency intensive care unit (ICU) who are scheduled to undergo electrolyte measuring.
Who may be able to join
- Patients who are scheduled to undergo electrolyte measuring in the emergency department (ED) or emergency intensive care unit (ICU).
Who cannot join
- Patients with a confirmed history of infectious diseases such as hepatitis B, syphilis, HIV/AIDS, etc.
- Missing residual sample types (serum or whole blood).
- Patients whose residual blood samples are not tested within the specified time frame after collection.
- Other patients deemed ineligible by the investigator.
Trial locations and contacts
1 site in 1 country. Contact the study team or a site near you to ask about taking part.
China1 site, 1 recruiting
- Recruiting
Zhujiang Hospital, Southern Medical University Organization
Guangzhou, Guangdong, 510000
Hua Xie, +86 19802050526, 52528316@qq.com
About this study
Official title: Validation of the Accuracy of a Novel POCT Dry Electrolyte Analysis System in the Acute Care Setting: A Cross-Sectional Study
The goal of this study is to compare the performance of a novel Point-of-Care Testing (POCT) dry electrolyte analyzer (P1) with the Nova Stat Profile pHOx Ultra Analyzer System (Nova) in an acute care setting. The main questions it aims to answer are:
- How consistent are the results between P1 and Nova in the emergency setting, including outliers, correlation, linearity, and bias?
- Whether P1 can serve as a suitable alternative to Nova in the acute care setting for electrolyte measurement in the emergency environment.
Read more
Participants will be patients in the emergency department who will undergo simultaneous measurement with P1 and Nova during their emergency care to assess the performance of both devices.
Additionally, the study will investigate electrolyte levels in emergency patients, focusing on the prevalence of abnormal ionized magnesium (iMg2+) levels, the correlation between iMg2+ and total magnesium (tMg), and the clinical significance of iMg2+ measurement.
Read the detailed description
This observational study aims to compare the performance of two electrolyte analysis systems, the electrolyte analyzer P1 and the Nova Stat Profile pHOx Ultra Analyzer System, in an emergency department (ED) or emergency intensive care unit (ICU) setting. The study will evaluate the consistency, correlation, and bias between the two electrolyte analysis systems. The study design is observational, with no intervention or treatment assigned to the participants. The comparison will be made using residual blood samples collected as part of routine care, following standard clinical procedures.
- Conditions
- Electrolyte Disorder, Arrhythmia, Stroke
- Being tested
- Nova Stat Profile pHOx Ultra Analyzer System(Nova), Novel POCT dry electrolyte analysis system(P1)
- Keywords
- Electrolyte Analyzer, Accuracy, P1, Eaglenos
- Medical subject headings
- Arrhythmias, Cardiac, Stroke
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Nova Stat Profile pHOx Ultra Analyzer System(Nova)Diagnostic test
This group involves the use of Nova for measuring serum and whole blood samples. The results from Nova will be compared with those obtained from P1.
Novel POCT dry electrolyte analysis system(P1)Diagnostic test
This group involves the use of P1 for measuring serum and whole blood samples. The results from P1 will be compared with those obtained from Nova.
| Study group | Role | What they receive |
|---|---|---|
| Group 1: Nova Group | This group involves the use of the Nova Stat Profile pHOx Ultra Analyzer System for measuring serum and whole blood samples. The results obtained from Nova will be compared with those obtained from P1. Receives: Diagnostic Test: Nova Stat Profile pHOx Ultra Analyzer System(Nova) | |
| Group 1: P1 Group | This group involves the use of the novel POCT dry electrolyte analysis system P1 for measuring serum and whole blood samples. The results obtained from P1 will be compared with those obtained from Nova. Receives: Diagnostic Test: Novel POCT dry electrolyte analysis system(P1) |
What is being measured
The primary outcome is the main question the study is designed to answer.
Consistency between P1 and Nova system for electrolyte parameters.
This outcome measure is to evaluate the consistency between the novel POCT dry electrolyte analysis system (P1) and the Nova Stat Profile pHOx Ultra Analyzer System (Nova). Consistency will be assessed by comparing the results of electrolyte parameters (e.g., Sodium (Na+), Potassium (K+), Chloride (Cl-), Ionized calcium (iCa2+), Ionized magnesium (iMg2+)) and will be evaluated using Bland-Altman analysis.
Measured: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
Correlation between P1 and Nova for electrolyte parameters.
This outcome measures the correlation between P1 and Nova for electrolyte parameters (e.g., Na+, K+, Cl-, iCa2+, iMg2+). The correlation will be quantified using Pearson's correlation coefficient
Measured: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
Bias at clinical decision levels between P1 and Nova.
This outcome measure is to evaluate the bias at clinical decision levels between P1 and Nova. Bias will be assessed by calculating the difference in the results of key electrolyte parameters (e.g., Na+, K+, Cl-, iCa2+, iMg2+) between the two analysis system at clinically relevant thresholds. These thresholds will be based on established clinical decision points where treatment decisions are made.
Measured: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
Outlier Detection in electrolyte Parameters
This outcome measure aims to identify and evaluate outliers in the electrolyte parameters measured by P1 and Nova. Outliers will be detected using statistical methods such as the Grubbs test or the Interquartile Range (IQR) method. The impact of these outliers on the consistency and accuracy of the measurements will also be assessed.
Measured: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
Secondary outcomes (2)
Prevalence of abnormal iMg2+ (ionized magnesium) levels in emergency patients.
This outcome will assess the prevalence of abnormal ionized magnesium (iMg2+) levels in patients admitted to the emergency department or emergency ICU.
Measured: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
Correlation between iMg2+ (ionized magnesium) and tMg (total magnesium) levels.
This outcome will evaluate the correlation between ionized magnesium (iMg2+) levels and total magnesium (tMg) levels, with tMg levels obtained from central laboratory test results, in emergency patients.
Measured: Within 30 minutes of serum sample collection
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted10 Dec 2024
- Study start17 Dec 2024
- Primary completion31 Aug 2025 (estimated)
- Study completion31 Oct 2025 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (15)Papers the sponsor has linked to this study.
- Other trials for Electrolyte DisorderStudies of the same condition, open and under way.
- More trials run by Zhujiang HospitalThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Other
- Time perspective
- Cross-sectional
- Enrolment
- 200 participants (estimated)
Sponsor and investigators
- Lead sponsor
- Zhujiang Hospital
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
- Investigators
- Hua Xie, Principal investigator, Zhujiang Hospital, Southern Medical University Organization
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT06726460
- Sponsor's protocol ID
- 2024-KY-405-01
- Status last verified
- Jun 2025
- First submitted
- 4 Dec 2024
- Last updated
- 18 Jul 2025
- Results posted
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- Yes: All collected IPD
Publications
Khan AI, Pratumvinit B, Jacobs E, Kost GJ, Kary H, Balla J, Shaw J, Milevoj Kopcinovic L, Vaubourdolle M, Oliver P, Jarvis PRE, Pamidi P, Erasmus RT, O'Kelly R, Musaad S, Sandberg S. Point-of-care testing performed by healthcare professionals outside the hospital setting: consensus based recommendations from the IFCC Committee on Point-of-Care Testing (IFCC C-POCT). Clin Chem Lab Med. 2023 Jun 2;61(9):1572-1579. doi: 10.1515/cclm-2023-0502. Print 2023 Aug 28.
PubMed 37267483Shi Q, Ba G, Li K, Mao Z. Is it practical to incorporate point-of-care testing into clinical prediction models for emergency? Intern Emerg Med. 2023 Oct;18(7):2155-2156. doi: 10.1007/s11739-023-03333-5. Epub 2023 Jun 6. No abstract available.
PubMed 37278909Johansson M, Whiss PA. Weak relationship between ionized and total magnesium in serum of patients requiring magnesium status. Biol Trace Elem Res. 2007 Jan;115(1):13-21. doi: 10.1385/BTER:115:1:13.
PubMed 17406070Escuela MP, Guerra M, Anon JM, Martinez-Vizcaino V, Zapatero MD, Garcia-Jalon A, Celaya S. Total and ionized serum magnesium in critically ill patients. Intensive Care Med. 2005 Jan;31(1):151-6. doi: 10.1007/s00134-004-2508-x. Epub 2004 Dec 17.
PubMed 15605229Oost LJ, van der Heijden AAWA, Vermeulen EA, Bos C, Elders PJM, Slieker RC, Kurstjens S, van Berkel M, Hoenderop JGJ, Tack CJ, Beulens JWJ, de Baaij JHF. Serum Magnesium Is Inversely Associated With Heart Failure, Atrial Fibrillation, and Microvascular Complications in Type 2 Diabetes. Diabetes Care. 2021 Aug;44(8):1757-1765. doi: 10.2337/dc21-0236. Epub 2021 Jun 18.
PubMed 34385344Rios FJ, Zou ZG, Harvey AP, Harvey KY, Nosalski R, Anyfanti P, Camargo LL, Lacchini S, Ryazanov AG, Ryazanova L, McGrath S, Guzik TJ, Goodyear CS, Montezano AC, Touyz RM. Chanzyme TRPM7 protects against cardiovascular inflammation and fibrosis. Cardiovasc Res. 2020 Mar 1;116(3):721-735. doi: 10.1093/cvr/cvz164.
PubMed 31250885Yogi A, Callera GE, Antunes TT, Tostes RC, Touyz RM. Transient receptor potential melastatin 7 (TRPM7) cation channels, magnesium and the vascular system in hypertension. Circ J. 2011;75(2):237-45. doi: 10.1253/circj.cj-10-1021. Epub 2010 Dec 7.
PubMed 21150127de Baaij JH, Hoenderop JG, Bindels RJ. Magnesium in man: implications for health and disease. Physiol Rev. 2015 Jan;95(1):1-46. doi: 10.1152/physrev.00012.2014.
PubMed 25540137Ye L, Zhang C, Duan Q, Shao Y, Zhou J. Association of Magnesium Depletion Score With Cardiovascular Disease and Its Association With Longitudinal Mortality in Patients With Cardiovascular Disease. J Am Heart Assoc. 2023 Sep 19;12(18):e030077. doi: 10.1161/JAHA.123.030077. Epub 2023 Sep 8.
PubMed 37681518Glasdam SM, Glasdam S, Peters GH. The Importance of Magnesium in the Human Body: A Systematic Literature Review. Adv Clin Chem. 2016;73:169-93. doi: 10.1016/bs.acc.2015.10.002. Epub 2016 Jan 13.
PubMed 26975973Olsson K, Ohlin B, Melander O. Epidemiology and characteristics of hyponatremia in the emergency department. Eur J Intern Med. 2013 Mar;24(2):110-6. doi: 10.1016/j.ejim.2012.10.014. Epub 2012 Nov 22.
PubMed 23176963Abensur Vuillaume L, Ferreira JP, Asseray N, Trombert-Paviot B, Montassier E, Legrand M, Girerd N, Boivin JM, Chouihed T, Rossignol P. Hypokalemia is frequent and has prognostic implications in stable patients attending the emergency department. PLoS One. 2020 Aug 4;15(8):e0236934. doi: 10.1371/journal.pone.0236934. eCollection 2020.
PubMed 32750075Ookuma T, Miyasho K, Kashitani N, Beika N, Ishibashi N, Yamashita T, Ujike Y. The clinical relevance of plasma potassium abnormalities on admission in trauma patients: a retrospective observational study. J Intensive Care. 2015 Aug 13;3(1):37. doi: 10.1186/s40560-015-0103-6. eCollection 2015.
PubMed 26269745Fleet JL, Shariff SZ, Gandhi S, Weir MA, Jain AK, Garg AX. Validity of the International Classification of Diseases 10th revision code for hyperkalaemia in elderly patients at presentation to an emergency department and at hospital admission. BMJ Open. 2012 Dec 28;2(6):e002011. doi: 10.1136/bmjopen-2012-002011. Print 2012.
PubMed 23274674Muschart X, Boulouffe C, Jamart J, Nougon G, Gerard V, de Canniere L, Vanpee D. A determination of the current causes of hyperkalaemia and whether they have changed over the past 25 years. Acta Clin Belg. 2014 Aug;69(4):280-4. doi: 10.1179/0001551214Z.00000000077. Epub 2014 Jun 18.
PubMed 24942977
Study at a glance
From the registry record, last updated by the sponsor on 18 Jul 2025. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Validation of the Accuracy of a Novel POCT Dry Electrolyte Analysis System in the Acute Care Setting: A Cross-Sectional Study
- ClinicalTrials.gov ID
- NCT06726460
- Acronym / sponsor ID
- 2024-KY-405-01
- Conditions
- Electrolyte Disorder, Arrhythmia, Stroke
- Keywords
- Electrolyte Analyzer, Accuracy, P1, Eaglenos
- Study type
- Observational
- Recruitment status
- Recruiting
- Enrolment
- 200 participants (estimated)
- Lead sponsor
- Zhujiang HospitalUniversity, hospital or other
- Study start
- 17 Dec 2024 (actual)
- Primary completion
- 31 Aug 2025 (estimated)
- Study completion
- 31 Oct 2025 (estimated)
- First posted
- 10 Dec 2024 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The goal of this study is to compare the performance of a novel Point-of-Care Testing (POCT) dry electrolyte analyzer (P1) with the Nova Stat Profile pHOx Ultra Analyzer System (Nova) in an acute care setting. The main questions it aims to answer are:
- How consistent are the results between P1 and Nova in the emergency setting, including outliers, correlation, linearity, and bias?
- Whether P1 can serve as a suitable alternative to Nova in the acute care setting for electrolyte measurement in the emergency environment.
Participants will be patients in the emergency department who will undergo simultaneous measurement with P1 and Nova during their emergency care to assess the performance of both devices.
Additionally, the study will investigate electrolyte levels in emergency patients, focusing on the prevalence of abnormal ionized magnesium (iMg2+) levels, the correlation between iMg2+ and total magnesium (tMg), and the clinical significance of iMg2+ measurement.
Detailed description
This observational study aims to compare the performance of two electrolyte analysis systems, the electrolyte analyzer P1 and the Nova Stat Profile pHOx Ultra Analyzer System, in an emergency department (ED) or emergency intensive care unit (ICU) setting. The study will evaluate the consistency, correlation, and bias between the two electrolyte analysis systems. The study design is observational, with no intervention or treatment assigned to the participants. The comparison will be made using residual blood samples collected as part of routine care, following standard clinical procedures.
Study design
- Observational model
- Other
- Time perspective
- Cross-sectional
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Group 1: Nova Group | – | Diagnostic Test: Nova Stat Profile pHOx Ultra Analyzer System(Nova)This group involves the use of the Nova Stat Profile pHOx Ultra Analyzer System for measuring serum and whole blood samples. The results obtained from Nova will be compared with those obtained from P1. |
| Group 1: P1 Group | – | Diagnostic Test: Novel POCT dry electrolyte analysis system(P1)This group involves the use of the novel POCT dry electrolyte analysis system P1 for measuring serum and whole blood samples. The results obtained from P1 will be compared with those obtained from Nova. |
Who may qualify
- Ages
- 28 days and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Child, Adult, Older adult
- Study population
- The study population consists of patients presenting to the emergency department (ED) or admitted to the emergency intensive care unit (ICU) who are scheduled to undergo electrolyte measuring.
Key inclusion criteria (1 listed)
- Patients who are scheduled to undergo electrolyte measuring in the emergency department (ED) or emergency intensive care unit (ICU).
Key exclusion criteria (4 listed)
- Patients with a confirmed history of infectious diseases such as hepatitis B, syphilis, HIV/AIDS, etc.
- Missing residual sample types (serum or whole blood).
- Patients whose residual blood samples are not tested within the specified time frame after collection.
- Other patients deemed ineligible by the investigator.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Nova Stat Profile pHOx Ultra Analyzer System(Nova)This group involves the use of Nova for measuring serum and whole blood samples. The results from Nova will be compared with those obtained from P1. | Diagnostic test | Group 1: Nova Group |
| Novel POCT dry electrolyte analysis system(P1)This group involves the use of P1 for measuring serum and whole blood samples. The results from P1 will be compared with those obtained from Nova. | Diagnostic test | Group 1: P1 Group |
Primary outcome
Consistency between P1 and Nova system for electrolyte parameters.
Time frame: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
This outcome measure is to evaluate the consistency between the novel POCT dry electrolyte analysis system (P1) and the Nova Stat Profile pHOx Ultra Analyzer System (Nova). Consistency will be assessed by comparing the results of electrolyte parameters (e.g., Sodium (Na+), Potassium (K+), Chloride (Cl-), Ionized calcium (iCa2+), Ionized magnesium (iMg2+)) and will be evaluated using Bland-Altman analysis.
Correlation between P1 and Nova for electrolyte parameters.
Time frame: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
This outcome measures the correlation between P1 and Nova for electrolyte parameters (e.g., Na+, K+, Cl-, iCa2+, iMg2+). The correlation will be quantified using Pearson's correlation coefficient
Bias at clinical decision levels between P1 and Nova.
Time frame: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
This outcome measure is to evaluate the bias at clinical decision levels between P1 and Nova. Bias will be assessed by calculating the difference in the results of key electrolyte parameters (e.g., Na+, K+, Cl-, iCa2+, iMg2+) between the two analysis system at clinically relevant thresholds. These thresholds will be based on established clinical decision points where treatment decisions are made.
Outlier Detection in electrolyte Parameters
Time frame: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
This outcome measure aims to identify and evaluate outliers in the electrolyte parameters measured by P1 and Nova. Outliers will be detected using statistical methods such as the Grubbs test or the Interquartile Range (IQR) method. The impact of these outliers on the consistency and accuracy of the measurements will also be assessed.
2 secondary outcomes
Prevalence of abnormal iMg2+ (ionized magnesium) levels in emergency patients.
Time frame: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection
Correlation between iMg2+ (ionized magnesium) and tMg (total magnesium) levels.
Time frame: Within 30 minutes of serum sample collection
Results and safety
This study is still under way. Results are not available in the source record yet.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 18 Jul 2025.
- Background: PubMed 37267483Khan AI, Pratumvinit B, Jacobs E, Kost GJ, Kary H, Balla J, Shaw J, Milevoj Kopcinovic L, Vaubourdolle M, Oliver P, Jarvis PRE, Pamidi P, Erasmus RT, O'Kelly R, Musaad S, Sandberg S. Point-of-care testing performed by healthcare professionals outside the hospital setting: consensus based recommendations from the IFCC Committee on Point-of-Care Testing (IFCC C-POCT). Clin Chem Lab Med. 2023 Jun 2;61(9):1572-1579. doi: 10.1515/cclm-2023-0502. Print 2023 Aug 28.
- Background: PubMed 37278909Shi Q, Ba G, Li K, Mao Z. Is it practical to incorporate point-of-care testing into clinical prediction models for emergency? Intern Emerg Med. 2023 Oct;18(7):2155-2156. doi: 10.1007/s11739-023-03333-5. Epub 2023 Jun 6. No abstract available.
- Background: PubMed 17406070Johansson M, Whiss PA. Weak relationship between ionized and total magnesium in serum of patients requiring magnesium status. Biol Trace Elem Res. 2007 Jan;115(1):13-21. doi: 10.1385/BTER:115:1:13.
- Background: PubMed 15605229Escuela MP, Guerra M, Anon JM, Martinez-Vizcaino V, Zapatero MD, Garcia-Jalon A, Celaya S. Total and ionized serum magnesium in critically ill patients. Intensive Care Med. 2005 Jan;31(1):151-6. doi: 10.1007/s00134-004-2508-x. Epub 2004 Dec 17.
- Background: PubMed 34385344Oost LJ, van der Heijden AAWA, Vermeulen EA, Bos C, Elders PJM, Slieker RC, Kurstjens S, van Berkel M, Hoenderop JGJ, Tack CJ, Beulens JWJ, de Baaij JHF. Serum Magnesium Is Inversely Associated With Heart Failure, Atrial Fibrillation, and Microvascular Complications in Type 2 Diabetes. Diabetes Care. 2021 Aug;44(8):1757-1765. doi: 10.2337/dc21-0236. Epub 2021 Jun 18.
- Background: PubMed 31250885Rios FJ, Zou ZG, Harvey AP, Harvey KY, Nosalski R, Anyfanti P, Camargo LL, Lacchini S, Ryazanov AG, Ryazanova L, McGrath S, Guzik TJ, Goodyear CS, Montezano AC, Touyz RM. Chanzyme TRPM7 protects against cardiovascular inflammation and fibrosis. Cardiovasc Res. 2020 Mar 1;116(3):721-735. doi: 10.1093/cvr/cvz164.
- Background: PubMed 21150127Yogi A, Callera GE, Antunes TT, Tostes RC, Touyz RM. Transient receptor potential melastatin 7 (TRPM7) cation channels, magnesium and the vascular system in hypertension. Circ J. 2011;75(2):237-45. doi: 10.1253/circj.cj-10-1021. Epub 2010 Dec 7.
- Background: PubMed 25540137de Baaij JH, Hoenderop JG, Bindels RJ. Magnesium in man: implications for health and disease. Physiol Rev. 2015 Jan;95(1):1-46. doi: 10.1152/physrev.00012.2014.
- Background: PubMed 37681518Ye L, Zhang C, Duan Q, Shao Y, Zhou J. Association of Magnesium Depletion Score With Cardiovascular Disease and Its Association With Longitudinal Mortality in Patients With Cardiovascular Disease. J Am Heart Assoc. 2023 Sep 19;12(18):e030077. doi: 10.1161/JAHA.123.030077. Epub 2023 Sep 8.
- Background: PubMed 26975973Glasdam SM, Glasdam S, Peters GH. The Importance of Magnesium in the Human Body: A Systematic Literature Review. Adv Clin Chem. 2016;73:169-93. doi: 10.1016/bs.acc.2015.10.002. Epub 2016 Jan 13.
- Background: PubMed 23176963Olsson K, Ohlin B, Melander O. Epidemiology and characteristics of hyponatremia in the emergency department. Eur J Intern Med. 2013 Mar;24(2):110-6. doi: 10.1016/j.ejim.2012.10.014. Epub 2012 Nov 22.
- Background: PubMed 32750075Abensur Vuillaume L, Ferreira JP, Asseray N, Trombert-Paviot B, Montassier E, Legrand M, Girerd N, Boivin JM, Chouihed T, Rossignol P. Hypokalemia is frequent and has prognostic implications in stable patients attending the emergency department. PLoS One. 2020 Aug 4;15(8):e0236934. doi: 10.1371/journal.pone.0236934. eCollection 2020.
- Individual participant dataSharing planned: Yes. All collected IPD
Related trials for Electrolyte Disorder
Recruiting
NIRAF-Guided Parathyroid Identification During Thyroidectomy
Hypocalcemia, Hypoparathyroidism
RecruitingPhase 3
Systematic Use of DDAVP to Prevent Serum Sodium Overcorrection in Severe Hyponatremia
Hyponatremia
Enrolling by invitationObservational study
AK+ Guard™ Pilot Study in Chronic Kidney Disease: Outpatient Diagnostic Accuracy and Remote Monitoring
Hyperkalemia, Chronic Kidney Disease (Stage 3-4)
RecruitingPhase 2
Rapid Rehydration for Severe Hypernatremia
Hypernatremia
RecruitingPhase 3
A Phase 3 Study Evaluating the Efficacy and Safety of WS016 for the Treatment of Hyperkalemia
Hyperkalemia
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part.
Hua Xie principal investigator
Zhujiang Hospital, Southern Medical University Organization
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Electrolyte Disorder and Arrhythmia, run by Zhujiang Hospital at 1 site in China.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 1 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: looking for participants now, as of the sponsor's last update on 18 Jul 2025. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
This study is still under way. Results are not available in the source record yet.
How can a doctor refer a patient or contact the research team?
Contact Hua Xie (+86 19802050526, 52528316@qq.com), or a site near the patient. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Consistency between P1 and Nova system for electrolyte parameters. (time frame: Within 30 minutes of serum sample collection and within 1 minute of whole blood sample collection). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT06726460, which also lists 15 linked publications.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 18 Jul 2025.