Neonatal Respiratory DistressNOBNCT06817850

Neonatal Oximeter Bias Study

An observational study of Neonatal Respiratory Distress and Oxygenation, run by Czech Technical University in Prague at 1 site in Czechia.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

The study aims to determine if there is a systematic difference (bias) between pulse oximeters used in control systems for automated oxygen delivery (A-FiO2) and those used in monitoring systems. When using A-FiO2 systems there are commonly two oximeter probes on the infant with two difference readouts. Read more about this study

Who is it for?
Any age
Conditions the study is for
Neonatal Respiratory Distress, Oxygenation
How many people
40 participants
Where
1 siteCzechia
Study dates
15 Mar 2025to 31 Oct 2025
Phase
Observational
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Any age
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Study population: All infants in the NICU monitored with two simultaneous oximeters with probes in preductal locations. The use of A-FiO2 is expected.

Who may be able to join

  • Observations of any infant in the neonatal ICU with two simultaneous oximeter monitors may be included. The sites' Ethics Committee must review the protocol, approve participation and determine if Informed Consent is required. If required, it might be prospective, that is prior to collecting any data, or retrospective.

Who cannot join

  • Observations are to be made at the convenience of the Investigative team. They should reflect routine situation, and not be selected to specifically capture problems. Thus, the photo should be taken immediately with both displays are on in the view, and not delayed until something interesting happens. Observations should not be made if any of the following exclusion criteria are meet. If after collection, any of these are meet, the data should not be included as an Observation on the Observation-CRF.
  • One oximeter monitoring post ductal SpO2 in infant with clinically relevant shunt.
  • Monitor and A-FiO2 Oximeter not reading between 70-100%
  • Less than 5 minutes between repeat Observations
  • No more than 5 measurements per day per subject with sensors in same position and same oximeters (i.e., every time either sensor is moved, 5 new Observations can be made that day).
  • Presence of motion artifact
  • Presence of sensor integrity alarms/alerts

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Jakub Rafl, PhD contact

+420728229991

rafl@fbmi.cvut.cz

Czechia1 site
  • Motol University Hospital, Neonatal Unit

    Prague, 15500

    Jana Dornakova, MD, +420607720060, Jana.Dornakova@fnmotol.cz

    Jana Dornakova, MD

Other trials in Czechia

About this study

Official title: Evaluation of the Differences Between Simultaneous SpO2s Readings in the NICU: a Multicenter Observational Study

The study aims to determine if there is a systematic difference (bias) between pulse oximeters used in control systems for automated oxygen delivery (A-FiO2) and those used in monitoring systems.

When using A-FiO2 systems there are commonly two oximeter probes on the infant with two difference readouts. Nurses report frustration that the two readings are often markedly different. It is understandable that physiological differences between sensor sites might reflect different regional oxygen saturation levels. It is also possible that there is a relevant systematic bias between difference monitors and sensors.

Therefore, a large systematic multicenter study is needed to determine whether these frequent differences should be ignored as physiological noise or considered clinically relevant.

Read the detailed description

The accuracy of clinical pulse oximeters is specified as 3% RMS (root mean squared of the bias and variance) compared to arterial SaO2, in the range of 70-100% SpO2. This combines errors from bias and from scatter (imprecision). These validations are carried out under ideal laboratory conditions; that is, on healthy volunteers who undergo experimental desaturations, and not in the routine clinical environment. In a large multicenter clinical observation study, Ross et al in 2013 identified important related concerns. Specifically, they found that many of the clinical measurements were outside the 3% accuracy envelop. Specifically, they reported that bias varied depending on the level of saturation, the oximeter brand, sensor, race and site perfusion. Others have identified relevant differences in oximeters, oximeter sensors [Maiwald], and skin pigmentation.

Our study would be the first study in the neonatal ICU evaluating bias to consider different oximeters, sensors and sensor sites. It is highly topical in that changes in average SpO2 of 3% have been associated with excess neonatal mortality and morbidity.

This study is planned to investigate the source and magnitude of differences in SpO2 readings from oximeter sensors at different sensor sites that are routinely noticed by clinicians. While these differences are most often ignored and attributed to sensor site perfusion, if systematic they could have marked impact on mortality and morbidity.

The aim of the investigation is to provide practical guidance relating to SpO2 bias among oximeters, oximeter sensors and sensor location. While taking advantage of the clinical need for 2 oximeter sensors, the results will be applicable to all neonatal oximeter monitoring.

An observational design was selected to take place in centers using automated FiO2 control systems (A-FiO2) that routinely require the use of one sensor for control and another for monitoring. This is a refinement of the approach used in the often-cited multicenter evaluation of neonatal SpO2 exposure. Thus, without an investigational intervention the observational design is well suited to collect comparable data from multiple centers.

Keywords
pulse oximetry, automated FiO2 control, SpO2
Medical subject headings
Pulmonary Atelectasis

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

Study groups in NCT06817850
Study groupRoleWhat they receive
Infants in the NICU with two oximetersInfants in the NICU monitored with two oximeters with probes in preductal locations. The use of A-FiO2 is expected.

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Bias between paired SpO2 measurements

    Bias between paired SpO2 in three ranges:

    1. Hypoxemia (SpO2: 70 - <89.5%),
    2. Normoxemia (SpO2: 89.5-95.5%)
    3. Hyperoxemia (SpO2: >95.5-100%)

    The ranges are determined as the mean of the two oximeters.

    Measured: 1 month (250 observations is expected from each participating center; on average 2 patients and 5 observations per day per patient assumed)

Secondary outcomes (6)
  1. Bias between paired SpO2 related to site

    Bias between paired SpO2 (in %) in three ranges related to independent variable: site (as factor).

    Measured: 1 month (250 observations is expected from each participating center; on average 2 patients and 5 observations per day per patient assumed)

  2. Bias between paired SpO2 related to sensor location

    Bias between paired SpO2 (in %) in three ranges related to independent variable: sensor location (as factor).

    Measured: 1 month (250 observations is expected from each participating center; on average 2 patients and 5 observations per day per patient assumed)

  3. Bias between paired SpO2 related to oximeter brand pairs

    Bias between paired SpO2 (in %) in three ranges related to independent variable: oximeter brand pairs (as factor).

    Measured: 1 month (250 observations is expected from each participating center; on average 2 patients and 5 observations per day per patient assumed)

  4. Bias between paired SpO2 related to sensor pairs

    Bias between paired SpO2 (in %) in three ranges related to independent variable: sensor pairs (as factor).

    Measured: 1 month (250 observations is expected from each participating center; on average 2 patients and 5 observations per day per patient assumed)

  5. Bias between paired SpO2 related to oximeter averaging setting

    Bias between paired SpO2 (in %) in three ranges related to independent variable: oximeter averaging setting (as factor).

    Measured: 1 month (250 observations is expected from each participating center; on average 2 patients and 5 observations per day per patient assumed)

  6. Bias between paired SpO2 related to skin shade

    Bias between paired SpO2 (in %) in three ranges related to independent variable: skin shade (as factor).

    Measured: 1 month (250 observations is expected from each participating center; on average 2 patients and 5 observations per day per patient assumed)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted10 Feb 2025
  2. Study start15 Mar 2025 (estimated)
  3. Primary completion31 Aug 2025 (estimated)
  4. Study completion31 Oct 2025 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
40 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06817850
Sponsor's protocol ID
NOB (PRICOIII)
Acronym
NOB
Status last verified
Mar 2025
First submitted
21 Jan 2025
Last updated
25 Mar 2025
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

An observational study of Neonatal Respiratory Distress and Oxygenation, run by Czech Technical University in Prague at 1 site in Czechia.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 25 Mar 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Czechia. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Jakub Rafl, PhD (+420728229991, rafl@fbmi.cvut.cz), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Bias between paired SpO2 measurements (time frame: 1 month (250 observations is expected from each participating center; on average 2 patients and 5 observations per day per patient assumed)). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 25 Mar 2025.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in