A Prospective, Single-site, Single-group Value Study on the Diagnosis of Functional Myocardial Ischemia Based on IVUS Plaque Characteristics

An observational study of Coronary Artery Disease, run by Insight Lifetech Co., Ltd. at 1 site in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study aims to evaluate the accuracy of IVUS-calculated percent atheroma volume (PAV) for diagnosing functional myocardial ischemia, using FFR as the gold standard. Read more about this study

Who is it for?
Ages 18 to 80 years
Conditions the study is for
Coronary Artery Disease
How many people
105 participants
Where
1 siteChina
Study dates
23 Sep 2025to 4 Jan 2026
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 80 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: Patients with multivessel coronary artery disease.

Who may be able to join

  • Study participants must be at least 18 years of age and younger than 80 years of age, regardless of gender;
  • Stable coronary artery disease, unstable angina, or acute myocardial infarction (AMI ≥ 7 days);
  • Coronary angiography showed primary lesions with a visual stenosis of 30% to 90% in at least two coronary arteries ≥ 2.5 mm in diameter;
  • Willing to participate in the study and able to understand, read, and sign the informed consent form.

Who cannot join

  • STEMI as clinical presentation within 7 days;
  • Chronic Total Occlusion (CTO) as target vessel;
  • Significant contraindication to adenosine administration (e.g., atrioventricular block, severe asthma, etc.);
  • Study participants are currently participating in other drug or device clinical studies and have not achieved the study endpoints;
  • Pregnant or nursing;
  • Other situations considered by the investigator as unsuitable for participation in this study.

Trial locations and contacts

1 site in 1 country.

China1 site
  • Fuwai Hospital,Cams & Pumc

    Beijing, Beijing Municipality, 100037

Other trials in China

About this study

This study aims to evaluate the accuracy of IVUS-calculated percent atheroma volume (PAV) for diagnosing functional myocardial ischemia, using FFR as the gold standard.

Read the detailed description

This prospective, single-site, single-group target-value study aims to investigate the accuracy of intravascular ultrasound (IVUS)-based plaque characteristics in diagnosing functional myocardial ischemia, using coronary fractional flow reserve (FFR) as the gold standard.

Keywords
IVUS, FFR, PCI, CAD
Medical subject headings
Coronary Artery Disease

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. The diagnostic accuracy of IVUS-derived percent atheroma volume (PAV) for identifying physiologically significant coronary stenoses

    The prespecified primary endpoint was the diagnostic accuracy of IVUS-derived percent atheroma volume (PAV) for identifying physiologically significant coronary stenoses, defined as fractional flow reserve (FFR) ≤0.80, using a PAV threshold of ≥39.0%.

    Measured: Intraoperative period(After coronary angiography, before the potential PCI)

Secondary outcomes (8)
  1. The diagnostic sensitivity and specificity of IVUS-derived percent atheroma volume (PAV) for identifying physiologically significant coronary stenoses

    The key secondary endpoints were the diagnostic sensitivity and specificity of IVUS-derived percent atheroma volume (PAV) for identifying physiologically significant coronary stenoses, defined as fractional flow reserve (FFR) ≤0.80, using a PAV threshold of ≥39.0%.

    Measured: Intraoperative period(After coronary angiography, before the potential PCI)

  2. The positive predictive value, negative predictive value, Receiver Operating Characteristic Curve (ROC) and Area Under the Curve (AUC) of IVUS-derived PAV for identifying physiologically significant coronary stenoses

    Using blood vessels as the research unit and FFR values measured by pressure microcatheters as reference (FFR ≤ 0.80 is positive, FFR>0.8 is negative), evaluate the positive predictive value, negative predictive value, Receiver Operating Characteristic Curve (ROC), and Area Under the Curve (AUC) of IVUS-derived PAV for identifying physiologically significant coronary stenoses (PAV ≥ 39.0% is positive, PAV < 39.0% is negative).

    Measured: Intraoperative period(After coronary angiography, before the potential PCI)

  3. Accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of Total Atheroma Volume (TAV) calculated from IVUS images on the diagnosis of functional myocardial ischemia

    Using blood vessels as the research unit and FFR values measured by pressure microcatheters as a reference (FFR ≤ 0.80 is positive, FFR>0.8 is negative), evaluate the accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of Total Atheroma Volume (TAV) calculated from IVUS images on the diagnosis of functional myocardial ischemia.

    Measured: Intraoperative period(After coronary angiography, before the potential PCI)

  4. Accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of Minimum Lumen Area (MLA) calculated from IVUS images on the diagnosis of functional myocardial ischemia

    Using blood vessels as the research unit and FFR values measured by pressure microcatheters as a reference (FFR ≤ 0.80 is positive, FFR>0.8 is negative), evaluate the accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of Minimum Lumen Area (MLA) calculated from IVUS images on the diagnosis of functional myocardial ischemia.

    Measured: Intraoperative period(After coronary angiography, before the potential PCI)

  5. Accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of Plaque Burden (PB) calculated from IVUS images on the diagnosis of functional myocardial ischemia

    Using blood vessels as the research unit and FFR values measured by pressure microcatheters as a reference (FFR ≤ 0.80 is positive, FFR>0.8 is negative), evaluate the accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of Plaque Burden (PB) calculated from IVUS images on the diagnosis of functional myocardial ischemia.

    Measured: Intraoperative period(After coronary angiography, before the potential PCI)

  6. Accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of Diameter Stenosis (DS%) calculated by Quantitative Coronary Angiography (QCA) on diagnosing functional myocardial ischemia

    Using blood vessels as the research unit and FFR values measured by pressure microcatheters as a reference (FFR ≤ 0.80 is positive, FFR>0.8 is negative), evaluate the accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of Diameter Stenosis degree (DS%) calculated by Quantitative Coronary Angiography (QCA) on diagnosing functional myocardial ischemia.

    Measured: Intraoperative period(After coronary angiography, before the potential PCI)

  7. Correlation between PAV, TAV, MLA, PB, DS% and FFR in pairwise comparisons

    Using blood vessels as the research unit, conducting Passing Bablok regression analysis and Pearson linear correlation analysis on Percent Atheroma Volume (PAV), Total Atheroma Volume (TAV), Minimum Lumen Area (MLA), Plaque Burden (PB), Diameter Stenosis degree (DS%) and Fractional Flow Reserve (FFR).

    Measured: Intraoperative period(After coronary angiography, before the potential PCI)

  8. Correlation between PAV gradient calculated from IVUS images and FFR Pullback Pressure Gradient (PPG)

    Using blood vessels as the research unit, PAV was calculated every 1mm along the vascular direction based on IVUS images, and FFR Pullback Pressure Gradient (PPG) distribution was calculated based on FFR pullback. The correlation between PAV gradient distribution along the vascular direction and FFR PPG was evaluated.

    Measured: Intraoperative period(After coronary angiography, before the potential PCI)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted16 Sep 2025
  2. Study start23 Sep 2025
  3. Primary completion4 Jan 2026
  4. Study completion4 Jan 2026

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
105 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07176585
Sponsor's protocol ID
CIP-0201
Status last verified
Oct 2026
First submitted
1 Sep 2025
Last updated
26 Oct 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

An observational study of Coronary Artery Disease, run by Insight Lifetech Co., Ltd. at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 26 Oct 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

The diagnostic accuracy of IVUS-derived percent atheroma volume (PAV) for identifying physiologically significant coronary stenoses (time frame: Intraoperative period(After coronary angiography, before the potential PCI)). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 26 Oct 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in
Clinical Trial NCT07176585 | AzTrials