Respiratory FailureNCT06726473

Validation of the Accuracy of a Novel POCT Dry Blood Gas Analyzer in Acute Care Settings

An observational study of Respiratory Failure and Acid-Base Balance Disorder, run by Zhujiang Hospital at 1 site in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The goal of this observational study is to validate the performance of the novel POCT dry blood gas analyzer EG-i30, against the currently widely used traditional wet blood gas analyzer, Radiometer ABL90, in acute care settings. The main questions it aims to answer are: 1. Read more about this study

Who is it for?
Ages 28 days and older
What is being tested?
Wet Blood Gas Analyzer (ABL90) and Dry Blood Gas Analyzer (EG-i30)
How many people
216 participants
Where
1 siteChina
Study dates
17 Dec 2024to 30 Jun 2025
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 28 days and older
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Study population: The study population consists of patients presenting to the emergency department (ED) or admitted to the emergency intensive care unit (ICU) who are scheduled to undergo arterial blood gas analysis.

Who may be able to join

  • Patients who are scheduled to undergo arterial blood gas analysis in the emergency department (ED) or emergency intensive care unit (EICU).

Who cannot join

  • Patients with a confirmed history of infectious diseases such as hepatitis B, syphilis, HIV/AIDS, etc.
  • Patients whose residual blood samples are not tested within the specified time frame after collection.
  • Other patients deemed ineligible by the investigator (e.g., samples with contamination or other factors affecting sample quality).

Trial locations and contacts

1 site in 1 country.

China1 site
  • Zhujiang Hospital, Southern Medical University Organization

    Guangzhou, Guangdong, 510000

Other trials in China

About this study

Official title: Validation of the Accuracy of a Novel POCT Dry Blood Gas Analyzer in Acute Care Settings:A Cross-Sectional Study

The goal of this observational study is to validate the performance of the novel POCT dry blood gas analyzer EG-i30, against the currently widely used traditional wet blood gas analyzer, Radiometer ABL90, in acute care settings. The main questions it aims to answer are:

  1. How consistent are the results between EG-i30 and ABL90 in acute care settings, including outliers, correlation, linearity, and bias?
  2. Whether EG-i30 can be considered a suitable alternative to ABL90 in acute care settings.

Participants will be patients in the emergency department who will undergo simultaneous testing with EG-i30 and ABL90 during their emergency care to assess the performance of EG-i30.

Read the detailed description

This observational study aims to compare the performance of two blood gas analyzers, the EG-i30 dry blood gas analyzer and the ABL90 wet blood gas analyzer, in an emergency department (ED) or emergency intensive care unit (ICU) setting. The study will evaluate the consistency, correlation, and bias between the two analyzers for various blood gas and biochemical parameters, such as Potential of hydrogen (pH), Partial pressure of oxygen (PaO2), Partial pressure of carbon dioxide (PaCO2), Sodium (Na+), Potassium (K+), Chloride (Cl-), Ionized calcium (iCa2+), Lactate (Lac), Glucose (Glu), and Hematocrit (Hct). The study design is observational, with no intervention or treatment assigned to the participants. The comparison will be made using residual blood samples collected as part of routine care, following standard clinical procedures.

Being tested
Wet Blood Gas Analyzer (ABL90), Dry Blood Gas Analyzer (EG-i30)
Keywords
blood gas analysis, accuracy, EG-i30, Eaglenos
Medical subject headings
Respiratory Insufficiency, Sepsis, Shock, Respiratory Distress Syndrome

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Wet Blood Gas Analyzer (ABL90)Diagnostic test

    This group involves the use of the ABL90 wet blood gas analyzer for measuring arterial whole blood samples. The results obtained from the ABL90 analyzer will be compared with those obtained from the EG-i30 dry blood gas analyzer

  • Dry Blood Gas Analyzer (EG-i30)Diagnostic test

    This group involves the use of the EG-i30 dry blood gas analyzer for measuring arterial whole blood samples. The results obtained from the EG-i30 analyzer will be compared with those obtained from the ABL90 wet blood gas analyzer.

Study groups in NCT06726473
Study groupRoleWhat they receive
Group 1: ABL90 GroupBlood gas and biochemical measurements are performed using the ABL90 blood gas analyzer. The results will be compared with those from the EG-i30 analyzer.
Receives: Diagnostic Test: Wet Blood Gas Analyzer (ABL90)
Group 2: EG-i30 GroupBlood gas and biochemical measurements are performed using the EG-i30 blood gas analyzer. The results will be compared with those from the ABL90 analyzer.
Receives: Diagnostic Test: Dry Blood Gas Analyzer (EG-i30)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Consistency between EG-i30 and ABL90 analyzers for blood gas and biochemical parameters.

    This outcome measures is to evaluate the consistency between the EG-i30 dry blood gas analyzer and the ABL90 wet blood gas analyzer. Consistency will be assessed by comparing the results of blood gas and biochemical parameters (e.g., pH, PaO2, PaCO2, Na+, K+, Cl-, iCa2+, Lac, Glu, and Hct) and will be evaluated using Bland-Altman analysis.

    Measured: Within 2 minutes after arterial blood sample collection

  2. Correlation between EG-i30 and ABL90 analyzers for blood gas and biochemical parameters.

    This outcome measures the correlation between the EG-i30 and ABL90 analyzers for blood gas and biochemical parameters (e.g., pH, PaO2, PaCO2, Na+, K+, Cl-, iCa2+, Lac, Glu, and Hct). The correlation will be quantified using Pearson's correlation coefficient.

    Measured: Within 2 minutes after arterial blood sample collection

  3. Bias at clinical decision levels between EG-i30 and ABL90 analyzers.

    This outcome measure is to evaluate the bias at clinical decision levels between the EG-i30 and ABL90 analyzers. Bias will be assessed by calculating the difference in the results of key blood gas and biochemical parameters (e.g., pH, PaO2, PaCO2, Na+, K+, Cl-, iCa2+, Lac, Glu, and Hct) between the two analyzers at clinically relevant thresholds. These thresholds will be based on established clinical decision points where treatment decisions are made.

    Measured: Within 2 minutes after arterial blood sample collection

  4. Outlier Detection in Blood Gas and Biochemical Parameters

    This outcome measure aims to identify and evaluate outliers in the blood gas and biochemical parameters (e.g., pH, PaO2, PaCO2, Na+, K+, Cl-, iCa2+, Lac, Glu, and Hct) measured by the EG-i30 and ABL90 analyzers. Outliers will be detected using statistical methods such as the Grubbs test or the Interquartile Range (IQR) method. The impact of these outliers on the consistency and accuracy of the measurements will also be assessed.

    Measured: Within 2 minutes after arterial blood sample collection

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted10 Dec 2024
  2. Study start17 Dec 2024
  3. Primary completion2 Jan 2025
  4. Study completion30 Jun 2025

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Other
Time perspective
Cross-sectional
Enrolment
216 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06726473
Sponsor's protocol ID
2024-KY-408-01
Status last verified
Jan 2025
First submitted
4 Dec 2024
Last updated
18 Jul 2025
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Yes: All collected IPD

Publications

  • Breathnach CS. The development of blood gas analysis. Med Hist. 1972 Jan;16(1):51-62. doi: 10.1017/s0025727300017257. No abstract available.

    PubMed 4558439
  • Shapiro BA. The history of pH and blood gas analysis. Respir Care Clin N Am. 1995 Sep;1(1):1-5.

    PubMed 9390846
  • Davis MD, Walsh BK, Sittig SE, Restrepo RD. AARC clinical practice guideline: blood gas analysis and hemoximetry: 2013. Respir Care. 2013 Oct;58(10):1694-703. doi: 10.4187/respcare.02786. Epub 2013 Jul 30.

    PubMed 23901131

Questions about this trial

What is this clinical trial studying?

An observational study of Respiratory Failure and Acid-Base Balance Disorder, run by Zhujiang Hospital at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 18 Jul 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Consistency between EG-i30 and ABL90 analyzers for blood gas and biochemical parameters. (time frame: Within 2 minutes after arterial blood sample collection). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT06726473, which also lists 3 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 18 Jul 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in