Vertebral Compression FractureEXPANDNCT07301749

A Clinical Study Evaluating the Safety and Efficacy of the VCFix Spinal System for Treatment of Vertebral Compression Fractures

A study testing VCFix Spinal System in stand-alone configuration in people with Vertebral Compression Fracture, run by Amber Implants B.V. at 3 sites in 2 countries.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

This study is being conducted to evaluate the safety and performance of the VCFix Spinal System, a medical device designed to treat certain types of spinal fractures called vertebral compression fractures (VCFs) in adults. Up to 8 hospitals in Europe will take part in the study. Read more about this study

Who is it for?
Ages 21 to 85 years
Conditions the study is for
Vertebral Compression Fracture
What is being tested?
VCFix Spinal System in stand-alone configuration
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
148 participants
Where
3 sitesin 2 countries
Study dates
4 Nov 2025to 31 Dec 2029
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 21 to 85 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Subject is ≥ 21 to ≤ 85 years old.
  • Single vertebral fracture which meets all of the following criteria:
  • Type A fractures from AO spine classification (compression injuries);
  • Vertebral fracture located in the thoracic and lumbar regions of the spine
  • Fracture age < 6 weeks or fractures with anatomic signs of reducibility (based on radiographic evidence as well as patient history)
  • Vertebral fracture shows an estimated height loss in the anterior or mid third of the vertebral body (VB) of at least 10% but not more than 60% based on radiographic evidence
Show all 12Show fewer
  • Target VB has appropriate anatomy, i.e. suitable pedicle diameter, pedicle length and chord length that would allow for selection of correct implant size, as described in the IFU and based on axial MRI/CT scan before surgery
  • Subject has a NPRS back pain score of ≥5
  • Subject is a candidate for surgical intervention based on investigator opinion
  • Subject has a body Mass Index (BMI) < 35
  • Subject is mentally capable of complying with trial protocol requirements for the duration of the study
  • Subject can understand the risks and benefits of participating in the study and is able to provide written informed consent

Who cannot join

  • Neoplasms with posterior involvement and/or presence of a mass within the spinal canal
  • Non-mobile fractures (i.e., fracture is not recent (>6 weeks), bone marrow edema or fluid or empty cleft are not visible in radiographic imaging, or fracture mode does not allow for fracture reduction in craniocaudal direction).
  • Spondylolisthesis > Grade 1 at target vertebral body(s)
  • Local kyphotic angle > 30°
  • Pre-existing vertebral fracture prior to the index fracture
  • Subjects that require anterior stabilization of the index fracture
Show all 27Show fewer
  • Fracture to the pedicle based on radiographic evaluation
  • Spinal cord compression or canal compromise requiring decompression
  • Severe back pain due to causes other than acute fracture with NPRS score>5
  • The subject has pain based on clinical diagnosis of herniated nucleus pulposus or severe spinal stenosis (progressive weakness or paralysis)
  • Pain due to any other condition that requires daily narcotic (opiates or opioids) medication
  • Pre-existing neurological deficit, radiculopathy or myelopathy
  • Pre-existing condition or significant co-morbidity:
  • Uncontrolled diabetes (HbA1c >8%)
  • Severe cardiopulmonary deficiencies
  • Any other condition that would impact study outcome in the opinion of the investigator
  • Contraindications to both MRI and radionuclide bone scan
  • Concurrent participation in another clinical study which could potentially interfere with the outcome of this study.
  • A life expectancy less than the study duration or undergoing palliative care
  • Subject non-ambulatory prior to fracture
  • Allergy to any components of the device/instruments used during the procedure
  • Active or incompletely treated infection of the vertebral column or active systemic infection, including unresolved urinary tract infection
  • Any underlying systemic bone disease other than osteoporosis (e.g., osteomalacia, osteogenesis imperfecta, Paget's disease, etc.)
  • Any evidence of alcohol or drug abuse.
  • Ongoing long-term steroid therapy (steroid dose ≥30 mg /day for >3 months)
  • The subject is currently on anti-cancer therapy or anti-HIV therapy
  • Pregnancy or subjects with child-bearing potential that are unwilling to use contraception throughout the study duration

Trial locations and contacts

3 sites in 2 countries. Contact the study team or a site near you to ask about taking part.

Clinical Research Manager contact

+31 15 200 2151

expand@amberimplants.com

3 sites in 2 countries

Germany2 sites, 2 recruiting
  • Wilhelmsburger Krankenhaus Groß-Sand

    Hamburg

    Neurosurgery Department, +49 40 75 205, sprechstunde@gross-sand.de

    Recruiting
  • Krankenhaus Mechernich

    Mechernich

    Orthopaedic Department, +49 2443 170, ortho-ambulanz@kkhm.de

    Recruiting

Other trials in Germany

France1 site, 1 recruiting
  • Centre Hospitalier Métropole de Savoie

    Chambéry

    Unité de Recherche Clinique, Dr., +33 4 79 96 50 30, info@ch-metropole-savoie.fr

    Not yet recruiting

Other trials in France

About this study

Official title: A Prospective Multicenter Clinical Investigation Evaluating the Safety and Efficacy of the VCFix Spinal System in Stand-alone Configuration for Vertebral Augmentation (EXPAND)

This study is being conducted to evaluate the safety and performance of the VCFix Spinal System, a medical device designed to treat certain types of spinal fractures called vertebral compression fractures (VCFs) in adults.

Up to 8 hospitals in Europe will take part in the study. All participants will receive treatment with the VCFix device. The treatment will be provided according to the intended use of the device and may be performed either with or without the use of a special bone cement.

Read more

Participants will be assigned to one of two study groups before treatment:

Cementless group: the VCFix device is implanted without bone cement. Cemented group: the VCFix device is implanted together with bone cement.

The study will assess how well the treatment works and how safe it is. Researchers will evaluate:

Changes in back pain 6 months after treatment compared with before treatment. Changes in daily functioning and ability to perform everyday activities 6 months after treatment compared with before treatment.

Whether any serious device-related complications occur during the first month after treatment.

The results will help determine whether treatment with the VCFix device provides benefits for patients with vertebral compression fractures while maintaining an acceptable safety profile.

Being tested
VCFix Spinal System in stand-alone configuration
Keywords
VCF, EXPAND, vertebral stabilization, fracture, trauma, osteoporosis, expandable implant
Medical subject headings
Fractures, Bone, Wounds and Injuries, Osteoporosis

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • VCFix Spinal System in stand-alone configurationDevice

    The distinguishing features of this inetrvention reside in:

    1. In situ adjustment and maintenance of the height and angle of the device that allows expansion of the fractured VB in the cranio-caudal direction to realign the vertebral endplates and correct kyphosis.
    2. The complex titanium perforated structure with optimized surface roughness that allows possible bone attachment, thereby permitting natural healing of the bone and potentially providing more stability to the device and the treated vertebra over time.
    3. The ability to likely secure and reinforce the vertebra without using PMMA bone cement, avoiding all the symptomatic and asymptomatic side effects of bone cement.
    4. The ability to connect the posterior to the anterior column of the spine through the pedicle screw component of the VCFix implant, therefore improving the biomechanical load distribution and consequently the ability to stabilize more severe fractures.
Study groups in NCT07301749
Study groupRoleWhat they receive
Cemented cohortExperimentalImplant of the VCFix device with the use of PMMA bone cement
Receives: Device: VCFix Spinal System in stand-alone configuration
Uncemented cohortExperimentalImplant of the VCFix device without the use of PMMA bone cement
Receives: Device: VCFix Spinal System in stand-alone configuration

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Reduction of more than 2 points in vertebral fracture-related pain intensity at 6 months after the index procedure compared to baseline as measured by a 11-point Numerical Pain Rating Scale (NPRS)

    Measured: 6 months

  2. Maintenance or improvement of function as measured by the 100-point Oswestry Disability Index (ODI) at 6 months compared to baseline

    Measured: 6 months

  3. Absence of serious adverse device effects (SADEs): No SADEs with causality rated as definite, monitored through 1 month post-procedure (safety)

    Measured: 1 month

Secondary outcomes (7)
  1. Vertebral height restoration (anterior, midline, and posterior aspects) at PO (Post-Operative), 1Months FU, 6 Months FU and 12Months FU.

    Measured: Post-Operative, 1Month FU, 6 Months FU, 12 Months FU

  2. Kyphotic angle and Cobb angle improvement at Post-Operative, 1 Month FU, 6 Months FU and 12 Months FU.

    Kyphotic angle and Cobb angle improvement

    Measured: Post-Operative, ost-Operative, 1 Month FU, 6 Months FU and 12 Months FU.

  3. Reduction in vertebral fracture-related pain (measured by the 11-point Numerical Pain Rating Scale (NPRS))

    Reduction in vertebral fracture-related pain

    Measured: Discharge, 1, 3, 6 and 12 Months FU compared to baseline.

  4. Improvement in function (measured by the 100-point Oswestry Disability Index (ODI)) at discharge, 1, 3, 6 and 12 months FU compared to baseline.

    Measured: Discharge, 1, 3, 6 and 12 months FU

  5. Improvement of Health-related quality of life (HRQoL) measured by EuroQol 5-Dimension Scale (EQ-5D-5L) at 1 Month, 3 Months, 6 Months and 12 Months post-procedure compared to baseline

    Improvement of Health-related quality of life (HRQoL)

    Measured: 1 Month, 3 Months, 6 Months and 12 Months post-procedure compared to baseline

  6. Device-related serious adverse events assessed at 12- and 24- months post-procedure.

    Device-related serious adverse events

    Measured: 12 months FU, 24 months FU

  7. All (per protocol reportable) adverse events assessed at 12- and 24- months post-procedure.

    Measured: 12 Months FU, 24 Months FU

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted24 Dec 2025
  2. Study start4 Nov 2025
  3. Primary completion31 Jul 2027 (estimated)
  4. Study completion31 Dec 2029 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Non-randomised
Intervention model
Parallel groups
Model description
The study comprises two predefined cohorts, assigned based on protocol-specific criteria applied prior to treatment: * a cementless cohort, where the device is implanted without polymethylmethacrylate (PMMA) bone cement * a cemented cohort, where the device is used in conjunction with PMMA bone cement Outcomes for each cohort will be evaluated independently against an external control. No direct comparison between the study cohorts is planned.
Masking
None (open label)
Enrolment
148 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07301749
Sponsor's protocol ID
CIP-VS02-EU
Acronym
EXPAND
Status last verified
Jun 2026
First submitted
19 May 2025
Last updated
22 Jun 2026
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing VCFix Spinal System in stand-alone configuration in people with Vertebral Compression Fracture, run by Amber Implants B.V. at 3 sites in 2 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 12 inclusion and 27 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 22 Jun 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 3 sites in Germany, France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Clinical Research Manager (+31 15 200 2151, expand@amberimplants.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Reduction of more than 2 points in vertebral fracture-related pain intensity at 6 months after the index procedure compared to baseline as measured by a 11-point Numerical Pain Rating Scale (NPRS) (time frame: 6 months). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 22 Jun 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in