AzTrials
AzTrials

Pulse OximetryNCT06912516

PULSE OXIMETRY Performance Test for a Software Medical Device 'Vitacam'

A study of Pulse Oximetry, run by Evondos Oy at 1 site in Finland.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This clinical investigation aims to collect data for algorithm development and testing of Vitacam as well as to later validate the system using a simultaneously collected, but segregated, hold-out validation dataset. Vitacam is a software medical device for measuring vital signs, manufactured by NE Device SW Oy. Read more about this study

Who is it for?
Ages 18 to 40 yearsHealthy volunteers can take part
Conditions the study is for
Pulse Oximetry
Type of treatment
Surgery or procedure
Could I receive a placebo?
No, everyone receives an active treatment
How many people
24 participants
Where
1 siteFinland
Study dates
22 Dec 2023to 1 Nov 2024
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 40 years
  • All sexes
  • Healthy volunteers accepted
  • Adult

Who may be able to join

  • Age: 18-40 years
  • ASA category 1
  • Positive Allen's test
  • Passing the physical examination
  • Ability to give informed consent
  • Volunteering for the study
  • Fluent Finnish or English

Who cannot join

  • Age under 18 years or over 40 years
  • Smoker
  • Pregnancy
  • Does not pass the physical examination
  • Negative Allen's test
  • Relevant medical illnesses
Show all 10Show fewer
  • Cardiovascular disease
  • Pulmonary disease
  • Inability to give informed consent
  • Denial

Trial locations and contacts

1 site in 1 country.

Finland1 site
  • Tampere University

    Tampere, Finland, FI-33014

Other trials in Finland

About this study

Official title: PULSE OXIMETRY Performance Test for Vitacam. Controlled Desaturation Test.

This clinical investigation aims to collect data for algorithm development and testing of Vitacam as well as to later validate the system using a simultaneously collected, but segregated, hold-out validation dataset. Vitacam is a software medical device for measuring vital signs, manufactured by NE Device SW Oy. It analyses digital video recorded using a smart device with a camera to locate a subject's face and chest region. From both these regions of interest, the video is further analysed to obtain the heart rate, heartbeat intervals, respiratory rate and oxygen saturation. These measurements are used as inputs to clinical decision making. Vitacam uses remote photoplethysmography to obtain the heart rate and heartbeat intervals, and will build on this to obtain oxygen saturation using information from red-green-blue channels.

The main focus of this study is to validate the oxygen saturation, heart rate and heartbeat intervals measured by Vitacam against gold standard references, applying the ISO 80601:2019 standard.

Conditions
Pulse Oximetry
Being tested
Controlled desaturation
Keywords
Oxygen saturation, remote photoplethysmography, pulse oximetry

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Controlled desaturationProcedure or surgery

    A subject rests comfortably on a hospital bed, so that they do not have to move during the desaturation process.

    The subject wears the altitude simulator head bivy and breathes freely without any oxygen regulation. During the measurement, the time in low O2 saturation plateaus is limited to ten minutes in each desaturation plateau.

    They are monitored with ECG, clinical pulse oximeter, EtCO2 from non-invasive capnography with nasal cannulas and periodically non-invasive blood pressure. The monitoring is supervised by two medical doctors, present during the entire duration of the study.

Study groups in NCT06912516
Study groupRoleWhat they receive
Desaturation subjectsOtherThis is a single arm study where healthy volunteers were subjected to a controlled desaturation protocol to ascertain if the investigational device could correctly track changes in oxygen saturation.
Receives: Procedure: Controlled desaturation

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Accuracy of oxygen saturation measurements - Root Mean Square Error

    Accuracy as measured by Root Mean Square Error should be less than or equal to 4% for SpO2 as per ISO 80601-2-61:2019. It is calculated from the difference of the SpO2 measured from the investigational device and that of the reference device.

    Measured: A 30-45 minute period during which the participant is subjected to the de-saturation process

  2. Accuracy of oxygen saturation measurements - Mean absolute error

    Accuracy as measured by Mean Absolute Error should be less than or equal to 4% for SpO2 as per ISO 80601-2-61:2019. It is calculated from the difference of the SpO2 measured from the investigational device and that of the reference device.

    Measured: A 30-45 minute period during which the participant is subjected to the de-saturation process

Secondary outcomes (1)
  1. Accuracy of interbeat intervals of the investigational device against successive R waves of the QRS complex from an Electrocardiogram

    To validate the accuracy of Vitacam interbeat intervals, both the intervals and heart rate variability metrics derived from these intervals, will be compared to the RR intervals and heart rate variability metrics derived from a 5-lead ECG

    Measured: A 30-45 minute period during which the participant is subjected to the de-saturation process

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Apr 2025
  2. Study start22 Dec 2023
  3. Primary completion15 Sep 2024
  4. Study completion1 Nov 2024

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
24 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06912516
Sponsor's protocol ID
R23022L
Status last verified
Mar 2025
First submitted
25 Mar 2025
Last updated
4 Apr 2025
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No: IPD cannot be shared due to GDPR

Questions about this trial

What is this clinical trial studying?

A study of Pulse Oximetry, run by Evondos Oy at 1 site in Finland.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 10 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 4 Apr 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Finland. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Accuracy of oxygen saturation measurements - Root Mean Square Error (time frame: A 30-45 minute period during which the participant is subjected to the de-saturation process). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 4 Apr 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in