Pulse Oximeter ValidationNCT06771206

Evaluation of Accuracy of the Owlet OSS 3.0 Sensor in the Neonatal Population

A study testing Owlet pulse oximeter neonatal accuracy validation in people with Pulse Oximeter Validation, run by Owlet Baby Care, Inc. at 2 sites in United States.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

The sponsor of this study, Owlet Baby Care, Inc., is comparing the accuracy of their pulse oximeter Sensor with the measurements of oxygen levels in newborns who are already being monitored with arterial blood samples. Read more about this study

Who is it for?
Ages 0 to 28 days
Conditions the study is for
Pulse Oximeter Validation
What is being tested?
Owlet pulse oximeter neonatal accuracy validation
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
20 participants
Where
2 sitesUnited States
Study dates
1 Aug 2025to Feb 2026
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 0 to 28 days
  • All sexes
  • No healthy volunteers
  • Child

Who may be able to join

  • Less than or equal to 44 weeks corrected gestational age
  • Weight equal or greater than 1500g
  • Subjects with standard of care (SOC) arterial blood sampling line already in place at time of enrollment, with anticipated arterial blood gas measurements as part of their plan of care

Who cannot join

  • Subjects with underdeveloped skin as assessed by a study investigator and/or reported by the clinical care team
  • Subjects with abnormalities (e.g., limb deformities, open wounds, severe edema) at the planned application sites that would interfere with system measurements
  • Potential subjects on extracorporeal membrane oxygenation (ECMO) or other mechanical circulatory support
  • Arterial line is placed in a preductal position with the potential of pulmonary to systemic shunting
  • Subjects undergoing cooling protocol/therapeutic hypothermia
  • Subjects who are deemed too clinically unstable by the Principal Investigator (PI), sub-investigator and/or clinical care team to participate in the study.

Trial locations and contacts

2 sites in 1 country. Contact the study team or a site near you to ask about taking part.

Alisa L Niksch, M.D. contact

(844) 334-5330

clinical@owletcare.com

Guy Alvarez, PhD contact

galvarez@owletcare.com

United States2 sites, 2 recruiting
  • University of Alabama Birmingham - Children's of Alabama

    Birmingham, Alabama, 35233

    Kristal Hock, MSN, (205) 996-3313, KristalHock@uabmc.edu

    Michael A Brock, M.D.

    Recruiting
  • University of Minnesota Medical Center

    Minneapolis, Minnesota, 55454

    Amy Kodet, MPP, (612) 625-6678, kodet006@umn.edu

    Gwenyth Fischer, M.D.

    Recruiting

Other trials in United States

About this study

The sponsor of this study, Owlet Baby Care, Inc., is comparing the accuracy of their pulse oximeter Sensor with the measurements of oxygen levels in newborns who are already being monitored with arterial blood samples. A minimum of 20, but up to 100 participants will be enrolled in a hospital environment and need to weigh 1500 grams or more. The purpose of the testing is to gather information on pulse oximeter readings taken by the Owlet OSS 3.0 Sensor and compare the readings to blood gas samples drawn by the medical staff. This will determine if the Owlet Sensor is accurate in newborns up to an age of 28 days after actual or expected delivery date.

The Owlet OSS 3.0 is the Sensor component within the Owlet Dream Sock, an over the counter, noninvasive pulse oximeter which is currently FDA cleared in infants 1-18 months and between 6 and 30lbs, meant to be used in the home environment. The Owlet Sock is a fabric wrap that secures the Sensor to a baby's foot. The Sensor estimates the baby's SpO2, pulse rate, and measures movement level, and transmits the wearer's readings to the paired Base Station via low energy BlueTooth energy.

Read the detailed description

This is a prospective, non-randomized, single arm study to evaluate SpO2 performance of the Owlet OSS 3.0 pulse oximetry Sensor in the neonate population up to 44 weeks corrected gestational age using convenience sampling of arterial blood gas samples as a part of routine clinical monitoring.

The Owlet OSS 3.0 Sensor is the component of the Owlet Dream Sock (DEN220091) and BabySat (K222597), both Food and Drug Administration (FDA) cleared pulse oximeters for infants 1-18 months in 2023. This component performs the signal acquisition and analysis to estimate pulse rate and SpO2 readings.

The study will enroll between 20-100 neonates weighing a minimum of 1500g in the hospital setting with available arterial line access and arterial blood gas samples being performed as part of planned patient management. No additional blood gas samples will be taken beyond clinical care indications.

The primary outcome of the study is the accuracy of the Owlet OSS 3.0 Sensor SpO2 readings compared to simultaneous arterial saturation (SaO2) measurements from arterial blood gas samples as determined by a root mean square (ARMS) calculation.

Secondary outcomes will include:

  • Bias in the Owlet OSS 3.0 Sensor SpO2 vs SaO2 defined as (Owlet OSS 3.0 Sensor SpO2 - SaO2).
  • Precision of the Owlet OSS 3.0 Sensor SpO2 vs SaO2 defined as ABS (Owlet OSS 3.0 Sensor SpO2 - SaO2).

Accuracy will be stratified by a number of demographic elements, including by a 10 point skin tone scale and Individual Typology Angle (ITA) as determined by a spectrophotometer to demonstrate that the sensor technology under test can be accurate across diverse user populations.

The primary hypothesis is that the Owlet OSS 3.0 Sensor will measure SpO2 values within +/- 3% ARMS compared to SaO2 measurements.

Each study participant will have a total study exposure time of 48 hours. Each participant will be assigned to a unique Owlet device with a specific serial number which will be paired with the study subject identifier. The Owlet OSS 3.0 Sensor will be attached to a study subject's foot and SpO2 readings will be acquired at a set frequency and time-aligned with arterial blood gas sampling performed during the course of routine patient care. Study operators will be blinded to Owlet device readings. Time stamping of Owlet SpO2 readings will be achieved by a prescribed tapping of the device's Base Station which will mark the point at which readings simultaneous to the blood gas sample were recorded. SpO2 readings from the available noninvasive reference pulse oximeter will also be recorded during the time of blood gas sampling to document the subject's baseline SpO2 plateau.

Being tested
Owlet pulse oximeter neonatal accuracy validation
Keywords
Pulse oximetry, Neonates, Owlet, Accuracy

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Owlet pulse oximeter neonatal accuracy validationDevice

    This study specifically evaluates the Owlet pulse oximetry sensor, which is specifically designed and FDA cleared for for infants, and whose accuracy is specifically being validated in the neonate population.

Study groups in NCT06771206
Study groupRoleWhat they receive
Owlet ArmExperimentalAll study participants meeting inclusion criteria with no stated exclusions will wear the Owlet OSS 3.0 Sensor during the study intervention. The Owlet OSS 3.0 Sensor may be placed continuously on a study subject for up to 48 hours with rotation to the opposite foot every 8 hours per device labeling to enable noninvasive data collection during the subject's standard of care blood draws. The Owlet base station will be activated in order to transmit readings but will not provide any external alarms or live displays during the monitoring period. No display of readings from the assigned Owlet device will be made available to the participant or the investigators during the time of data acquisition.
Receives: Device: Owlet pulse oximeter neonatal accuracy validation

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Pulse oximetry Accuracy - Root Mean Square

    Accuracy of the Owlet OSS 3.0 Sensor SpO2 readings compared to simultaneous arterial saturation (SaO2) measurements from arterial blood gas samples as determined by a root mean square (ARMS) calculation.

    Measured: 48 hours

Secondary outcomes (2)
  1. Pulse oximetry Accuracy - Bias

    Bias in the Owlet OSS 3.0 Sensor SpO2 vs SaO2 defined as (Owlet OSS 3.0 Sensor SpO2 - SaO2).

    Measured: 48 hours

  2. Pulse oximetry Accuracy - Precision

    Precision of the Owlet OSS 3.0 Sensor SpO2 vs SaO2 defined as ABS (Owlet OSS 3.0 Sensor SpO2 - SaO2).

    Measured: 48 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 Jan 2025
  2. Study start1 Aug 2025
  3. Primary completionDec 2025 (estimated)
  4. Study completionFeb 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
20 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06771206
Sponsor's protocol ID
OWL PR0007
Status last verified
Aug 2025
First submitted
7 Jan 2025
Last updated
14 Aug 2025
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
Undecided

Publications

  • Okunlola OE, Lipnick MS, Batchelder PB, Bernstein M, Feiner JR, Bickler PE. Pulse Oximeter Performance, Racial Inequity, and the Work Ahead. Respir Care. 2022 Feb;67(2):252-257. doi: 10.4187/respcare.09795. Epub 2021 Nov 12.

    PubMed 34772785
  • Sjoding MW, Dickson RP, Iwashyna TJ, Gay SE, Valley TS. Racial Bias in Pulse Oximetry Measurement. N Engl J Med. 2020 Dec 17;383(25):2477-2478. doi: 10.1056/NEJMc2029240. No abstract available.

    PubMed 33326721
  • Wackernagel D, Blennow M, Hellstrom A. Accuracy of pulse oximetry in preterm and term infants is insufficient to determine arterial oxygen saturation and tension. Acta Paediatr. 2020 Nov;109(11):2251-2257. doi: 10.1111/apa.15225. Epub 2020 Mar 6.

    PubMed 32043645

Questions about this trial

What is this clinical trial studying?

A study testing Owlet pulse oximeter neonatal accuracy validation in people with Pulse Oximeter Validation, run by Owlet Baby Care, Inc. at 2 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 14 Aug 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Alisa L Niksch, M.D. ((844) 334-5330, clinical@owletcare.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Pulse oximetry Accuracy - Root Mean Square (time frame: 48 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT06771206, which also lists 3 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 14 Aug 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in