Hypoxemic Acute Respiratory FailureCLOC-INNCT06829732

Closed Loop Oxygen Control in INtubated Critically Ill Patients

A study testing 2. manual oxygen titration (MOT) in people with Hypoxemic Acute Respiratory Failure, run by Parc Taulí Hospital Universitari at 1 site in Spain.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

The deleterious effects of alveolar hyperoxia and (severe) hyperoxaemia are well described. Achieving safe and efficient oxygenation may be challenging. The hypothesis of the present study is that closed-loop oxygen control in intubated and mechanically ventilated critically ill patients improves oxygen administration… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Hypoxemic Acute Respiratory Failure
What is being tested?
2. manual oxygen titration (MOT)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
46 participants
Where
1 siteSpain
Study dates
1 Apr 2023to Mar 2025
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patients admitted to the ICU who are expected to require invasive mechanical ventilation for at least 8 hours and are expected to stay respiratory stable´ in the upcoming 8 hours.
  • Acute respiratory failure with PaO2/FiO2 <300 with FiO2 ≥ 0.4 at inclusion
  • Age older than 18 years old.
  • Written informed consent signed and dated by the patient or one relative in case that the patient is unable to consent, after full explanation of the study by the investigator and prior to study participation.

Who cannot join

  • Low quality on the SpO2 measurement using finger and ear sensor (quality index <60%).
  • Hemodynamic instability defined as a need of continuous infusion of epinephrine or norepinephrine > 1 mg/h.
  • Severe acidosis (pH ≤ 7.30).
  • Chronic or acute dyshemoglobinemia: methemoglobin, CO poisoning, sickle cell disease.
  • Patient under guardianship or deprived of liberties.
  • Impossibility to give informed consent by both patient and family (i.e. language barrier).
  • Patient included in another interventional research study under consent with similar outcome.

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Spain1 site, 1 recruiting
  • Parc Tauli Hospital Universitary

    Sabadell, Barcelona, 08028

    Marina García-de-Acilu, PhD, +34661739188, mggacilu@gmail.com

    Recruiting

Other trials in Spain

About this study

Official title: Closed Loop Oxygen Control in INtubated Critically Ill Patients - CLOC-IN

The deleterious effects of alveolar hyperoxia and (severe) hyperoxaemia are well described. Achieving safe and efficient oxygenation may be challenging. The hypothesis of the present study is that closed-loop oxygen control in intubated and mechanically ventilated critically ill patients improves oxygen administration compared with standard manual oxygen titrations. A single-blind, randomised crossover clinical trial assessing the efficacy and safety of the use of a closed-loop oxygen control versus manual oxygen titration in patients receiving mechanical ventilation has been designed. Patients will be randomised to receive first either closed-loop oxygen control (CLOC) or manual oxygen titration (MOT). The percentage of time spent in optimal and sub-optimal SpO2 (oxygen saturation by pulse oximetry) ranges will be calculated for each period.

Read the detailed description

The deleterious effects of alveolar hyperoxia and (severe) hyperoxaemia are well described. A safe and efficient oxygenation should be achieved. However, from a practical point of view, this may require frequent manual adjustments of the inspired oxygen, which might be unfeasible, especially in high-demand periods. Several automatic oxygen systems have been tested in non-intubated patients. The hypothesis of the present study is that closed-loop oxygen control in intubated and mechanically ventilated critically ill patients improves oxygen administration compared with standard manual oxygen titrations. Thus, the objective of the study is to compare the percentage of time spent in a predefined optimal SpO2 range between closed-loop oxygen control and manual oxygen titration.

A single-blind, randomised crossover clinical trial assessing the efficacy and safety of the use of a closed-loop oxygen control versus manual oxygen titration in patients receiving mechanical ventilation has been designed. Once included, patients will be randomised to receive first either closed-loop oxygen control (CLOC) or manual oxygen titration (MOT). Every period will last for 4h. Clinical and respiratory variables will be recorded. A memory box will be connected to the patient's ventilator. This device will record ventilator parameters, including FiO2 and SpO2. To assess the outcome, the percentage of time spent in optimal and sub-optimal SpO2 ranges will be calculated for each period.

Being tested
closed-loop oxygen control (CLOC), 2. manual oxygen titration (MOT)
Medical subject headings
Respiratory Insufficiency

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • closed-loop oxygen control (CLOC)Device

    HAMILTON-C6 (Hamilton Medical, Bonaduz, Switzerland) ventilators will be used. It allows for O2 Assist software, that will be installed on all C6 ventilators before the start of this study. Automatic Oxygen management sets the Oxygen control according to the patient's measured SpO2 and the operator-set SpO2 target range. The Oxygenation controller will be activated for 4h and adjusts the Oxygen according to patient's measured SpO2.

  • 2. manual oxygen titration (MOT)Device

    O2 will be manually set according to SpO2 target during a 4h period.

Study groups in NCT06829732
Study groupRoleWhat they receive
1. closed-loop oxygen control (CLOC)ExperimentalAutomatic Oxygen management sets the Oxygen control according to the patient's measured SpO2 and the operator-set SpO2 target range during 4h.
Receives: Device: closed-loop oxygen control (CLOC)
2. manual oxygen titration (MOT)Active comparatorConventional oxygen manual management according to SpO2 target range set by clinician, during 4h.
Receives: Device: 2. manual oxygen titration (MOT)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Efficacy outcome:

    To compare the percentage of time spent in a predefined optimal SpO2 (oxygen saturation by pulse oximetry) range between closed-loop oxygen control and manual oxygen titration

    Measured: Hour 1 to 4

Secondary outcomes (13)
  1. Percentage of time with SpO2 (oxygen saturation by pulse oximetry) in sub-optimal ranges

    To compare the percentage of time with SpO2 in sub-optimal ranges, above and below the optimal range, during closed-loop oxygen control and manual oxygen titration

    Measured: Hour 1 to 4

  2. Percentage of time with SpO2 (oxygen saturation by pulse oximetry) out of range

    To compare the percentage of time with SpO2 out of range, above and below the optimal and sub-optimal range.

    Measured: Hour 1 to 4

  3. Number of manual adjustments

    To compare the number of manual adjustments

    Measured: Hour 1 to 4

  4. Mean SpO2/FiO2 (oxygen saturation by pulse oximetry to fraction of inspired oxygen ratio) between the two methods

    To compare the mean SpO2/FiO2 between the two methods of oxygen titration

    Measured: Hour 1 to 4

  5. Mean FiO2 (fraction of inspired oxygen) between the two methods of oxygen titration

    To compare the mean FiO2 between the two methods of oxygen titration

    Measured: Hour 1 to 4

  6. Time with FiO2 (fraction of inspired oxygen) below 40%

    To compare the percentage of time with FiO2 below 40%

    Measured: Hour 1 to 4

  7. Time with FiO2 (fraction of inspired oxygen) above 60%

    To compare the percentage of time with FiO2 above 60%

    Measured: Hour 1 to 4

  8. Time with FiO2 (fraction of inspired oxygen) at 100%

    To compare the percentage of time with FiO2 at 100%

    Measured: Hour 1 to 4

  9. SpO2 (oxygen saturation by pulse oximetry) below 88 and 85%.

    To compare the percentage of time with SpO2 below 88 and 85%.

    Measured: Hour 1 to 4

  10. SpO2 (oxygen saturation by pulse oximetry) below 88 and 85% (Event > 10s and 60 s)

    To compare the number of events with SpO2 below 88 and 85% (Event > 10s and 60 s)

    Measured: Hour 1 to 4

  11. Number of alarms

    To compare the number of alarms

    Measured: Hour 1 to 4

  12. Time with SpO2 (oxygen saturation by pulse oximetry) signal available.

    To compare the percentage of time with SpO2 signal available.

    Measured: Hour 1 to 4

  13. Total amount of oxygen used

    To compare the total amount of oxygen used (in L)

    Measured: Hour 1 to 4

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted17 Feb 2025
  2. Study start1 Apr 2023
  3. Primary completionMar 2025 (estimated)
  4. Study completionMar 2025 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Supportive care
Allocation
Randomised
Intervention model
Crossover
Model description
Every patient included will be randomize to recieved both arm in different order. Every period will last for 4 h.
Masking
None (open label)
Enrolment
46 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06829732
Sponsor's protocol ID
PT_UCI_RESP1
Acronym
CLOC-IN
Status last verified
Jan 2025
First submitted
15 Jan 2024
Last updated
17 Feb 2025
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study testing 2. manual oxygen titration (MOT) in people with Hypoxemic Acute Respiratory Failure, run by Parc Taulí Hospital Universitari at 1 site in Spain.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 17 Feb 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Spain. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Each site lists its own contact: choose the nearest one under Trial locations. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Efficacy outcome: (time frame: Hour 1 to 4). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 Feb 2025.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in