Post-Surgical Contamination of Rigid Endoscopic InstrumentsNCT07077499

Cleaning Efficacy of Pulsed Vacuum Device on Rigid Endoscopic Instruments

A study testing Pulsed Vacuum Cleaning and Disinfection Device and Vacuum Ultrasonic Cleaner in people with Post-Surgical Contamination of Rigid Endoscopic Instruments, run by Xue Wang at 1 site in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study is a prospective, comparative trial designed to evaluate the cleaning effectiveness of a novel pulsed vacuum cleaning and disinfection device against a conventional vacuum ultrasonic cleaner for rigid endoscopic instruments. Read more about this study

Who is it for?
Any age
What is being tested?
Pulsed Vacuum Cleaning and Disinfection Device and Vacuum Ultrasonic Cleaner
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
800 participants
Where
1 siteChina
Study dates
1 Jul 2024to 31 Dec 2024
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Any age
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Rigid endoscopic instruments scheduled for post-operative cleaning.
  • Types of instruments included: laparoscopes, hysteroscopes, arthroscopes, nephroscopes, prostate resection scopes, and thoracoscopes.

Who cannot join

None listed.

Trial locations and contacts

1 site in 1 country.

China1 site
  • Shijiazhuang Maternal and Child Health Hospital

    Shijiazhuang, Hebei, 050000

Other trials in China

About this study

Official title: Analysis of the Cleaning Effect of Pulsed Vacuum Cleaning and Disinfection Device on Rigid Endoscopic Instruments

This study is a prospective, comparative trial designed to evaluate the cleaning effectiveness of a novel pulsed vacuum cleaning and disinfection device against a conventional vacuum ultrasonic cleaner for rigid endoscopic instruments. Key outcomes measured include cleaning time, cleaning quality (via visual, light source, and ATP inspection), protein residue levels, and instrument damage rate.

Read the detailed description

Rigid endoscopic instruments, due to their complex structure, pose significant challenges for cleaning and sterilization, which is critical for preventing hospital-acquired infections. Traditional methods like ultrasonic and spray cleaning have limitations in efficiency and thoroughness. This study was conducted to assess a newer technology, the pulsed vacuum cleaning and disinfection device, which integrates cleaning, rinsing, disinfection, and drying. A total of 800 rigid endoscopic instruments were divided into two groups: one cleaned with the pulsed vacuum device (Observation Group) and the other with a conventional vacuum ultrasonic cleaner (Control Group). The study aims to provide evidence-based insights into the efficiency, efficacy, and safety of the pulsed vacuum device for clinical application in sterilization supply centers.

Being tested
Pulsed Vacuum Cleaning and Disinfection Device, Vacuum Ultrasonic Cleaner

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Pulsed Vacuum Cleaning and Disinfection DeviceDevice

    After manual pre-treatment, instruments were placed into the pulsed vacuum cleaning and disinfection machine (Shandong Xinhua Medical, Model PC-150L). The automated program included pulsed vacuum cleaning, rinsing, thermal disinfection (93°C for 2.5 minutes), and vacuum drying.

  • Vacuum Ultrasonic CleanerDevice

    After manual pre-treatment, instruments were cleaned in a vacuum ultrasonic cleaner (Shandong Xinhua Medical, Model super-6000) with a multi-enzyme solution for a minimum of 5 minutes. This was followed by a separate process in a spray cleaning and disinfection machine for rinsing and drying.

Study groups in NCT07077499
Study groupRoleWhat they receive
Pulsed Vacuum Cleaning GroupExperimentalA total of 400 rigid endoscopic instruments were cleaned using a pulsed vacuum cleaning and disinfection device following manual pre-treatment. The process integrated cleaning, rinsing, disinfection, and vacuum drying into a single automated cycle.
Receives: Device: Pulsed Vacuum Cleaning and Disinfection Device
Ultrasonic Cleaning GroupActive comparatorA total of 400 rigid endoscopic instruments were cleaned using a conventional vacuum ultrasonic cleaner, followed by a separate spray cleaning and drying process, after manual pre-treatment.
Receives: Device: Vacuum Ultrasonic Cleaner

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Protein Residue Detection

    The level of residual protein on the instrument surface was measured using the biuret method. Outcomes were categorized as clean, mild, moderate, or severe contamination. The rate of clean instruments was compared between groups.

    Measured: Assessed once per instrument, immediately following the completion of the single cleaning procedure (Day 0)

  2. Cleaning Qualification Rate

    The cleaning effect was evaluated based on the "Hospital Sterilization Supply Center Part 3: Cleaning, Disinfection, and Sterilization Effect Monitoring Standards". Qualification was determined by a composite of three methods:

    • Visual Inspection: No visible contaminants.
    • Magnifying Light Source Observation: No visible stains or residues under magnification.
    • ATP Fluorescence Detection: Relative Light Unit (RLU) value ≤2000. The percentage of instruments passing all three checks was calculated.

    Measured: Assessed once per instrument, immediately following the completion of the single cleaning procedure (Day 0)

Secondary outcomes (2)
  1. Cleaning Time

    Total time required to process all 400 instruments in each group, from the start of manual pre-treatment to the completion of the final drying step. Unit: Minutes.

    Measured: Assessed during a single time point event (the cleaning procedure on Day 0). The duration of the measurement is approximately 60 to 90 minutes.

  2. Instrument Damage Rate

    The number and percentage of instruments found to be damaged (e.g., deformation, breakage) after the complete cleaning and drying cycle in each group.

    Measured: Assessed at a single time point: immediately after the completion of the single cleaning procedure on Day 0.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted22 Jul 2025
  2. Study start1 Jul 2024
  3. Primary completion31 Dec 2024
  4. Study completion31 Dec 2024

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
800 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07077499
Sponsor's protocol ID
2024039
Status last verified
Jul 2025
First submitted
5 Jul 2025
Last updated
22 Jul 2025
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study testing Pulsed Vacuum Cleaning and Disinfection Device and Vacuum Ultrasonic Cleaner in people with Post-Surgical Contamination of Rigid Endoscopic Instruments, run by Xue Wang at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 0 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 22 Jul 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Protein Residue Detection (time frame: Assessed once per instrument, immediately following the completion of the single cleaning procedure (Day 0)). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 22 Jul 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in