HemodialysisNCT06950944
Efficacy and Safety of Repeated-use Middle Cut Off Membrane Dialyzers in Hemodialysis Patients
A study testing Theranova 500 in people with Hemodialysis and Hemodialysis Treatment, run by Khon Kaen University at 1 site in Thailand.
Trial summary
The goal of this clinical trial is to compare the efficacy of uremic toxin and inflammatory cytokine clearance by using reprocessing middle cutoff membrane (MCO) dialyzer in chronic hemodialysis patients. Read more about this study
- Who is it for?
- Ages 18 to 80 yearsHealthy volunteers can take part
- Conditions the study is for
- Hemodialysis, Hemodialysis Treatment, Dialysis Adequacy
- What is being tested?
- Theranova 500
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 48 participants
- Where
- 1 siteThailand
- Study dates
- 1 Jan 2024to 30 Apr 2024
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 80 years
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- chronic hemodialysis with dialysis vintage more than 3 months
- hemodialysis via permanent vascular access (fistula or graft)
- adequate hemodialysis define with spKt/V > 1.8 and > 1.2 per session in twice and thrice weekly hemodialysis, respectively
Who cannot join
- preganancy
- hemodialysis vascular access dysfunction
- blood transmitted infection
- advanced malignancy or severe morbidity with life expectancy less than 1 year
- severe dialyzer reaction
- receive immunosuppressive drug
Trial locations and contacts
1 site in 1 country.
Thailand1 site
Faculty of Medicine, Khon Kaen University
Khon Kaen, Changwat Khon Kaen, 40002
About this study
Official title: Efficacy and Safety of Repeated-use Middle Cut Off Membrane Dialyzers on Middle Molecule Uremic Toxin and Inflammatory Cytokine Clearances in Hemodialysis Patients
The goal of this clinical trial is to compare the efficacy of uremic toxin and inflammatory cytokine clearance by using reprocessing middle cutoff membrane (MCO) dialyzer in chronic hemodialysis patients. The main questions it aims to answer are:
- To evaluate efficacy of uremic toxin and inflammatory cytokine clearance and safty of hemodialysis with reprocessing MCO dialyzer compared with new MCO dialyzer
- To compare efficacy of uremic toxin and inflammatory cytokine clearance between 2 types of MCO dialyzer which availble in Thailand (Theranova 500 and Elisio 21Hx).
Participants will randomized to 2 types of MCO groups after that they will receive hemodialysis by using study dialyzer with similar hemodialysis prescription. The dialyzer in this study were reprocessing by standard hospital protocol. Researchers will compare 2 types of dialyzer groups to see efficacy of uremic toxin and inflammatory cytokine clearance and safety of use reprocessing dialyzer.
Read the detailed description
Study design and population The single - center prospective open-label randomize control trial was conducted in prevalence HD patients at the Srinagarind Hospital, Khon Kaen University between January 2024 and March 2024. The inclusion criteria were participants aged 18-80 years old, with adequate HD (spKt/V > 1.2 and 1.8 per session for patients undergoing thrice- and twice-weekly HD, respectively) for more than 3 months, utilizing arteriovenous fistula or graft (AVF or AVG) for vascular access. The exclusion criteria were pregnancy, vascular access dysfunction, contraindicated to reprocess dialyzer (e.g., blood-transmitted infections such as HBV, HIV, and MDR pathogens), advanced-stage cancer or cardiovascular diseases, severe comorbidities with less than a 1-year life expectancy, prior severe infections or hospitalization within 3 months or receiving immunosuppressive drugs, and history of severe dialyzer allergy.
Study Intervention All enrolled 48 participants were randomized to hemodialysis using Theranova 500 dialyzer or Elisio 21HX dialyzer stratified by age, sex, causes of ESRD, dialysis vintage, dialysis sessions per week and type of vascular accesses.Each patient received twice and thrice weekly 4-hour dialysis sessions with their usual prescriptions including blood and dialysate flow rates, dialysate composition, intradialytic coagulant and dry weight. Both groups utilizing the Fresenius 4008H machine for dialysis treatment, with the study concluding when complete of 15th dialysis session or study dialyzer was disqualified for reused adhering to the standard dialyzer reprocessing protocol in Srinagarind hospital. Intradialytic monitoring and treatment of intradialytic complications were performed by trained HD nurses. The ultrapure quality dialysis water defined by total microbial counts < 0.1 CFU/mL and endotoxin concentrations < 0.03 EU/mL was used in both groups. Both study dialyzers underwent disinfection with 2% peracetic acid and reprocessed using an automatic dialyzer reprocessing machine. In cases where abnormalities or blood clots were detected by the automatic machine, trained staff conducted a reassessment and rinsed the dialyzer with reverse osmosis water until the clot dissipated. The dialyzers exhibiting a reduction TCV exceeding 20% or unable to clear blood clots were deemed unsuitable for reprocessing and were disposed.
Data collection and definition of outcomes Baseline characteristics and demographic data, e.g., age, sex, weight, height, body mass index, comorbid diseases, causes of ESKD, physical status, and socioeconomic data were collected from the interview, physical examination, and review of medical records. Baseline laboratory data were collected 2-week prior to the study period. The efficacy of various-size uremic toxins removal including phosphorus, urea, creatinine, homocysteine, PTH, β2-MG, κ-FLC and λ-FLC together with inflammatory markers including IL-6 and C-reactive protein (CRP) were evaluated by pre- to post-dialysis RRs at 1st, 2nd, 5th, 10th, 15th session of dialysis intervention. The RR was calculated by the following equations according to Bergström and Wehle. Techniques for measurement of serum solute levels composed of molecular absorption spectrometry (for urea and creatinine), immunoturbidimetric assay (for β2-MG and CRP), electrochemiluminescence immunoassay (for IL-6), enzymatic assay (for homocysteine), and latex particle enhance immunoturbidimetric method (for κ-FLC and λ-FLC). Regarding of safety outcomes, data of dialyzer related complications including dialyzer reaction, rupture of dialyzer membranes, and intradialytic dialyzer clot were collected.
- Conditions
- Hemodialysis, Hemodialysis Treatment, Dialysis Adequacy
- Being tested
- Theranova 500
- Keywords
- Middle cutoff membrane dialyzer, Uremic toxin clearance, Reused membrane dialyzer
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Theranova 500Device
Participants will dialysis by Theranova 500 dialyzer during study period, the dialyzer will reprocessing with standard hospital protocol and dialyzer will check before use in dialysis session.
| Study group | Role | What they receive |
|---|---|---|
| Theranova 500 | Experimental | The participant underwent dialysis using the Theranova 500 dialyzer Receives: Device: Theranova 500 |
| Elisio 21Hx | Active comparator | The participant underwent dialysis using the Elisio 21Hx dialyzer Receives: Device: Theranova 500 |
What is being measured
The primary outcome is the main question the study is designed to answer.
Efficacy of uremic toxin clearances from repeated-use middle cut off membrane dialyzers
The efficacy of various-size uremic toxins removal including phosphorus, urea, creatinine, homocysteine, PTH, β2-MG, κ-FLC and λ-FLC together with inflammatory markers including IL-6 and C-reactive protein (CRP) were evaluated by pre- to post-dialysis reduction ratio (RRs) at 1st, 2nd, 5th, 10th, 15th session of dialysis intervention. The RR was calculated by the following equations according to Bergström and Wehle.
Measured: up to 3 months or 12 weeks
Secondary outcomes (2)
Albumin loss from hemodialysis
The loss of albumin from dialysis was calculated from area under the dialysate albumin concentration-time curve (AUC) of measurement of albumin quantity in the dialysis fluid at initial, 60, 120 min and at the end of dialysis sessions.
Measured: up to 3 months or 12 weeks
Adveres event from reused middle cutoff membrane dialyzer
Data of dialyzer related complications including, intradialytic hypotension (nadir systolic blood pressure less than 90 mmHG or drop from predialysis more than 40 mmHg), dialyzer reaction, rupture of dialyzer membranes, and intradialytic dialyzer clot were collected.
Measured: up to 3 months or 12 weeks
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted30 Apr 2025
- Study start1 Jan 2024
- Primary completion31 Mar 2024
- Study completion30 Apr 2024
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for HemodialysisStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Model description
- Randomized controlled study
- Masking
- Single blind
- Who is masked
- Participant
- Enrolment
- 48 participants (actual)
Sponsor and investigators
- Lead sponsor
- Khon Kaen University
- Sponsor type
- University, hospital or other
- Responsible party
- Principal investigator, Eakalak Lukkanalikitkul, Khon Kaen University
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT06950944
- Sponsor's protocol ID
- HE661417
- Status last verified
- Apr 2025
- First submitted
- 6 Feb 2024
- Last updated
- 4 May 2025
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- No
Study at a glance
From the registry record, last updated by the sponsor on 4 May 2025. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Efficacy and Safety of Repeated-use Middle Cut Off Membrane Dialyzers on Middle Molecule Uremic Toxin and Inflammatory Cytokine Clearances in Hemodialysis Patients
- ClinicalTrials.gov ID
- NCT06950944
- Acronym / sponsor ID
- HE661417
- Conditions
- Hemodialysis, Hemodialysis Treatment, Dialysis Adequacy
- Keywords
- Middle cutoff membrane dialyzer, Uremic toxin clearance, Reused membrane dialyzer
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 48 participants (actual)
- Lead sponsor
- Khon Kaen UniversityUniversity, hospital or other
- Study start
- 1 Jan 2024 (actual)
- Primary completion
- 31 Mar 2024 (actual)
- Study completion
- 30 Apr 2024 (actual)
- First posted
- 30 Apr 2025 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The goal of this clinical trial is to compare the efficacy of uremic toxin and inflammatory cytokine clearance by using reprocessing middle cutoff membrane (MCO) dialyzer in chronic hemodialysis patients. The main questions it aims to answer are:
- To evaluate efficacy of uremic toxin and inflammatory cytokine clearance and safty of hemodialysis with reprocessing MCO dialyzer compared with new MCO dialyzer
- To compare efficacy of uremic toxin and inflammatory cytokine clearance between 2 types of MCO dialyzer which availble in Thailand (Theranova 500 and Elisio 21Hx).
Participants will randomized to 2 types of MCO groups after that they will receive hemodialysis by using study dialyzer with similar hemodialysis prescription. The dialyzer in this study were reprocessing by standard hospital protocol. Researchers will compare 2 types of dialyzer groups to see efficacy of uremic toxin and inflammatory cytokine clearance and safety of use reprocessing dialyzer.
Detailed description
Study design and population The single - center prospective open-label randomize control trial was conducted in prevalence HD patients at the Srinagarind Hospital, Khon Kaen University between January 2024 and March 2024. The inclusion criteria were participants aged 18-80 years old, with adequate HD (spKt/V > 1.2 and 1.8 per session for patients undergoing thrice- and twice-weekly HD, respectively) for more than 3 months, utilizing arteriovenous fistula or graft (AVF or AVG) for vascular access. The exclusion criteria were pregnancy, vascular access dysfunction, contraindicated to reprocess dialyzer (e.g., blood-transmitted infections such as HBV, HIV, and MDR pathogens), advanced-stage cancer or cardiovascular diseases, severe comorbidities with less than a 1-year life expectancy, prior severe infections or hospitalization within 3 months or receiving immunosuppressive drugs, and history of severe dialyzer allergy.
Study Intervention All enrolled 48 participants were randomized to hemodialysis using Theranova 500 dialyzer or Elisio 21HX dialyzer stratified by age, sex, causes of ESRD, dialysis vintage, dialysis sessions per week and type of vascular accesses.Each patient received twice and thrice weekly 4-hour dialysis sessions with their usual prescriptions including blood and dialysate flow rates, dialysate composition, intradialytic coagulant and dry weight. Both groups utilizing the Fresenius 4008H machine for dialysis treatment, with the study concluding when complete of 15th dialysis session or study dialyzer was disqualified for reused adhering to the standard dialyzer reprocessing protocol in Srinagarind hospital. Intradialytic monitoring and treatment of intradialytic complications were performed by trained HD nurses. The ultrapure quality dialysis water defined by total microbial counts < 0.1 CFU/mL and endotoxin concentrations < 0.03 EU/mL was used in both groups. Both study dialyzers underwent disinfection with 2% peracetic acid and reprocessed using an automatic dialyzer reprocessing machine. In cases where abnormalities or blood clots were detected by the automatic machine, trained staff conducted a reassessment and rinsed the dialyzer with reverse osmosis water until the clot dissipated. The dialyzers exhibiting a reduction TCV exceeding 20% or unable to clear blood clots were deemed unsuitable for reprocessing and were disposed.
Data collection and definition of outcomes Baseline characteristics and demographic data, e.g., age, sex, weight, height, body mass index, comorbid diseases, causes of ESKD, physical status, and socioeconomic data were collected from the interview, physical examination, and review of medical records. Baseline laboratory data were collected 2-week prior to the study period. The efficacy of various-size uremic toxins removal including phosphorus, urea, creatinine, homocysteine, PTH, β2-MG, κ-FLC and λ-FLC together with inflammatory markers including IL-6 and C-reactive protein (CRP) were evaluated by pre- to post-dialysis RRs at 1st, 2nd, 5th, 10th, 15th session of dialysis intervention. The RR was calculated by the following equations according to Bergström and Wehle. Techniques for measurement of serum solute levels composed of molecular absorption spectrometry (for urea and creatinine), immunoturbidimetric assay (for β2-MG and CRP), electrochemiluminescence immunoassay (for IL-6), enzymatic assay (for homocysteine), and latex particle enhance immunoturbidimetric method (for κ-FLC and λ-FLC). Regarding of safety outcomes, data of dialyzer related complications including dialyzer reaction, rupture of dialyzer membranes, and intradialytic dialyzer clot were collected.
Study design
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- Single blind (participant)
- Number of arms
- 2
- Model description
- Randomized controlled study
| Arm | Type | Receives |
|---|---|---|
| Theranova 500 | Experimental | Device: Theranova 500The participant underwent dialysis using the Theranova 500 dialyzer |
| Elisio 21Hx | Active comparator | Device: Theranova 500The participant underwent dialysis using the Elisio 21Hx dialyzer |
Who may qualify
- Ages
- 18 to 80 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (3 listed)
- chronic hemodialysis with dialysis vintage more than 3 months
- hemodialysis via permanent vascular access (fistula or graft)
- adequate hemodialysis define with spKt/V > 1.8 and > 1.2 per session in twice and thrice weekly hemodialysis, respectively
Key exclusion criteria (6 listed)
- preganancy
- hemodialysis vascular access dysfunction
- blood transmitted infection
- advanced malignancy or severe morbidity with life expectancy less than 1 year
- severe dialyzer reaction
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Theranova 500Participants will dialysis by Theranova 500 dialyzer during study period, the dialyzer will reprocessing with standard hospital protocol and dialyzer will check before use in dialysis session. | Device | Elisio 21Hx, Theranova 500 |
Primary outcome
Efficacy of uremic toxin clearances from repeated-use middle cut off membrane dialyzers
Time frame: up to 3 months or 12 weeks
The efficacy of various-size uremic toxins removal including phosphorus, urea, creatinine, homocysteine, PTH, β2-MG, κ-FLC and λ-FLC together with inflammatory markers including IL-6 and C-reactive protein (CRP) were evaluated by pre- to post-dialysis reduction ratio (RRs) at 1st, 2nd, 5th, 10th, 15th session of dialysis intervention. The RR was calculated by the following equations according to Bergström and Wehle.
2 secondary outcomes
Albumin loss from hemodialysis
Time frame: up to 3 months or 12 weeks
Adveres event from reused middle cutoff membrane dialyzer
Time frame: up to 3 months or 12 weeks
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 4 May 2025.
- Individual participant dataSharing planned: No
Related trials for Hemodialysis
Active, not recruitingPhase 2
Safety and Efficacy of Empagliflozin in Hemodialysis
End Stage Renal Disease
Not yet recruiting
Manual Acupuncture for Intradialytic Nausea and Vomiting in Patients Undergoing Hemodialysis
Intradialytic Nausea and Vomiting, Chronic Kidney Disease
RecruitingPhase 1
A Study to Test the Safety of the Study Drug Fosmanogepix and How the Body Handles It in Adults With Severe Kidney Failure on Dialysis
End Stage Renal Disease on Dialysis, Adult
Not yet recruitingPhase 1/2
Bupropion for Fatigue in End-stage Kidney Disease Patients on Hemodialysis
End Stage Renal Disease, Fatigue
Active, not recruiting
Group Exercise Program Supported By Audiovisual Media During Hemodialysis
Chronic Kidney Failure, Hemodialysis
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Theranova 500 in people with Hemodialysis and Hemodialysis Treatment, run by Khon Kaen University at 1 site in Thailand.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 4 May 2025. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Thailand. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Efficacy of uremic toxin clearances from repeated-use middle cut off membrane dialyzers (time frame: up to 3 months or 12 weeks). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 4 May 2025.