Hip Arthroplasty ReplacementCUPUSURENCT07207005

Comparison of Wear Between Two Dual-mobility Acetabular Cups in Patients Undergoing Total Hip Arthroplasty

A study testing ECIMA CORIN and CORIN in people with Hip Arthroplasty Replacement, run by GCS Ramsay Santé pour l'Enseignement et la Recherche at 1 site in France.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

In 2022, data from the UK's National Joint Registry (NJR) and the American Joint Replacement Registry (AJRR) indicated that Total Hip Replacement (THR) is the most commonly performed orthopaedic joint surgery. The NJR has observed that between 10 and 20% of hip implants require revision within the first 10 years. Read more about this study

Who is it for?
Ages 55 to 80 years
Conditions the study is for
Hip Arthroplasty Replacement
What is being tested?
ECIMA CORIN and CORIN
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
96 participants
Where
1 siteFrance
Study dates
30 Jul 2026to 31 Dec 2038
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 55 to 80 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patient, male or female, aged between 55 and 80.
  • Patient with a life expectancy > 5 years at the time of inclusion.
  • Patient scheduled to undergo primary total hip arthroplasty with a dual-mobility cup prosthesis.
  • Patient affiliated with or beneficiary of a social security scheme.
  • Patient who speaks French and has signed an informed consent form.

Who cannot join

  • Patients with a contraindication mentioned in the instructions for use of the medical device under investigation
  • Patients with an ASA score of 4
  • Protected patients: adults under guardianship, curatorship or other legal protection, deprived of their liberty by judicial or administrative decision
  • Patients participating in another clinical study whose objectives could interfere with this study

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

JEAN LANGLOIS, MD contact

+ 33 6 03 29 45 33

jeangast@gmail.com

France1 site, 1 recruiting
  • Hôpital Privé Jean Mermoz

    Lyon, France, 69008

    Jean LANGLOIS, MD, +33 6 03 29 45 33, jeangast@gmail.com

    Recruiting

Other trials in France

About this study

In 2022, data from the UK's National Joint Registry (NJR) and the American Joint Replacement Registry (AJRR) indicated that Total Hip Replacement (THR) is the most commonly performed orthopaedic joint surgery.

The NJR has observed that between 10 and 20% of hip implants require revision within the first 10 years. The main reasons for these revisions include implant fractures, joint incompatibility, aseptic loosening, dislocation and wear of the polyethylene (PE) insert.

Read more

To address this problem, highly cross-linked polyethylene (HXLPE) inserts have been developed. Cross-linking helps to limit wear on polyethylene prostheses in total hip arthroplasty. However, there are concerns about the loss of elasticity of HXLPE and oxidation after implantation.

A new generation of inserts based on highly cross-linked polyethylene (irradiation dose ≥100 Kgy), enriched with vitamin E, has been developed. This PE, called ECIMA, was developed to maintain mechanical properties, minimise PE wear and improve long-term oxidation resistance.

In the HAS evaluation report on highly cross-linked polyethylene inserts dated 24 January 2023, data from the literature show less wear of highly cross-linked polyethylene compared to conventional polyethylene.

Nevertheless, the use of highly cross-linked polyethylene is not yet the gold standard. Furthermore, the data in the literature mainly concern single-mobility cups but not double-mobility cups. Finally, new imaging protocols now make it possible to assess the wear of prosthesis materials more accurately.

In this context, it seems appropriate to conduct a French study comparing double-mobility cups containing highly cross-linked polyethylene or conventional polyethylene inserts.

Read the detailed description

In 2022, data from the UK's National Joint Registry (NJR) and the American Joint Replacement Registry (AJRR) indicated that total hip arthroplasty or Total Hip Replacement (THR) is the most commonly performed orthopaedic joint surgery.

THR involves replacing worn and damaged areas of cartilage in the hip with artificial parts of similar shape and size. A total hip replacement consists of two parts:

  • A part designed to replace the head of the femur: a spherical prosthesis mounted on a stem that is inserted into the femur;
  • A part designed to replace the acetabulum: the acetabular cup. There are single-mobility cups and double-mobility cups, where an insert is positioned on the metal head inside a single-mobility cup. Introduced in the 1960s by Sir John Charnley, UHMWPE (Ultra High Molecular Weight Polyethylene) remains the most widely used material for inserts in 2024.

The NJR has observed that between 10 and 20% of hip implants require revision within the first 10 years. The main reasons for these revisions include implant fractures, joint incompatibility, aseptic loosening, dislocation and wear of the polyethylene (PE) insert.

To address this problem, highly cross-linked polyethylene (HXLPE) inserts have been developed. Cross-linking helps to limit wear on polyethylene prostheses in total hip arthroplasty. However, there are concerns about the loss of elasticity of HXLPE and oxidation after implantation.

A new generation of inserts based on highly cross-linked polyethylene (irradiation dose ≥100 Kgy), enriched with vitamin E, has been developed. This PE, called ECIMA, was developed to maintain mechanical properties, minimise PE wear and improve long-term oxidation resistance.

In the HAS evaluation report on highly cross-linked polyethylene inserts dated 24 January 2023, data from the literature show less wear of highly cross-linked polyethylene compared to conventional polyethylene.

Nevertheless, the use of highly cross-linked polyethylene is not yet the gold standard. Furthermore, the data in the literature mainly concern single-mobility cups but not double-mobility cups. Finally, new imaging protocols now make it possible to assess the wear of prosthesis materials more accurately.

In this context, it seems appropriate to conduct a French study comparing double-mobility cups containing highly cross-linked polyethylene or conventional polyethylene inserts.

In the principal investigator's current practice, prostheses with conventional PE inserts are used for patients aged 70 or over, while prostheses with highly cross-linked PE inserts are used for patients under 70. However, this age-based choice is modulated favourably or unfavourably by two other parameters: the patient's medical history and general condition.

Being tested
ECIMA CORIN, CORIN
Keywords
Hip Arthroplasty Insert

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • ECIMA CORINDevice

    Mobile cup (Corin) made of highly cross-linked PE doped with vitamin E (ECIMA, Corin).

  • CORINDevice

    Mobile cupule (Corin) is made of conventional PE

Study groups in NCT07207005
Study groupRoleWhat they receive
Highly cross-linked PEExperimentalThe hip prosthesis will be a MOBILIT acetabular cup (Corin), MEIJE femoral stem (Corin), 28 mm diameter metal ball (Corin), and mobile cup (Corin) made of highly cross-linked PE doped with vitamin E (ECIMA, Corin).
Receives: Device: ECIMA CORIN
Conventional PEActive comparatorThe hip prosthesis will be a MOBILIT (Corin) acetabular prosthesis, MEIJE (Corin) femoral stem, 28 mm diameter metal ball (Corin), and mobile cup (Corin) in conventional PE.
Receives: Device: CORIN

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Femoral head penetration into the acetabulum

    Femoral head penetration into the acetabulum will be measured in mm/year, based on a hip X-ray and PolyWare® software, Rev 8, Draftware Inc, North Webster, IN, USA

    Measured: Year 5

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted3 Oct 2025
  2. Study start30 Jul 2026
  3. Primary completion31 Dec 2038 (estimated)
  4. Study completion31 Dec 2038 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
Single blind
Who is masked
Participant
Enrolment
96 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07207005
Sponsor's protocol ID
2024-A02615-42
Acronym
CUPUSURE
Status last verified
Aug 2025
First submitted
26 Sep 2025
Last updated
20 Aug 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing ECIMA CORIN and CORIN in people with Hip Arthroplasty Replacement, run by GCS Ramsay Santé pour l'Enseignement et la Recherche at 1 site in France.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 20 Aug 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact JEAN LANGLOIS, MD (+ 33 6 03 29 45 33, jeangast@gmail.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Femoral head penetration into the acetabulum (time frame: Year 5). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 20 Aug 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in