Preterm InfantBubble CPAPNCT07060833

Acoustics as a Metric of Airway Pressure in Premature Infants Using Bubble Continuous Positive Airway Pressure

An observational study of Preterm Infant, run by McGill University Health Centre/Research Institute of the McGill University Health Centre at 1 site in Canada.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

This is an observational, proof-of-concept, feasibility study where 30 preterm infants on bubble CPAP with gestational age < 32+0 weeks will be recruited from the neonatal intensive care unit (NICU) at the Montreal Children's Hospital. The study's main goals are: 1. Read more about this study

Who is it for?
Ages 28 to 32 weeks
Conditions the study is for
Preterm Infant
What is being tested?
External pressure transducer, Standard microphone and Wireless acoustic sensor
How many people
30 participants
Where
1 siteCanada
Study dates
3 Apr 2025to Dec 2026
Phase
Observational
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 28 to 32 weeks
  • All sexes
  • No healthy volunteers
  • Child

Study population: Preterm infants admitted to the neonatal intensive care unit at the Montreal Children's Hospital.

Who may be able to join

  • Infants on bCPAP with gestational age < 32+0 weeks
  • Postmenstrual age between 28+0 and 36+6 weeks at the time of the study
  • Postnatal age greater than 168 hours (7 days) at the time of the study
  • On the bubble CPAP device with binasal prongs at the time of the study
  • Receiving bubble CPAP levels of 5 to 7 cm H2O with gas flows between 6L/min and 10L/min at the time of the study

Who cannot join

  • Infants with known major congenital anomalies
  • Infants with known congenital heart disorders
  • Infants with known neuromuscular disease
  • Infants receiving ventilator-derived CPAP at the time of the study
  • Infants receiving CPAP via a nasal mask interface at the time of the study
  • Infants receiving inotropes, narcotics or sedative agents at the time of the study
  • Infants deemed clinically unstable for the study by the attending neonatologist

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Canada1 site, 1 recruiting
  • McGill University Health Center

    Montreal, Quebec, H4A 3J1

    Wissam Shalish, MD, PhD, 514-412-4400 ext. 22341, wissam.shalish@mcgill.ca

    Wissam Shalish, MD PhD

    Recruiting

Other trials in Canada

About this study

Official title: The Use of Bubble Continuous Positive Airway Pressure in Premature Infants: Acoustics as a Metric of Effective Pressure Delivery

This is an observational, proof-of-concept, feasibility study where 30 preterm infants on bubble CPAP with gestational age < 32+0 weeks will be recruited from the neonatal intensive care unit (NICU) at the Montreal Children's Hospital.

The study's main goals are:

  1. To determine the relationship between ambient bubbling sounds and delivered pressures in preterm infants on bCPAP.
  2. To determine the relationship between transmitted bubbling sounds and airway pressures transmitted to the lungs of preterm infants on bCPAP.
  3. To develop models to predict delivered and transmitted bCPAP pressures from the acoustic properties of bubbling sounds.
Read the detailed description

Continuous positive airway pressure (CPAP) is an essential, non-invasive therapy for treating various respiratory conditions in the Neonatal Intensive Care Units (NICU). CPAP is an effective treatment for respiratory distress syndrome, apneas, or after extubation, exerting its physiological benefits by maintaining upper airway patency and functional residual capacity. Bubble CPAP (bCPAP) is the most widely used CPAP due to its low cost and ease of use. It consists of an inspiratory tube carrying heated and humidified air, a nasal interface, and an expiratory tube immersed in a water chamber. The generation of bubbles in the water chamber by exhaled gas creates low amplitude and high-frequency pressure oscillations that are transmitted back to the chest. Successful CPAP requires constant transmission of the pressure via an unobstructed circuit. However, this is difficult to achieve in practice due to inadequate interface, leaks from an open mouth, and obstructed airway. As a result, bCPAP requires frequent manual checks by nurses and respiratory therapists to ensure that the circuit is secure and unobstructed.

As a proposed solution, bCPAP sounds heard in the patient room or upon auscultation are routinely used to assess the effectiveness of CPAP therapy. This sound can be heard both from the water tank creating the vibrations and during auscultation with a stethoscope, as the sound vibration is transmitted to the neonatal lungs. In the current era of digital technology, acoustic sounds can be converted to electronic signals for further processing and analysis.

We hypothesize that continuous recording and analysis of bCPAP sounds could be used as a proxy for real-time objective monitoring of the pressure transmitted to infants' lungs.

Conditions
Preterm Infant
Being tested
External pressure transducer, Standard microphone, Wireless acoustic sensor, Internal pressure transducer
Keywords
Bubble CPAP, Airflow, Neonatal Intensive Care, Respiratory acoustics, Acoustic monitoring, Airway pressure monitoring, Bubbling sounds
Medical subject headings
Premature Birth

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • External pressure transducerDevice

    Also called External catheter pressure transducer (Millar Mikro-tip ® SPR-524, 3.5F Single straight, Texas)

    The delivered CPAP pressure will be measured using an ultra-thin, multi-use catheter pressure transducer inserted into a port in the expiratory limb of the bubble CPAP circuit.

  • Standard microphoneDevice

    Also called Lavalier condenser microphone, Tosuny

    The bubble sound of the water tank will be collected with a standard condenser microphone directly affixed to the pole holding the water tank, with a secure clip.

  • Wireless acoustic sensorDevice

    The wireless acoustic sensor contains a dual microphone capable of capturing target sounds as well as ambient noise. The frequencies associated with ambient noise will be subtracted to maximize the signal-to-noise ratio of the bubble sound waveform. The wireless sensor will be placed on the suprasternal notch of the infant for monitoring the bubble sounds transmitted to the lungs and secured using a silicone-based tape approved for use in neonates. Data will be transmitted in real-time to a research-dedicated tablet using the Bluetooth Communication Controller and stored for future analysis.

  • Internal pressure transducerDevice

    The transmitted CPAP pressure will be measured using an ultra-thin, single-use catheter pressure transducer inserted through the mouth to the level of the infant's nasopharynx. The data will be acquired with a sampling rate of 10kHz and stored for later analysis.

Study groups in NCT07060833
Study groupRoleWhat they receive
Phase I: 15 preterm infants on Bubble Continuous Positive Airway Pressure (bCPAP)Group 1 will consist of 15 preterm infants on bCPAP in whom the external bubble CPAP sounds and the pressures from the CPAP system itself will be recorded for 3 hours. To do so, we will use a standard microphone placed on the water tank and an external pressure transducer inserted into the expiratory limb of the bubble CPAP circuit.
Receives: Device: External pressure transducer; Device: Standard microphone
Phase II: 15 preterm infants on Bubble Continuous Positive Airway Pressure (bCPAP)Group 2 will consist of 15 preterm infants on bCPAP in whom the external bubble CPAP sounds and the pressures from the CPAP system itself will be recorded. To do so, we will use a standard microphone placed on the water tank and a pressure transducer inserted into the expiratory limb of the bubble CPAP circuit. In addition, the internal bubble CPAP sound and pressure transmitted to the infant's lungs will be measured. For these, a wireless acoustic sensor will be secured on the infant's suprasternal notch to capture transmitted CPAP sounds, and an internal pressure transducer will be inserted from the infant's mouth into the nasopharynx to record the pressure delivered in the airway. Data will be recorded for 3 hours.
Receives: Device: External pressure transducer; Device: Standard microphone; Device: Wireless acoustic sensor; Device: Internal pressure transducer

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Pressure

    The mean pressure and the standard deviation of the pressure will be computed for each segment.

    Measured: 3 hours

  2. External bubble CPAP sounds

    Two metrics will be computed:

    1. The root mean square (RMS)
    2. The power contained between a pre-determined range within each segment; We will determine the range that contains 80% of the signal power, in order to minimize other noises from bubble CPAP signal.

    Measured: 3 hours

  3. Internal bubble CPAP sounds

    1. We will apply biomedical signal processing methods to separate the bubbling sounds from breathing sounds. Then, RMS and power will be computed.
    2. We will use the Pearson correlation coefficient to compute the relationship between bubbling sound and pressure metrics. We will evaluate the linear regression models to identify the combination of sound metrics and covariates with the highest predictive accuracy.

    Measured: 3 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted11 Jul 2025
  2. Study start3 Apr 2025
  3. Primary completionSep 2026 (estimated)
  4. Study completionDec 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
30 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07060833
Sponsor's protocol ID
2025-10708
Other IDs
2025
Acronym
Bubble CPAP
Status last verified
Mar 2026
First submitted
2 Jul 2025
Last updated
30 Mar 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

An observational study of Preterm Infant, run by McGill University Health Centre/Research Institute of the McGill University Health Centre at 1 site in Canada.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 30 Mar 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Canada. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Wissam Shalish, MD, PhD (514-412-4400 ext. 22341, wissam.shalish@mcgill.ca), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Pressure (time frame: 3 hours). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 30 Mar 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in